Olympus Medical Systems RU2020 Endoscope Reprocessor User Manual GT9882 0100 fm10

Olympus Medical Systems Corp. Endoscope Reprocessor GT9882 0100 fm10

Operation Manual 5

11.3 Log management with PC543OER-Elite OPERATION MANUALCh.11Table 11.35th level (Reprocessing) & 6th level (the others): Output rangeThe “ALL” folder is created when the data of all records are saved.The “SetMonth” folder is created when the record data is saved by selecting a month.The “SetPeriod” folder is created when the record data is saved by selecting a period.The “Differential” folder is created when the record data that has not been output to the portable memory is saved.4th Level 5th Level DescriptionReprocessing – The folder for Reprocessing record.Function Air Purge The folder for air purge record.Alcohol Flush The folder for alcohol flush record.Alcohol Line Disinfection The folder for alcohol line disinfection record.ALT Self-Check The folder for ALT self-check record.Detergent Line Disinfection The folder for detergent line disinfection record.Heat LCG The folder for heat LCG record.Leaking Scope Decontamination The folder for leaking scope decontamination record.Leak Test The folder for auto and manual leak test record.Maintenance The folder for periodic maintenance and repair record.Mix LCG The folder for Mix LCG record.MRC Check Result The folder for MRC check result record.Portable Memory The folder for portable memory record.Rinse The folder for rinse record.Self Disinfection The folder for self disinfection record.WaterLine Disinfection The folder for water line disinfection record.Error – The folder for error record.ReplacementOf ConsumableItemsReplace Air Filter The folder for replace air filter record.Replace Detergent The folder for replace detergent record.Replace Gas Filter (lid) The folder for replace gas filter (lid) record.Replace Gas Filter (tank) The folder for replace gas filter (tank) record.Replace LCG The folder for replace LCG record.Replace Water Filter The folder for replace water filter record.
54411.3 Log management with PCOER-Elite OPERATION MANUALCh.11(b) File typeThe record data in the CSV (Comma-Separated Values) format is saved.(c) File namesThe files are named based on the range selected when data is saved in the portable memory, the type of record, etc.Figure 11.18 ExampleSelected period Reprocessor nameReprocessor serial numberProcess namePortable memory save count
11.3 Log management with PC545OER-Elite OPERATION MANUALCh.11Table 11.4Items constituting the file namesItem DescriptionSelected period This item indicates the output mode selected when downloading the records.ALL (20170601) ALL data output setting. The figures inside the parentheses indicate the date of downloading.20170501 – 20170615 Selected period (when more than one day is specified)20170501 Selected period (when one day is specified)201706 Selected month (displayed according to the date display format setting)Diff (20170601) Data not output to portable memoryReprocessor name OER-EliteReprocessor serial number1234567Data type Type of the saved records.Displayed record namesReprocessing Reprocessing recordLeakTest Leak test recordReplaceLCG Disinfectant replacement recordMRCCheckResult MRC check result recordHeatLCG Heat LCG recordMixLCG Mix LCG recordReplaceWaterFilter Water filter replacement recordReplaceAirFilter Air filter replacement recordReplaceGasFilter(lid) Gas filter (lid) replacement recordReplaceGasFilter(tank) Gas filter (tank) replacement recordReplaceDetergent Detergent replacement recordWaterLineDisinfection Water line disinfection recordAlcoholLineDisinfection Alcohol line disinfection recordDetergentLineDisinfection Detergent line disinfection recordRinse Rinse recordAirPurge Air purge recordAlcoholFlush Alcohol flush recordSelfDisinfection Self disinfection recordPortableMemory USB memory output recordError Error recordALTSelf-Check Auto Leak Test Self-check recordLeakingScopeDecontamination Leaking scope decontamination recordMaintenance Periodic Maintenance and repair recordsPortable memory save countTotal number of portable memory saving operations in 5 digits.Range: 00001 to 65535. The count after 65535 becomes 00001.
54611.3 Log management with PCOER-Elite OPERATION MANUALCh.11NOTEIf the same filename is already there in the folder, a number is appended to the end of the filename of the new file (Example: 2120_02-01_1700_OER-Elite_2700000_Reprocessing_00010(2).csv). This number is appended to prevent the edited file in the PC from being overwritten.(d) Information in each fileEach file is saved in a portable memory in the CSV format.Every file is composed of the header, data and footer sections. The header and footer sections are common to all record data. The data section varies depending on the data type. For detail of record data, refer to Figure 11.19. There is no color in each column of record data.Figure 11.19Header section Data section Memo sectionFooter sectionDescriptionMinor itemMajor item
11.3 Log management with PC547OER-Elite OPERATION MANUALCh.11Table 11.5Header sectionLine No. Information Example1 Reprocessor name OER-EliteReprocessor serial number 12345672 File output date 06/01/2017File output time 9:00:30 PM3 Output mode AllSet periodSet monthDifferential: Blank.4 Output range Blank(When output mode is All)2017/05/01 – 2017/05/31(When output mode is Set period and more than one day is specified)2017/05/01(When output mode is Set period and one day is specified)2017/05(When output mode is Set month)5 – 7 Blanks8 Process name ReprocessingLeak TestReplace LCGMRC Check ResultHeat LCGMix LCGReplace Water FilterReplace Air FilterReplace Gas Filter (lid)Replace Gas Filter (tank)Replace DetergentWater Line DisinfectionAlcohol Line DisinfectionDetergent Line DisinfectionRinseAir PurgeAlcohol FlushSelf Disinfection & Water SamplingPortable MemoryErrorALT Self-CheckLeaking Scope DecontaminationMaintenance
54811.3 Log management with PCOER-Elite OPERATION MANUALCh.11Data section• Reprocessing recordsMajor item Minor item Description ExamplesReprocessing Record # Record number 5000, 4999, 4998, …Reprocess Result Results of reprocessing CompletedE XXX (in case of error)Date/Time Started (Day) Reprocessing start date 06/01/2017Started (Time) Reprocessing start time 15:30:30Completed (Day) Reprocessing completion date 06/01/2017Completed (Time) Reprocessing completion time 16:00:30ID/Name Scope No Endoscope number 1 (First endoscope)2 (Second endoscope)Scope Model Model number of endoscope GIF-H190Master (Master card input)– (Not input)Scope ID Endoscope serial number 12800542– (Not input)User ID ID number of user installing the endoscope(s)12345678– (Not input)User Name Name of user installing the endoscope(s) UserMaster (Master card input)– (Not input)Physician ID ID number of physician – (Not input)12345678Physician Name Name of physician – (Not input)PhysicianPatient ID ID number of patient – (Not input)12345678Patient Name Name of patient *****– (Not input)PatientProcedure ID ID number of procedure – (Not input)0123456Procedure Name Name of procedure – (Not input)ERCPUser3 ID ID number of user removing the endoscope(s)– (Not input)1User3 Name Name of user removing the endoscope(s) – (Not input)User
11.3 Log management with PC549OER-Elite OPERATION MANUALCh.11Table 11.6Program Info. Cln time Cleaning time (minutes) 3Dis Time Disinfection time (minutes) 7ALT Auto leak test setting ONOFFHeat LCG Heat LCG setting ONAlcohol Flush Alcohol flush setting ONChannel Monitor Channel Monitoring setting FullPartialProgram No. Program number 1, 2, 3, 4Check Leak Test (Sink) Input result of leak test before manual cleaning– (Not input)PassFailManual Cleaning Input result of manual cleaning – (Not input)DoneFunction Result Leak Test Result Result of leak test during the reprocessing PassFailLCG Info. Expiry Expiry of disinfectant solution 08/30Lot # Lot number of the disinfectant solution AA123MRC Check Result Result of MRC check – (Not input)FailPassTemp Disinfectant solution temperature 20Cycle # Usage count of the disinfectant solution – (Not input)1Days Number of days that have elapsed since preparation of the disinfectant solution– (Not input)1Detergent Info. Expiry Expiry of the detergent solution – (Not input)09/15Lot # Lot number of the detergent solution – (Not input)DD1234AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Total Count Total number of reprocessing processes performed1Memo Memorandum (Free input by user) – (Not input)Free input by userMajor item Minor item Description Examples
55011.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.7• Leak test recordsMajor item Minor item Description ExamplesLeak Test Record # Record number 5000, 4999, 4998, …Process Result Result of leak test CompleteE XXX (in case of error)Option Method of leak test Auto (Auto leak test)Manual (Manual leak test)Date/Time Started (Day) Start date of leak test 2017/06/01Started (Time) Start time of leak test 15:30:30Completed (Day) Completion date of leak test 2017/06/01Completed (Time) Completion time of leak test 15:45:10ID/Name Scope No Endoscope number 1 (First endoscope)2 (Second endoscope)Scope Model Model number of endoscope GIF-H190Master (Master card input)– (Not input)Scope ID Endoscope serial number 12800542– (Not input)User ID ID number of user installing the endoscope(s)1– (Not input)User Name Name of user installing the endoscope(s) 1Master (Master card input)– (Not input)Function Result Leak Test Result Result of leak test PassFail– (in case of error)AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Total Leak Test # Total number of performing leak test 10Memo Memorandum (Free input by user) – (Not input)Free input by user
11.3 Log management with PC551OER-Elite OPERATION MANUALCh.11Table 11.8• LCG replacement recordsMajor item Minor item Description ExamplesReplace LCG Record # Record number 5000, 4999, 4998, …Date/Time Loaded (Day) Completion date of preparation of disinfectant solution06/01/2017Loaded (Time) Completion time of preparation of disinfectant solution08:45:30Drained (Day) Completion date of discharge of disinfectant solution06/05/2017Drained (Time) Completion time of draining disinfectant solution17:30:30ID/Name (Load LCG)User ID ID number of user who prepared disinfectant solution– (Not input)1User Name Name of user who prepared disinfectant solution– (Not input)UserID/Name (Drain LCG)User ID ID of user who drained disinfectant solution – (Not input)1User Name Name of user who drained disinfectant solution– (Not input)UserOption Method that discharged the disinfectant solutionDrainCollectLCG Info. Cycle # Usage count of the disinfectant solution – (Not input)20Days Number of days that have elapsed since preparation of the disinfectant solution– (Not input)5Lot # Lot number of disinfectant solution – (Not input)123456789Expiry Expiration date of disinfectant solution – (Not input)2013/09AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
