Welch Allyn FN802FH Propaq 802 LTRN User Manual part 7 of 10

Welch Allyn, Inc. Propaq 802 LTRN Users Manual part 7 of 10

Users Manual part 7 of 10

585Alarms and AlertsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85Responding to an Alarm  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90Silencing an Alarm or Alert Tone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86Suspending the Alarm Tone. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87Customizing Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89About ParamSet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90Responding to an Alert  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90About Battery Charge Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91Alert Messages and Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92OverviewAn Alarm warns of a patient condition, such as a vital-sign reading that is outside of acceptable limits. When an alarm occurs, the red light on the monitor flashes and the numerics of the violating alarm limits on the display turn red. If tones are not suspended, the alarm tone sounds.An Alert warns of an equipment condition, such as a low battery or a detached lead. When an alert occurs, the yellow light on the monitor flashes and a message describing the error condition appears on the display. If tones are not suspended, the alert tone sounds.Alarms have priority over alerts. If an alarm and an alert are detected simultaneously, the monitor notifies you of the alarm. It then notifies you of the alert only if the alert condition still exists after the alarm condition is removed.
86 Alarms and Alerts Welch Allyn Propaq LT Vital Signs MonitorSilencing an Alarm or Alert ToneA tone sounds whenever the monitor detects an alarm or alert condition. To Silence the Currently Sounding Tone for 90 Seconds1. P r e s s .  • The red light (alarm) or the yellow light (alert) flashes. • After 90 seconds, if the condition is not corrected, the tone starts again. • If the condition is corrected within 90 seconds of silencing the tone, the monitor resets the tones for the next alarm or alert. If a new alarm or alert condition occurs while an earlier alarm or alert is silenced, the tone sounds again.Figure 84.  Sample Alarm ScreenFigure 85.  Sample Alert Screen2. Check the patient and provide appropriate care.Note Silencing the tone does not affect the other alarm or alert indicators.%140/78140/7812512531318888HALL, ROBERT E.HALL, ROBERT E.345618734561873:00:06P3:00:06PAdultAdultRm 239Rm 239HR/minHR/minNIBP mmHg (102)NIBP mmHg (102)@2:47P Manual@2:47P ManualResp/minResp/minSpO2SpO2II  1mV/cmII  1mV/cmXResp/minResp/minSpO2SpO2I      1mV/cmI      1mV/cmHALL, ROBERT E.HALL, ROBERT E.3456187345618712:41:3212:41:32AdultAdultRm 239Rm 239Audible Alarms SilencedAudible Alarms Silenced8080140/78140/7812129797 %HR/minHR/minNIBP mmHg (102)NIBP mmHg (102)@2:47P Manual@2:47P ManualEQUIPMENT ALERTEQUIPMENT ALERTECG FaultECG FaultLL Lead FailedLL Lead FailedACKNOWLEDGEACKNOWLEDGE
Directions for Use Alarms and Alerts 873. Press  to silence the audible tone for 90 seconds at the monitor and at Acuity.• Silencing the audible tone does not remove visual alarm or alert indications.• Silencing the audible tone causes an alarm or alert suspend at Acuity.• When the alarm or alert condition is corrected, all alarm or alert indicators cease and all alarms are immediately rearmed.4. After caring for the patient, verify that alarm limits are enabled and correctly set.Suspending the Alarm ToneIf this feature is enabled in the monitor configuration (See “Monitor Configuration” on page 111), you can suspend all alarm tones for all parameters—preventing the alarm tone from sounding if an alarm condition occurs—while monitoring a patient. If an alarm condition occurs while the alarm tones ars suspended, the monitor presents visual alarm indicators but does not sound the tone.In the monitor configuration, the alarm tone suspension period can be set to Disabled, to Always On, or to a period: 90 sec or 2, 3, 4, 5, 10, 15, 30, or 60 minutes.• If it is configured to Disabled, you cannot suspend the alarm tone at the monitor.• If it is configured to Always On and you set Suspend Audible Alarms to On, then the alarm tone remains suspended until:•you set Suspend Audible Alarms to Off or • monitor power is turned off and then turned on again or• the monitor is reconfiguredTo Suspend the Alarm Tone1. Access the Setup menu. (See “To Access the Setup Menus” on page 38.)2. Highlight Suspend Audible Alarms (Figure 86). The configured suspension period—90 seconds in this example—is displayed to the right of the line.WARNING   If you turn off or modify any alarm limits while responding to an alarm, restore the appropriate alarm limits before you resume monitoring.Note The factory default suspension period is 4 minutes.Note Suspend Audible Alarms does not affect the behavior of the alarm silence/reset feature ( ). Pressing   always either silences a sounding alarm tone for 90 seconds or resets the audible alarm if it was already silenced.
