Rayence RY1012WCA Medical Image Processing Unit User Manual

Rayence Co., Ltd. Medical Image Processing Unit User Manual

User_Manual

       Release Version 1.0.6 Medical Image Processing Unit 1012WCA User Manual
  R-M-011-121030                          © Copyright 2012, Rayence Co. Ltd.   All  pages  of  this  document  contain  proprietary  and  confidential information of Rayence Corporation and are intended for exclusive use by Rayence Corporation personnel or customers. Copying, disclosure to others  or  other  use  is  prohibited  without  the  express  written authorization from Rayence Corporation. Please report any violations of requirement to Rayence Corporation.
  R-M-011-121030      Attention For improvement of product performance, supplementation, or follow-up of information; the contents of this manual are subject to change without separate prior notice.   Please note that our company has neither responsibility for any accidents nor obligation to do free repair service for any damage of the equipment due to user's mistake, which resulted from failure  to follow  the  contents  in  this  manual.  Make  sure  to  be familiar  with  the  safety precautions  and  usage procedures. Also  note  that  the  product  may  slightly differ  from  the contents of this manual depending on specification.  The following marks are used for the effective use of the product in this manual.            Attention, consult accompanying documents.       This is used to emphasize essential information. Be sure to read this information to avoid incorrect operation. This indicates hazardous situation which, if not heeded, may result in minor or moderate injury to you or others, or may result in machine damage.  This indicates a potentially hazardous situation which, if not heeded, could result in death or serious injury to you or others.   Federal Law restricts this device to sale by or the order of a radiologist  or any other practitioners licensed by the law of the state in which that person practices to use or order the use of the device.
  R-M-011-121030      Contents Introduction ........................................................................................ 1 Intended use ...................................................................................... 1 Product features ................................................................................ 2 Product components ......................................................................... 3 Components Description................................................................... 5 Warning .............................................................................................. 8 Caution ............................................................................................. 10 Marking and labeling symbols ........................................................ 12 Notes for Using the Detector ............................................................ 14 Preparing .......................................................................................... 14 Handling ........................................................................................... 15 Handle Assembly ............................................................................. 15 Before Exposure .............................................................................. 17 During Exposure .............................................................................. 17 Limit of Load .................................................................................... 19 Disinfection and Cleaning ............................................................... 20 Others ............................................................................................... 20 Technical Features ........................................................................... 21 Mechanical Features ........................................................................ 21 Electrical Features ........................................................................... 21 Environmental requirement ............................................................. 24 PC Requirements ............................................................................. 24 Installation........................................................................................ 25 General interface ............................................................................. 25 Connection ....................................................................................... 25 IP set up ........................................................................................... 31 Using Web Manager (IP, SSID Change / Upgrade FW) ................... 32 Calibration ........................................................................................ 36 General Principle ............................................................................. 36 Calibration ........................................................................................ 38 Image Acquisition Test ..................................................................... 43 Program setup ................................................................................. 43 Get Image ......................................................................................... 44 View Images ..................................................................................... 45 Operation .......................................................................................... 50 Recommend ..................................................................................... 50 Switching power on / off .................................................................. 50 Storage ............................................................................................. 50 Information ....................................................................................... 51 Safety standard ................................................................................ 51 Electromagnetic Compatibility Information .................................... 53 FCC Notice ....................................................................................... 53 Maintenance ..................................................................................... 60 Maintenance ..................................................................................... 60 Cleaning ........................................................................................... 60 Inspection ........................................................................................ 61 Disposal or Recycling...................................................................... 62 Appendix ........................................................................................... 63
 1  R-M-011-121030    Introduction  Overview The 1012WCA is  a wireless digital X-ray flat panel detector that can  generate  images  of  any  part  of  the  body.  The  wireless LAN((IEEE  802.11a/g/n)  communication  feature  improves  the operability,  and  high-speed  processing.  This  X-ray  imaging system  consists  of  a  scintillator  directly  coupled  to  an  a-Si  TFT sensor. It makes high-resolution, high-sensitive digital images.  Intended use For U.S.A. 1012WCA  Digital  Flat  Panel  X-Ray  Detector  is  indicated  for digital  imaging  solution  designed  for  human  anatomy  including head,  neck,  spinal  column,  arm,  leg  and  peripheral  (foot,  hand, wrist,  fingers,  etc.).  It  is  intended  to  replace  film  based radiographic  diagnostic  systems  and  provide  a  case  diagnosis and  treatment  planning  for  physicians  and  other  health  care professionals. Not to be used for mammography. For European Union This  device  provides  digital  X-ray  imaging  for  diagnosis  of disease,  injury,  or  any  applicable  health  problem.  The  image  is obtained  as  the  result  of  imaging  X-rays  passed  through  the human  body  with  an  X-ray  flat  panel  detector  and  importing  a digital signal output from the detector into the image processor.