55211.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.9MRC Check recordsMajor item Minor item Description ExamplesMRC Check Result Record # Record number 5000, 4999, 4998, …Date/Time Completed (Day) Completion date of MRC check 06/01/2017Completed (Time) Completion time of MRC check 14:00:30ID/Name User ID ID number of user who performed MRC check– (Not input)1User Name Name of user who performed MRC check – (Not input)UserLCG Info. Expiry Expiration date of disinfectant solution – (Not input)09/2013Lot # Lot number of disinfectant solution – (Not input)123456789MRC Check Result Result of MRC check FailPassTemp Disinfectant solution temperature – (Not input)20Cycle # Disinfectant cycle count – (Not input)1Days Disinfectant day count – (Not input)5AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
11.3 Log management with PC553OER-Elite OPERATION MANUALCh.11Table 11.10Heat LCG recordsMajor item Minor item Description ExamplesHeat LCG Record # Record number 5000, 4999, 4998, …Process Result Result of Heat LCG CompleteE XXX (in case of error)Option Method of Heat LCG ManualTimerDate/Time Started (Day) Start date of Heat LCG 06/01/2017Started (Time) Start time of Heat LCG 08:30:30Completed (Day) Completion date of Heat LCG 06/01/2017Completed (Time) Completion time of Heat LCG 10:00:30ID/Name User ID ID number of user performed Heat LCG – (Not input)1User Name Name of user performed Heat LCG – (Not input)UserTimer Info. Setting Date Setting date of Heat LCG –06/01/2017 (In case Heat LCG Timer)Setting Time Setting time of Heat LCG –10:00 (In case Heat LCG Timer)LCG Info. Temp Disinfectant solution temperature 20AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
55411.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.11Table 11.12Mix LCG recordsMajor item Minor item Description ExamplesMix LCG Record # Record number 5000, 4999, 4998, …Process Result Result of mix LCG CompleteE XXX (in case of error)Date/Time Started (Day) Start date of mix LCG 06/01/2017Started (Time) Start time of mix LCG 15:30:30Completed (Day) Completion date of mix LCG 06/01/2017Completed (Time) Completion time of mix LCG 15:45:10ID/Name User ID ID number of user performed mix LCG – (Not input)1User Name ID number of user performed mix LCG – (Not input)UserAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by userWater filter replacement recordsMajor item Minor item Description ExamplesReplace Water FilterRecord # Record number 5000, 4999, 4998, …Process Result Result of Water filter replacement CompleteE XXX (in case of error)Date/Time Started (Day) Start date of Water filter replacement 06/01/2017Started (Time) Start time of Water filter replacement 08:30:00Replaced (Day) Completion date of Water filter replacement 06/01/2017Replaced (Time) Completion time of Water filter replacement 09:15:30ID/Name User ID ID number of user who replaced water filter – (Not input)1User Name Name of user who replaced water filter – (Not input)UserFilter Info. Lot # Lot number of water filter – (Not input)2017/05AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
11.3 Log management with PC555OER-Elite OPERATION MANUALCh.11Table 11.13Table 11.14Air filter replacement recordsMajor item Minor item Description ExamplesReplace Air Filter Record # Record number 5000, 4999, 4998, …Date/Time Replaced (Day) Completion date of air filter replacement 06/01/2017Replaced (Time) Completion time of air filter replacement 11:30:45ID/Name User ID ID number of user who replaced air filter – (Not input)1User Name Name of user who replaced air filter – (Not input)UserFilter Info. Lot # Lot number of air filter – (Not input)987654321AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by userGas filter (lid) replacement recordsMajor item Minor item Description ExamplesReplace Gas Filter (lid)Record # Record number 5000, 4999, 4998, …Date/Time Replaced (Day) Completion date of gas filter replacement 06/01/2017Replaced (Time) Completion time of gas filter replacement 12:15:30ID/Name User ID ID number of user who replaced gas filter – (Not input)1User Name Name of user who replaced gas filter – (Not input)UserFilter Info. Lot # Lot number of gas filter – (Not input)2017/05AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
55611.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.15Table 11.16Gas filter (tank) replacement recordsMajor item Minor item Description ExamplesReplace Gas Filter (tank)Record # Record number 5000, 4999, 4998, …Date/Time Replaced (Day) Completion date of gas filter replacement 06/01/2017Replaced (Time) Completion time of gas filter replacement 15:30:30ID/Name User ID ID number of user who replaced gas filter – (Not input)1User Name Name of user who replaced gas filter – (Not input)UserFilter Info. Lot # Lot number of gas filter – (Not input)2017/05AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by userDetergent replacement recordsMajor item Minor item Description ExamplesReplace Detergent Record # Record number 5000, 4999, 4998, …Date/Time Replaced (Day) Completion date of detergent replacement 06/01/2017Replaced (Time) Completion time of detergent replacement 09:15:30ID/Name User ID ID number of user who replaced detergent – (Not input)1User Name Name of user who replaced detergent – (Not input)UserDetergent info. Expiry Expiration date of detergent – (Not input)09/2017Lot # Lot number of detergent – (Not input)0987654321AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
11.3 Log management with PC557OER-Elite OPERATION MANUALCh.11Table 11.17Water line disinfection recordsMajor item Minor item Description ExamplesWater Line DisinfectionRecord # Record number 5000, 4999, 4998, …Process Result Result of water line disinfection CompleteE XXX (in case of error)Date/Time Started (Day) Start date of water line disinfection 06/01/2017Started (Time) Start time of water line disinfection 08:30:30Completed (Day) Completion date of water line disinfection 06/01/2017Completed (Time) Completion time of water line disinfection 10:00:30ID/Name User ID ID number of user performed water line disinfection– (Not input)1User Name Name of user performed water line disinfection– (Not input)UserLCG Info. Expiry Expiration date of disinfectant solution – (Not input)09/2013Lot # Lot number of disinfectant solution – (Not input)123456789MRC Check Result Result of MRC check – (Not input)FailPassTemp Disinfectant solution temperature – (Not input)20Cycle # Disinfectant cycle count – (Not input)1Days Disinfectant day count – (Not input)5AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
55811.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.18Alcohol line disinfection recordsMajor item Minor item Description ExamplesAlcohol Line DisinfectionRecord # Record number 5000, 4999, 4998, …Process Result Result of alcohol line disinfection CompleteE XXX (in case of error)Date/Time Started (Day) Start date of alcohol line disinfection 06/01/2017Started (Time) Start time of alcohol line disinfection 08:30:30Completed (Day) Completion date of alcohol line disinfection 06/01/2017Completed (Time) Completion time of alcohol line disinfection 10:00:30ID/Name User ID ID number of user performed alcohol line disinfection– (Not input)1User Name Name of user performed alcohol line disinfection– (Not input)UserProgram Dis Time Disinfection time – (Not input)10LCG Info. MRC Check Result Result of MRC check – (Not input)FailPassAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
11.3 Log management with PC559OER-Elite OPERATION MANUALCh.11Table 11.19Detergent line disinfection recordsMajor item Minor item Description ExamplesDetergent Line DisinfectionRecord # Record number 5000, 4999, 4998, …Process Result Result of detergent line disinfection CompleteE XXX (in case of error)Date/Time Started (Day) Start date of detergent line disinfection 06/01/2017Started (Time) Start time of detergent line disinfection 08:30:30Completed (Day) Completion date of detergent line disinfection06/01/2017Completed (Time) Completion time of detergent line disinfection 10:00:30ID/Name User ID ID number of user performed detergent line disinfection– (Not input)1User Name Name of user performed detergent line disinfection– (Not input)UserProgram Dis Time Disinfection time – (Not input)10LCG Info. MRC Check Result Result of MRC check – (Not input)FailPassAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
56011.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.20Rinse recordsMajor item Minor item Description ExamplesRinse Record # Record number 5000, 4999, 4998, …Process Result Result of rinse CompleteE XXX (in case of error)Date/Time Started (Day) Start date of rinse 06/01/2017Started (Time) Start time of rinse 08:30:30Completed (Day) Completion date of rinse 06/01/2017Completed (Time) Completion time of rinse 10:00:30ID/Name User ID ID number of user performed rinse – (Not input)1User Name Name of user performed rinse – (Not input)UserAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
11.3 Log management with PC561OER-Elite OPERATION MANUALCh.11Table 11.21Table 11.22Air purge recordsMajor item Minor item Description ExamplesAir Purge Record # Record number 5000, 4999, 4998, …Process Result Result of air purge CompleteE XXX (in case of error)Date/Time Started (Day) Start date of air purge 06/01/2017Started (Time) Start time of air purge 08:30:30Completed (Day) Completion date of air purge 06/01/2017Completed (Time) Completion time of air purge 10:00:30ID/Name User ID ID number of user performed air purge – (Not input)1User Name Name of user performed air purge – (Not input)UserAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by userAlcohol flush recordsMajor item Minor item Description ExamplesAlcohol flush Record # Record number 5000, 4999, 4998, …Process Result Result of alcohol flush CompleteE XXX (in case of error)Date/Time Started (Day) Start date of alcohol flush 06/01/2017Started (Time) Start time of alcohol flush 08:30:30Completed (Day) Completion date of alcohol flush 06/01/2017Completed (Time) Completion time of alcohol flush 10:00:30ID/Name User ID ID number of user performed alcohol flush – (Not input)1User Name Name of user performed alcohol flush – (Not input)UserAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
56211.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.23Self disinfection and water sampling recordsMajor item Minor item Description ExamplesSelf Disinfection & Water SamplingRecord # Record number 5000, 4999, 4998, …Process Result Result of self-disinfection and water samplingCompleteE XXX (in case of error)Water Sampling Whether water sampling was performed or not.– (Not input)DoneDate/Time Started (Day) Start date of self-disinfection 06/01/2017Started (Time) Start time of self-disinfection 08:30:30Completed (Day) Completion date of self-disinfection 06/01/2017Completed (Time) Completion time of self-disinfection 10:00:30ID/Name User ID ID number of user performed self-disinfection– (Not input)1User Name Name of user performed self-disinfection – (Not input)UserLCG Info. Expiry Expiration date of disinfectant solution – (Not input)09/2013Lot # Lot number of disinfectant solution – (Not input)123456789MRC Check Result Result of MRC check – (Not input)FailPassTemp Disinfectant solution temperature – (Not input)20Cycle # Disinfectant cycle count – (Not input)1Days Disinfectant day count – (Not input)5AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