88 Alarms and Alerts Welch Allyn Propaq LT Vital Signs MonitorFigure 86.  Suspend Audible Alarms: Off3. Highlight On (Figure 87).Figure 87.  Suspend Audible Alarms: On• The alarm tone is suspended immediately. If an alarm condition occurs during the suspension period, the alarm tone does not sound.• A count-down timer appears below the line to indicate the time remaining in the suspension period.• ‘Audio alarms suspended’ appears in yellow in the upper left corner of the screen.• When the suspension period elapses, the alarm tone is again enabled.Note If audible alarms are suspended, pressing   cancels the suspension.II      1mV/cmII      1mV/cmExitExitTrendsTrendsSnapshotsSnapshotsSetupSetupLowLowMedMedHR/minHR/minNIBP mmHgNIBP mmHgResp/minResp/minSpO2SpO2OffOffOnOnAlarm ToneAlarm ToneSuspend Audible AlarmsSuspend Audible AlarmsSetupSetup(90 sec)(90 sec)%3:00:06P3:00:06PAdultAdultRm 239Rm 239AlarmsAlarmsECGECGNIBPNIBPTimingsTimingsHighHighServiceServiceHALL, ROBERT E.HALL, ROBERT E.34561873456187Right (   ) to suspend audible alarms.Right (   ) to suspend audible alarms.II      1mV/cmII      1mV/cmExitExitTrendsTrendsSnapshotsSnapshotsSetupSetupLowLowMedMed8080HR/minHR/minNIBP mmHgNIBP mmHgResp/minResp/minSpO2SpO2OffOffOnOnAlarm ToneAlarm ToneSuspend Audible AlarmsSuspend Audible AlarmsSetup:Setup:(90 sec)(90 sec)12129797%876543287654323:00:06P3:00:06PAdultAdultRm 239Rm 239AlarmsAlarmsECGECGNIBPNIBPTimingsTimingsHighHighServiceService0:01:30 remaining0:01:30 remainingAudio alarms suspendedAudio alarms suspendedID: 01018VDO9PBHID: 01018VDO9PBH‘Tone suspended’ indicatorSuspension time remainingConfigured suspension period
Directions for Use Alarms and Alerts 89Customizing Alarm LimitsAt the MonitorTypically, each institution defines the patient alarm limits for adult, pediatric, and neonatal patients, and then configures the monitor with those alarm limits before putting the monitor into service. Alarm limits can be temporarily customized for an individual patient while the monitor is in use. Later, when the monitor is turned off in preparation for monitoring another patient, the temporary alarm limits are lost and the configured alarm limits are restored.To Temporarily Customize Alarm Limits for the Current Patient1. Highlight the vital sign for which you want to set custom limits.2. Press .3. Highlight the limit you want to change.4. Set a new alarm limit.Figure 88.  Customizing Alarm Limits for the Current Patient5. Press .At AcuityFor a wireless monitor, patient alarm limits can also be customized from Acuity. (See the user manual for any Acuity Central Monitoring Station.)II      1mV/cm II      1mV/cm 1 1 2 2 3 3 5 5 10 10 15 15 30 30 60 60 Exit Exit Trends Trends Snapshots Snapshots Setup Setup Auto Auto Manual Manual Turbo Turbo HR/min HR/min NIBP mmHg  NIBP mmHg  Resp/min Resp/min SpO2 SpO2 Off Off Off Off On On On On 220 220 Lower Sys Limit Lower Sys Limit NIBP Mode NIBP Mode Auto Interval (min) Auto Interval (min) Upper Sys Alarm Upper Sys Alarm Upper Sys Limit Upper Sys Limit Lower Sys Alarm Lower Sys Alarm NIBP   Manometer NIBP   Manometer Systolic Systolic Diastolic Diastolic Mean Mean 75 75 % % STEWART, ANN STEWART, ANN 7762940 