Sensors and Software PE-PRO-TLF Ground Penetrating Radar User Manual pE100TX Manual

Sensors & Software Inc. Ground Penetrating Radar pE100TX Manual

Users Manual

USER’S GUIDECopyright 2003 Sensors & Software Inc.                                                                                2003-00376-00            pulseEKKO PRO TLF                      Transmitter
SENSORS & SOFTWARE INC. PRODUCT LICENCE, COPYRIGHT, LIABILITY AND WARRANTY INFORMATIONImportantPlease read this document carefully before removing the SOFTWARE PRODUCT storage media from their protective cover or assembling the HARDWARE PRODUCT. By removing the storage media or assembling the hardware, you are agreeing to be bound by the terms of this agreement. If you do not agree to the terms of this agreement, promptly contact Sen-sors & Software, Inc. at the address indicated at the end of this document.DefinitionThe word PRODUCT as used herein defines any tangible item sold by Sensors & Software, Inc. and may be comprised of HARDWARE PRODUCT which consists of physical objects and SOFTWARE PRODUCT which means computer programs, codes and related support materials.Software Product Licence AgreementIn order to preserve and protect its rights under the applicable laws, Sensors & Software, Inc. (hereafter S&S) does not sell any rights to its Software product. Rather, S&S grants the right to use its software, diskettes (or other storage media) and documentation (hereafter collectively called SOFTWARE PRODUCT) by means of a SOFTWARE PRODUCT licence. You acknowledge and agree that S&S retains worldwide title and rights to all its software and that the SOFT-WARE PRODUCT contains proprietary materials protected under copyright, trademark and trade secret laws.Grant of Software Product LicenceIn consideration of payment of the licence fee which is the price you pay for the SOFTWARE PRODUCT and your agree-ment to abide by the terms and conditions of this Licence Agreement, S&S grants to you, the Licensee, a non-exclu-sive right to use the SOFTWARE PRODUCT under the following conditions:You  may:• use the SOFTWARE PRODUCT on a single workstation owned, leased or otherwise controlled by you• copy the SOFTWARE PRODUCT for backup purposes in support of  your use of the product on a single workstationYou may not:• copy, distribute or sell copies of the SOFTWARE PRODUCT or accompanying written materials, including modified or merged SOFTWARE PRODUCT to others• sell, licence, sublicence, assign or otherwise transfer this licence to anyone without the prior written consent of S&S• modify, adapt, translate, decompile, disassemble or create derivative works based on the SOFTWARE PRODUCTTerminationThis licence is effective until terminated. You may terminate the licence at any time by returning the SOFTWARE PROD-UCT and all copies to S&S. The licence will automatically terminate without notice by S&S if you fail to comply with any terms or conditions of this agreement. Upon termination, you agree to return all copies of the SOFTWARE PRODUCT to S&S.Update PolicyS&S may create, from time to time, updated versions of its SOFTWARE PRODUCT. At its option, S&S will make such updates available to licencees who have paid the update fee.Product Warranty, Limited Remedy and Limited Liability S&S warrants the PRODUCT to be free from defect in material and workmanship under normal use for a period of one year (365 days) from the date of shipment. Any third party computer systems or other items not manufactured directly by S&S purchased with any PRODUCT or independently from S&S are subject to the original manufacturer's warranty and not the responsibility of S&S.S&S makes no other warranties including, but not limited to, any implied warranty of merchantability or fitness for a particular purpose.  If this product is defective within the warranty period stated above, your exclusive remedy shall be, at S&S’s option to replace or repair the S&S product or refund the purchase price of the S&S product.  Except where prohibited by law, S&S will not be liable for any loss or damage arising from this S&S product, whether direct, indirect, special, incidental or consequential regardless of the legal theory asserted.
