Transtek Medical Electronics TMB995-BS Blood Pressure Monitor User Manual bf 1256 b 20121229

Guangdong Transtek Medical Electronics Co.,Ltd Blood Pressure Monitor bf 1256 b 20121229

Users Manual

version:1.0User ManualBlood Pressure Monitor TMB-995-BSTo use the monitor correctly and safely, please read the manual thoroughly.Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB-995-BS.Please keep well this manual in order to reference in future.Arm TypeGuangdong Transtek Medical Electronics Co., Ltd.Zone B, No.105 ,Dongli Road, Torch Development District, Zhongshan,528437,Guangdong,China  STARTSTOPFCC ID:OU9TMB995-BS
1Table of ContentsCATALOGUE CATALOGUEINTRODUCTION...................................................................................................................2General DescriptionIndications for UseContraindicationsMeasurement PrincipleSafety InformationLCD Display SignalMonitor ComponentsList BEFORE YOU START...........................................................................................................9The Choice of Power SupplyInstalling and Replacing the BatteriesSetting Date and TimeSelect the User MEASUREMENT.................................................................................................................13Tie the CuffStart the MeasurementDATA MANAGEMENT..........................................................................................................15Recall the RecordsDelete the RecordsINFORMATION FOR USER................................................................................................19Tips for measurementMaintenancesABOUT BLOOD PRESSURE..............................................................................................21What are systolic pressure and diastolic pressure?What is the standard blood pressure classification?Irregular heartbeat detectorWhy does my blood pressure fluctuate throughout the day?Why do I get a different blood pressure at home compared to the hospital?Is the result the same if measuring on the right arm?TROUBLESHOOTING........................................................................................................23SPECIFICATIONS...............................................................................................................24AUTHORIZED COMPONENT ............................................................................................25CONTACT INFORMATION..................................................................................................25COMPLIED STANDARDS LIST...........................................................................................26FCC STATEMENT................................................................................................................27EMC GUIDANCE.................................................................................................................28
2 3INTRODUCTION INTRODUCTIONThank you for selecting TRANSTEK arm type blood pressure Monitor (TMB-995-BS). The monitor features blood pressure measurement, pulse rate measurement and the result storage. The design provides you with two years of reliable service.Readings taken by the TMB-995-BS are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method.This manual contains important safety and care information, and provides step by step instructions for using the product.Read the manual thoroughly before using the product.Features:80*60mm Digital LCD display Maximum 60 records per each userMeasuring during inflation technologyGeneral Description1.The device is not suitable for use on may be pregnant women or pregnant women.2.The device is not suitable for use on patients with implanted,electrical devices,such as cardiac pacemakers, defibrillators.ContraindicationsThis product uses the Oscillometric Measuring Method to detect blood pressure.Before every measurement, the unit establishes a “zero point” equivalent to theatmospheric pressure. Then it starts inflating the cuff. Meanwhile, the unit detectspressure oscillation generated by beat-to-beat pulsatile, which is used to deter- mine the systolic pressure and diastolic pressure as well as pulse rate.  Measurement PrincipleThe signs below might be in the user manual, labeling or other components. They are the requirement of standard and using. Safety InformationIndications for Use    The Transtek Blood Pressure Monitor is digital monitors intended for use in measuring blood pressure and heartbeat rate with arm circumference ranging from 22 cm to 32 cm ( about 8¾˝-12½˝ ) or 22cm to 42cm(about 8¾˝-16½˝).It is intended for adult indoor use only.The Green Dot is the license symbol of a European network of industry-fundedsystems for recycling the packaging materials of consumer goods.Symbol for “Recycle”Symbol for “THE OPERATION GUIDE MUST BE READ”Symbol for “MANUFACTURER”Symbol for “SERIAL NUMBER”Symbol for “TYPE BF APPLIED PARTS”Symbol for “DIRECT CURRENT”Symbol for “ENVIRONMENT PROTECTION - Electrical waste products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local authority or retailer for recycling advice”For indoor use onlyF1T1A/250V Φ3.6*10CCCSymbol for “Class II Equipment”Caution: These notes must be observed to prevent any damage to the device.SNSymbol for “MANUFACTUREDATE”