11.3 Log management with PC563OER-Elite OPERATION MANUALCh.11Table 11.24Portable memory recordsMajor item Minor item Description ExamplesPortable Memory Record # Record number 5000, 4999, 4998, …Process Result Result of portable memory output CompleteE XXX (in case of error)Date/Time Started (Day) Start date of portable memory output 06/01/2017Started (Time) Start time of portable memory output 08:30:30Completed (Day) Completion date of portable memory output 06/01/2017Completed (Time) Completion time of portable memory output 10:00:30ID/Name User ID ID number of user performed portable memory output– (Not input)1User Name Name of user performed portable memory output– (Not input)UserAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
56411.3 Log management with PCOER-Elite OPERATION MANUALCh.11Error recordsMajor item Minor item Description ExamplesError Record # Record number 5000, 4999, 4998, …Error code Error code of error – (Not input)EXXX (in case of error)Date/Time Started (Day) Start date of rinse 06/01/2017Started (Time) Start time of rinse 08:30:30Canceled (Day) Date of occurrence of error 06/01/2017Canceled (Time) Time of occurrence of error 10:00:30ID/Name Scope No Endoscope number 1 (First endoscope)2 (Second endoscope)Scope Model Model number of endoscope GIF-H190– (Not input)Scope ID Endoscope serial number 12800542– (Not input)User ID ID number of user installing the endoscope(s)– (Not input)1User Name Name of user installing the endoscope(s) – (Not input)UserPhysician ID ID number of physician – (Not input)1Physician Name Name of physician – (Not input)DoctorPatient ID ID number of patient – (Not input)1Patient Name Name of patient – (Not input)PatientProcedure ID ID number of procedure – (Not input)0123456Procedure Name Name of procedure – (Not input)ERCPUser3 Name Name of user removing the endoscope(s) – (Not input)UserUser3 ID ID number of user removing the endoscope(s)– (Not input)1
11.3 Log management with PC565OER-Elite OPERATION MANUALCh.11Program Info. Cln time Cleaning time (min.) 3Dis Time Disinfection time (min.) 7ALT Auto leak test setting – (Not input)ONOFFHeat LCG Heat LCG setting – (Not input)ONOFFAlcohol Flush Alcohol flush setting – (Not input)ONOFFChannel Monitor Channel Monitoring setting – (Not input)Full,PartialProgram No. Program number – (Not input)1, 2, 3, 4Check Leak Test (Sink) Input result of leak test before manual cleaning– (Not input)PassFailManual Cleaning Input result of manual cleaning –DoneFunction Result Leak Test Result Result of leak test during the reprocessing – (Not input)FailPassLCG Info. Expiry Expiration date of disinfectant solution – (Not input)2017/05Lot # Lot number of disinfectant solution – (Not input)123456789MRC Check Result Result of MRC check – (Not input)FailPassTemp Disinfectant solution temperature – (Not input)20Cycle # Disinfectant cycle count – (Not input)1Days Disinfectant day count – (Not input)1Detergent Info. Expiry Expiry of the detergent solution – (Not input)09/15Lot # Lot number of the detergent solution – (Not input)DD1234Major item Minor item Description Examples
56611.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.25ErrorDetails Process Name Process in which error is generated Cleaning, Heat LCG, etc.Process No. Number associated with the process (number is hexadecimal)– (Not input)0001, 0002Process Time Elapsed time of the process – (Not input)00:03:00Error Information 1 – 38Error value 1 – 38 – (Not input)PassFailOnOffNo ConnectConnectUnknownFlow SideWater SideMovingActual measurement value of each error(e.g., decimal place from first to third)AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Total Count Total accumulated of reprocessing process count4000Memo Memorandum (Free input by user) – (Not input)Free input by userMajor item Minor item Description Examples
11.3 Log management with PC567OER-Elite OPERATION MANUALCh.11Periodic Maintenance and RepairMajor item Minor item Description ExamplesMaintenance Record # Record number  5000, 4999, 4998, …Date/Time Working (Day) Working date of Maintenance 06/01/2017Working (Time) Working start time of Maintenance 8:30:30Service Program Type Detail for working Periodic Maintenance RepairPerson Working person name – (Not input), Person nameScheduled PM CycleNext Periodic Maintenance date 25000Scheduled PM DateNumber of Periodic Maintenance 06/01/2018Maintenance Total CountNumber of Periodic Maintenance 0AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number  1234567Total Count Total number of reprocessing processes performed1Parts Operation Count/TimeWater Supply Valve Water Supply Valve count  1Channel Valve A1 Channel Valve A1 count 1Channel Valve B1 Channel Valve B1 count 1Channel Valve C1 Channel Valve C1 count 1Channel Valve D1 Channel Valve D1 count 1Channel Valve A2 Channel Valve A2 count 1Channel Valve B2 Channel Valve B2 count 1Channel Valve C2 Channel Valve C2 count 1Channel Valve D2 Channel Valve D2 count 1Bypass Valve Bypass Valve count 1Adjust Valve Adjust Valve count 1Relief Shutoff Valve Relief Shutoff Valve count  1Pump Valve Pump Valve count 1ALT Valve (Front) ALT Valve (Front) count 1ALT Valve (Back) ALT Valve (Back) count 1Drain Valve Drain Valve count  1Collection Valve Collection Valve count 1Three Port Valve Three Port Valve count 1Alcohol Valve Alcohol Valve count 1Channel Pump Channel Pump count 1Channel Pump Channel Pump time (min)  1
56811.3 Log management with PCOER-Elite OPERATION MANUALCh.11Parts Operation Count/TimeDisinfectant Solution PumpDisinfectant Solution Pump count 1Disinfectant Solution PumpDisinfectant Solution Pump time (min) 1Drain Pump Drain Pump count 1Drain Pump Drain Pump time (min) 1Circulation Pump Circulation Pump count 1Circulation Pump Circulation Pump time (min) 1Alcohol Pump Alcohol Pump count 1Alcohol Pump Alcohol Pump time (min) 1Alkali Detergent PumpAlkali Detergent Pump count 1Alkali Detergent PumpAlkali Detergent Pump time (min) 1Compressor Compressor count 1Compressor Compressor time (min) 1Ultrasonic TransducerUltrasonic Transducer count 1Ultrasonic TransducerUltrasonic Transducer time (min) 1Basin Heater Basin Heater count 1Basin Heater Basin Heater time (min) 1Tank Heater Tank Heater count 1Tank Heater Tank Heater time (min) 1Solenoid Lock for LidSolenoid Lock for Lid count 1Solenoid Lock for CassetteSolenoid Lock for Cassette count 1Print Print count 1Start Button Start Button count 1Stop Button Stop Button count 1Heat LCG Timer ButtonHeat LCG Timer Button count 1Touch Screen OperationTouch Screen Operation count 1ALT Function ALT Function count 1MLT Function MLT Function count 1MLT Function MLT Function time (min) 1ALT Self-Check ALT Self-Check count 1LCD Backlight LCD Backlight time (min) 1Battery Working Battery Working Days 1Barcode Scan Barcode Scan count 1Major item Minor item Description Examples
11.3 Log management with PC569OER-Elite OPERATION MANUALCh.11Parts Replacement Water Supply Valve With or without replacement for water supply valve0, 1Channel Valve 1 With or without replacement for Channel Valve 10, 1Channel Valve 2 With or without replacement for Channel Valve 20, 1Bypass Valve With or without replacement for Bypass Valve0, 1Adjust Valve With or without replacement for Adjust Valve 0, 1Relief Shutoff Valve With or without replacement for Relief Shutoff Valve0, 1Pump Valve With or without replacement for Pump Valve 0, 1ALT Valve (Front) With or without replacement for ALT Valve (Front)0, 1ALT Valve (Back) With or without replacement for ALT Valve (Back)0, 1Switching Valve With or without replacement for Switching Valve0, 1Three Port Valve With or without replacement for Three Port Valve0, 1Alcohol Valve With or without replacement for Alcohol Valve0, 1Channel Pump  With or without replacement for Channel Pump 0, 1Disinfectant Solution PumpWith or without replacement for Disinfectant Solution Pump0, 1Drain Pump With or without replacement for Drain Pump 0, 1Circulation Pump With or without replacement for Circulation Pump0, 1Alcohol Pump With or without replacement for Alcohol Pump0, 1Alkali Detergent PumpWith or without replacement for Alkali Detergent Pump0, 1Compressor With or without replacement for Compressor 0, 1Ultrasonic TransducerWith or without replacement for Ultrasonic Transducer0, 1Basin Heater With or without replacement for Basin Heater 0, 1Tank Heater With or without replacement for Tank Heater 0, 1Solenoid Lock for Lid With or without replacement for Solenoid Lock for Lid 0, 1Solenoid Lock for CassetteWith or without replacement for Solenoid Lock for Cassette0, 1Printer With or without replacement for Printer 0, 1Major item Minor item Description Examples
57011.3 Log management with PCOER-Elite OPERATION MANUALCh.11Table 11.26Parts Replacement Panel Switch With or without replacement for Panel Switch 0, 1Touch Screen With or without replacement for Touch Screen0, 1ALT With or without replacement for ALT 0, 1LCD With or without replacement for LCD 0, 1Battery With or without replacement for Battery 0, 1Barcode Scanner With or without replacement for Barcode Scanner0, 1Major item Minor item Description Examples
11.3 Log management with PC571OER-Elite OPERATION MANUALCh.11Table 11.27ALT self check recordsMajor item Minor item Description ExamplesALT Self-Check Record # Record number 5000, 4999, 4998, …Process Result Result of ALT self check CompleteE XXX (in case of error)Option “Periodic (Auto)” or “Manual” is represented Periodic (Auto), ManualAtmospheric PressureAtmospheric pressure during ALT self check – (Not input)101.3 kPaDate/Time Started (Day) Start date of ALT self check 06/01/2017Started (Time) Start time of ALT self check 08:30:30Completed (Day) Completion date of ALT self check 06/01/2017Completed (Time) Completion time of ALT self check 10:00:30ID/Name User ID ID number of user performed ALT self check – (Not input)1User Name Name of user performed ALT self check – (Not input)UserFunction Result Self-Check Result Result of ALT self check – (Not input)PassFailAER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Memo Memorandum (Free input by user) – (Not input)Free input by user
