Sorrel Medical SORRELWIP Wearable Infusion Pump User Manual

Sorrel Medical Ltd. Wearable Infusion Pump Users Manual

Users Manual

 Page 1 of 38           USER MANUAL WEARABLE INFUSION PUMP      For use with Wearable Infusion Pump Software Rev1.2.0 15126-048-0001 Rev.00|Rev.1.2.0/01.2019
 Page 2 of 38   Table of Contents   1.  INTRODUCTION ........................................................................................................................ 5 1.1.  Product Intended Use and Indications for Use .................................................................... 5 1.2.  Terms and Abbreviations ..................................................................................................... 5 1.3.  Document Conventions ........................................................................................................ 6 1.4.  Safety and Compliance Information ..................................................................................... 6 1.4.1  Symbols and Labeling .............................................................................................................................6 1.4.2  Compliance and Classification ................................................................................................................9 1.4.3  FCC and IC Information for EMC and Radio Frequency ....................................................................... 10 1.4.4  Biocompatibility ................................................................................................................................... 12 1.4.5  Degree of Protection against Ingress of Water and Dust .................................................................... 12 1.5.  Warnings and Safety Precautions ......................................................................................12 1.5.1  General Warnings Precautions ............................................................................................................ 12 1.5.1.1  Waste Disposal ................................................................................................................................ 13 1.5.1.2  Explosion Hazard ............................................................................................................................. 14 1.5.1.3  Electromagnetic Compatibility ........................................................................................................ 14 2.  WEARABLE INFUSION PUMP OVERVIEW ...............................................................................16 2.1.  Top View .............................................................................................................................16 2.2.  Side View ............................................................................................................................16 2.3.  Pre-filled Medication Cartridge ..........................................................................................17 2.4.  Single Unit Pack ..................................................................................................................17 3.  INSTRUCTIONS FOR USE .........................................................................................................18 The patient is an intended operator of the Wearable Infusion Pump. ..........................................18 3.1.  Step 1 - Prepare ..................................................................................................................18 3.2.  Step 2 - Inject ......................................................................................................................21 3.3.  Step 3 - Finish .....................................................................................................................24 4  ALARMS ..................................................................................................................................28 4.1  Alarm Information ..............................................................................................................28
 Page 3 of 38 5  MAINTENANCE AND STORAGE ...............................................................................................29 5.1  Battery Classification ..........................................................................................................29 5.2  Transport and Storage ........................................................................................................29 6  TECHNICAL SPECIFICATIONS ...................................................................................................30 6.1  Pump Accuracy ...................................................................................................................30 6.1.1  Start-up and Trumpet Graphs ........................................................................................30 6.2  Pump Specifications ...........................................................................................................32 6.3  Pump Ranges ......................................................................................................................34 6.4  Electromagnetic Compatibility Statement .........................................................................34 6.4.1  Electromagnetic Emission ..............................................................................................34 6.4.2  Electromagnetic Immunity .............................................................................................34 7  COMPLIANCE INFORMATION STATEMENTS ..........................................................................36 7.1  FCC Declaration of Conformity ...........................................................................................36 7.2  EU Declaration of Conformity ............................................................................................37