All statements, technical information, and recommendations related to S&S products are based on information believed to be reliable, but the accuracy or completeness is not guaranteed. Before using this product, you must eval-uate it and determine if it is suitable for your intended application.  You assume all risks and liability associated with such use.  Any statements related to the product which are not contained in S&S current publications, or any contrary statements contained on your purchase order shall have not force or effect unless expressly agreed upon, in writing, by an authorized officer of S&S.S&S warrants the diskettes or other storage media on which the SOFTWARE PRODUCT is furnished to be free from defects in material and workmanship under normal use for a period of ninety (90) days from the date of purchase as evidenced by a copy of your invoice.Except as specified above, any SOFTWARE PRODUCT is provided "as is" without warranty of any kind, either expressed or implied, including, but not limited to, the use or result of use of the product in terms of correctness, accuracy, reliabil-ity, currentness or otherwise. The entire risk as to the results and performance of the PRODUCT is assumed by you. If the PRODUCT is defective or used improperly, you, and not S&S or its dealers, distributors, agents, or employees, assume the entire cost of all necessary servicing, repair or correction.S&S's entire liability and your exclusive remedy for SOFTWARE PRODUCT shall be, at S&S's option, either• the replacement of any diskette or hardware components which do not meet S&S's Limited Warranty and which are returned to S&S postage prepaid with a copy of the receipt, or• if S&S is unable to deliver a replacement diskette which is free of defects in material or workmanship, Licensee may terminate this agreement and have the licence fee refunded by returning all copies of the SOFTWARE PRODUCT postage prepaid with a copy of the receipt.If failure of any PRODUCT resulted from accident, abuse or misapplication, S&S shall have no responsibility to replace the SOFTWARE PRODUCT, refund the licence fee, or replace or repair the HARDWARE PRODUCT.Do not tamper with any PRODUCT.  PRODUCT contains no user serviceable parts.  If tampering is evident in S&S’s opinion, warranty is void and null.No oral or written information or advice given by S&S, its dealers, distributors, agents or employees shall create a war-ranty or in any way increase the scope of this warranty and you may not rely on any such information or advice.Neither S&S nor anyone else who has been involved in the creation, production or delivery of the PRODUCT shall be lia-ble for any direct, indirect, special, exemplary, incidental or consequential damages, claims or actions including lost information, lost profits, or other damages arising out of the use or inability to use this PRODUCT even if S&S has been advised of the possibility of such damages.This warranty gives you specific rights.  You may have other rights which vary from province to province, territory to ter-ritory and certain limitations contained in this limited warranty may not apply to you.GeneralpulseEKKO®, Noggin®, SpiView®, Conquest® and SnowScan® are registered trademarks of S&S. No right, licence, or interest to such trademarks is granted hereunder with the purchase of the PRODUCT or the SOFTWARE PRODUCT licence.Governing LawIn the event of any conflict between any provision in this licence agreement and limited warranty and any applicable provincial legislation, the applicable provincial legislation takes precedence over the contravening provision. This agreement shall be governed and construed in accordance with the laws of the Province of Ontario, Canada.ServiceabilityShould any term of this agreement be declared void or not enforceable by any court of competent jurisdiction, the remaining terms shall remain in full effect.WaiverFailure of either party to enforce any of its rights in this agreement or take action against any other party in the event of a breach of this agreement shall not be considered a waiver of the right to subsequent enforcement of its rights or actions in the event of subsequent breaches by the other party.
AcknowledgementYou acknowledge that you have read this agreement, understand it and agree to be bound by its terms and conditions. You further agree that this agreement is the complete and exclusive statement of agreement between the parties and supersedes all proposals or prior agreements oral or written between the parties relating to the subject matter of this agreement.Should you have any questions concerning this agreement, please contact in writing:Sensors & Software Inc.1040 Stacey CourtMississauga, OntarioCanada L4W 2X8Tel:(905) 624-8909Fax:(905) 624-9365E-mail: sales@sensoft.capulseEKKO, Noggin, SpiView, SnowScan, Conquest, SmartCart, SmartHandle, Conquest for Windows, Win_EKKO, EKKO_Mapper, EKKO_3D and EKKO_Pointer are Trademarks of Sensors & Software Inc.
pulseEKKO PRO TLF Transmitter iTable of Contents 1 Overview ................................................................................................1 2 System Assembly and Startup .......................................................32.1 Antenna Assembly ...........................................................................32.2 Connecting up the Radar .................................................................42.3 Adding Optional Items ......................................................................62.3.1 Fiber Optic Remote Trigger and Beeper Unit..............................62.3.2 Odometer................................................................................7 3 Troubleshooting ..................................................................................93.1 Transmitter Problem: No Signal on Screen......................................9 4 Care and Maintenance ..................................................................... 114.1 General........................................................................................... 114.2 Radar Unit ...................................................................................... 114.3 Battery Power Requirements ......................................................... 114.4 Transmitter and Receiver Battery Maintenance ............................. 114.5 Testing Batteries ............................................................................124.6 Fiber Optics Cables........................................................................13 5 Helpful Hints .......................................................................................155.1 Handling Fiber Optic Cables ..........................................................155.2 Connecting the Fiber Optics...........................................................155.3 Batteries .........................................................................................15
ii pulseEKKO PRO TLF Transmitter Appendix A  Health & Safety Certification ....................................... A-1 Appendix B  FCC Regulations ............................................................. B-1 Appendix C  Instrument Interference ................................................C-1 Appendix D  Safety Around Explosive Devices .............................D-1
pulseEKKO PRO TLF Transmitter 1-Overview1 1 OverviewThis manual describes how to use the pulseEKKO PRO Ground Penetrating Radar (GPR)system transmitter. System Assembly Section 2 on page 3 discusses in detail the step by step procedure for& Startup:  assembling the pulseEKKO PRO system and connecting the transmit-ter to it. Troubleshooting:  Section 3 on page 9 presents some simple steps the user should go through when the transmitter is not working as it should.Care and Maintenance:  Section 4 on page 11 discusses procedures for the care and mainte-nance of your pulseEKKO transmitter.Helpful Hints:  Based on years of experience, Section 5 lists some helpful hints to help make data collection and field operations run as smooth as possi-ble.Users should read Sensors & Software’s statement on issues regarding health and safety in Appendix A - Health & Safety Certification.