57211.3 Log management with PCOER-Elite OPERATION MANUALCh.11Leaking scope decontamination recordsMajor item Minor item Description ExamplesLeaking Scope DecontaminationRecord # Record number 5000, 4999, 4998, …Process Result Result of leaking scope decontamination CompleteE XXX (in case of error)Date/Time Started (Day) Start date of leaking scope decontamination 06/01/2017Started (Time) Start time of leaking scope decontamination 08:30:30Completed (Day) Completion date of leaking scope decontamination06/01/2017Completed (Time) Completion time of leaking scope decontamination10:00:30ID/Name Scope Model Model number of endoscope. GIF-H190Master (Master card input)– (Not input)Scope ID Endoscope serial number 12800542– (Not input)User ID ID number of user performed portable memory output– (Not input)1User Name Name of user performed portable memory output– (Not input)UserPhysician ID ID number of physician – (Not input)1Physician Name Name of physician – (Not input)PhysicianPatient ID ID number of patient – (Not input)1Patient Name Name of patient – (Not input)PatientProcedure ID ID number of procedure – (Not input)0123456Procedure Name Name of procedure – (Not input)ERCPUser 2 Name Name of user removing the endoscope – (Not input)UserUser 2 ID ID number of user removing the endoscope – (Not input)1Check Leak Test (Sink) Input result of leak test before manual cleaning– (Not input)PassFailManual Cleaning Input result of manual cleaning – (Not input)Done
11.3 Log management with PC573OER-Elite OPERATION MANUALCh.11Table 11.28Managing records on PCCAUTIONThe reprocessing data copied and saved in the “BackUp” folder in the PC becomes the master data. Do not alter the master data.NOTE• For the operating environment of the portable memory, see the instructions provided with it. To manage the record at the facility the use of a PC installed with application software such as spreadsheet software that enables editing of a CSV file is recommended.• If a file exists in the “Work” folder in the portable memory from which data is downloaded, a number is appended to the end of the filename (Example: 2120_02-01_1700_OER-Elite_2700000_Reprocessing_00010(2).csv). This number is appended to prevent the edited file in the PC from being overwritten.• After transferring data in the PC, delete the “OER-Elite” folder in the portable memory. Otherwise, there would be a possibility that a file having a name identical to a file already existing in the “Work” folder in the PC is saved with the same name except for the number appended to the end of the filename.LCG Info. Expiry Expiration date of disinfectant solution – (Not input)07/01/2017Lot # Lot number of disinfectant solution – (Not input)123456789MRC Check Result Result of MRC check – (Not input)FailPassTemp Disinfectant solution temperature – (Not input)20Cycle # Disinfectant cycle count – (Not input)1Days Disinfectant day count – (Not input)1AER Info. Model Name Reprocessor name OER-EliteSerial No. Reprocessor serial number 1234567Total Count Total accumulated of reprocessing process count4000Memo Memorandum (Free input by user) – (Not input)Free input by userMajor item Minor item Description Examples
57411.3 Log management with PCOER-Elite OPERATION MANUALCh.11(a) Copying data in PCNOTEAlso, refer to the instruction manual for the PC.(b) Managing the record dataTo manage the record data, use the record data in the “Work” folder copied to the PC. The record data should be managed according to the purpose of application of data.The records can be managed as shown in the following example.NOTEThe following example uses Microsoft Excel for managing the CSV files. The management tool can be selected from the desired application software including statistical software according to the purpose and application of data.NOTEWhen editing a CSV file using application software such as statistical software, there is a risk that the operation above cannot save the edited data. It is therefore recommended to save the edited data in the dedicated file format of the software in use.When saving data in a PC for the backup purpose1Attach the portable memory containing record data to the PC.2Copy the “OER-Elite” folder in the portable memory to the PC.NOTEIf a message confirming overwriting of a folder or file is displayed when you attempt to copy data in the PC, select “Overwrite” in any case.3After completion of the copying, remove the portable memory from the PC.Example of record management (Reference)1Open the “Work” folder copied to the PC.2Open the file to be edited, check the header and data sections, and enter desired information in the edit section.3When the entry completes, save the edited file. Do not need rename file name.
11.4 Printing records575OER-Elite OPERATION MANUALCh.11The records of the following processes can be printed with the MAJ-1937 printer included in the optional MAJ-2144 printer set.• Reprocessing process: The record of reprocessing process, including errors.• Leak test processes (auto, manual): The record of auto and manual leak tests performed on the reprocessor.• Leaking scope decontamination process: The record of leaking scope decontamination process.CAUTIONPay attention to handle the portable memory and printed paper that include patient information exported from the OER-Elite.NOTEPrinting records is also possible on the record details screens. Refer to the record details in 11.2 Log display for the details.PrintingThis section is defined method for printing of daily basis. For other printing method, refer to page 232.11.4 Printing records1Turn this reprocessor and the printer ON.2Press the “Log” button on the Menu screen.Figure 11.20
57611.4 Printing recordsOER-Elite OPERATION MANUALCh.113Press the “Printout” button.Figure 11.214Select the record to be printed.No. Button Description1 Reprocessing RecordsPrints the records of the reprocessing processes.2 Leak Test Records Prints the records of the leak tests.3 Leaking Scope Decontamination RecordsPrints the records of the leaking scope decontamination.4 Menu Returns to the Menu screen.5 LCG info. Opens the disinfectant information screen.6 Cancel Returns to the previous screen.4 61235
11.4 Printing records577OER-Elite OPERATION MANUALCh.115Select whether or not the error records are to be printed.No. Button Description1 Include error results Records are printed including error records.2 Excludes error resultsRecords are printed without error records. (Only the records of successfully completed operations are printed.)3 Menu Returns to the menu.4 Return Returns to the previous screen.6Set the date of the records to be printed by pressing the “+” or “–” button.Figure 11.224132
57811.4 Printing recordsOER-Elite OPERATION MANUALCh.11NOTE• Characters “PRINT END” are printed at the end of each print session. If they are not printed, not all the records of the selected period are printed out. Restart printing from the beginning.• Records are printed according to the time.Figure 11.257Press the “Start” button. Printing starts and the touch screen displays the progress.Figure 11.238When the printing completes, the following screen appears. Press the “OK” button.Figure 11.24Progress bar
11.4 Printing records579OER-Elite OPERATION MANUALCh.11Print formatReprocessing – Normal (Separate)Figure 11.26Serial number of this reprocessorDate and time of start of processDate and time of end of processInformation of first endoscopeUser ID (remove) of the end of process of first endoscopeTotal accumulated of reprocessing process countNumber given to each record in the order of occurrenceSetting detailsLCG Info.Detergent Info.Serial number of this reprocessorDate and time of start of processDate and time of end of processInformation of second endoscopeUser ID (remove) of the end of processTotal accumulated of reprocessing process countNumber given to each record in the order of occurrenceSetting detailsLCG Info.Detergent Info.
58011.4 Printing recordsOER-Elite OPERATION MANUALCh.11Reprocessing – Normal (Combine)Figure 11.27Serial number of this reprocessorDate and time of start of processDate and time of end of processTotal accumulated of reprocessing process countNumber given to each record in the order of occurrenceSetting detailsLCG Info.Detergent Info.Information of first endoscopeInformation of second endoscopeUser ID (remove) of the end of process
11.4 Printing records581OER-Elite OPERATION MANUALCh.11Reprocessing – Error (Separate)Figure 11.28Serial number of this reprocessorDate and time of start of processDate and time of end of processInformation of first endoscopeNumber given to each record in the order of occurrenceError codeProcess number of occurrence of errorSerial number of this reprocessorDate and time of start of processDate and time of end of processInformation of second endoscopeNumber given to each record in the order of occurrenceError codeProcess number of occurrence of error
58211.4 Printing recordsOER-Elite OPERATION MANUALCh.11Reprocessing – Error (Combine)Figure 11.29Serial number of this reprocessorDate and time of start of processDate and time of end of processProcess number of occurrence of errorInformation of first endoscopeError codeNumber given to each record in the order of occurrenceInformation of second endoscope
11.4 Printing records583OER-Elite OPERATION MANUALCh.11Leak test – Normal (Separate)Figure 11.30Serial number of this reprocessorDate and time of start of processDate and time of end of processInformation of first endoscopeNumber given to each record in the order of occurrenceType of the leak testSerial number of this reprocessorDate and time of start of processDate and time of end of processInformation of second endoscopeNumber given to each record in the order of occurrenceType of the leak test
58411.4 Printing recordsOER-Elite OPERATION MANUALCh.11Leak test – Normal (Combine)Figure 11.31Serial number of this reprocessorDate and time of start of processDate and time of end of processInformation of first endoscopeNumber given to each record in the order of occurrenceType of the leak testInformation of second endoscope
11.4 Printing records585OER-Elite OPERATION MANUALCh.11Leak test – Error (Separate)Figure 11.32Serial number of this reprocessorDate and time of start of processDate and time of end of processProcess number of occurrence of errorInformation of first endoscopeError codeNumber given to each record in the order of occurrenceSerial number of this reprocessorDate and time of start of processDate and time of end of processProcess number of occurrence of errorInformation of second endoscopeError codeNumber given to each record in the order of occurrence
58611.4 Printing recordsOER-Elite OPERATION MANUALCh.11Leak test – Error (Combine)Figure 11.33Serial number of this reprocessorDate and time of start of processDate and time of end of processProcess number of occurrence of errorInformation of first endoscopeError codeNumber given to each record in the order of occurrenceInformation of second endoscope
11.4 Printing records587OER-Elite OPERATION MANUALCh.11Leaking scope decontamination – NormalFigure 11.34Serial number of this reprocessorDate and time of start of processDate and time of end of processInformation of endoscopeUser ID (remove) of the end of processTotal accumulated of reprocessing process countNumber given to each record in the order of occurrenceSetting detailsLCG Info.