 Page 4 of 38 Important Notice The  Wearable Infusion Pump User Manual is  delivered subject to  the conditions  and restrictions  listed in this section. Clinicians and users should read the entire User Manual prior to operating the Wearable Infusion Pump in order to fully understand the functionality and operating procedures of the pump. Prescription Notice Federal United States law restricts this device for sale by or on the order of a physician only {21 CFR 801.109(b) (1)}. The  Wearable  Infusion  Pump  is  for  use  at  the  direction  of,  or  under  the  supervision  of,  licensed  physicians and/or licensed healthcare professionals who are trained in the use of the pump and in the administration of subcutaneous  infusions.  The  instructions  for  use  presented  in  this  manual  should  in  no  way  supersede established medical protocol concerning patient care. Copyright, Trademark and Patent Information © 2018, Sorrel Medical Ltd. All rights reserved. Sorrel (with or without logos) is a trademark of Sorrel Medical Ltd. The design, pumping mechanism and other features of the Sorrel Wearable Infusion Pump are protected under one or more US and Foreign Patents. Disclaimer The information in this manual has been carefully examined and is believed to be reliable. No responsibility is assumed for any inadvertent inaccuracies. Sorrel Medical Ltd. reserves the right to make changes to any of its products  in  order  to  improve  reliability,  design  and  performance.  The  instructions  presented  in  this  manual should in no way supersede established medical protocol concerning patient care. The text and drawings herein are for the purposes of illustration and reference only; the specifications on which they are based are subject to change without notice. Warning Sorrel  Medical  Ltd.  will  assume  no  responsibility  for  incidents  which  may  occur  if  the  product  is  not  used  in accordance with product labeling. Refer to Warnings and Safety Precautions on page 12 for a complete list of warnings and cautions. Technical Assistance For  technical  questions,  troubleshooting  assistance  and  reporting  of  unexpected  events,  please  contact  your healthcare  provider.  You  may  also  contact  Sorrel  Medical  Ltd.  support  via  email  to  the  following  address: support@sorrelmedical.com
 Page 5 of 38 1. INTRODUCTION 1.1. Product Intended Use and Indications for Use The Wearable Infusion Pump is intended for use in subcutaneous infusion of prescribed liquid medication of up to 6 hours. The pump is medication agnostic (no specific medication) and designed for pediatric and adults subcutaneous treatment. 1.2. Terms and Abbreviations Term/Abbreviation Meaning Continuous Infusion Infusion is continuously administered in one programmable rate EMC Electromagnetic Compatibility EMI Electromagnetic Interference FCC Federal Communications Commission mL Milliliters MRI Magnetic Resonance Imaging RH Relative Humidity RF Radio Frequency ROW Rest of World (i.e. non-US) Volume To Be Infused The amount of fluid programmed or remaining to be infused. The value range is 0.1 to 3 mL
 Page 6 of 38 1.3. Document Conventions The following messages in this manual prompt reader to pay special attention to specific points:  Warnings alert the user about situations to be avoided, the result of which could lead to death or serious injury. They may also describe potential serious adverse reactions and safety hazards.  Cautions caution the user of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or damage to the equipment or other property. It may also be used to alert against unsafe practices.   Notes provide additional information to help obtain optimal equipment performance.  1.4. Safety and Compliance Information The following section presents important labeling, safety and compliance information: 1.4.1 Symbols and Labeling The following table describes the symbols that appear on the Wearable Infusion Pump's labels, and identifies their locations on the labels: Symbol Description Location  CE certification mark Single unit pack label (ROW) & shipping package (ROW)  BluetoothTM logo Single unit pack label &  shipping package  Do not use if package is damaged Single unit pack label
 Page 7 of 38  Waste Electrical and Electronic Equipment (WEEE) Disposal This symbol indicates that used batteries and electronic equipment must not be disposed of as unsorted municipal waste, and must be collected separately. Contact an authorized representative for information concerning the decommissioning of your equipment. Single unit pack label  Non-DEHP tubing Single unit pack label & shipping package  Latex free Single unit pack label & shipping package  Do not re-use Single unit pack label  Non-pyrogenic Single unit pack label
 Page 8 of 38  Refer to instruction manual/ booklet Single unit pack label  Keep away from sunlight Single unit pack label & shipping package  Keep dry Shipping package  Storage temperature range Single unit pack label & shipping package  Storage humidity range Single unit pack label & shipping package  Storage atmospheric pressure range Single unit pack label & shipping package  Degree of Protection against Ingress of Water and Dust Single unit pack label   Defibrillation proof  type CF applied part Single unit pack label  Rx Only US federal law restricts this device to prescription only Single unit pack label (US/Canada)  Name and address of manufacturer Single unit pack label & shipping package