 2  R-M-011-121030      Product features       Wi-Fi (802.11a/g/n)   Based on  a-Si TFT active matrix   Compact (18mm thickness) and light weight (Typ. 3.15kg)   Limiting Resolution : 3.9 lp/mm   16-bit digital output   Easy integration
 3  R-M-011-121030    Product components   Medical Image Processing Unit Photo Item Part Name Quantity  Detector  SD1012WCA  1 Handle  -  1  Battery pack  RB37WH  2 Battery charger  RC120W  1  Charger adapter PMP120-13-3  1  AGI  RA001A  1
 4  R-M-011-121030        Cables  Item Part Name Length Quantity  Link cable VRH076A  6m  1  P-Interface cable VRH017A  8m  1  USB cable  (A to B) VRH078A  1.8m  1  AC Power cord  VRH018A/019A  1.8m  1    Installation CD o  Manual o  Detector Library    Option o  Additional Battery o  Charger for the Car o  Cover Bag o  AP package
 5  R-M-011-121030    Components Description The detector is designed to capture radiographic images. Captured images are transmitted to PC using the wireless/wired connection. (1)  Detector             1.  LED indicator: The LED indicates the state of detector. i.  Green on : Power on. ii.  Blue blinking : Wireless Connection. iii.  Orange on : Low battery. 2.  Handle  :  Hold  this  handle  when  carrying  the  sensor  unit.  It  is removable. (Horizon / Vertical)   3.  Handle unlock-lever : This is an unlock-lever to remove handle. 4.  Link cable connector : This is a connector for Wire communication
 6  R-M-011-121030      and  power  supplying.  Connect  the  detector  to  PC  and  SMPS(not provided) using Link cable. 5.  CFRP(Carbon Fiber Reinforced Plastic) : The part of the patient’s body to which an image is to be taken should be placed against this plate. 6.  Power button : Power on /  Power off switch. 7.  Shock  sensor  :  Detector  has  built-in  4  Shock  sensors.  It  detects and records impact and mishandling of fragile 8.  Battery unlock-lever : This is an unlock-lever to remove battery.  (2)  Battery & Charger   1.  Battery  : Lithium ion  battery. The number  of  times being  acquired image  is  600  images(@  cycle  time  :  15s).  The  batteries  last  2.5 hours and are rechargeable. 2.  Charger : Two port cradle type.  3.  LED indicator :  i.  Orange on : charging ii.  Green on : Charging is completed.
 7  R-M-011-121030    (3)  AGI  1.  Trigger  connector:  This  is  a  connector  for  synchronization between detector and generator. Connect the AGI to the generator using P-interface cable.. 2.  USB  connector:  This  is  a  connector  for  communication  between AGI and PC. Connect the AGI to the PC using USB cable.