4 5CAUTION* This device is intended for adult use in homes only. * The device is not suitable for use on neonatal patients, pregnant women,patients with implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy. Please consult your doctor prior to using the unit if you suffer from illnesses.* The device is not suitable for measuring the blood pressure of children. Ask your doctor before using it on older children.* The device is not intended for patient transport outside a healthcare facility.* The device is not intended for public use.* This device is intended for no-invasive measuring and monitoring of arterial blood pressure.It is not intended for use on extremities other than the arm or for functions other than obtaining a blood pressure measurement.* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood pressure.Do not begin or end medical treatment without asking a physician for treatment advice.* If you are taking medication,consult your physician to determine the most appropriate time to measure your blood pressure. Never change a prescribed medication without consulting your physician.* Do not take any therapeutic measures on the basis of a self measurement. Never alter the dose of a medicine prescribed by a doctor. Consult your doctor if you have any question about your blood pressure.* When the device was used to measure patients who have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, the best result may occur with deviation. Please consult your physician about the result.* Don't kink the connection tube during use, otherwise, the cuff pressure may continuously increase which can prevent blood flow and result in harmful injury to the PATIENT.* When using this device, please pay attention to the following situation which may interrupt blood flow and influence blood circulation of the patient, thus cause harmful injury to the patient: connection tubing kinking too frequent and consecutive multiple measurements; the application of the cuff and its pressurization on any arm where intravascular access or therapy, or an arterio-venous (A-V) shunt, is present; inflating the cuff on the side of a mastectomy.* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around simultaneously, because this could cause temporary loss of function of those simultaneously-used monitoring ME equipment.*On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure > 300mmHg or constant pressure > 15mmHg for more than 3 minutes) applied to the arm may lead to an ecchymosis.*Please check that operation of the device does not result in prolonged impairment of patient blood circulation.* When measurement, please avoid compression or restriction of the connection tubing.* The device cannot be used with HF surgical equipment at the same time.CAUTION* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060-2:2013.* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact the manufacturer.* This device is contraindicated for any female who may be suspected of, or is pregnant. Besides providing inaccurate readings, the effects of this device on the fetus are unknown.* Too frequent and consecutive measurements could cause disturbances in blood circulation and injuries.* This unit is not suitable for continuous monitoring during medical emergencies or operations.Otherwise, the patient’s arm and fingers will become anaesthetic, swollen and even purple due to a lack of blood.* When not in use, store the device with the adapter in a dry room and protect it against extreme moisture, heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.* This device may be used only for the purpose described in this booklet. The manufacturer cannot be held liable for damage caused by incorrect application.*This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in this booklet.* The maximum temperature that the applied part can be achieved is 42.8℃ while the environmental temperature is 40℃.* The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide.* Warning: No servicing/maintenance while the ME equipment is in use.* The patient is an intended operator.* The patient can measure,transmit data and charge power under normal circumstances and maintain the device and its accessories according to the user manual.* To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated interference signal or electrical fast transient/burst signal.* The blood pressure monitor, its adaptor, and the cuff are suitable for use within the patient environment. If you are allergic to polyester, nylon or plastic, please don't use this device. * During use, the patient will be in contact with the cuff. The materials of the cuff have been tested and found to comply with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction.* Adaptor is specified as a part of ME EQUIPMENT.* If you experience discomfort during a measurement, such as pain in the arm or other complaints, press any button to release the air immediately from the cuff. Loosen the cuff and remove it from your arm.* If the cuff pressure reaches 40 kPa (300 mmHg), the unit will automatically deflate. Should the cuff not deflate when pressures reaches 40 kPa (300 mmHg), detach the cuff from the arm and press any button to stop inflation.* Before use, make sure the device functions safely and is in proper working condition. Check the device, do not use the device if it is damaged in any way. The continuous use of a damaged unit may cause injury, improper results, or serious danger.* Do not wash the cuff in a washing machine or dishwasher!INTRODUCTION INTRODUCTION