58811.4 Printing recordsOER-Elite OPERATION MANUALCh.11Leaking scope decontamination – ErrorFigure 11.35Serial number of this reprocessorDate and time of start of processDate and time of end of processProcess number of occurrence of errorInformation of endoscopeError codeNumber given to each record in the order of occurrence
12.1 RFID data check589OER-Elite OPERATION MANUALCh.12Chapter 12 Information Menu ScreenThis function is to check the RFID data of Scope, User, and Physician.WARNINGIf electromagnetic interference is caused by the radio wave radiated from the RFID reader of this reprocessor, immediately move the interfered device apart from the RFID reader or turn the reprocessor OFF. If you still do not feel well, consult a physician. The radio wave radiated from the RFID reader of this reprocessor may cause malfunction of medical devices such as a pacemaker or other electronic equipment.CAUTIONAn electromagnetic interference with other devices may shorten the communications distance of the designated ID tag and cause signals to become unreadable. Try to take mitigation measures such as keeping the affecting device away from this reprocessor.12.1 RFID data check1Press the “Information” button on the Menu screen.Figure 12.1
59012.1 RFID data checkOER-Elite OPERATION MANUALCh.122Press the “RFID Information” button on the Information menu.Figure 12.23Hold the internal ID endoscope connector, the external ID tag of endoscope, the user ID card, or the physician ID card to the RFID reader of the reprocessor, and scan the tag with the reader until a short beep sounds.Figure 12.3For internal ID endoscope (For example 180 series endoscope)For external ID tagFor internal ID endoscope (190 series endoscope)Scope ID markFor user ID card or physician ID card
12.1 RFID data check591OER-Elite OPERATION MANUALCh.124The touch screen displays the RFID data.Figure 12.45Repeat Step 4 to remove current display information if check another RFID data.Scope model numberScope ID dataScope serial numberUser ID dataPhysician ID dataID typeUser nameUser IDID typePhysician namePhysician IDID type
59212.2 Reprocessor information checkOER-Elite OPERATION MANUALCh.12This function is used to check the information on the reprocessor.6Press the “End” button to go back to the Information menu.Figure 12.512.2 Reprocessor information check1Press the “Information” button on the Menu screen.Figure 12.62Press the “OER-Elite Information” button on the Information menu.Figure 12.7
12.2 Reprocessor information check593OER-Elite OPERATION MANUALCh.123The touch screen displays the OER-Elite Information.No. Item/Button Description1 Body number Body number of the reprocessor.2 Installation Date Date of installation of the reprocessor.3 Total Cycle Counts Total count of reprocessing cycles performed on the reprocessor.4 Total LCG Loads Total count of loading disinfectant solution on the reprocessor.5 Total Detergent Loads Total count of loading detergent solution on the reprocessor.6 Total Water Filter Loads Total count of loading water filter on the reprocessor.7 Total Air Filter Loads Total count of loading air filter on the reprocessor.8 Total Gas Filter (Lid) Loads Total count of loading gas filter on the lid on the reprocessor.9 Total Gas Filter (Tank) LoadsTotal count of loading gas filter on the tank on the reprocessor.10 Cancel button Returns to information menu screen.11 Next button Goes to the next information screen.12 Software version Version number of the software used in the reprocessor.13 Hardware version Version number of the hardware used with the reprocessor.14 MAC address MAC address of the reprocessor.15 Scheduled PM Cycles/DateInformation of the next periodic maintenance.16 Last PM Date Date of the last periodic maintenance.17 Last Repair Date Date of the last repair.18 Back button Goes to the previous information screen.19 OK button Returns to information menu screen.1102345678911 18 19121314151617
59412.2 Reprocessor information checkOER-Elite OPERATION MANUALCh.124Press the “Cancel” button or “OK” button to go back to the information menu screen.Figure 12.8
12.3 List management595OER-Elite OPERATION MANUALCh.12This section is to edit the ID list. The IDs edited here can be used in the step “(b) Recalling the pre-registered ID” on page 54.NOTEThe ID list contains the following IDsScope IDUser IDPhysician IDProcedure ID12.3 List management1Press the “Information” button on the Menu screen.2Press the “List Management” button.Figure 12.93After selecting the ID list to edit, the ID list appears.Figure 12.104On the ID select screen, press the “Edit” button to be input. The ID input screen (software keyboard) will appear.
59612.3 List managementOER-Elite OPERATION MANUALCh.12NOTEFifty IDs can be registered for each ID type.No. Item/Button Description1 ID type Type of the currently edited ID.2 Model # or Name Scope ID list: Displays the model numbers of the endoscopes.User ID list: Displays the user names.Physician ID list: Displays the physician names.Procedure ID list: Displays the procedure names.3 Serial # or ID Scope ID list: Displays the serial numbers of the endoscopes.User ID list: Displays the ID number of users.Physician ID list: Displays the ID number of physicians.Procedure ID list: Displays the ID number of procedures.4 Edit buttons Press one of the buttons to open the ID information edit screen.5 Previous page button Press to go to the next page. When there is no further page to go forward to, the button turns gray and becomes unavailable.6 Next page button Press to return to the next page. When there is no further page to turn back to, the button turns gray and becomes unavailable.7 Back Press to close the ID select screen.5Enter the name on the software keyboard displayed on the touch screen.125346 7
12.3 List management597OER-Elite OPERATION MANUALCh.12No. Button Note1 Alphabet/Numeral keyEnter an alphabet or a numeral.2 Uppercase/LowercasePress “Uppercase/lowercase” button to switch alphabet character on the soft keyboard between uppercase characters and lowercase characters.3 Numeric/AlphabeticPress “Numeric or Alphabetic” button to switch the input mode between a numeral and the alphabet.4 Backspace Press the “Backspace” button to delete the left character of a cursor.When a cursor is on the left edge, this button turns gray and becomes unavailable.5 Space Press the “Space” button to insert a space character.6 Cursor move Press the “Cursor move” button to move the cursor left or right.7 Delete Press the “Delete” button to delete about all information.8 Cancel Return to the previous screen without saving the setting value.9 Save Return to the previous screen and save the entered value.6Press the “Save” button to save the selected setting.7Press the “Back” button.4123568 94123568 9Alphabet input Numeral input7 7
59812.3 List managementOER-Elite OPERATION MANUALCh.12
13.1 Emergency stop and automatic processing after stopping599OER-Elite OPERATION MANUALCh.13Chapter 13 Troubleshooting and RepairWARNINGIf the reprocessing process is interrupted, the endoscopes will not be properly reprocessed. In this case, reprocessing in the OER-Elite should be started again from the beginning.If any irregularity is detected during an inspection or if the reprocessor is clearly malfunctioning, do not use it. Contact Olympus for repair.Some malfunctions may be corrected by referring to the Section 13.2, “Troubleshooting guide”. If the problem cannot be resolved by the described remedial action, do not use the reprocessor and contact Olympus.The OER-Elite can be stopped at any time during operation by pressing the STOP button on the main control panel. The touch screen will display error code [E000] at this time. When the reprocessor is stopped, it identifies the situation in which it was stopped and automatically performs the required operations (including draining of the reprocessing basin), to return the unit to an idle state.WARNING• When a process is interrupted, be sure to execute it again from the beginning. An endoscope that has not been subjected to a complete, error-free process is not safe for patient use. Otherwise, the endoscope reprocessing may be insufficient.• If the process is stopped because of a problem with the reprocessor, contact Olympus.CAUTIONDo not press the power switch OFF during operation. Otherwise, fluid will remain in the reprocessing basin without automatic draining and Air Purge, and the reprocessor may malfunction.Treatment after automatic processing has completedRemove the cause of the reprocessor stoppage, and start the process again from the beginning.13.1 Emergency stop and automatic processing after stopping
60013.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13WARNING• Before handling the disinfectant solution, read the cautions carefully and use it as instructed. It is especially important to know what to do if the disinfectant solution comes in contact with your skin and eyes.• When handling the disinfectant solution, wear personal protective equipment to prevent any disinfectant from getting on your skin and eyes or being inhaled. Avoid direct physical contact and inhalation of vapors. If any disinfectant solution gets in your eyes, immediately rinse with a large amount of fresh water and then consult a medical specialist. Personal protective equipment, such as eyewear, face mask, moisture-resistant clothing, and chemical-resistant gloves that fit properly and are long enough so that your skin and eyes is not exposed. All personal protective equipment should be inspected before use and replaced periodically before it is damaged.• Do not put your hand behind the disinfectant solution bottle drawer. Irritation of skin due to contact with concentrated disinfectant solution, injury by touching a projection, or malfunction of this reprocessor may result.• In the event, in case the process stops and fluid remains in the reprocessing basin due to a power failure or malfunction, keep the lid closed and do not contact with fluid in the basin. Also, do not put the fluid close to fire. Otherwise, adverse physical effects caused by inhalation of disinfectant vapor or physical contact with chemicals, or fire hazard from the residual alcohol may occur.13.2 Troubleshooting guide
13.2 Troubleshooting guide601OER-Elite OPERATION MANUALCh.13WARNING• When using disinfectant solution and alcohol, Olympus recommends the use of gas filters and operating this reprocessor in a well-ventilated area.Wear a face mask, gloves, and protective clothes to minimize aspiration and skin contact.Wear goggles for eye protection.Refer to the following association’s guidelines related to ventilation:If the person operating the reprocessor exhibits an allergic reaction or symptoms, no matter how slight, they should discontinue the task they are performing and vacate the room.• In the event, in case the process stops and fluid remains in the reprocessing basin due to a power failure or malfunction, keep the lid closed and do not contact with fluid in the basin. Also, do not put the fluid close to fire. Otherwise, adverse physical effects caused by inhalation of disinfectant vapor or physical contact with chemicals, or fire hazard from the residual alcohol may occur.CAUTION• If a rinse is performed while an error is actively being displayed and disinfectant solution is still in the reprocessing basin, the rinse process will drain the disinfectant out of the reprocessor. To prevent this, refer to “Error codes and remedial actions” on page 602 for instructions for collecting the disinfectant.• Do not press the power switch OFF when an error code is displayed. Doing so may result in malfunction. The reprocessor will automatically start a process to address the error code. (The error code will blink during automatic processing.) After the error code starts blinking, follow the instructions in “Error codes and remedial actions” below.SGNA (Society of Gastroenterology Nurses and Associates)ASGE (American Society of Gastroenterological Endoscopy)APIC (Association for Professionals of Infection Control and Epidemiology)AORN (Association of Preoperative Registered Nurses)ASTM (American Society for Testing and Materials)OSHA (Occupational Safety and Health Administration)ACGIH (American Conference of Governmental Industrial Hygienists)NIOSH (National Institute for Occupational Safety and Health)AIA (American Institute of Architects)
60213.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13NOTETo complete the process properly, some process cannot be stopped with STOP button.Error codes and remedial actions1If the reprocessor detects irregularity, buzzer beeps for three seconds and the error code will be displayed on the touch screen. While the touch screen displays the following screen, the reprocessor executes the automatic processing.Figure 13.12When the automatic processing is completed, the buzzer beeps for three second and the touch screen displays the possible cause or the remedial action. The error code remains on the touch screen and the buzzer beeps intermittently until the error code is cleared. If the touch screen displays possible causes, press the “OK” button to display the remedial actions.