1-Overview pulseEKKO PRO TLF Transmitter2
pulseEKKO PRO TLF Transmitter 2-System Assembly and Startup3 2 System Assembly and StartupThe modular design of the pulseEKKO Pro GPR makes the system very flexible and readily field-portable.  There are four essential components to the radar system: the control or console unit,the transmitter assembly, the receiver assembly and the Digital Video Logger (DVL).  Thetransmitter and receiver assemblies are connected to the console unit via appropriate fiber opticscables and the console is connected to the DVL via a RS232 cable (Figure 2-1).  This sectiondiscusses the detailed steps to follow to assemble the whole pulseEKKO Pro system. Figure:  2-1  2.1 Antenna AssemblyThe procedures for assembling the transmitter and receiver antennas are identical.  A detaileddiagram of this assembly can be found in Figure 2-2. The assembly steps are as follows:a) Check the two male brass antenna connector pins for damage; replace if necessary.  Insert the pins into the two threaded holes in the center of the antennas.  Tighten the pins finger-tight.  DO NOT APPLY UNDUE FORCE !b) Insert and tighten the two female brass antenna sockets into the bottom of the transmitter and receiver electronic boxes. Tighten the sockets finger-tight. DO NOT APPLY UNDUE FORCE !c) Attach the antenna mounting block to the antenna by using the flathead screwdriver to tighten the 4 (four) quarter-turn fasteners, ensuring that the male brass antenna pins protrude up the center holes of the mounting blocks. Quarter-turn fasteners work by aligning the screw in the socket and the pressing downward and tightening a
2-System Assembly and Startup pulseEKKO PRO TLF Transmitter4quarter of a turn. DO NOT APPLY UNDUE FORCE. It is usually best to have all the screws properly aligned in their socket before tightening each one. d) Carefully place the electronic boxes down onto the mounting block such that both brass pins fully connect.  Then use the 2 plastic draw latch connectors to hold the electronics boxes on the mounting block.e) With the transmitter and receiver power switched OFF, unlatch the 2 battery covers on the sides of the electronics and open.  Place one 12-volt battery on each side on the electronics boxes making sure the positive (+) terminal faces inward toward the electronics (the battery only fits properly in this orientation).  Close and latch the battery covers.  Note that the system will run with only one 12 volt battery but using two batteries is recommended.f) Attach the adjustable handle to the antenna using the flat head screwdriver to tighten the 4 quarter-turn fasteners as with the mounting block.  This handle can then be adjusted for height by loosening the 2 knurled knobs by hand, moving the handle to the desired height and retightening the knobs (Figure 2-2). Figure:  2-2  2.2 Connecting up the RadarOnce the antennas are assembled, the next step is to connect the antennas to the console andthe console to the DVL. Refer to Figure 2-1.a) Check the fiber optic cables for damage by holding one end towards a light source and looking into the other end.  If light is not transmitted through the cable or appears dim, then replace or repair the cable.  Inspect the cable for any kinks or signs of dam-age and, again, repair if necessary.