 Page 9 of 38  Sterilized using ethylene oxide Single unit pack label & shipping package  Serial number Single unit pack label & pump side label & shipping package  Catalogue Number Single unit pack label & shipping package  Authorized representative in the European Community Single unit pack label (ROW)   Sorrel Medical company logo Single unit pack label & pump top label & shipping package  Use by date: YYYY-MM-DD Single unit pack label & pump side label  Instructions for use regarding RF data (includes a table of relevant countries with English language) Single unit pack label (ROW)   1.4.2 Compliance and Classification This manual has been written in conjunction with the requirements of the following International Standards:  IEC 60601-1 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance  IEC  60601-1-2  -  Medical  electrical  equipment  -  Part  1-2:  General  requirements  for  basic  safety  and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests  IEC 60601-2-24 - Medical Electrical Equipment - Part 2-24: Particular Requirements for Safety of Infusion Pumps  and  Controllers.  Data  presented  in  the  Technical  Specification  section  reflect  specific  test
Page 10 of 38 conditions defined in this standard. Other external factors, such as varying back pressure, temperature, head height, set usage, fluid restrictions, solution viscosity, or combinations of these factors may result in deviations from the performance data presented IEC  60601-1-8  -  Medical  electrical  equipment  --  Part  1-8:  General  requirements  for  basic  safety  andessential performance - Collateral standard: General requirements, tests and guidance for alarm systemsin medical electrical equipment and medical electrical systemsIEC 60601-1-11 - Medical electrical equipment -- Part 1-11: General  requirements for basic safety andessential  performance  –  Collateral  Standard:  Requirements  for  medical  electrical  equipment  andmedical electrical systems used in the home healthcare environmentISO 11608-1 - Needle-based injection systems for medical use — Requirements and test methods — Part1: Needle-based injection systemsFCC:  The  Federal  Communications  Commission  (FCC)  regulates  interstate  and  internationalcommunications  by  radio,  television,  wire,  satellite  and  cable.  The  Wearable  Infusion  Pump  complieswith Part 15 of the FCC Rules.Classification according to IEC 60601-1:oInternally poweredoType CFoContinuous operationoNot suitable for use in the presence of flammable aesthetic mixture with air or with oxygen ornitrous oxideClassification according to IEC 60601-2-24:oType 1 Ambulatory pump1.4.3 FCC and IC Information for EMC and Radio Frequency FCC ID: 2AR6L-SORRELWIP IC: 24632-SORRELWIP CAN ICES-3 (B) /NMB-3 (B) HVIN 04 FVIN 1.2.0 Sorrel  Medical  Ltd.  has  not  approved  any  changes  or  modifications  to  this  device  by  the  user.  Any changes  or  modifications  not  expressly  approved  by  Sorrel  Medical  Ltd.  could  void  the  user’s  authority  to operate the equipment.
Page 11 of 38 Sorrel Medical Ltd. n’approuve aucune modification apportée à l’appareil par l’utilisateur, quelle qu’en soit  la  nature.  Tout  changement  ou  modification  peuvent  annuler  le  droit  d’utilisation  de  l’appareil  par l’utilisateur. Class B digital device warnings The FCC Wants You to Know This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential  installation.  This  equipment  generates,  uses  and  can  radiate  radio  frequency  energy  and,  if  not installed  and  used  in  accordance  with  the  instructions,  may  cause  harmful  interference  to  radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: a) Reorient or relocate the receiving antenna.b) Increase the separation between the equipment and receiver.c) Connect the equipment to an outlet on a circuit different from that to which the     receiveris connected. d) Consult the dealer or an experienced radio/TV technician.CAN ICES-3 (B) / NMB-3 (B)  This Class B digital apparatus complies with Canadian ICES-003.  Cet appareil numérique de classe B est conforme à la norme canadienne ICES-003. Interference statement This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation,Science and Economic Development Canada’s licence-exempt RSS(s). Operation is subject to the following two conditions:(1) This device may not cause interference.(2) This device must accept any interference, including interference that may cause undesired operation of the device.
Page 12 of 38 Wireless notice This device complies with FCC/ISED radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines and RSS‐102 of the ISED radio frequency (RF) Exposure rules. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. Cet appareil contient des émetteurs / récepteurs exemptés de licence conformes aux RSS (RSS)d'Innovation, Sciences et Développement économique Canada. Le fonctionnement est soumis auxdeux conditions suivantes :(1) Cet appareil ne doit pas causer d'interférences.(2) Cet appareil doit accepter toutes les interférences, y compris celles susceptibles deprovoquer un fonctionnement indésirable de l'appareil.1.4.4 Biocompatibility All materials in the pump casing and in the fluid path have been tested for biocompatibility, and are in compliance with applicable international standard ISO 10993-1 for biocompatibility. 1.4.5 Degree of Protection against Ingress of Water and Dust The Sorrel Wearable Infusion Pump meets the IP65 splash/dust requirements according to IEC 60529. 1.5. Warnings and Safety Precautions All warnings and safety precautions should be read carefully before operating the Wearable Infusion Pump. WARNING: No modification of this equipment is allowed. Safety information specific to particular pump functions appear in the relevant sections of this manual. 1.5.1 General Warnings Precautions To ensure safety and proper operation, read the User Manual before operating this device. In addition, adhere to the following safety guidelines:
Page 13 of 38 It is important that you do not try to give yourself the injection unless you have received training fromyour healthcare provider.Do not shake the Wearable Infusion Pump.Do not use the Wearable Infusion Pump and pre-filled cartridge if either has been dropped onto a hardsurface. Part of the Wearable Infusion Pump and pre-filled cartridge may be broken even if you cannotsee the break. Use a new Wearable Infusion Pump and pre-filled cartridge.Do not reuse the Wearable Infusion Pump and pre-filled cartridge. The Wearable Infusion Pump andpre-filled cartridge are for single use only.The Wearable Infusion Pump is only to be used with a 3 mL pre-filled cartridge.Do not use the Wearable Infusion Pump if its single pack unit is damaged.Do not use the Wearable Infusion Pump after the expiration date that appears on the single pack.To avoid damage to the pump, keep the Wearable Infusion Pump away from unattended children andpets.1.5.1.1 Waste Disposal Waste Disposal Safety Precautions Keep used pumps and packaging and tubing out of the reach of children.Pumps should be disposed of in a proper manner, considering the nature of residual fluid that maybe contained within, in accordance with disposal practices.Do not dispose of the pump in or near fire.The Wearable Infusion Pump contains a needle.When disposing of the device, use a FDA-cleared sharps disposal container.If you do not have a FDA-cleared sharps disposal container, you may use a household container thatis:omade of a heavy-duty plastic,ocan be closed with a tight-fitting, puncture-resistant lid, without sharps being able to comeout,oupright and stable during use,oleak-resistant, andoproperly labeled to warn of hazardous waste inside the container.When your sharps disposal container is almost full, you will need to follow your communityguidelines for the right way to dispose of your sharps disposal container. There may be state or locallaws about how you should throw away used needles and syringes. For more information about safe
Page 14 of 38 sharps disposal, and for specific information about sharps disposal in the state that you live in, go to the FDA’s website at: http://www.fda.gov/safesharpsdisposal . Do not dispose of your used sharps disposal container in your household trash unless yourcommunity guidelines permit this. Do not recycle your used sharps disposal container.Do not recycle the Wearable Infusion Pump without specific guidance from the device manufacturerImportant:  Always keep the sharps disposal container out of the reach of children.1.5.1.2 Explosion Hazard The equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or nitrous oxide. 1.5.1.3 Electromagnetic Compatibility The  Wearable  Infusion  Pump  is  designed  to  conform  to  the  electromagnetic  compatibility  (EMC) standard  IEC  60601-1-2  and  to  operate  accurately  in  conjunction  with  other  medical  equipment  which  also meets the requirements of this standard. To avoid electromagnetic interference that may affect the operation of the pump, do not use the pump near sources of strong electric and magnetic interference (EMI), such as MRI, CT, diathermy, electromagnetic security systems (e.g., metal detectors), and large electric motors. Portable  and  mobile  RF  communication  equipment,  such  as  RF  emitters,  cellular  telephones,  2-way  radios, BluetoothTM devices, and microwave ovens in close proximity to this device may affect wireless communications with the Infusion pump and/or the operation of the Infusion pump. This device has been tested for compliance with FCC RF exposure limits in a portable configuration. This device must not be used with any other antenna or transmitter that has not been approved to operate in conjunction with this device. Special precautions need to be exercised regarding EMC. These include: Maintaining a minimum separation distance of 2 1⁄2 ft. (3⁄4 m) between the Infusion pump system andportable/mobile RF communications equipment.Managing  the  electromagnetic  environment  to  permit  the  device  to  perform  as  intended  withoutdisturbing other equipment.Separating  the  device  from  all  other  electronic  equipment.  If  the  device  must  be  used  near  otherelectrical equipment, monitor the equipment to ensure there is no electromagnetic interference.Devices should  not be  used adjacent to  or  stacked with other equipment.  If the device must  be usedadjacent to or stacked with other equipment, monitor the device to verify normal operation.
 Page 15 of 38  Some of  these EMI sources (mostly RF emitters) may  not be visible and the  device  can  potentially  be exposed to fields from these EMI sources without the user’s awareness.  If you identify or suspect that external RF sources or other equipment are influencing device operation (from  known  or  unknown  sources),  try  to  (as  applicable)  increase  the  pump’s  distance  from  the  EMI source, re-orient the device, or relocate the device,  The  EMC  limits  for  the  Medical  Device  Directive  93/42/EEC  (EN301489-1/-17  IEC/EN  60601-1-2:2014)  are designed to provide reasonable protection against harmful interference in a typical home use installation. The equipment  generates,  and  can  radiate,  radio  frequency  energy,  and  if  not  used  in  accordance  with  the instructions, may  cause  harmful  interference  to other  devices  in  the  vicinity.  However,  there  is  no  guarantee that interference  will not  occur in  a  particular installation.  If  this  equipment  does  cause harmful  interference with other  devices, the user is encouraged to try  to correct the interference by one  or more of the  following measures:  Reorient or relocate the receiving device  Increase the distance separating between the equipment parts  Consult the manufacturer or field service technician for help  Electromagnetic Safety Precautions  Do not expose the pump to therapeutic levels of ionizing radiation, as permanent damage to the pump electronic circuitry may occur. Remove the pump from the patient during therapeutic radiation sessions.  Do not expose the pump to magnetic resonance imaging (MRI) equipment and do not use the pump in the vicinity of it, as magnetic fields may adversely affect the operation of the pump. Remove the pump from the patient during MRI procedures, and keep it at a safe distance from magnetic energy.