13.2 Troubleshooting guide603OER-Elite OPERATION MANUALCh.13No. Button Description1 Error Info. Press to display the error info screen.2 OK Press to go to the next screen.3 Detail Press to display the details of remedial actions.3Perform remedial actions according to the instruction described in the table on page 605 through 613 or the instructions displayed on the touch screen.4Press the “OK” button.Figure 13.25Read the instructions displayed on the touch screen and conduct the processes. Then Press the “OK” button to clear the error code.1Possible causes screen Remedial action screen2 1 2 3
60413.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13NOTEFor the error info screen, see the following figure.No. Button Description1 Error Info. Press to display the error info screen.2 Remedial Actions Press to display the remedial actions.3 Possible Causes Press to display the possible causes.4 OK Press to clear the error.No. Item/Button Description1 Serial number Serial number of the reprocessor.2 Software Ver. Version number of the software used in the reprocessor.3 Process Process in which error is generated.4 Error Code Error code of error.5 Date & Time Date and time of occurrence of error.6 Error Info. Information required for contacting Olympus.7 Print button Press to print information of the error.8 OK button Press to close the error info screen.1 2 3 47 8123456
13.2 Troubleshooting guide605OER-Elite OPERATION MANUALCh.13Error code Problem Possible causes Remedial actionsE000 Process was interrupted. Stop button was pressed during the process.Perform the process again.E001 Filling the basin with water took too long (water supply time is beyond maximum setting).• The water faucet is not open enough.• Water flow rate is too low.• Water is leaking from the water supply hose.• Float switch (long) is stuck.• The water filter is clogged.• The mesh filter in the water supply hose connector is clogged.1 Perform the following.Open the water faucet all the way.Check if water is leaking from the water supply hose.Power off the OER-Elite and check the float switch (long) for a proper movement. If not, clean that.2 Perform the process again. If the error repeats, perform the following.Replace the water filter.oRefer to Section 8.4, “Replacing the water filter (MAJ-824 or MAJ-2318)”.Clean the mesh filter in the water supply hose connection port.oRefer to Section 9.7, “Cleaning the mesh filter in the water supply hose connector”.E002 Cleaning fluid is not discharged.• Drain hose is improperly installed.• Drain port mesh filter in basin is clogged.1 Check the drain hose for proper installation. (Height of the drain hose should be less than 60 cm.)2 Wait until the fluid in the basin drained completely.3 Wear appropriate protective gear and clean the drain port mesh filter basin. oRefer to Section 10.3, “Cleaning the mesh filters”.4 Close the lid by pushing until it clicks and perform the air purge to remove water.oRefer to Section 7.6, “Air purge”.5 Perform “Rinse”. oRefer to Section 7.5, “Rinse”.6 Perform the process again.E004 Cleaning fluid decreases during the cleaning process.Internal problem with the reprocessor.Contact Olympus.E005 Basin fluid level is too high.Irregularity in the fluid level sensor.1 Close the water faucet.2 Contact Olympus.E006 Fluid level sensor malfunctions.Float switch (short) or float switch (long) is stuck.1 Clean the float switches.oRefer to Section 10.4, “Cleaning the float switches”.2 Perform “Rinse”.oRefer to Section 7.5, “Rinse”.3 Perform the process again.
60613.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13E007 Fluid level sensor malfunctions.Float switch (short) is stuck.1 Clean the float switches.oRefer to Section 10.4, “Cleaning the float switches”.2 Perform “Rinse”. oRefer to Section 7.5, “Rinse”.3 Perform the process again.E008 Inappropriate water temperature.Water temperature is too high.Check the water temperature is in the range of 5 – 28 qC (41 – 82 qF. )E011 There is too much LCG in the disinfectant solution tank.• Irregularity in the drain hose.• Clogging of the drain port of the reprocessing basin.• Close the lid if it is open. Automatic Processing will start after the lid is closed.• Perform the following.1 Check the drain hose for proper installation (height of the drain hose should be less than 60 cm.)2 Wear appropriate protective gear and clean the mesh filter in the drain port of the basin. oRefer to Section 10.3, “Cleaning the mesh filters”.3 Perform “Rinse”. oRefer to Section 7.5, “Rinse”.4 Perform MRC check. oRefer to Section 4.3, “MRC check setting”.5 Perform the process again.E012 There is insufficient amount of LCG in the disinfectant solution tank.Blockage of the ventilation openings on the gas filter case (lid).1 Check that the ventilation openings on the gas filter case (lid) are not blocked.2 Perform “Drain LCG”. oRefer to Section 8.2, “Replacing the disinfectant solution”.3 Perform “Load LCG”. oRefer to Section 8.2, “Replacing the disinfectant solution”.4 Perform the process again.E013 Basin exists a rinse water.• Drain hose is improperly installed.• Drain port mesh filter in basin is clogged.1 Check the drain hose for proper installation. (Height of the drain hose should be less than 60 cm.)2 Wait until the fluid is drained completely.3 Wear appropriate protective gear and clean the drain port mesh filter in basin.4 Perform “Rinse”. oRefer to Section 7.5, “Rinse”.5 If the quantity of LCG in the disinfectant solution tank is not sufficient, perform “Drain LCG”, then “Load LCG”. oRefer to Section 8.2, “Replacing the disinfectant solution”.6 Perform the process again.Error code Problem Possible causes Remedial actions
13.2 Troubleshooting guide607OER-Elite OPERATION MANUALCh.13E014 LCG cannot be returned to the tank.• Blockage of the ventilation openings on the gas filter case (lid).• Drain port mesh filter in basin is clogged.1 Check that the ventilation openings on the gas filter case (lid) are not blocked.2 Wait until the fluid in the basin collected completely.3 Wear appropriate protective gear and clean the drain port mesh filter in basin.4 Perform “Rinse”. oRefer to Section 7.5, “Rinse”.5 If the quantity of LCG in the disinfectant solution tank is not sufficient, perform “Drain LCG”, then “Load LCG”. oRefer to Section 8.2, “Replacing the disinfectant solution”.6 Perform the process again.E015 Fluid level sensor malfunctions.Float switch (short) is stuck.1 Power off the OER-Elite. Wear appropriate protective gear and check the float switch (short) for proper function. If not clean, clean the fluid level sensor”.2 Perform “Rinse”. oRefer to Section 7.5, “Rinse”.3 Perform the process again.E016 It takes too long to fill the basin with LCG.Internal problem with the reprocessor.Contact Olympus.E017 LCG cannot be heated. Internal problem with the reprocessor.Contact Olympus.E018 Temperature sensor malfunctions.Internal problem with the reprocessor.Contact Olympus.E019 LCG temperature out of range.Internal problem with the reprocessor.1 Power off the OER-Elite.2 Contact Olympus.E020 Heat LCG Timer could not be executed.• OER-Elite was powered off.• Power loss.Perform “Heat LCG”. oRefer to Section 7.2, “Heat LCG”.E021 Insufficient air purge pressure.• Air filter is not installed properly.• Air filter is clogged.1 Check if the air filter is firmly attached.2 If the air filter is firmly attached, the air filter may be clogged. Perform “Replacing the air filter”. oRefer to Section 8.5, “Replacing the air filter (MAJ-823)”.3 Perform the air purge. oRefer to Section 7.6, “Air purge”.4 Perform the process again.E022 Insufficient fluid pressure. The circulation port mesh filter in basin is clogged.1 Wear appropriate protective gear and clean the circulation port mesh filter in basin. oRefer to Section 10.3, “Cleaning the mesh filters”.2 Perform the process again.Error code Problem Possible causes Remedial actions
60813.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13E023 Excessive fluid pressure. Internal problem with the reprocessor.Contact Olympus.E024 Channel Monitor Error • Connecting tubes are not connected.• Leak test air tube is not connected.• Endoscope’s suction channel is clogged.• Unnecessary connecting tubes are connected to connectors in the reprocessing basin.Take optimum remedial actions as described in “When the error code [E024] is displayed during the reprocessing process” on page 614.E026 Insufficient fluid flow rate. The circulation port mesh filter in basin is clogged.1 Wear appropriate protective gear and clean the circulation port mesh filter in basin. oRefer to Section 10.3, “Cleaning the mesh filters”.2 Perform the process again.E027 Internal valves malfunction.Internal problem with the reprocessor.Contact Olympus.E031 The lid is open. Internal problem with the reprocessor.Contact Olympus.E041 The power was lost during the process.• Power loss.• OER-Elite was powered off.1 Check that the power cord is connected firmly.2 Perform the process again.Note:Do not power off the OER-Elite during the process.E051 Fluid leakage inside the OER-Elite.Internal problem with the reprocessor.1 Close the water faucet.2 Turn the power off.3 Disconnect the power cord from the outlet.4 Contact Olympus.E052 Fluid leakage inside the OER-Elite.Internal problem with the reprocessor.1 Close the water faucet.2 Turn the power off.3 Disconnect the power cord from the outlet.4 Check the O-ring of leak test connector is firmly attached.5 Contact Olympus.E061 Ultrasonic cleaning is not functioning.Irregularity in the ultrasonic oscillator.Contact Olympus.E071 Abnormalities with the disinfectant bottle drawer sensor.Internal problem with the reprocessor.Contact Olympus.If the odor of disinfectant solution increases, wear appropriate personal protective reprocessor.Error code Problem Possible causes Remedial actions
13.2 Troubleshooting guide609OER-Elite OPERATION MANUALCh.13E072 LCG in the tank cannot be discharged.• Disinfectant collection hose is improperly connected.• Drain port mesh filter in basin is clogged.• Drain hose is improperly installed.• When draining through the disinfectant collection hose is selected1 Check the disinfectant collection hose for proper connection.2 Perform “Drain LCG” again. oRefer to Section 8.2, “Replacing the disinfectant solution”.• When draining through the drain hose is selected.1 Check the drain hose for proper installation. (Height of the drain hose should be less than 60 cm.)2 Wait until the fluid in the basin drained completely.3 Wear appropriate protective gear and clean the drain port mesh filter in basin. oRefer to Section 10.3, “Cleaning the mesh filters”.4 Perform “Drain LCG” again. oRefer to Section 8.2, “Replacing the disinfectant solution”.E073 Water supply irregularity during the Load LCG process.• The water faucet is not open enough.• Drain port mesh filter in basin is clogged.• Water flow rate is too low.• Water is leaking from the water supply hose.• The water filter is clogged.• The mesh filter in the water supply hose connector is clogged.1 Open the water faucet all the way.2 Wear appropriate protective gear and clean the drain port mesh filter in basin.3 Check if water is leaking from the water supply hose.4 Close the lid by pushing until it clicks and restart “Load LCG” according to the guide on the touch screen.E074 LCG in the cassette bottles is not supplied to the disinfectant solution tank.Internal problem with the reprocessor.Contact Olympus.Error code Problem Possible causes Remedial actions