pulseEKKO PRO TLF Transmitter 2-System Assembly and Startup5b) Plug the black end of the single fiber optics cable (or dual fiber optics cable if you are using a remote trigger or odometer; see below) into the INPUT (black) connector in the transmitter and the other end into the single red receptacle labelled “Transmitter” on the console.c) Plug the dual fiber optics cable into the receiver, black to black and grey to grey, and the other ends likewise into the double yellow receptacles labelled “Receiver” on the radar console.d) Using the RS232 cable, connect the console unit (white receptacle) to the serial port on the DVL (the cable will only fit in one of the receptacles on the back of the DVL).  It is suggested that these connections be secured with proper screws to prevent accidental disconnection during operation.e) Connect the DVL to Power Supply Cable to the 9-socket connector on the back of the DVL. Then connect the Power Cable Extension with Alligator Clips to the end of the DVL to Power Supply Cable. The alligator clips can then be connected to a 12 Volt battery. Make sure that the black clip is attached to the negative (-) and the red clip is attached to the positive (+) battery terminals. If the alligator clips are connected to the wrong terminals of the battery, the DVL will not be powered. When the DVL is receiving power the upper red light on the front of the DVL will be illuminated.f) Turn the transmitter and receiver ON by pressing the button on the top of each unit. The red Power light should come on to indicate that power is being received. If not, check that the batteries inside the Transmitter and Receiver are fully charged and have been inserted the right way. g) Turn the DVL ON by pressing any button on the front. As the DVL boots up, the lower red light will come on and it will beep until it has fully booted up.h) Connect the console power supply to the white POWER receptacle and turn power supply ON, if necessary. If a battery cable is being used make sure that the black clip is attached to the negative (-) and the red clip is attached to the positive (+) battery terminals.i) Before actually storing data, allow the console to run to reach ambient operating temperature.  This time varies depending on outside temperature; however, 5 to 10 minutes is generally sufficient.j) When not collecting data, the transmitter, receiver and console should be turned OFF to increase the life of the batteries.Note that it is possible for the pulseEKKO Pro console and the DVL to share a common 12 Volt power source like a battery. Simply connect the alligator clips on the Console Power Supply cable and the Power Cable Extension with Alligator Clips to the proper terminals of the battery.
2-System Assembly and Startup pulseEKKO PRO TLF Transmitter62.3 Adding Optional Items2.3.1 Fiber Optic Remote Trigger and Beeper UnitTo attach the Fiber Optic Remote unit to the pulseEKKO Pro deluxe handles, remove the screwnear the top of one of the handles. Then attach the two pieces of the handle attachmentassembly (Figure 2-3) and replace the screw as shown. Now press the Fiber Optic Remote ontothe handle and twist it a quarter-turn into place.  Figure:  2-3  For PVC handles, attach the Fiber Optic Remote to the handle as shown in Figure 2-4. Figure:  2-4  To connect up the fiber optic remote trigger and beeper unit you should have a dual fiber opticcable for the console to transmitter connection plus a short single fiber optic cable (Figure 2-5).One of the two cables on the dual fiber optic cable will, if the above directions were followed,already be connected from the red transmitter receptacle on the console to the black (INPUT)receptacle on the transmitter.
pulseEKKO PRO TLF Transmitter 2-System Assembly and Startup7 Figure:  2-5  The steps necessary to complete the connection of the remote trigger and beeper unit are:1) The black fiber optic cable of the dual cable connects from the red REMOTE receptacle on the console to the OUTPUT (grey) fiber optic connector on the remote trigger and beeper unit. This connection is necessary for the trigger part of the remote trigger and beeper unit to work. 2) The short single fiber optic cable connects from the INPUT (black) fiber optic connector on the remote trigger and beeper unit to the OUTPUT (grey) receptacle on the transmitter. This connection is necessary for the beeper part of the remote trigger and beeper unit to work. During data acquisition, the beeper will emit a beep as data are being collected. As well, when the radar system is run in Step mode, data acquisition can be controlled using thebutton on the remote trigger and beeper unit.2.3.2 OdometerThe pulseEKKO Pro odometer is useful for collecting data at a user defined stepsize (1 cm to 10m) without the necessity of a tape measure or survey makers.
2-System Assembly and Startup pulseEKKO PRO TLF Transmitter8The odometer has 2 settings: string and wheel.  The wheel odometer setting can be used insituations where a wheel or cart can easily be pulled across the surface.  The string odometersetting is useful when conditions are too rugged for the use of a wheel.The odometer has both a fiber optic connection and an electrical connection (Figure 2-6).  Eitherconnection can be used to connect the odometer to the console. Figure:  2-6  1) Electrical Connection:  To connect up the odometer using the electrical connection you should have the proper cable for the console to odometer connection.  Connect the cable from the red electrical REMOTE receptacle on the console to the electrical receptacle on the odometer. (Figure 2-6a)2) Fiber Optic Connection:  To connect up the odometer using the fiber optic connection you should have a dual fiber optic cable for the console to transmitter connection.  One of the two cables on the dual fiber optic cable will, if the above directions were followed, already be connected from the red transmitter receptacle on the console to the INPUT (black) receptacle on the transmitter. (Figure 2-6b)To complete the connection of the odometer the black fiber optic cable of the dual cable connectsfrom the red REMOTE receptacle on the console to the grey fiber optic connector on theodometer.With either of these connections made, when the radar system is run in Step mode, dataacquisition is controlled by the odometer.