 8  R-M-011-121030       Warning Environment of Use and Storage Follow the specified process of operational instructions written in this manual for the safety of the users and patients. Does  not  use  or  store  the  instrument  near  any  flammable chemicals such as thinner, benzene, etc. Also, this instrument is not a category AP or APG equipment. If chemicals are spilled or evaporate,  it may  result  in fire or  electric  shock  through  contact with  electric  parts  inside  the  instruments.  Also,  some disinfectants  are  flammable.  Be  sure  to  take  care  when  using them. Connection Do not connect the instrument with anything other than specified. Otherwise, it may result in fire or electric shock. Handling Always be sure to keep checking the condition of the system and the  patient  to  ensure  they  are  normal  during  the  use  of  the instrument.  If  any problem  is  found,  take  appropriate  measures, such as stopping the operation of the instrument, as required. Never disassemble or modify the product as  it may result  in fire or  electric  shock.  Also,  since  the  instrument  incorporates  parts that  may  cause  electric  shocks  and  other  hazardous  parts, touching them may cause death or serious injury. Do  not  hit  or  drop  the  instrument.  The  instrument  may  be damaged  if  it  receives  a  strong  jolt,  which  may  result  in  fire  or electric shock if the instrument is used without being repaired.                                    Make sure                  to observe the following right
 9  R-M-011-121030    When Problem Occurs Should  any  of  the  following  occur,  immediately  turn  OFF  the power  of  each  instruments,  unplug  the  power  supply  cord  from the AC outlet, and contact Rayence representative or distributor.    When there is smoke, odd smell or abnormal sound.    When liquid  has been spilled  into the  instrument or  a  metal object has entered through an opening.    When the instrument has been dropped and it is damaged.  Maintenance and Inspection For  safety  reasons,  be  sure  to  turn  OFF  the  power  of  each instrument  when  the  inspections  indicated  in  this  manual  are going to be performed. Otherwise, it may result in electric shock. When the instrument is going to be cleaned, be sure to turn OFF the power of each instrument, and unplug the power supply cord from the AC outlet.  The instrument must be repaired by a qualified engineer only. If it is  not  repaired  properly,  it  may  cause  fire,  electric  shock,  or accident.                                      Make sure                  to observe the following right.
 10  R-M-011-121030      Caution Environment of Use and Storage Do  not  install  the  instrument  in  a  location  with  the  conditions listed  below.  Otherwise,  it  may  result  in  failure  or  malfunction, cause fire or injury.   Close to facilities where water is used.   Where it will be exposed to direct sunlight.   Close to air-conditioner or ventilation equipment.   Close to heat source such as a heater.   Prone to vibration.   Insecure place.   Dusty environment.   Saline or sulfurous environment.   High temperature or humidity.   Freezing or condensation. Do  not  place  the  storage  case  in  a  location  with  the  conditions listed below.   Where  the  cable  of  the  sensor  unit  will  be  strongly  pulled when the sensor unit is put into the case, otherwise, the cable may be damaged, resulting in fire or electric shock.    Where someone might get their foot caught in the cable of the sensor unit. Otherwise they could trip over, resulting in injury
 11  R-M-011-121030    Handling Do  not  spill  liquid or  chemicals onto the  instrument  or,  in  cases where the patient is injured, allow it to become wet with blood or other body fluids, as doing so may result in fire or electric shock. In such situation, protect the instrument with disposable covering as necessary. Wipe  the  CFRP  plate  of  the  sensor  unit  with  ethanol  or glutaraldehyde solution to disinfect it each time a different patient uses the instrument, in order to prevent infection. Turn  off  the  power of  each  instrument  for  safety  when  they are not going to be used. Maintenance and Inspection For  safety  reasons,  be  sure  to  inspect  the  instrument  before using it. In addition, carry out a regular inspection at least once a year. Modifications Any changes or modifications in construction of this device which are  not  expressly  approved  by  the  party  responsible  for compliance  could  void  the  user's  authority  to  operate  the equipment.
 12  R-M-011-121030      Marking and labeling symbols Symbols  Meaning  Caution : “Attention, see instructions for use”  Manufacturer  Date of manufacture  Serial number  WEEE  :  Waste  Electrical  and  Electronic Equipment  Authorized  representative  in  the  European Community  CE  symbol  grants  the  product  compliance  to the  European  Directive  for  Medical  Devices 93/42/EEC as a class Ⅱa device. Authorized by Notified Body SGS (code no.:0120) of British  non-ionizing radiation
 13  R-M-011-121030    Labels Detector label   Battery Pack   Battery Charger
 14  R-M-011-121030      Notes for Using the Detector  Preparing Fully charge the battery pack.  Charge the  battery on the day of examination or on the previous day.   Battery  slowly  discharges  even  of  not  in  use.  The  battery pack  may  have  expired  if  it  discharges  immediately  after being  fully  charged.  You  can  purchase  an  optional  battery pack to replace an exhausted one.   Be sure to fully charge the battery before use. ※ The  battery  charger,  RC120W  is  designed  for  the dedicated battery pack. ※ When the detector will not be used for some time, remove the battery pack. Attach  the battery pack.  Align  the  claw on  the  battery pack and the groove on the battery bay. Insert the battery pack fully. Push down the battery pack. Slide the lock lever toward (lock) side and lock it.