 Page 16 of 38 2. WEARABLE INFUSION PUMP OVERVIEW 2.1. Top View   2.2. Side View    Pump Side Label Door Cartridge Adhesive Window Adhesive Liner Initiation Button Plastic Shell Pump Top Label
 Page 17 of 38 2.3. Pre-filled Medication Cartridge   2.4. Single Unit Pack      Cartridge Plunger Cartridge Cap Medication
 Page 18 of 38 3. INSTRUCTIONS FOR USE  The patient is an intended operator of the Wearable Infusion Pump. 3.1. Step 1 - Prepare 3.1.1 Gather all materials in a clean, well-lit area:   Wearable Infusion Pump single unit kit   Pre-filled medication cartridge    Adhesive bandage   Alcohol wipe   Cotton ball or gauze pad   Sharps disposal container*  *In case you do not have a sharps disposal container – please refer to section 1.5.1.1       Handle the medication according to your healthcare provider's instructions, including but not limited to: refrigeration of the pre-filled cartridge.       Review the information written on the paper cover of the single unit pack including the expiration date and the medication information.
 Page 19 of 38   3.1.2 Visually inspect the cartridge to verify cartridge integrity.     If there are any cracks or leaks coming from the cartridge- do not use the cartridge, and proceed to consult with your healthcare provider.    3.1.3 Peel the paper cover from the single unit pack and proceed to remove the plastic cover.  3.1.4 Now that the Wearable Infusion Pump is visible, remove the pump from the single unit pack.      It is recommended to keep the single unit pack for later; it can be used for easy disposal of the pump.    3.1.5 Visually inspect the pump to verify integrity prior to use.       If there are any cracks in the plastic shell of the pump, indication that the pump is damaged or impaired, or if the needle is visible - do not use the pump, and proceed to consult with your healthcare provider.
 Page 20 of 38 3.1.6 Open the pump door.   3.1.7 Load the cartridge into the pump, with the cartridge cap head first, and push inside until secure in place.   Correct insertion:                         Incorrect insertion:
 Page 21 of 38 3.1.8 Close the pump door; you should hear a click when the door closes successfully.   3.1.9 Choose an injection site that is appropriate for subcutaneous injections; a relatively smooth skin surface, without hair, open wounds or scars:  Abdomen:  at  or  under the  level  of the  belly  button,  about  two inches away from the navel  Arm: back or side of the upper arm  Thigh: front of the thigh  Lower back  Buttocks 3.1.9.1 Prepare the injection site by swabbing it with an alcohol wipe.     Swabbing of the skin and application of the pump to the skin should be done consecutively, within a reasonable timeframe.   3.2. Step 2 - Inject 3.2.1 Remove the adhesive liner from the Wearable Infusion Pump by pulling at the adhesive liner tab. Peel one side of the adhesive liner, and then the other. The adhesive liner, once separated from the Wearable Infusion Pump, can be discarded.  Once the adhesive liner has been removed, a beep will sound and the initiation button will light up momentarily
 Page 22 of 38  3.2.2 Prepare your injection area for application of the pump; ensure there are no folds of the skin, and place the pump on your skin firmly.  Abdomen placement:   Thigh placement:    Beep
 Page 23 of 38 Arm placement:   3.2.3 Press the initiation button to begin delivery of the medication   The pump shall beep with initiation of delivery and the green light shall begin blinking, and will continue blinking throughout the delivery.      If at this point you do not see a green blinking light illuminating the initiation button, try pressing the initiation button once more. Make sure the cartridge has been inserted into the pump, and that the pump is adhered to the skin.  If still there is no green light, an error has occurred. Do not continue using the pump. Remove the pump from your skin and contact your healthcare provider for further assistance.    Beep
 Page 24 of 38  3.2.4 During delivery:  Do not open the pump door during delivery  Take notice if alarms sound during delivery (refer to section 4)  Take notice if there is leakage from the pump during delivery       If there is leakage from the pump during delivery, an error has occurred. Do not continue using the pump. Remove the pump from your skin and contact your healthcare provider for further assistance.    3.3. Step 3 - Finish     When the injection is finished you should hear a beeping sequence and the green light will stop blinking. At this point no light illuminating the initiation button should be seen.      A slight stinging sensation in the injection site is expected.
