Spacelabs Healthcare 010-1471ABG 010-1471-00 User Manual 061 2050 00 Rev D

Spacelabs Healthcare, Inc. 010-1471-00 061 2050 00 Rev D

User Manual

www.spacelabshealthcare.com 061-2050-00 Rev. D 9/09SummaryThe Ultraview SL™ 2200 is a compact monitoring system with a 10.4-inch resistive touchscreen display. Dual-battery slots allow this monitor to be used in transport, as well as at the bedside. The SL2200 includes one single-high module slot that is compatible with all Spacelabs modules. When used with the 91493 Integrated Module Housing (IMH), two additional parameter modules are supported.FeaturesUser Interface  All controls are on-screen touch keys, with the exception of the power (ON/OFF) key; optional controls include mouse, keyboard, and remote controlDisplayColor thin film transistor (TFT) resistive touchscreen liquid crystal display (LCD)Resolution640 × 480 pixelsSize21.11 cm (8.31 inches) wide15.85 cm (6.24 inches) highTrace Height4.2 cm (1.65 inches)6 cm (2.36 inches)Sweep SpeedA variety of speeds are available under module controlParameter Capacity  Provides three times the number of waveforms purchased, up to a maximum of 12 parametersAlarm Limit Review  Provides a snapshot view of bedside alarm limits for all active parameters; this feature is only available with specific Ultraview SL modulesDepth: 16.51 cm (6.5 inches)Weight: 5.5 kg (12.1 pounds) with two batteries and recorder27.05 cm(10.65 inches)31.24 cm(12.3 inches)26.42 cm10.4 in91367 SL2200 Compact Monitor
www.spacelabshealthcare.com 061-2050-00 Rev. D 9/0991367 SL2200 Compact MonitorTrends  Graphic  1-, 2-, 6-,12-, or 24-hour segments; data is stored in 1-minute resolutionTabular  Time increments of 1, 5, 10, 15, or 30 minutes; 1, 1.5, or 3 hoursPower  An external direct current (DC) power supply provides power to the monitors and up to three Flexport® system interfacesAdvanced Power Management  Maximizes battery performance during transport; includes battery “fuel-gauge”Connectors  Three standard USB 1.1 connectors (keyboard, mouse or barcode scanner), synchronous data link control (SDLC for Flexport systems or module housings), RJ45 (10/100BaseT Ethernet), remote alarm light, VGA output (video), RS-232 serial connectorNetwork Communication  10/100 BaseT modular connector (RJ45) provided; wireless is optional; will send data, including all available waveforms and numerics, to the Ultraview SL networkBarcode Scanner Support  Provides support for using third-party USB-powered barcode scanners; for entering patient demographic data into a patient’s Admit Record (limited to 128 ASCII characters); this feature is only available in the bedside mode. USB-powered bar code scanners must be purchased from the manufacturerOptions03  Three display zones04  Four display zonesD  Perioperative; provides customizable, user-specific display setups and start case/end case functionsNote  The SL2200 does not support the 91388 Remote Display Controller.N  Vital Signs Calculations (Hemodynamics, Oxygenation, Ventilation, and Renal)O  Drug Dose CalculationsQ  Data Shuttle® (for use with selected modules only) supports transfer for up to 24 hours of the database from the monitor, including continuous and episodic events, and trend information for all parameters monitored by modules and Flexport system interfacesR  Patient Data Logger; presents an ASCII data stream of patient name and vital sign data to the serial port in a pre-defined formatU  Dual Channel Internal RecorderPrinting Method  Thermal array print headResolution  Eight dots per mm (vertical) and 32 dots per mm (horizontal) at 25 mm per second sweep speedPaper  Heat-sensitive roll of paper, 50 mm wide × 30 m longPrints  Manual and automatic alarm recordings for waveforms and vital sign data, trends, calculations; full annotations are includedFrequency Response  Determined by the parameter recordedPaper Speed  1.56, 3.12, 6.25, 12.5, 25, and 50 mm per second (depending on the monitor sweep speed selected)Alarm Record  Records any parameters in an alarm state when alarm recording is activeAuto Run  20 seconds or duration of alarm violation (whichever is longer)Controls  CONVERT TO CONTINUOUS, STOP RECORDING, SLOW ON/OFF, PRINTER ON/OFFIndicators  Paper out, unit offRecord  Enables selection of up to two active monitor channels plus trendsX  Wireless Ethernet Communication; provides 802.11a/b/g wireless ethernet with WEP, WPA, and WPA2 security; requires site qualification survey to ensure adequate performance