pulseEKKO PRO TLF Transmitter 3-Troubleshooting9 3 Troubleshooting In this section, a thorough discussion of the error conditions will be presented together with asystematic step-by-step procedure for tracking down the source of the problem.3.1 Transmitter Problem: No Signal on Screen a) In Graph mode try the Automatic timezero search.  Keep the antennas about one antenna length apart.b) If the error comes up “No Signal Found” then try using the Full Search option in Graph mode timezero search to look for the pulse.If the pulse still cannot be found do the following checks:Make sure the transmitter battery is properly charged and connected.Check the condition of the single fiber optic cable to the transmitter, looking for any signs ofstress or damage.  Replace or repair if needed.Check that the brass antenna sockets in the bottom of the transmitter and the brass antenna pinsin the top of the antenna are properly tightened. Also check that there is a good contact betweenthem.Set the system up with a time window of 1500 ns, stacks at 1024 and sampling interval at 800ps.Unplug the grey transmitter fiber optics at the console.With the system running, look for a red light coming from the grey hole in the console (singlehole). Plug this cable back in and unplug the INPUT (black) cable at the transmitter.  Check for ared light coming from the cable.If the above process fails while checking the light in the cable then replace the cable.If the above process fails while checking the light in the console, carefully remove the back fromthe console unit and check the short fiber optic cables inside the unit.  As a test, these shortsections of cable may be bypassed by plugging the long transmitter fiber optic cable directly intothe electronics.  Replace or polish the short cable if needed.c) If the signal is still not found, set the system up with the maximum time window (2048 ns on older systems, 32000 ns on newer systems) and stacks at one.  Run in Graph mode and use the Full Search option to look for signal.  If the pulse is visible on the trace, adjust it to the shaded region.  Now you should be able to change the time window back to the desired length and have timezero visible.d) If the signal is still not found, contact Sensors & Software Inc.
3-Troubleshooting pulseEKKO PRO TLF Transmitter10
pulseEKKO PRO TLF Transmitter 4-Care and Maintenance11 4 Care and Maintenance4.1 GeneralTo operate the pulseEKKO Pro equipment in a geophysical field environment, the user shouldexercise the normal care afforded other field instrumentation.  Items that require specificmaintenance procedures are listed below together with detailed descriptions of the procedures.All the equipment is ‘shower proof’ but will not tolerate total immersion.  If the equipment isimmersed in water, open up the affected box, and dry it out before operating.  If the equipment isoperated in very humid conditions, open up the boxes occasionally and allow the components todry out.  Condensation inside the console may occur in cold, wet climates.  If condensation isseen, open up the console and dry out before applying power.The interior of the console can become very hot very quickly if exposed to strong, direct sunlight.High temperatures can affect system electronics and may lead to failures. Try to keep theconsole well insulated against extreme temperatures. If a suspected heat failure does occur, stopthe system and allow it to cool down before retrying.The equipment is housed in heavy duty boxes and can withstand a reasonable amount ofphysical abuse.  Items that are particularly sensitive to damage are the following.4.2 Radar UnitIf operated in dusty conditions the console receptacles and fiber optic sockets may become dustyinside.  In such conditions open the unit up on a regular basis and blow out or vacuum all thedust. It is best to place the plastic connector plugs into the receptacles when not in use.4.3 Battery Power RequirementsThe pulseEKKO Pro console is designed to operate from a 12 volts DC, 3 to 4 amp source suchas batteries or power supply.  The input voltage tolerance is 10.2 volts to 15 volts.  The consoledraws 1 amp.The transmitter employs one or two 12-volt batteries and draw less than 1 amp.  They will, undernormal conditions, run for 4 hours or more on a pair of 2.3 AH batteries.  Sets of rechargeablelead-acid batteries are provided with the system for the transmitter.4.4 Transmitter and Receiver Battery MaintenanceThe pulseEKKO Pro transmitter both use one or two 12-volt sealed lead acid batteries.  If theyare maintained in a charged condition they will give long life and reliable service. Improper useand maintenance will greatly reduce their life.Sealed lead acid batteries should NEVER be left in a discharged condition for any period of time.Charge the batteries as soon as possible after use.