 Page 25 of 38  3.3.1 Remove the pump from the skin by pulling at the adhesive.     There should not be a needle visible when the pump is removed from the body. If this is not the case - take extra care when handling and disposing of the pump.    3.3.2 Look through the pump window to verify that the medication was delivered, by viewing that the cartridge is empty.
 Page 26 of 38 3.3.3 Optional: Return the used pump into its original single unit pack for easy handling and disposal of the used pump.  3.3.4 Optional: For downloading the treatment summary from your Wearable Infusion Pump prior to disposal, open the Sorrel Wearable Infusion Pump smartphone application and follow the instructions.      The Sorrel Wearable Infusion Pump smartphone application allows you to download the treatment summary from our Wearable Infusion Pump and provides a summary of your treatment; date, time, flow rate, volume infused, pump serial number, and alarm information if relevant. This information can then be shared from your smartphone.
 Page 27 of 38 3.3.5 Dispose of the pump into a designated sharps disposal container.      This device is for single use only. Do not try to reuse it.    3.3.6 Check your injection site.   If there is blood, press a cotton ball or gauze pad on your injection site. Do not rub the injection site. Apply an adhesive bandage if needed.  Some irritation or redness of the skin may be present briefly after the injection has ended. This should disappear shortly. If it continues, contact your healthcare provider for further assistance.
 Page 28 of 38 4 ALARMS 4.1 Alarm Information    The Wearable Infusion Pump has only one type of alarm (high-priority), which may occur due to several reasons. In case of alarm, there will be consecutive beeping in two cycles, and then a 5 minute break before the next cycle. In addition, the initiation button will turn from green to red. Removing the device from the body will stop the alarm beeping sounds.  Alarm information can be seen in the treatment summary, via the smartphone application.      4.1.1 If an alarm occurs, the use of the Wearable Infusion Pump has ended, and the device must be removed from the body. Once removed, consult with your healthcare provider.  4.1.2    Alarm volume is always Max. Maximum alarm volume is 55 dBA    Beep
 Page 29 of 38 5 MAINTENANCE AND STORAGE     This device is for single use only! Do not try to reuse, clean or disinfect it.   5.1 Battery Classification The  UL  1642  Standard  for  Lithium  batteries  classifies  the  Lithium-Manganese  battery  used  in  the  Wearable Infusion Pump as primary battery (non-rechargeable). 5.2 Transport and Storage The pump should always be transported in its single pack. During handling and transport, protect the pump and the single pack from water, excessive humidity, and heat sources.  To safeguard the pump against prolonged exposure to dust and moisture, the pump must be stored in a clean and dry environment.  Specific recommendations for transport and long term storage conditions are listed in the following table. Condition Parameters Transport (transient – up to 24 hours) Long term storage Temperature -40ºC (-40ºF) to +70ºC (+158ºF) 2ºC (36ºF) to 27ºC (81ºF) Relative humidity 20% RH to 85% RH 20% RH to 85% RH Atmospheric pressure 69.6 kPa to 106 kPa (696 hPa to 1060 hPa) 69.6 kPa to 106 kPa (696 hPa to 1060 hPa)
 Page 30 of 38 6 TECHNICAL SPECIFICATIONS 6.1 Pump Accuracy  The dose accuracy of the Wearable Infusion per ISO 11608-1 under normal conditions is ±10% for dose of 0.1mL and ±5% for doses of up to 3mL. The rated accuracy of the Wearable Infusion per IEC 60601-2-24 under normal conditions is ±5%.  Testing was performed under normal conditions at room temperature (25oC, 72 oF). Normal conditions to ensure optimal accuracy  Barometric pressure of sea level altitude (101kPa)  Subcutaneous medication with water like fluid characteristics In  the  Wearable  Infusion  Pump,  as  in  all  infusion  systems,  external  factors  may  cause  fluctuations  in  rate accuracy. Conditions that can cause flow fluctuations include:  Fluid  characteristics  that  deviate  from  water-like  characteristics,  such  as  density,  viscosity  and homogeneity  Barometric pressure  below 101kPa  6.1.1 Start-up and Trumpet Graphs  The  following  graphs  and  curves  were  derived  from  the  pump  accuracy  testing  procedures  described  in  the IEC60601-2-24 standard. The start-up graphs represent startup flow versus operating time for the first half of the 3 mL cartridge (filled with 3 mL water) from the start of the infusion ("Stabilization period"), while the trumpet curve represents the percent flow rate deviation from the programmed rate over time of the second half of the cartridge ("Analysis period"). The  horizontal axis  represents the  observation time  intervals, which were adapted  to  represent  the time of the delivery. Over long observation windows, short-term fluctuation has little effect on accuracy, as represented by the flat part  of  the  curve.  As  the  observation  window  is  reduced,  short-term  fluctuations  have  a  greater  effect,  as represented by the "mouth" of the trumpet.