7762940 3:00:06P 3:00:06P Adult Adult Rm 263Rm 263125 125 140/78 140/78 31 31 88 88 (102) (102) II      1mV/cm II      1mV/cm 1 1 2 2 3 3 5 5 10 10 15 15 30 30 60 60 Exit Exit Trends Trends Snapshots Snapshots Setup Setup Auto Auto Manual Manual Turbo Turbo HR/min HR/min NIBP mmHg  NIBP mmHg  Resp/min Resp/min SpO2 SpO2 Off Off Off Off On On On On 210 210 Lower Sys Limit Lower Sys Limit NIBP Mode NIBP Mode Auto Interval (min) Auto Interval (min) Upper Sys Alarm Upper Sys Alarm Upper Sys Limit Upper Sys Limit Lower Sys Alarm Lower Sys Alarm NIBP   Manometer NIBP   Manometer Systolic Systolic Diastolic Diastolic Mean Mean 65 65 % % STEWART, ANN STEWART, ANN 7762940 7762940 3:00:06P 3:00:06P Adult Adult Rm 263Rm 263(102) (102) 125 125 140/78 140/78 31 31 88 88
90 Alarms and Alerts Welch Allyn Propaq LT Vital Signs MonitorAbout ParamSetUsing ParamSet, you can quickly widen the alarm limits by a configured percentage (relative to the patient’s alarming reading) for any vital sign. For information about ParamSet, see “ParamSet” on page 154.Responding to an AlarmAn alarm condition is indicated on the monitor in the following ways:• The RED indicator (rightmost of three) above the display screen flashes.• The numerics for the vital sign in alarm are displayed in RED.• The alarm-tone sequence sounds repeatedly—3 short tones, a short pause, 2 short tones, and a long pause.Responding to an AlertAn alert condition (“Over view” on page 85) is indicated on the monitor in the following ways:• A flashing yellow light above the monitor display.• A yellow alert message on the monitor display (for example, NO ECG CABLE DETECTED).• Repeated sounding of the alert-tone sequence (if tones are not suspended): 3 long tones and a pause.Figure 89.  Example Alert Screen14040/7878808012129797%HALL, ROBERT E.HALL, ROBERT E.345618734561873:00:06P3:00:06PAdult,Adult,Rm 239Rm 239HR/minHR/minNIBP mmHg (102)NIBP mmHg (102)@2:47P Manual@2:47P ManualResp/minResp/minSpO2SpO2SpO2SpO22x2xII  1mV/cmII  1mV/cmPropaq LTThe yellow light flashes during an alert and while the alert is silenced.Press   to silence the alert tone for 90 seconds.Text on a yellow field identifies the alert condition. XResp/minResp/minSpO2SpO2II      1mV/cmII      1mV/cmHALL, ROBERT E.HALL, ROBERT E.3456187345618712:41:3212:41:32AdultAdultRm 239Rm 239Audible Alarms SilencedAudible Alarms Silenced8080140/78140/7812129797 %HR/minHR/minNIBP mmHg (102)NIBP mmHg (102)@2:47P Manual@2:47P ManualEQUIPMENT ALERTEQUIPMENT ALERTECG FaultECG FaultLL Lead FailedLL Lead FailedACKNOWLEDGEACKNOWLEDGE
Directions for Use Alarms and Alerts 91To Respond at the Monitor to an Alert1. Press  to silence the tone for 90 seconds; press  to clear the alert.2. Troubleshoot and correct the problem. (See “Alert Messages and Status Messages” on page 92.)To Respond at Acuity to an AlertFor a monitor in communication with Acuity, alarms and alerts can be detected by either the monitor or Acuity, and are displayed in both places. See Acuity Directions For Use.About Battery Charge StatusTable 8.  