 15  R-M-011-121030    Handling  Handle  the  instrument  carefully,  as  it  may  be  damaged  if something is hit against it, dropped, or receives a strong jolt.  Handle Assembly  ※ Insert the handle always in the same direction. B A
 16  R-M-011-121030        Handle Assembly Type 1      Handle Assembly Type 2  Insert the handle in the same direction Insert the handle in the same direction A B Please confirm lever position. When the handle is locked, the lever should be same position with the right picture.                                  If the  handle is  not locked, the detector can be dropped. Please confirm lever position. When the handle is locked, the lever should be same position with the right picture.
 17  R-M-011-121030    Before Exposure Be  sure  to  check  the  equipment  daily  and  confirm that  it  works properly. Sudden  heating  of  the  room  in  cold  areas  will  cause condensation  to  form  on  the  instrument.  In  this  case,  wait  until condensation  disappears  before  performing  exposure.  If  the instrument is used with condensation formed on it, problems may occur in the quality of the instrument. When an air-conditioner is going  to  be  used,  be  sure  to  raise/lower  the  temperature gradually  so that a difference in temperature in  the  room and  in the  instrument  does  not  occur,  to  prevent  forming  of condensation.  During Exposure Do  not  use  the  detector  near  devices  generating  a  strong magnetic field. Doing so may produce image noise or artifacts. Do not apply excessive weight to the sensor unit. Otherwise, the sensor may be damaged.
 18  R-M-011-121030        Sleep Mode/Wake up   Control “e” parameter : sleep time - 0 : disabled (default) - 1 : 1 sec - 2 : 2 sec … - 600 : 600 sec  If  you  set  sleep  time,  the  detector  goes  to  the  sleep  mode  after  “N”  sec  after acquiring image  If you want to wake up the detector, Press the power button(less than 1 second)                                  Do not  change  parameters except “Control e”
 19  R-M-011-121030    Limit of Load  Uniform load: 150 kg over the whole area of sensor window.  Local load: 100 kg on an area 40 mm in diameter.
 20  R-M-011-121030       Be sure to use the sensor unit on a flat place so it will not bend. Otherwise, the sensor may be damaged.  Disinfection and Cleaning Do not spray the detector directly with disinfectants or detergents.  Do not use anything other than neutral detergent for cleaning the cover of the instrument. Otherwise, the coating will be corroded.  Others Be  sure  to  reconnect  the  cables  to  the  proper  connectors. Otherwise, the instrument may malfunction or may be damaged.
 21  R-M-011-121030    Technical Features Mechanical Features Size  395 x 337 x 18 mm Weight  2.85 kg (incl. Battery pack) Encapsulation Material Mg Window Material  Carbon fiber plate with 1.4 mm thickness  Electrical Features Detector Sensor Type  Amorphous Silicon with TFT (Single Panel) X-ray Converter  CsI:Tl Total Pixel Number  2080 × 2560 pixels Active Pixel Area  25.9 cm × 32.0 cm Active Pixel Number  2040 × 2520 pixels Pixel Pitch  127 μm Limiting Resolution  Max. 3.9 lp/mm Energy Range  40 - 150 kV A/D Conversion  16 bits Frame Rate  240 fph  Preview Time  < 2.0 sec Charge Integration Time  upto 4 sec Max. Linear Dose  Typ. 60 μGy @ SID 1500 mm  Saturation Dose  Typ. 70 μGy @ SID 1500 mm  Data Interface - Wireless - Wire  Wi-Fi (802.11a/g/n) Gigabit Ethernet   Under RQA5 condition (70kVp, 21mmAl)   Preview time may vary by complex factors
 22  R-M-011-121030      Battery Size  232.5 x 132.8 x 7 mm Weight  Typ. 0.3 kg Input  12.5 VDC Output  11.1 VDC Charging time  Typ. 2.5 hours Capacity  Typ. 3400 mAh The number of times being acquired image  600 images  The Battery level can be displayed on the LED status of battery.  If  the  battery  level  goes  down  under  25%,  please  charge  the battery Display  Battery level  75~100 %  50~75 %  25~50 %  0~25 %  Charger Size  267.5 x 100 x 77 mm Weight  Typ. 0.86 kg Input  20 VDC Output  11.1 VDC  Adapter Size  160 x 76 x 43.7mm (cable length: 1.28m) Weight  Typ. 0.92 kg  Input  100~240 VAC, 47~63 Hz, 1.4~0.6A Output  20 VDC, Max 6.0A