 Page 31 of 38  Figure 6.1. Delivery Startup Graph ("Stabilization period"), first 1.5 hours of Test Period, 1 mL/h   Figure 6.2. Trumpet Graph ("Analysis period"), second (and last) 1.5 hours of Test Period, 1 mL/h  00.20.40.60.811.2010 20 30 40 50 60 70 80 90Flow rate [mL/h] Time [min] -6.000%-5.000%-4.000%-3.000%-2.000%-1.000%0.000%1.000%2.000%0 5 10 15 20 25 30Error [%] Observation Window [min] maxminaverage
 Page 32 of 38  Figure 6.3. Delivery Startup Graph ("Stabilization period"), first hour of Test Period, 1.5 mL/h   Figure 6.4. Trumpet Graph ("Analysis period"), second (and last) hour of Test Period, 1.5 mL/h   6.2 Pump Specifications The following table lists and describes pump specifications. Parameter Description GENERAL  Type of Pump Container, type 1 ambulatory Operating Principle Piston 00.20.40.60.811.21.41.61.8010 20 30 40 50 60Flow rate [mL/h] Time [min] -5.000%-4.000%-3.000%-2.000%-1.000%0.000%1.000%2.000%3.000%0 5 10 15 20 25 30Error [%] Observation Window [min] maxminaverage
 Page 33 of 38 Patient population Adult and pediatric Delivery Mode Continuous Delivery route Subcutaneous System delivery accuracy (Nominal) ±5% Sensors Air sensor, cartridge sensor, temperature sensor Occlusion alarm threshold  4 bar Maximum  time  for activation  of  the  occlusion alarm 1 mL/h: 02:27 [min:sec], 1.5 mL/h: 02:05 [min:sec] History log Yes ELECTRICAL SAFETY  Electrical Safety Compliant with IEC 60601-1 - Edition 3.1 Electromagnetic compatibility Compliant with IEC 60601-1-2 – Edition 4.0 Recovery time for defibrillation-proof applied parts < 1 sec MECHANICAL AND POWER SPECIFICATIONS Pump Size 23 x 95 x 46 mm HxWxD (0.91 x 3.74 x 1.81 inches) Pump Weight   70 grams Ingress Protection IP65 Power Source Non-rechargeable Li- Manganese battery 3.0 V/1000 mA Battery consumption 6 hours of work at 0.5 mL/h OPERATING ENVIRONMENT Temperature +5°C to 40°C (41°F to 104°F) Relative Humidity 15% to 95% Atmospheric pressure 50 kPa to 106 kPa (500 hPa to 1060 hPa) TRANSPORT ENVIRONMENT Temperature -40°C to + 70°C (-40°F to 158°F) Relative Humidity 20% to 85% Atmospheric pressure 69.6 kPa to 106 kPa (696 hPa to 1060 hPa) STORAGE ENVIRONMENT  Temperature 2°C to 27°C (36°F to 81°F) Relative Humidity 20% to 85% Atmospheric pressure 69.6 kPa to 106 kPa (696 hPa to 1060 hPa)
 Page 34 of 38  The  following  list  provides  guidelines  about  environmental  conditions  and  situations  to  be  avoided  when working with or storing the Wearable Infusion Pump:  Avoid locations where there is inadequate ventilation.  Avoid locations where sudden impact or vibration may occur.  Avoid damp locations or locations where moisture level may increase considerably.  Avoid locations with large temperature fluctuations.  Avoid locations near an electrical heating apparatus.  Avoid locations exposed to chemicals or explosive gases. 6.3 Pump Ranges Parameter Range Increments Volume To Be Infused (mL) 0.1-3 0.1 Rate (mL/h) 0.01-60 0.01 Time 00:01-05:59  1 min / 1 hour  6.4 Electromagnetic Compatibility Statement  The following sections provide information about testing of and recommendations for electromagnetic compatibility statement. 6.4.1 Electromagnetic Emission The  Wearable  Infusion  Pump  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The customer or the user of the pump should ensure that it is used in such an environment. Emission Test Compliance Electromagnetic Environment Guidance RF emission  CISPR 11 class B The  pump  is  suitable  for  use  in  home healthcare environment.  6.4.2 Electromagnetic Immunity  The Wearable Infusion Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should ensure that it is used in the following environment.