4-Care and Maintenance pulseEKKO PRO TLF Transmitter12Charge batteries at room temperature whenever possible.The receiver and transmitter contain a voltage monitoring circuit that will turn off the unit when theinput voltage drops below 10.2 volts.  If the voltage of the discharged batteries is measured sometime after they have been changed it will be higher than 10.2 volts (this is normal).  Do not mixcharged and discharged batteries and DO NOT attempt to use discharged batteries a secondtime.If a battery has been deeply discharged or left in a discharged condition for some period of time itmay not accept charge immediately when it is connected to the charger (The fast charge LED willnot light).  If the fast charge light does not come on within 6 hours the battery should beconsidered damaged and discarded.Do not assume that a battery that is still charging after 8 hours is nearing the end of its charge cycle.  Typical charging time for an empty battery is 12 hours from start of fast charge. When charging a depleted battery it is normal for the charge light to go out after 6-8 hours.  The charger has gone into trickle mode only and the battery must remain in the charger for the minimum 12 hours required. Ensure that the batteries are fully charged before storing.  If practical, store the batteries in a coolplace, 10oC (a refrigerator is ideal), but make sure the temperature is not likely to drop below -30oC or the electrolyte may freeze and possibly split the case.4.5 Testing BatteriesIf you are experiencing reduced operation time with either your pulseEKKO Pro Transmitter theproblem will be the result of a damaged 12 V Battery or a poor battery connection in one of thebattery compartments. The pulseEKKO Pro Transmitter is designed to operate while using a 12Vbattery in one battery compartment or both battery compartments.  This feature can be used toload test the batteries, as outlined below:a) Label all batteries so that they can be distinguished from one another.b) Insert fully charged suspect battery into the battery compartment of either the Transmitter or Receiver unit.   c) Open the empty top battery compartment and measure the D.C. voltage at the two contact points on the battery holder plate.  A suspect battery will show a  substantial drop in voltage when the battery is loaded down (POWER ON) and will slowly drop to less than 10 V after 2 hours of monitoring voltage.  A good battery will remain at 12 V or higher after 2 hours of monitoring voltage.
pulseEKKO PRO TLF Transmitter 4-Care and Maintenance13All battery compartments should be checked periodically for proper operation.  Take a knowngood battery and install into the four battery compartments.  Ensure the power light activates withno intermittent operation.  If the power light will not activate look in the battery compartment andcheck the battery contact for signs of excessive stress.  Also check the condition of the 2A fusewiring.  If the problem persists, remove the six screws holding the battery plate in place and liftupwards.  On the bottom of the battery plate you will see a small circuit board that extends thebattery input to the power switch.  Check the connections to and from this board.  Please contactSensors & Software Inc.  should all the above fail to resolve the problem.  4.6 Fiber Optics CablesThese cables can stand a great deal of abrasive wear but the user should avoid sharp bends inthe cable, pulling the cable at right angles close to the fiber optics connectors, or allowing theunprotected ends to get dirty or scratched.  Erratic timezero jumps usually indicate a crackedcable.  If problems are encountered with the fiber optics cables, repair is very straightforward andis outlined in detail below.  The best policy is to carry spares.  Short fiber optic cables are locatedinside each of the units.  These short cables may need cleaning and polishing from time to time.
4-Care and Maintenance pulseEKKO PRO TLF Transmitter14
pulseEKKO PRO TLF Transmitter 5-Helpful Hints15 5 Helpful Hints5.1 Handling Fiber Optic CablesThe fiber optic cables are the most fragile component in the system. They can be easilydamaged by misuse. When handling the fiber optic cables always roll and unroll them as if they were on a spool.  DONOT roll them around your arm like a rope or extension cord. Also, these cables do not twist,therefore laying the roll on the ground and pulling the cable off will cause the cables to tangle.5.2 Connecting the Fiber OpticsAlways push in and remove the cable by holding the strain relief jacket or the plastic endconnector.  Pulling on the cable will loosen the crimp and damage the cable.  After connecting thefiber optic cables to the antenna electronic boxes, loosely loop the cable and tape off some slackonto the handles.  Most breaks occur right at the cable ends when they are accidentally pulled.5.3 BatteriesReplace batteries before they go dead.  Nothing is more annoying than losing data in the middleof a traverse.  Only operate continuously for about 3 hours before replacing unless you havenew, well charged batteries.  Carry plenty of spares.