76LCD Display SignalSYMBOL DESCRIPTIONSystolic blood pressure High pressure resultDiastolic blood pressure Low pressure resultEXPLANATIONPulse Pulse/minuteDeflating CUFF air is exhausting of deflatingCurrent Time  Time(year:month:day:hour:minute)Memory If “MEM” shows, the displayed measurement  values is from the memory. mmHg Measurement Unit of the blood pressureBattery Indicator Indicate the current battery Irregular heartbeatIrregular heartbeatGradeHeartbeat Heartbeat dectetion during measurementThe grade of the blood pressureUser 1User 2Start measurement and transmit the results for User 1Start measurement and transmit the results for User 2Data pending to transmit Measurement data stored in the equipmentShocking reminder Shocking will result in inaccurateINTRODUCTIONCAUTION    * The service life of the cuff may vary by the frequency of washing, skin condition, and storage state. The typical service life is 10000 times.* It is recommended that the performance should be checked every 2 years and after maintenance and repair, by retesting at least the requirements in limits of the error of the cuff pressure indication and air leakage (testing at least at 50mmHg and 200mmHg).* Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to the local guidelines.* Manufacturer will make available on request circuit diagrams, component part lists, descriptions, calibration instructions,etc., to assist to service personnel in parts repair.* The plug/adapter plug pins insulates the device from the main supply. Do not position the device in a position where it is difficult to disconnect from the supply mains to safely terminate operation of ME equipment.* The operator shall not touch output of batteries /adapter and the patient simultaneously.* Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth to clean the whole unit before and after use. Don’t use any abrasive or volatile cleaners.* The device doesn’t need to be calibrated within two years of reliable service.* If you have any problems with this device, such as setting up, maintaining or using, please contact the SERVICE PERSONNEL of Transtek. Don’t open or repair the device by yourself in the event of malfunctions. The device must only be serviced, repaired and opened by individuals at authorized sales/service centers.* Please report to Transtek if any unexpected operation or events occur.* Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of small parts. It is dangerous or even fatal.* Be careful to strangulation due to cables and hoses, particularly due to excessive length.* At least 30 min required for ME equipment to warm from the minimum storage temperature between uses until it is ready for intended use. At least 30 min required for ME equipment to cool from the maximum storage temperature between uses until it is ready for intended use.* This equipment needs to be installed and put into service in accordance with the information provided in the ACCOMPANYING DOCUMENTS;* Wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, walkie-talkies can affect this equipment and should be kept at least a distance d away from the equipment. The distance d is calculated by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and Table 9 of IEC 60601-1-2:2014, as appropriate.* Please use ACCESSORIES and detachable partes specified/ authorised by MANUFACTURE. Otherwise, it may cause damage to the unit or danger to the user/patients.* There is no luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood vessel.* Please use the device under the environment which was provided in the user manual. Otherwise, the performance and lifetime of the device will be impacted and reduced.INTRODUCTIONBluetooth icon The bluetooth icon blinks when the bluetooth is working
STARTSTOP98Component list of pressure measuring system:1 Cuff 2 Air pipe3 PCBA4 Pump5 ValveMonitor ComponentsBATTERY COMPARTMENTLCD DISPLAY AIR HOSEAIR CONNECTOR PLUGCUFFList1. Blood Pressure Monitor    (TMB-995-BS)3. User manual4. 4*AA batteries2. Cuff (22~32cm or 22~42cm)    (Type BF applied part)5. AC adaptor (BLJ06L060100P-U)The Choice of Power Supply1.Slide off the battery cover.2.Install the batteries by matching the correct polarity, as shown.3.Replace the cover.1.Battery powered mode:    6VDC 4*AA batteries2.AC adaptor powered mode:    6V     1A(Please use the AC adaptor which authorized by the manufacturer!)   Right picture is the hole in for power adaptor.Installing and Replacing the BatteriesReplace the batteries whenever the below happenThe                     showsThe display dimsThe display does not light upBEFORE YOU STARTINTRODUCTIONIn order to get the best effect and protect your monitor,please use the the right batteries and special power adapter which complies with local safety standard.CAUTIONCAUTIONDo not use new and used batteries together. Do not use different types of batteries together.Do not dispose the batteries in fire. Batteries may explode or leak.Remove batteries if the device is not likely to be used for some time.Worn batteries are harmful to the environment. Do not dispose with daily garbage.Remove the old batteries from the device following your local recycling guidelines.STARTSTOPMEM/UP BUTTONSTART/STOP BUTTONSET/DOWN BUTTON