 23  R-M-011-121030    Wireless Spec Standard  802.11a/g/n compliance Peak Rate  300Mbps Frequency  2.4 GHz / 5 GHz Bandwidth  20MHz/40MHz MIMO  2x2    5.15~5.25 GHz band is restricted to indoor operations only. (for FCC) 5.15~5.350 GHz band is restricted to indoor operations only. (for CE) 5.15~5.350 GHz is indoor use only. (for Japan)  ※ Recommended Maximum operable distance : 7m (From the Access Point)
 24  R-M-011-121030      Environmental requirement Item  Min.  Typ.  Max.  Unit  Note Temperature  -10    50  ℃  Humidity  10    80  % H.R.   Pressure  70    106  kPa   Shocks (Wrapping condition)  10G (25G)     16ms, 1000times, in 6directions, non-Driving Vibrations (Wrapping condition)  2G (10G)     10-150Hz, 10Sweeps, 1min/Octave, XYZ axis  ※ Regularly changed parts : Battery (warranty 6 months)  PC Requirements   Processor  :  At  least  Intel  Pentium  IV  HT  with  2.8GHz,  Intel Core Duo / Core 2 or comparable AMD Dual Core processor   At least 2 GB RAM   At  least  40  GB hard  disk for  the  software,  in addition  to  the required archive capacity for image storage   Upper that Intel® PRO/100VE Network card   Graphics card/monitor  : Resolution of at  least 1,280 x  1,024 pixels in True Color mode   OS : Windows XP, vista, 7(32/64 bit)
 25  R-M-011-121030    Installation  The  Detector  is  composed  of  sensitive  electronic  parts  and components.  It  is  recommended  to  use  the  product  in  a  clean place and to exercise caution to ensure that it is not affected by dust or liquids. It is recommended to Use a dry and soft cloth to clean the detector housing. General interface   Connection  Power Connection  A. Connect the battery pack or power cable to the equipment. B. Connect the USB cable from your PC to AGI. ※ Be sure to sure only the dedicated battery pack, RB37WH for 1012WCA.                 Portable                  Imaging                 processing unit must be installed in a way that enables the user to achieve optimal use
 26  R-M-011-121030      Wireless Communication  A. AP Router(Line sharer) setting -  SSID : Griffon -  Internal network   - IP address : 2.2.2.1   - Subnet mask : 255.255.255.0   - Dynamic IP allocation range : 2.2.2.2~2.2.2.254 - Pre-Shared Key(Password) : project302   - Authentication methods : WPAPSK  or WPA2PSK   - Password methods : TKIP/AES - AP IP : 2.2.2.1 -  Channel (Frequency) - Avoid crowded channel (Using wireless detector under crowded channel result in low performance) -  If  available,  Use  ‘Auto-Channel  Selection’  function of router to find optimal channel  B. Reception Indicator Link LED flickering Blink Speed : Slow – Low link quality     Fast – High link quality
 27  R-M-011-121030    C. Checking Link Quality - After wireless connection is established, perform ‘Get Bright’ in ‘Calibration’ tap. - Check the value named ‘Wireless Signal’ in black log screen. Wireless Signal = Link Quality (Max. 100)     The value ‘Wireless Signal’ will be displayed here
 28  R-M-011-121030      Trigger Connection A. Connect the P-interface cable to the generator     X-ray Generator Connection Connect  the  P-interface  cable  between  the  AGI  box  and  X-ray generator.  A. Wiring Mode 1 : P-interface cable mode      Make assurance doubly sure   SIGNAL RATING before connection.
 29  R-M-011-121030     Operating description   The window time can be changed. Refer to the following pictures
 30  R-M-011-121030                          Exposure                 Time(*clk)                 1333333 is designate to 0.5sec
 31  R-M-011-121030    IP set up  [My  Network  Places]  →  [Properties]  →  [Local  Area  Connection] → [Properties] → [Internet Protocol (TCP/IP)]  → [Use the following IP address] IP address : Obtain an IP address automatically                      IP address : Obtain an IP address automatically