Battery Status IndicatorsDisplay Status/AlertTextBattery Monitor Recommended ActionGreen(none) Fully charged Okay NoneGreenPartially full Okay NoneGreenPartially full; charging Okay NoneYellowEQUIPMENT ALERTLow Battery.Charge battery soon.Low Can function for up to 30 minutes, but NIBP is disabled.Prepare to discontinue monitoring. If possible, insert the monitor into a cradle.YellowLow; charging Normal function. Do not remove the monitor from the cradle.RedEQUIPMENT ALERTBattery Too Low.Shutting down.Almost completely dischargedShutting down soon. Can function for up to 5 minutes, but NIBP is disabled. Prepare to discontinue monitoring. If possible, insert the monitor into a cradle.If this indicator appears when the monitor is in a powered cradle, then the battery is damaged and must be replaced.In this case, all stored patient data will be deleted when the monitor is removed from the cradle.RedAlmost completely discharged; chargingNormal function.  Do not remove the monitor from the cradle.
92 Alarms and Alerts Welch Allyn Propaq LT Vital Signs MonitorAlert Messages and Status MessagesTable 9.  Alert Messages and Status MessagesAlert Type Message Possible Cause and Suggested ResponseECG ECG Fault.XX lead failed.Lead XX (LA, LL, RA, C, or RL) has very poor contact or no contact with the patient. Check for proper connection and replace the electrode if needed.ECG Fault.XX, XX leads failed.Leads XX and XX (any two leads on a 5-lead cable) have very poor contact or no contact with the patient. Check for proper connection; replace electrodes if needed.ECG Fault.Multiple lead fail.At least three leads of a 5-lead cable or at least two leads of a 3-lead cable have very poor contact or no contact with the patient. Check for proper connection; replace electrodes if needed.ECG Fault.Excessive offset.At least one channel has excessive offset. At least one electrode is old, contaminated, or defective. Replace the electrodes.ECG Fault.Cable disconnected.The ECG cable is unplugged.NIBP NIBP Fault.Air leak. Check hose.The monitor could not properly inflate the cuff. Check the hose and cuff for leaks.NIBP Fault.Kinked hose.Check hose.The monitor could not properly inflate the cuff. Check for a hose kink between the monitor and the patient.NIBP Fault.Overpressure condition.The pressure in the cuff exceeded the acceptable limits for the current patient mode. Check the hose and retry the measurement.NIBP Fault.Weak Pulses. Can’t find Sys/Dia.Not enough pulses to determine the systolic or diastolic pressures, but a mean pressure is available. Squeeze all air from the cuff and reapply it.NIBP Fault.Artifact. Can’t find Sys/Dia.The systolic or diastolic pressures are unreliable due to artifact, but a mean pressure is available. Usually caused by patient motion.NIBP Fault.No pulses detected.The cuff might not be properly applied to the patient, or the patient might not have detectable pulses due to shock or arrhythmias.WARNING The monitor cannot determine whether this alert has a physiologic cause or a cuff application cause. Always evaluate the patient for presence of life-threatening conditions when this message occurs.NIBP Fault.Connect ECG to reduce NIBP artifact.NIBP artifact prevents a valid reading. Connect ECG electrodes to improve NIBP measurements. (See “Improving NIBP Accuracy with Smartcuf” on page 71.)