61013.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13E075 Water supply irregularity during the Load LCG process.• The water faucet is not open enough.• Drain port mesh filter in basin is clogged.• Water flow rate is too low.• Water is leaking from the water supply hose.• The water filter is clogged.• The mesh filter in the water supply hose connector is clogged.1 Open the water faucet all the way.2 Wear appropriate protective gear and clean the drain port mesh filter in basin.3 Check if water is leaking from the water supply hose.4 Close the lid by pushing until it clicks and restart “Load LCG” according to the guide on the touch screen.E076 Irregularity of the fluid level sensor in the disinfectant solution tank.Internal problem with the reprocessor.Contact Olympus.E077 Incomplete positioning of the disinfectant bottle drawer.Disinfectant bottle drawer is not completely closed.Ensure that the disinfectant bottle drawer is completely closed.E078 Cassette bottles are empty.Empty cassette bottles were installed.Perform the “Load LCG” with new cassette bottles.E081 The process cannot be properly controlled.Irregularity in the valve inside the reprocessor.Contact Olympus.E082 Internal irregularity in the OER-Elite.Irregularity in the electrical circuitry inside the reprocessor.Contact Olympus.E083 The process cannot be properly controlled.Irregularity in the electrical circuitry inside the reprocessor.Contact Olympus.E084 Malfunction of the RFID reader.Irregularity in the electrical circuitry inside reprocessor.Contact Olympus.To continue using the OER-Elite, and manually input the ID information via the touch screen. RFID will be disabled.E086 Internal irregularity in the OER-Elite.Irregularity in the pump inside the reprocessor.Contact Olympus.E087 Internal irregularity in the OER-Elite.Irregularity in the flow sensor inside the reprocessor.Contact Olympus.E088 Insufficient fluid flow. The circulation port mesh filter in the basin is clogged.1 Wear appropriate protective gear and clean the circulation port mesh filter in basin. oRefer to Section 10.3, “Cleaning the mesh filters”.2 Perform the process again.Error code Problem Possible causes Remedial actions
13.2 Troubleshooting guide611OER-Elite OPERATION MANUALCh.13E089 Internal irregularity in the OER-Elite.Irregularity in the electrical circuitry inside the reprocessor.Contact Olympus.E092 Manual leak test time has been exceeded.Manual leak test time of 10 minutes has been exceeded.Perform manual leak test again.E093 Alcohol cannot be supplied during the process.Clogging of the alcohol piping.Refer to Section 7.14, “Alcohol flush”.E094 Date cannot be printed. • Printer is not properly connected to the OER-Elite.• Printer paper roll has run out.• Malfunction of printer.1 If the printer paper runs out, replace the printer paper roll.2 If the paper cover is open, close the cover.3 Check the power LED of the printer lights up. If not, press power button on the operational panel of the printer to ON.4 Check the interface cable is connected properly.5 Print the data again.E095 Detergent cannot be supplied during the process.Clogging of the detergent piping.• Refer to “If the message screen “Message 093” is displayed again after replacing the detergent tank and restarting the reprocessing process:” on page 638.E111 Internal irregularity in the OER-Elite.Internal problem with the reprocessor.Contact Olympus.E112 ALT Self-Check Error Leak test air tube was connected during the ALT Self-Check.1 Confirm the leak test air tube is not connected.2 Perform “ALT Self-Check” again.oRefer to Section 7.13, “Self-check of auto leak test”.Error code Problem Possible causes Remedial actions
61213.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13E114 Leak detected during ALT process.• There is a leak within the endoscope.• Leak test air tube is broken.When the ALT is incorporated in the beginning of the reprocessing process:1 If warm water was used in manual cleaning, the auto leak test may be erroneous because the temperature of endoscope(s) may change drastically after the manual cleaning. If the auto leak test is to be performed at the start of reprocessing, begin the reprocessing after the endoscope temperature has returned to the room temperature. Wait more than 15 minutes after manual cleaning for the endoscope temperature to return to room temperature.2 If warm water was not used in manual cleaning, perform the following.Identify the endoscope with leak from the scope IDs displayed on the touch screen.Start the process again for the non-leaking endoscope.Be sure to perform “Leaking Scope Decontamination” with the leaking endoscope before returning for repair.When the ALT is incorporated in the end of the reprocessing process:1 Identify the endoscope with leak from the scope IDs displayed on the touch screen.2 Remove the leaking endoscope from the OER-Elite, then start the process again for the non-leaking endoscope.3 Return the leaking endoscope for repair.Error code Problem Possible causes Remedial actions
13.2 Troubleshooting guide613OER-Elite OPERATION MANUALCh.13E115 Leak test air pressure is low.• Irregularities of leak test air tube.• Irregularities of the O-ring of the leak test connector.• Excessive leak exists within the endoscope.Perform the manual leak test to check if the endoscope or the leak test air tube leaks.If a leak point exists on the leak test air tube, replace it with new one and retry the process.If a leak point exists between the connector of the leak test air tube and the leak test connector (E1 or E2), or between the metallic connector and the venting connector of the endoscope, contact Olympus.If a leak point exists on the outer surface, stop it with a piece of tape and perform the leaking scope decontamination. After it, send the scope for servicing. If a leak cannot be reduced by taping and this error occurs, contact Olympus.If the leak point exists in a position where taping is impossible, do not execute this process but contact Olympus. E117 Excessive leak test air pressure.Internal problem with the reprocessor.1 Contact Olympus.2 If you wish to continue reprocessing, disable the ALT function.E118 Cannot release the pressure of endoscope.Internal problem with the reprocessor.Contact Olympus.E119 Auto leak test system malfunction.Irregularity in the electrical circuitry inside the reprocessor.1 Contact Olympus.2 If you wish to continue reprocessing, disable the ALT function.E124 Auto leak test system malfunction.Irregularity in the electrical circuitry inside the reprocessor.1 Contact Olympus.2 If you wish to continue reprocessing, disable the ALT function.E131 Portable memory was disconnected.Portable memory was disconnected while the data was exported.Re-insert the portable memory and export data again.E132 Cannot write to portable memory.• Improper connection of the portable memory.• Abnormalities of the portable memory.Use another portable memory and export data again.E134 Portable memory port malfunction.Irregularity in the electrical circuitry inside the reprocessor.Contact Olympus.Error code Problem Possible causes Remedial actions
61413.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13When the error code [E024] is displayed during the reprocessing processIf an error such as clogging of an endoscope channel or connection irregularity of the connecting tube or leak test air tubes occurs during the process, the error code [E024] is displayed on the touch screen and the reprocessor will stop the process. After the auto processing, the touch screen displays the following screen.No. Item/Button Description1 Connector Show Connector in the reprocessing basin.2 Detection result Displays a detection result if any irregularity is detected to a connector. Blank cell means that there are not any irregularities with the connector.3 Detail button Press to display the remedial actions of a connector that detected an irregularity. The button is unavailable if there are not any irregularities with the connector.4 Error Info. button Press to display the error information.5 Connection Guide buttonPress to display the connection guide screen. For detail, refer to “Connection guide of first endoscope” on page 179, “Connection guide of second endoscope” on page 206.6 Next button Press to display the next screen.7 Back button Press to return to the first page.8 OK button Press to go to the next page.41st page 2nd page62314 7 85 5
13.2 Troubleshooting guide615OER-Elite OPERATION MANUALCh.13Tab le 1 3. 11Step on the foot pedal to open the lid.2According to the detection result displayed on the touch screen, corrects as follows.Detection Result Remedial actionsDisconnection (OER-Elite side)1 Check the connecting tube is not kinked.2 Check the connectivity between the connecting tube and OER-Elite.3 Remove the endoscope and perform manual cleaning. Refer to “Reprocessing manual” of the endoscope for details.Blocked Channel 1 Check the correct connecting tube is used.2 Check the connecting tube is connected to the correct connector in the basin.3 Check the connecting tube is not kinked.4 Remove the endoscope and perform manual cleaning. Refer to “Reprocessing manual” of the endoscope for details.Wrong connection 1 Disconnect the connecting tube and confirm the correct connecting tube is used.Disconnection (Scope side) 1 Check the correct connecting tube is used.2 Check the connecting tube is connected to the correct connector in the basin.3 Check the connectivity between the connecting tube and the endoscope.4 Check the connectivity between the leak test air tube and the OER-Elite.NOTEIf the Detail button is pressed, the remedial actions can be displayed on the touch screen.3Press the “Next” button on the first page to display the second page.Figure 13.34According to the detection result displayed on the touch screen, corrects as Step 2.
61613.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.135Press the “OK” button on the second page.Figure 13.46Press the “OK” button to finish remedial actions. If the endoscope will be removed from the reprocessing basin, select “Yes” button and press “OK”. Then, the IDs for this scope will be deleted.Figure 13.5Figure 13.67Start the reprocessing process and rinse from the beginning.