www.spacelabshealthcare.com 061-2050-00 Rev. D 9/0991367 SL2200 Compact MonitorZ  Wireless Ethernet Communication; provides 802.11b  wireless ethernet with WEP security; requires site qualification survey to ensure adequate performanceProduct SpecificationsClassificationMDD  Class IIbEN 60601-1Class 1  Requires outlet with safety ground (Protective Earth) conductor. Rated for continuous operation.CISPR11, Group 1, Class B  Suitable for use in domestic establishments connected to a low-voltage supply networkElectrical SpecificationsMains Power  100 to 240 VAC, 50 to 60 Hz, 2 ABatteries  One or two 12-volt (2.45 Ah NiMH) batteries may be used, providing 2.5 to 4 hours of operation with two batteries (TFT display), depending on monitor configuration and setup, for up to 600 charge/discharge cycles; operating time is dependent on configuration and usage; 2 hours are required to charge batteries to 100% of capacity with AC connected and the monitor powered OFF; 3-hour charge time required with monitor powered ON.Isolation  Chassis leakage current meets AAMI, UL 60601-1, CSA Std. C22.2 No. 601.1, and IEC 60601-1 standardsPhysical DimensionsHeight27.05 cm (10.65 inches)Depth16.51 cm (6.5 inches)Width31.24 cm (12.3 inches)Weight5.5 kg (12.1 pounds) withtwo batteries and recorderEnvironmental RequirementsStorageTemperature-25° to 60° C (-13° to 140° F) Humidity95% (non-condensing)Altitude0 to 12,192 meters(0 to 40,000 feet)OperatingTemperature0° to 50° C (32° to 122° F)Humidity95% (non-condensing)Altitude0 to 4,572 meters(0 to 15,000 feet)Electromagnetic CompatibilityEN 60601-1-2: 2001Emissions  (CISPR 11) EN 55011, Class BEN 61000-3-2HarmonicsEN 61000-3-3FlickerImmunity IEC 61000-4-2ESD, 8 kV contact/15 kV airIEC 61000-4-3RF fields, 20 volts/meter, 80 MHz to 2.5 GHzIEC 61000-4-4Burst, 1 kV data and I/O ports/2 kV power portsIEC 61000-4-5Surge, 1 kV differential/2 kV common modeIEC 61000-4-6Conducted RF, 3 Vrms 150 kHz to 80 MHz
www.spacelabshealthcare.com 061-2050-00 Rev. D 9/0991367 SL2200 Compact MonitorIEC 61000-4-8Magnetic field (50/60 Hz), 60 Amps/meterIEC 61000-4-11Power quality, voltage, and frequency variationsAccessoriesP/N 010-1114-01Dual external NiMH/SLA battery charger, 100 to 240 VP/N 146-0055-00Rechargeable NiMH battery90360-01Remote keypadP/N 016-0369-00Bedrail hook/quick release mounting plateP/N 010-1622-00Mouse, USBP/N 010-1621-00Keyboard (U.S. English only; other keyboards are available for use outside the U.S.)P/N 119-0480-00Power supply*Height7.14 cm (2.81 inches)Length14.4 cm (5.67 inches)Depth4.19 cm (1.65 inches)Weight0.455 kg (1.0 pounds)* This accessory is included.A variety of GCX-brand mounting and mobility solutions are available from Spacelabs or GCX directly.DocumentationUltraview® Operations Documents CD-ROM (P/N 084-1101-xx)Spacelabs Healthcare Service Documents CD-ROM (P/N 084-0700-xx)Spacelabs Healthcare Supplies and Accessories Catalog CD-ROM (P/N 084-1201-xx)Regulatory Approvals® C  USCSA certified: Meets IEC 60601-1, CSA C22.2 No. 601.1, and UL 60601-1 for electrical safety. IECEE CB Scheme.0123Meets EN 60601-1. CE marked in accordance with the Medical Device Directive 93/42/EEC.
Ultraview, Ultraview SL, Ultraview Care Network, Flexport, and Data Shuttle are trademarks of Spacelabs Healthcare.Brands and product names are trademarks of their respective owners.www.spacelabshealthcare.com061-2050-00 Rev. D 9/0991367 SL2200 Compact Monitor0981 !FCCID: CM6010-1471ABGIC: 2434A-01047100Option X, private wireless 802.11a/b/g  0168 !FCC ID: H9PLA4137IC: 1549104431AOption Z, private wireless 802.11b  The embedded wireless options 802.11a/b/g (X) and 802.11b (Z) of this device comply with part 15 of the FCC Rules, with RSS-210 of Industry Canada, and are CE marked in accordance with the RTTE Directive 1999/5/EC. Operation of wireless options X and Z are subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.The radio-transceiver may only be used for Wireless Local Area Network (WLAN) operation within a medical facility. It is not intended for home or vehicle use. Changes or modifications not expressly approved by Spacelabs Healthcare will void the user’s authorization to operate this equipment.To comply with the FCC’s RF safety Specific Absorption Rate (SAR) requirements, the user must ensure that the monitor, which contains the radiating element of the antenna, is located at least 20 cm (8 inches) away from a person’s head or body.