1110Setting Date and Time   It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (year :2018—2058 time:12 H)1.When the unit is off, hold pressing “SET ” button for 3 seconds to enter the mode for year setting.2.Press the “  MEM    ” button to change the [YEAR].3.When you get the right year, press “ SET  ” button to set down and turn to next step.4.Repeat steps 2 and 3 to set the [MONTH] and [DAY].5.Repeat steps 2 and 3 to set the [HOUR] and [MINUTE].6.After the [UNIT] is set, the LCD will display “dOnE”, and then turn off.BEFORE YOU STARTBEFORE YOU STARTSTARTSTOPSTARTSTOPSTARTSTOPSTARTSTOPSTARTSTOPSTARTSTOP STARTSTOPSTARTSTOP
1312Tie the CuffMEASUREMENTBEFORE YOU START1.When the monitor is off , press the  “SET” buttonto enter user setting mode. Select the User2.Then press “SET”  button again, select the user ID user 1 and user 2 .3. After selecting the suitable user ID, press “START/STOP” button toconfirm. Then the LCD will turn off.2~3cm1.4.Hold your arm with your palm facing up and tie the cuff on your upper arm, then position the tube off-center toward the inner side of arm in line with the little finger. Or position the artery mark      over  the main artery (on the inside of your arm).  Note: Locate the main artery by pressing with 2 fingers approximately 2 cm above the bend of your elbow on the inside of your left arm. Identify where the pulse can be felt the strongest. This is your main artery.The cuff should be snug but not too tight. You should be able to insert one finger between the cuff and your arm.Remove all jewelry, such as watches and bracelets from your left arm.Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right arm.Roll or push up your sleeve to expose the skin. Make sure your sleeve is not too tight.2.3.5. Sit comfortably with your tested arm resting on a flat surface. Place your elbow on a table so that the cuff is at the same level as your heart. Turn your palm upwards. Sit upright in a chair, and take 5-6 deep breaths.Rest for 5 minutes before first measuring.Wait at least 3 minutes between measurements. This allows your blood circulation to recover.The patient must relax as much as possible and do not move and talk during the measurement procedure.For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same arm, or as directed by a physician.6.Take the measurement in a silent room.The cuff should maintain at the same level as the right atrium of the heart.Please sit comfortably. Do not cross your legs and keep your feet flat on the ground.Keep your back against the backrest of the chair.Helpful tips for Patients, especially for Patients with Hypertension:STARTSTOPSTARTSTOP
15141.When the monitor is off,press the START/STOP button to turn on the monitor, and it will finish the whole measurement,save and transmit the measurement data for the desired user. (Take User 1 for example.)Adjust the zero .LCD displayInflating and measuring. Display and save the results. The data transmission will proceed.Start the Measurement4.Press “START/STOP” button to power off, otherwise it will turn off within 1 minute.DATA MANAGEMENTMEASUREMENTTips: A.  You can press “START/STOP” button at any time to stop measuring during the process of measurement.B. Maximum 60 records are both for USER 1 and USER 2.C. If the measurement result is out of the measurement range (SYS: 60mmHg to    230mmHg; or DIA: 40mmHg to 130mmHg;or Pulse: 40-199 pulse/minute), the LCD will display “out”.2.This device will proceedto data transmissionafter measurement.The Bluetooth symbolblinks on the LCDindicates data istransmitting.3.If the data is successfullytransmitted, the Bluetoothsymbol doesn’t blink any more,and the         will disappear.If the data transmission fails,the          will proceed.STARTSTOPSTARTSTOPCAUTION 1. 2.  Interference may occur in the vicinity of equipment marked with the following symbol     . And TMB-995-BS may interfering vicinity electrical equipment.Sensitive people, including pregnant women pre-eclamptic and those who implanted medical electronic instruments, should avoid using the unit whenever possible.Keep the monitor at least 20 centimeters away from the human body (especially the head) when the data transmission is proceeding after measurement.To enable the data transmission function, this product should be paired to Bluetooth end at 2.4 GHz.How to mitigate  possible interference?The range between the device and BT end should be reasonably close, from 1 meter to 10 meters. Please ensure no obstacles between the device and BT end so as to obtain quality connection and to lower the RF output range.To avoid interference, other electronic devices (particularly those with wireless transmission / Transmitter) should be kept at least 1 meter away from the monitor.