 Page 35 of 38 Use  of  this  equipment  adjacent  to  or  stacked  with  other  equipment  (see  below  table)  should  be avoided because it could result in improper operation. If such use is necessary, this  equipment and the other equipment should be observed to verify that they are operating normally.  Test Method Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge (ESD) per IEC 61000-4-2 ± 8 kV contact; ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air*,** ±8 kV contact;  ±15 kV air Floors  should  be  wood,  concrete  or ceramic  tile.  If  floors  are  covered with  synthetic  material, the  relative humidity should be at least 30%. Radiated RF EM field immunity per IEC 61000-4-3 10 V/m, 80 MHz – 2.7 GHz, 80 % AM at 1 kHz* 10 V/m, 80 MHz – 2.7 GHz, 80 % AM at 1 kHz NA 10 V/m, 26 MHz – 2.5 GHz** 10 V/m, 26 MHz – 2.5 GHz NA Power frequency (50/60 Hz) magnetic field immunity per IEC 61000-4-8 30 A/m* 30 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. RF wireless communications equipment fields Immunity per IEC 61000-4-3 Frequencies and levels as specified at IEC 60601-1-2, Table 9* Max. 28 V/m NA * The pump was tested according to the EMC requirements of IEC 60601-1-2 (fourth edition) ** The pump was tested according to the EMC requirements of ISO 11608-1 (third edition)
 Page 36 of 38 7 COMPLIANCE INFORMATION STATEMENTS 7.1 FCC Declaration of Conformity We Sorrel Medical Ltd. declare: Type of equipment: Infusion Pump  Brand name or trademark: Wearable Infusion Pump Product Identification number: SORREL V04 | FCC ID: 2AR6L-SORRELWIP Applicable Compliance Statements: (e.g. for part 15 devices see §15.19(a)(3)) Country of origin: Israel Manufacturer: Sorrel Medical Ltd.  Responsible Party name (IN USA): Z & B Enterprises, Inc. Address: 12154 Darnestown Road, #236, Gaithersburg, MD 20878, USA Telephone: 301-251-9570  Internet E-Mail:  rhonashanker07@verizon.net Standards applied:  ● FCC Part 15B - For Unintentional radiators; ( test report number: XXX)  ● FCC Part 15C - For Intentional radiators; ( test report number: XXX)   Test reports/ certificates issued by: TBD Telecom Certification Body by: TBD As manufacturer/ manufacturer's authorized representative within the USA, we declare under our sole of responsibility that the equipment follows the provisions of FCC Equipment Authorization Procedures under CERTIFICATION (47 CFR Section 2.907) and / or SUPPLIER’S DECLARATION OF CONFORMITY (47 CFR Section 2.906) as stated above.   Thus,    is placed on the product Date of issue: December 28, 2018 Place of issue: Israel (Signature & Name of authorized person)                      (Company Stamp)
 Page 37 of 38 7.2 EU Declaration of Conformity We Sorrel Medical Ltd. declare under our sole responsibility: Type of equipment: Infusion Pump Brand name or trademark: Wearable Infusion Pump Product Identification number: V04 Traceability Identification: 15126-000-0001 Country of origin: Israel Manufacturer: Sorrel Medical Ltd. The authorized representative located within the Community is: Mrs. Dr. Stephanie Vorwerk Company: MedNet GmbH Address: Borkstrasse 10, Munster 48163, Germany Essential requirements according to directives:  ● 2014/30/EU EMCD ● 2014/35/EU LVD ● 1999/5/EC R&TTE or 2014/53 RED ● 2011/65/EU RoHS Standards applied:  ● EN 301489-17, EN 301489-3 ● EN 300 328 V2.1.1 , EN 300 330 V2.1.1 ● EN 62479 Other normative references: CEPT Radio Resolution 4561 Radio Equipment Class:  Class 2 Test reports/ certificates issued by: Global United Technology Services Co., Ltd. (GTS) China. As manufacturer/ manufacturer's authorized representative within the EEA, we declare under our sole of responsibility that the equipment follows the provisions of the Directive(s) as stated above.  Thus,     is placed on the product  Date  of issue:  December 28, 2018 Place of issue: Israel (Signature & Name of authorized person)                      (Company Stamp)
 Page 38 of 38    MEDNET GmbH  Borkstrasse 10  48163 Munster, Germany Sorrel Medical Ltd. 29 Yad Haruzim St.  P.O.Box 8639 Netanya 4250529, Israel 0123 eng Manufactured for: 15126-048-0001 Rev.00|Rev1.2.0/01.2019