5-Helpful Hints pulseEKKO PRO TLF Transmitter16
pulseEKKO PRO TLF Transmitter Appendix A - Health & Safety CertificationA-1 Appendix A  Health & Safety CertificationRadio frequency electromagnetic fields may pose a health hazard when the fields are intense.Normal fields have been studied extensively over the past 30 years with no conclusive epidemiol-ogy relating electromagnetic fields to health problems.  Detailed discussions on the subject arecontained in the references and the web sites listed below.The USA Federal Communication Commission (FCC) and Occupational Safety and HealthAdministration (OSHA) both specify acceptable levels for electromagnetic fields.  Similar powerlevels are mandated by corresponding agencies in other countries.  Maximum permissible expo-sures and time duration specified by the FCC and OSHA vary with excitation frequency.  The low-est threshold plane wave equivalent power cited is 0.2 mW/cm2 for general population over the30 to 300 MHz frequency band.  All other applications and frequencies have higher tolerances asshown in graphically in Figure A-1. Figure A-1:  FCC limits for maximum permissible exposure (MPE) plane-wave equivalent power density mW/cm2.All Sensors & Software Inc. pulseEKKO, Noggin and Conquest products are normally operatedat least 1 m from the user and as such are classified as “mobile” devices according to the FCC.Typical power density levels at a distance of 1 m or greater from any Sensors & Software Inc.product are less than 10-3 mW/cm2 which are 200 to 10,000 times lower than mandated limits.As such, Sensors & Software Inc. products pose no health and safety risk when operated in thenormal manner of intended use.
Appendix A - Health & Safety Certification pulseEKKO PRO TLF TransmitterA-2References1. Questions and answers about biological effects and potential hazards of radio-frequency electromag-netic fieldUSA Federal Communications Commission, Office of Engineering & TechnologyOET Bulletin 56(Contains many references and web sites)2. Evaluation Compliance with FCC Guidelines for Human Exposure to Radio Frequency Electromag-netic Fields.USA Federal Communications Commission, Office of Engineering & TechnologyOET Bulletin 56(Contains many references and web sites)3. USA Occupational Safety and Health Administration regulations paragraph 1910.67 and 1910.263.Web Siteswww.fcc.gov/Bureau/EngineeringTechnlogy/Documents/bulletin www.osha-slc.gov/SLTC (see radio frequency)
pulseEKKO PRO TLF Transmitter Appendix B - FCC RegulationsB-1 Appendix B FCC RegulationsThis device complies with Part 15 of the USA Federal Communications Commission (FCC)Rules.  Operation in the USA is subject to the following two conditions:(1) this device may not cause harmful interference and(2) this device must accept any interference received, including interference that may cause undesired operation.Part 15 – User InformationThis equipment has been tested and found to comply with the limits for a Class A digital  device,where applicable, and for an ultrawide bandwidth (UWB) device where applicable, pursuant toPart 15 of the FCC Rules. These limits are designed to provide reasonable protection againstharmful interference when the equipment is operated in a commercial environment. This equip-ment generates, uses and can radiate radio frequency energy and, if not installed and used inaccordance with the instruction manual, may cause harmful interference to radio communica-tions. Operation of this equipment in a residential area is likely to cause harmful interference inwhich case the user will be required to correct the interference at his own expense.WARNINGChanges or Modifications not expressly approved by Sensors & Software Inc. could void theuser’s authority to operate the equipment.Certification of this equipment has been carried out using approved cables and peripheraldevices.  The use of non-approved or modified cables and peripheral devices constitutes aChange or Modification outlined in the warning above.Operating RestrictionsOperation of this device is limited to purposes associated with law enforcement, fire fighting,emergency rescue, scientific research, commercial mining, or construction.  Parties operatingthis equipment must be eligible for licensing under the provisions of Part 90 of this chapter.  FCC Interpretation of Operation Restrictions issued July 12, 2002(FCC Order DA02-1658, paragraph 9)The regulations contain restrictions on the parties that are eligible to operate imaging systems. 1Under the new regulations, GPRs and wall imaging systems may be used only by law enforce-ment, fire and emergency rescue organizations, by scientific research institutes, by commercialmining companies, and by construction companies.  Since the adoption of the Order, we havereceived several inquiries from the operators of GPRs and wall imaging systems noting thatthese devices often are not operated by the users listed in the regulations but are operated undercontract by personnel specifically trained in the operation of these devices.  We do not believethat the recent adoption of the UWB rules should disrupt the critical safety services that can be1.    See 47 C.F.R. §§15.509(b), 15.511(b), and 15.513(b)