1716Recall the Records2. Press the “  MEM  ” or “ SET  ” button again to rotate the records.   The order of the record, date and time will be displayed alternatively.The most recent record (1) is shown first.  Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (60) is dropped from the list.The current is the No. 1.Twenty records in total.The corresponding time is A.M.9:00.CAUTION1.Hold pressing “  MEM  ” button  for 3 seconds when the monitor is under thememory enquiry mode. The LCD will display the blinking “Use ID + dEL ALL”.If you did not get the correct measurement, you can delete all results of the selected user by following steps below .Delete the RecordsDATA MANAGEMENTThe corresponding date is July 30th. 2.Press “ MEM  ” button to confirm deleting all the memories , the LCD willdisplay “Use ID + dEL dOnE” and the monitor will turn off.1.When the monitor is off, please press “  MEM ” button to show theaverage value of the latest three records.If the records are less than 3 groups, it will display the latest record instead.Displays different colors at different levels when the monitor is under the memory enquiry mode.(Take user 1 for example.)DATA MANAGEMENTBluetooth Module No.: Transmitting Distance: 1-10 metersRF Frequency Range: 2.402 - 2.480 GHzMax Output Power:Supply Voltage: LS51802  -0.53dBm1.8V - 3.6V
19185.If you don’t want to delete the records, press the “START/STOP” button, it will turn off and quit the deletion. Or when it is in the deleting memory mode, it will turn off and quit the deletion if there is no operation in one minute.  6. If there is no record, the LCD will display as the following picture:Tips for Measurementwait at least 1 hour after dinner or drinkingWait at least 20 minutes after taking a bathIn a very cold environmentImmediate measurement after tea, coffee, smoking When talking or moving your fingersWhen you want to discharge urineMeasurements may be inaccurate if taken in the following circumstances.INFORMATION FOR USERDATA MANAGEMENT3.Hold pressing “  SET  ” button  for 3 seconds when the monitor is under the memory enquiry mode(excluding the LCD displays AVG). The LCD will display the blinking “ Use ID + dEL ONE”.4.Press “ MEM ” button to confirm deleting the last record, the LCD will display “Use ID + dEL dOnE” and the monitor will turn off.
2120What are systolic pressure and diastolic pressure? pressartery veinblood dischargingSystolicrelaxblood enteringDiastolic  When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.Irregular Heartbeat DetectorCAUTIONThe appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heartbeat was detected during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.ABOUT BLOOD PRESSUREINFORMATION FOR USERAn irregular heartbeat is detected when a heartbeat rhythm varies while the unit is measuringthe systolic and diastolic blood pressure.During each measurement, the monitor records all thepulse intervals and calculate the average ; if there are two or more pulse intervals ,the differencebetween each interval and the average is more than the average value of ±25% , or there are fouror more pulse intervals ,the difference between each interval and the average is more than theaverage value of ±15%,the irregular heartbeat symbol appears on the display when themeasurement results are appear.MaintenanceIn order to get the best performance, please follow the  instructions below.STARTSTOPSTARTSTOPSTARTSTOPSTARTSTOPSTARTSTOPPut in a dry place and avoid the sunshineAvoid intense shakingand collisionsUsing wet cloths to remove dirtAvoid touching water,clean it with a dry cloth in case.Avoid dusty and unstabletemperature environmentAvoid washing the cuffWhat is the standard blood pressure classification?The chart on the right is the standard blood pressure classifi-cation published by AmericanHeart Association (AHA).Please consult a physician if your measuring result falls outside the range. Please note that only a physician can tell whether your blood pressure value has reached a dangerous point.CAUTIONBlood Pressure CategoryNormalElevatedHigh Blood Pressure(Hypertension) Stage 1High Blood Pressure(Hypertension) Stage 2Hypertensive Crisis(Consult your doctor immediately)SystolicmmHg (upper#)DiastolicmmHg (lower#)less than 120120-129130-139140 or higherHigher than 180andororand/orless than 8080-8990 or higherHigher than 120This chart reflects blood pressure categories defined by American Heart Association.and less than 80