Directions for Use Alarms and Alerts 93NIBP Fault.No valid blood pressure found.The patient mode setting is incorrect or the wrong hose or cuff is being used for the current patient mode.NIBP Fault.Calibrating. Please wait.The monitor periodically calibrates (zeroes) the NIBP channel to make sure it can properly make NIBP measurements. No NIBP monitoring can be done until the calibration is completed. Other normal monitor operation continues during NIBP calibration.NIBP Fault.Calibrating. Minimize motion.Motion is detected during a periodic NIBP calibration. Minimize patient motion or motion on the cuff, or disconnect the cuff. Motion-generated noise on the pressure transducer can cause the calibration to continue indefinitely.NIBP Fault.Low battery. NIBP disabled.The battery is too far discharged to operate the NIBP channel. Insert the monitor into a powered cradle.NIBP Fault.Service required. NIBP disabled.Have the monitor serviced.NIBP Fault.Kinked or neonate hose.A hose is kinked or a neonate hose is detected in the adult patient mode. Check the hose and the patient mode selection.NIBP Fault.Artifact present.Minimize motion.The monitor has detected too much artifact to allow accurate readings. Take steps to reduce artifact. Position the patient’s limb away from the body so the applied cuff is not in contact with the patient’s body or any other object such as a bed rail.Network Communication Comms Fault.Check Acuity/network connection.The monitor detects a network communication problem.Comms Fault.Check USB connection.The monitor detects a problem in communication with the cradle. Detach and reattach the USB cable.Not on Network.Patient info entry not allowed.You attempted to select the Name, ID, or Rm field on an Acuity-enabled monitor that is not connected to the network.Battery Low Battery.Charge battery soon.The monitor battery charge is low, and the monitor will shut down in 30 minutes or less. Insert the monitor into the cradle. If no cradle is available, find an alternative method of monitoring the patient before the monitor shuts down.Very Low Battery.Charge battery now.The monitor battery charge is very low; the monitor will shut down in 5 minutes or less. Insert the monitor into a cradle or find another way to monitor the patient before the monitor shuts down.Battery Too Low.Shutting down.The monitor battery charge is too low to support monitor function. Monitor operation can continue only after the battery is recharged or replaced or until the monitor is inserted in a powered cradle.Table 9.  Alert Messages and Status Messages (continued)Alert Type Message Possible Cause and Suggested Response
94 Alarms and Alerts Welch Allyn Propaq LT Vital Signs MonitorCharger Charger Fault.Service charger.Service required.Charger Disabled.Battery temperature too high or low.The battery is too cold or too hot to charge. Normalize the battery temperature before attempting to charge it.Battery Fault.Replace battery.The battery is missing; the battery is discharged too far to be charged; the charger has timed out; a cell in the battery pack is overcharged due to cell imbalance; the fuse is blown. Service required.SpO2SpO2 Fault.No sensor detected.An SpO2 sensor has been disconnected from the monitor after being connected for more than a few seconds.SpO2 Fault.Defective SpO2 sensor.Replace the sensor.Resp Resp Fault.Lead fail.One or more electrodes have very poor or no contact. Check for proper connection; replace electrodes if needed.Resp Fault.Noisy signal.Check electrodes.Electrodes have poor contact and might be dried out. Replace electrodes.Resp Fault.Inappropriate ECG cable.The ECG cable does not contain 1 kΩ current-limiting resistors, which are required for Resp operation and to protect the monitor from damage during defibrillation. Replace the cable with one of the proper type.General Multiple Faults. Multiple equipment alerts have been triggered simultaneously. WARNING If you acknowledge this alert message before determining which alerts are triggered, you cannot identify individual alerts. Table 9.  Alert Messages and Status Messages (continued)Alert Type Message Possible Cause and Suggested Response

Navigation menu