Appendix B - FCC Regulations pulseEKKO PRO TLF TransmitterB-2performed effectively only through the use of GPRs and wall imaging systems.  We viewed theseoperating restrictions in the broadest of terms.  For example, we believe that the limitation on theuse of GPRs and wall imaging systems by construction companies encompasses the inspectionof buildings, roadways, bridges and runways even if the inspection finds no damage to the struc-ture and construction does not actually result from the inspection; the intended purpose of theoperation of the UWB device is to determine if construction is required.  We also believe that theGPRs and wall imaging systems may be operated for one of the purposes described in the regu-lations but need not be operated directly by one of the described parties.  For example, a GPRmay be operated by a private company investigating forensic evidence for a local police depart-ment.  FCC Permitted Mode of UsageThe GPR antenna must be kept on the surface to be in compliance with FCC regulations.  Use ofthe antenna is not permitted if it is lifted off the surface.  Use as a through-the-wall imagingdevice is prohibited.  GPR Use CoordinationFCC regulation 15.525(c) requires users of GPR equipment to coordinate the use of their GPRequipment as described below: a) UWB imaging systems require coordination through the FCC beforethe equipment may be used. The operator shall comply with any con-straints on equipment usage resulting from this coordination. b) The users of UWB imaging devices shall supply operational areas tothe FCC Office of Engineering and Technology, which shall coordinatethis information with the Federal Government through the NationalTelecommunications and Information Administration.  The informationprovided by the UWB operator shall include the name, address, andother pertinent contact information of the user, the desired geographi-cal area(s) of operation, and the FCC ID number and other nomencla-ture of the UWB device.  If the imaging device is intended to be usedfor mobile applications, the geographical area(s) of operation may bethe state(s) or county(ies) in which the equipment will be operated.The operator of an imaging system used for fixed operation shall sup-ply a specific geographical location or the address at which the equip-ment will be operated.  This material shall be submitted to the followingaddress: Frequency Coordination Branch., OETFederal Communications Commission445 12th Street, SWWashington, D.C.  20554ATTN:  UWB CoordinationThe form given on the following page is a suggested format for performing the coordination.
pulseEKKO PRO TLF Transmitter Appendix B - FCC RegulationsB-3FCC GROUND PENETRATING RADAR COORDINATION NOTICE NAME:ADDRESS:CONTACT INFORMATION [CONTACT NAME AND PHONE NUMBER]:AREA OF OPERATION [COUNTIES, STATES OR LARGER AREAS]:FCC ID: [E.G. QJQ-PE-PRO-TLF)]EQUIPMENT NOMENCLATURE: [ E.G. PULSEEKKO PRO TLF] Send the information to:Frequency Coordination Branch., OETFederal Communications Commission445 12th Street, SWWashington, D.C.  20554ATTN:  UWB CoordinationFax:       202-418-1944INFORMATION PROVIDED IS DEEMED CONFIDENTIAL
Appendix B - FCC Regulations pulseEKKO PRO TLF TransmitterB-4
pulseEKKO PRO TLF Transmitter Appendix C - Instrument InterferenceC-1 Appendix C Instrument InterferenceImmunity regulations place the onus on instrument/apparatus/device manufacturers to assurethat extraneous interference will not unduly cause an instrument/apparatus/device to stop func-tioning or to function in a faulty manner.Based on independent testing house measurements, Sensors & Software Inc. systems complywith such regulations in Canada, USA, European Community and most other jurisdictions. GPRdevices can sense electromagnetic fields.  External sources of electromagnetic fields such as TVstations, radio stations and cell phones, can cause signals detectable by a GPR which maydegrade the quality of the data that a GPR device records and displays.Such interference is unavoidable but sensible survey practice and operation by an experiencedGPR practitioner can minimize such problems. In some geographic areas emissions from exter-nal sources may be so large as to preclude useful measurements. Such conditions are readilyrecognized and accepted by the professional geophysical community as a fundamental limitationof geophysical survey practice.  Such interference being present in the GPR recordings is notconsidered as an equipment fault or as a failure to comply with immunity regulations.
Appendix C - Instrument Interference pulseEKKO PRO TLF TransmitterC-2
pulseEKKO PRO TLF Transmitter Appendix D - Safety Around Explosive DevicesD-1 Appendix D Safety Around Explosive DevicesConcerns are expressed from time to time on the hazard of GPR products being used near blast-ing caps and unexploded ordnance (UXO). Experience with blasting caps indicates that thepower of Sensors & Software Inc.’s GPR products are not sufficient to trigger blasting caps.Based on a conservative independent testing house analysis, we recommend keeping the GPRtransmitters at least 5 feet (2m) from blasting cap leads as a precaution.  Some customers doexperimental trials with their particular blasting devices to confirm with safety.  We strongly rec-ommend that GPR users routinely working with explosive devices develop a systematic safetymethodology in their work areas.  The UXO issue is more complex and standards on fuses do not exist for obvious reasons. Todate, no problems have been reported with any geophysical instrument used for UXO. Sinceproximity and vibration are also critical for UXO, the best advice is to be cautious and understandthe risks.
Appendix D - Safety Around Explosive Devices pulseEKKO PRO TLF TransmitterD-2

Navigation menu