Why does my blood pressure fluctuate throughout the day?Is the result the same if measuring on the right arm?Why do I get a different blood pressure at home compared to the hospital?231. Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.2.If the person takes medicine, the pressure will vary more.3.Wait at least 3 minutes for another measurement.The blood pressure is different even throughout the day due to weather, emotion, exercise etc, Also, there is the “white coat” effect, which means blood pressure usually increases in clinical settings.It is ok for both arms, but there will be some different results for different people. We suggest you measure the same arm every time.What you need to pay attention to when you measure your blood pressure at home: If the cuff is tied properly. If the cuff is too tight or too loose.If the cuff is tied on the upper arm.If you feel anxious.Taking 2-3 deep breaths before beginning will be better for measuring.        Advice: Relax yourself for 4-5 minutes until you calm down.    This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the products not operating as you think it should, check here before arranging for servicing.TROUBLESHOOTINGPROBLEM SYMPTOM CHECK THIS REMEDYNo powerLowbatteriesErrormessageDisplaywill not light up.Batteries are exhausted. Replace with new batteriesInsert the batteries correctlyReplace with new batteriesBatteries are insertedincorrectly.Batteries are low.E 1 shows Communication error Check if the APP is on,operate and send the data again.E 3 shows The cuff is not secure. Readjust the cuff and relax for a momentand then measure again.E10 or E11 showsMovement can affect themeasurement.Relax for a moment and then measure again.E20 showsThe measurement process does not detectthe pulse signal.Loosen the clothing on thearm and then measure againEExx,shows on the display.A calibration erroroccurred.Retake the measurement.If the problem persists,contact the retailer or ourcustomer service department for further assistance.Refer to the warranty for contact information and return instructions.E21 shows The treatment of themeasurement failed.Relax for a moment andthen measure again.AC adaptor is insertedincorrectly.Insert the AC adaptor tightlyDisplay is dim orshowsThe monitor detectedmotion,talking or thepluse is too poorwhile measuring.ABOUT BLOOD PRESSURE22 Warningmessage“out ” shows Out of measurementrangeRelax for a moment.Refasten the cuff and thenmeasure again. If theproblem persists, contactyour physician.
Authorized Component1.please use the TRANSTEK authorized adapter.Contact InformationFor more information about our products, please visit www.transtek.cn.you can get customer service, usual problems and customer download, transtek will serve you anytime.24 25Manufactured by:Company:Address:Guangdong Transtek Medical Electronics Co., Ltd.Zone B, No.105 ,Dongli Road, Torch Development District, Zhongshan,528437,Guangdong,China  Guangdong Transtek Medical Electronics Co., Ltd.AUTHORIZED COMPONENTDigital  LCD  V.A.60mm*80mmApprox.300g(Excluding the batteries)Approx.102mm*143mm*73mmSPECIFICATIONS6VDC 4×AA batteriesUnknow4×AA batteries,user manual,AC adapter  6V      1ABattery powered mode: AC adaptor powered mode:(Please only use the recommended  AC adaptor model). About 22cm~32cm or 22cm~42cmType BF applied partWARNING: No modification of this equipment is allowed.Power supplyDisplay modeMeasurement mode Oscillographic testing modeMeasurement rangeMeasurement perimeterof the upper armWeightExternal dimensionsAttachmentMode of operation Continuous operationDegree of protectionProtection against ingress of waterAccuracyNormal working conditionStorage & transportationconditionSoftware VersionPressure:5℃-40℃within±3mmHg(0.4kPa) Pulse value:±5%Rated cuff pressure: 0mmHg~299mmHg(0kPa ~ 39.9kPa)Measurement pressure: SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa) DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)Pulse value: (40-199)beat/minuteIP21 It means the device could protected against solid foreign objects of 12.5mm and greater, and protect against vertically falling water drops.Device ClassificationBattery Powered Mode: Internally Powered ME EquipmentAC Adaptor Powered Mode: Class II ME EquipmentA temperature range of :+5°C to +40°C A relative humidity range of 15% to 90%, non-condensing, but not requiring a water vapour partial pressure greater than 50 hPaAn atmospheric pressure range of : 700 hPa to 1060 hPaTemperature:-20°C to +60°CA relative humidity range of ≤ 93%, non-condensing, at a water vapour pressure up to 50hPaBLJ06L060100P-U100-240V 50-60Hz,0.2Amax6V      1000mAAdaptorType:Input:Output:
26 27Complied Standards ListEN ISO 14971:2012 / ISO 14971:2007 Medical devices - Application of risk management to medical devicesEN 1041:2008 Information supplied by the manufacturer of medical devicesEN 60601-1:2006/ IEC 60601-1:2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performanceEN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environmentEN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and testsEN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement typeEN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systemsEN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometersEN 60601-1-6:2010/IEC 60601-1-6:2010+A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: UsabilityEN 62366-1:2015/ IEC 62366-1:2015 Medical devices - Part 1: Application of usability engineering to medical devicesEN 62304:2006/AC: 2008 / IEC 62304:2006   Medical device software - Software life-cycle processesRisk managementLabelingUser manualGeneral Requirements for SafetyElectromagneticcompatibilityPerformancerequirementsClinical investigationUsabilitySoftware life-cycle processesBio-compatibilityISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management processISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationEN ISO 15223-1:2016 / ISO 15223-1:2016  Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. Part 1 : General requirementsIEC 80601-2-30:2013 Medical electrical equipment- Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometersISO 81060-2:2013  Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement typeCOMPLIED STANDARDS LISTThis device complies with Part 15 of the FCC Rules. Operation is subjectto the following two conditions: (1) this device may not cause harmfulinterference, and (2) this device must accept any interference received,including interference that may cause undesired operation.Caution: The user is cautioned that changes or modifications notexpressly approved by the party responsible for compliance could voidthe user's authority to operate the equipment.NOTE: This equipment has been tested and found to comply with thelimits for a Class B digital device, pursuant to Part 15 of the FCC Rules.These limits are designed to provide reasonable protection againstharmful interference in a residential installation. This equipment generates,uses and can radiate radio frequency energy and, if not installed and used inaccordance with the instructions, may cause harmful interference toradio communications. However, there is no guarantee that interference willnot occur in a particular installation.If this equipment does cause harmful interference to radio or televisionreception, which can be determined by turning the equipment off and on,the user is encouraged to try to correct the interference by one or moreof the following measures:-- Reorient or relocate the receiving antenna.-- Increase the separation between the equipment and receiver.-- Connect the equipment into an outlet on a circuit different from that towhich the receiver is connected.-- Consult the dealer or an experienced radio/TV technician for help.FCC Radiation Exposure Statement:This equipment complies with FCC radiation exposure limits set forth foran uncontrolled environment.This transmitter must not be co-located or operating in conjunction withany other antenna or transmitter.FCC StatementFCC STATEMENTFCC ID:OU9TMB995-BS
EMC Guidance1)This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided,and this unit can be affected by portable and mobile RF communications equipment.2)* Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.3)Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!4)* Caution: This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.Table 1 Guidance and manufacturer’s declaration – electromagnetic emissionsRF emissions CISPR 11 Group 1 Class B Class ACompliesComplianceHarmonic emissions IEC 61000-3-2Voltage fluctuations/ flicker emissions  IEC 61000-3-3  RF emissions CISPR 11Emissions test Electromagnetic environment - guidanceThe device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.The device is suitable for use in all establishments,other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.EMC GUIDANCE28 29 Guidance and manufacturer’s declaration – electromagnetic immunity Immunity test ±8 kV contact±15 kV air ±8 kV contact±15 kV air  ±2 kV power supply lines: line(s) to line(s): ±1 kV  line(s) to earth: ±2 kV 0%UT; 0.5 cycleAt 0°, 45°, 90°, 135°,180°,225°,270° and 315°0%UT ;  1 cycle and 70%UT ; 25/30 cyclesSingle phase: at 0°0% UT ; 300 cycle30 A/m50Hz/60HzNOTE      UT is the a.c. mains voltage prior to application of the test level. Table 2 Compliance levelElectrostatic discharge (ESD) IEC 61000-4-2 Electromagnetic environment - guidance30 A/m50Hz/60HzPower frequency (50Hz/60Hz) magnetic field IEC 61000-4-8 Voltage dips, short interruptionsand voltagevariations on power supply input linesIEC 61000-4-11 Electrical fast transient/burst IEC 61000-4-4 SurgeIEC61000-4-5input/output lines:±1 kVThe device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.IEC 60601 test levelFloors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.±2 kV power supply lines: 100 kHz repetitionfrequencyline(s) to line(s): ±1 kV  100 kHz repetitionfrequencyMains power quality should be that of a typical commercial or hospital environment.Mains power quality should be that of a typical commercial or hospital environment.Mains power quality should be that of a typical commercial or hospital environment.Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.0% UT ; 0.5 cycleAt 0°, 45°, 90°, 135°,180°,225°,270° and 315°0% UT ;  1 cycle and 70% UT ; 25/30 cyclesSingle phase: at 0°0% UT ;300 cycleEMC GUIDANCE
Table 3Guidance and manufacturer’s declaration – electromagnetic immunityImmunity test Compliance levelIEC 60601 Test levelConducted RFIEC 61000-4-6 Radiated RF IEC 61000-4-310V/m, 80% Amat 1kHz150 kHz to80 MHz: 3 Vrms6Vrms (in ISM and amateur radio bands)80% Am at 1kHz Electromagnetic environment - guidanceNOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.aField strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.bThe device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.150 kHz to80 MHz: 3 Vrms6Vrms (in ISM and amateur radio bands)80% Am at 1kHz Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter.Recommended separation distances:d=0.35        ;d=1.210V/m, 80% Amat 1kHz80 MHz to 800 MHz:d=1.2800 MHz to 2.7 GHz:d=2.3where, P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer, d is the recommended separation distance in meters (m).  Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.Interference may occur in the vicinity of equipment marked with the followingsymbol:PPPPEMC GUIDANCE30 31 150 kHz to 80 MHz  80 MHz to 800 MHz  800 MHz to 2.7 GHz Table 4Recommended separation distances between portable and mobile RF communications equipment and the device.Rated maximum outputpower of transmitter(W) Separation distance according to frequency of transmitter (m)0.010.11101000.12 0.120.381.23.8120.230.732.37.323=d =d 1.2 2.3=d0.371.23.8123.5The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmittters) and the device as recommended below, according to the maximum output power of the communications equipment.For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.NOTE 1    At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.PPPEMC GUIDANCE
 Table 5Guidance and manufacturer’s declaration - electromagnetic immunityTest Frequency (MHz)385 0.3 27The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device, should assure that it is used in such an environment.NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.Radiated RF IEC61000-4-3 (Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment) Band a)(MHz)Service a) Modulation b) Modulation b)(W)Distance (m)IMMUNITY TEST LEVEL (V/m)380-390 TETRA 400Pulse modulation b) 18Hz1.8450 380-390GMRS 460,FRS 460FM c) ± 5kHzdeviation 1kHzsine20.3 28710 704-787745780LTE Band13,17Pulse modulation b)217Hz0.2 0.3 9810870930800-960GSM 800/900,TETRA 800,iDEN 820,CDMA 850,LTE Band 5Pulse modulation b)18Hz20.3 281720184519701700-1990GSM 1800;CDMA 1900;GSM 1900;DECT;LTE Band 1,3,4,25; UMTSPulse modulation b)217Hz20.3 2820.3 282450 2400-2570Bluetooth,WLAN,802.11 b/g/n, RFID2450, LTEBand 7Pulse modulation b)217 Hz5240524057855100-5800WLAN802.11a/nPulsemodulation b)217 Hz0.2 0.3 9a) For some services, only the uplink frequencies are included.b) The carrier shall be modulated using a 50% duty cycle square wave signal.c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation:                                                                E=Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in V/m.EMC GUIDANCE32