13.2 Troubleshooting guide617OER-Elite OPERATION MANUALCh.13When any leaks are detectedWhen the error code [E114] or [E115] is displayed on the touch screen, the result of auto leak test is “Leaked”, or any leaks are detected during manual leak test, perform following.NOTE• When the error code [E114] is displayed, leak detected during ALT process at the reprocessing process.• When the error code [E115] is displayed, leak test air pressure is low.• When these errors frequently occur, perform self-check of auto leak test. For detail, refer to Section 7.13, “Self-check of auto leak test”.Workflow when the error code [E114] is displayed, or the result of auto leak test is “Leaked”Figure 13.7The error code [E114] is displayed during reprocessing process. (The auto leak setting is “Beginning of the process”).The result of auto leak test is “Leaked”.The error code [E114] is displayed during reprocessing process. (The auto leak setting is “End of the process”).Did each process perform immediately after completing manual cleaning using warm water?Wait for more than 15 minutes (In the case of the ultrasonic endoscope, wait for more than 30 minutes) after completing manual cleaning.YesRetry reprocessing process or auto leak test.Is the following occurred again?• The error code [E114] is displayed during reprocessing process.• The result of auto leak test is “Leaked”.NoThe endoscope has no leaks. Perform the leaking scope decontamination. Refer to Section 7.15, “Leaking scope decontamination”.Contact Olympus to repair the endoscope.NoYes
61813.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13Workflow when the error code [E115] is displayedFigure 13.8The error code [E115] is displayed during reprocessing process.Perform the manual leak test.The error code [E115] is displayed during auto leak test function.The error code [E115] is displayed during manual leak test function.Contact Olympus to repair the reprocessor.• Replace the leak test air tube with a new one.• If there are no more leak test tube, contact Olympus.Contact Olympus to replace a new O-ring.Refer to “Workflow when any leaks are detected during manual leak test” on page 619. Then contact Olympus to repair the endoscope.Contact Olympus to repair the reprocessor.Is the error code [E115] displayed again?Is there leak on the tube section of the leak test air tube?Is there leak between the connector of the leak test air tube and the leak test connector (E1 or E2), or between the metallic connector and the venting connector of the endoscope?Is there leak from the endoscope?YESYESYESYESNONONONO
13.2 Troubleshooting guide619OER-Elite OPERATION MANUALCh.13Workflow when any leaks are detected during manual leak testManual decontamination for leaking endoscope(a) Manual cleaning for leaking endoscope1Perform the leaking scope decontamination. For the detail, refer to Section 7.15, “Leaking scope decontamination”.2If the leaking scope decontamination is completed without errors, contact Olympus to repair the endoscope.If any errors occur, perform manual decontamination for leaking endoscope below. Then, contact Olympus to repair the endoscope.1For a leak detected in the covering of the insertion tube, bending section, or universal cord, dry the leaking area thoroughly and wipe with alcohol. Tape over the location of the leak with a piece of electrical tape or other waterproof tape prior to immersing in detergent solution. Wrap the tape tightly. For leaks detected in other locations (e.g., internal channel), proceed with the instructions provided below.2Fill a basin with detergent solution at the temperature and concentration recommended by the detergent manufacturer. Use a basin that is at least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the endoscope to be completely immersed.3Insert the leakage tester connector into the output socket of the maintenance unit such as MU-1 or the light source and turn the maintenance unit or the light source ON. Set the light source’s airflow regulator switch to “HIGH” or “3”.4Connect the leakage tester’s connector cap to the venting connector of the water-resistant cap. 5Immerse the endoscope in the detergent solution.6Perform manual cleaning according to the instructions provided in the reprocessing manual. Minimize unnecessary flexion of the insertion tube and universal cord during cleaning.
62013.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13(b) Sterilization for leaking endoscope(c) High-level disinfection for leaking endoscopeWARNINGThis process is intended to treat the endoscope with water leak before sending it for servicing. It does not guarantee the reprocessing of the endoscope and its accessories after this process. After completing this process, do not use the endoscope and its accessories in examination but send it for servicing.Following manual cleaning, the preferred method of rendering a leaking endoscope safe to handle is ethylene oxide sterilization. Ethylene oxide sterilization should be performed according to the instruction provided in the reprocessing manual. If electrical tape was applied to a leak detected in the endoscope’s external surface, remove the tape and wipe with 70% ethyl or isopropyl alcohol prior to ethylene oxide sterilization. If ethylene oxide sterilization is not possible, perform high-level disinfection according to the instructions provided below.1Fill a basin with disinfectant solution at the temperature and concentration recommended by the disinfectant manufacturer. Use a basin that is at least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the endoscope to be completely immersed.2Insert the leakage tester connector into the output socket of the maintenance unit or the light source and turn the maintenance unit or the light source ON. Set the light source’s airflow regulator switch to “HIGH” or “3”. 3Connect the leakage tester’s connector cap to the venting connector of the water-resistant cap.4Immerse the endoscope in the disinfectant solution.5Perform high-level disinfection according to the instructions provided in the reprocessing manual. Minimize unnecessary flexion of the insertion tube and universal cord during reprocessing.
13.2 Troubleshooting guide621OER-Elite OPERATION MANUALCh.13Message, caution, and warning screensNOTE• The button for resuming a process varies based on message screens displayed on the touch screen.• The message cannot record to logs.1If the reprocessor requires operator action, the buzzer beeps and the message screen will be displayed on the touch screen.Figure 13.92Perform remedial actions according to the instruction described in the table below or the instructions displayed on the touch screen.3Press the “OK” button to resume a process.Figure 13.10
62213.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13No. Message Remedial actionsMessage 001 Scope ID has not been entered. Enter scope ID before starting the process.Message 002 User ID has been entered. Enter user ID before starting the process.Message 003 OER-Elite cannot accept more than two user IDs per reprocessing cycle.1 Verify the entered user ID information is correct.2 Delete the incorrect user information. Re-enter user ID.Message 004 ID entry for 1st endoscope has not been completed.Before entering the 2nd endoscope ID, enter the remaining IDs for the 1st endoscope.Message 005 User ID Setting has not been activated. To enter user ID, go to Menu screen, then change the User ID Setting.Message 005 Patient ID Setting has not been activated.To enter patient ID, go to Menu screen, then change the Patient ID Setting.Message 005 Physician ID Setting has not been activated.To enter physician ID, go to Menu screen, then change the Physician ID Setting.Message 006 OER-Elite cannot accept more than two Scope IDs per reprocessing cycle.1 Verify the entered scope ID information is correct.2 Delete the incorrect scope information. Re-enter scope ID.Message 008 Scope ID master has been entered. Channel Monitoring, auto leak testing and Connection Guide will be disabled.Refer to the “List of compatible Endoscopes/Connecting Tubes <OER-Elite>” to confirm the correct connecting tubes for this endoscope.Message 009 Scope ID has been entered from the touch screen. Channel monitoring, auto leak testing and connection guide will be disabled.Refer to the “List of compatible Endoscopes/Connecting Tubes <OER-Elite>” to confirm the correct connecting tubes for this endoscope.Message 010, 011This endoscope is incompatible for auto leak testing. Auto leak testing will be disabled.No additional action is required.Message 014 Patient ID has not been entered. Enter patient ID before starting the process.Message 015 Physician ID has not been entered. Enter physician ID before starting the process.Message 016 OER-Elite cannot accept more than two patient IDs per reprocessing cycle.1 Verify the entered patient ID information is correct.2 Delete the incorrect patient information. Re-enter patient ID.Message 017 OER-Elite cannot accept more than two physician IDs per reprocessing cycle.1 Verify the entered physician ID information is correct.2 Delete the incorrect physician information. Re-enter physician ID.Message 018 Malfunction of the RFID reader. 1 Manually input the ID information via the touch screen.2 Contact Olympus.Message 019 Procedure ID has not been entered. Enter procedure ID before starting the process.Message 020 This endoscope is incompatible for channel connectivity monitoring.Channel connectivity monitoring will be disabled.No additional action is required.
13.2 Troubleshooting guide623OER-Elite OPERATION MANUALCh.13Message 021 This endoscope is incompatible for channel blockage monitoring.Channel blockage monitoring will be disabled.No additional action is required.Message 022 This endoscope is incompatible for channel monitoring. Channel monitoring will be disabled.No additional action is required.Message 023 Refer to the Message 008 Refer to the Message 008.Message 024 Scope ID of endoscope that cannot be reprocessed simultaneously has been entered. OER-Elite cannot accept more than one User ID per reprocessing cycle.1 Verify the entered user ID information is correct.2 Delete the incorrect user information. Re-enter user ID.Message 025 This endoscope combination is incompatible for auto leak testing.Perform “Auto Leak Test” separately for second scope.Message 031 LCG cycle count/usage exceeds the set value.Perform MRC check before performing the process.Message 032 There is no LCG in the disinfectant solution tank.Perform “Load LCG”.Message 034 Load LCG cannot be performed with LCG in the disinfectant solution tank.Perform “Drain LCG”, then “Load LCG”.Message 035 Drain LCG cannot be performed with no LCG in the disinfectant solution tank.No additional action is required.Message 036 An invalid set time for Heat LCG Timer has been entered.1 Enter a minimum time of at least 2.5 hours.2 Enter a date of less than 5 days.Message 037 The LCG temperature is already above the specified temperature.No additional action is required.Message 038 Cannot activate the Heat LCG Timer. System clock battery low.Contact Olympus.Message 047 More than 1 hour has passed since the MRC check result has been entered.Perform MRC check again and enter the result.Message 048 MRC check result has not been entered.Perform MRC check again and enter the result.Message 050 There is no LCG in the disinfectant solution tank.Perform “Load LCG” before MRC check.Message 061 Portable memory is full, cannot Export data.Insert portable memory that has sufficient capacity. At least 100MB capacity is necessary.Message 062 No record exists for the selected period. Confirm the dates for the period selected.Message 070 The lid is open. Close the lid to perform the process.Message 071 Cannot drain LCG through the disinfectant collection hose with the lid closed.1 Verify the collection method selected is correct.2 Follow the prompts on the OER-Elite.Message 080 Time for periodic maintenance. The time for regularly scheduled maintenance is near. Contact Olympus to perform the maintenance.Message 081 Water filter usage exceeds the set value.Perform “Replace Water Filter”.Message 081 Air filter usage exceeds the set value. Perform “Replace Air Filter”.No. Message Remedial actions
62413.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13Message 081 Gas filter usage exceeds the set value. Perform “Replace Gas Filter on the Lid” or perform “Replace Gas Filter on the Tank”.Message 082 The OER-Elite has not been used for more than 14 days.Perform “Self-Disinfection” and “Water Line Disinfection”.Message 083 ALT Self-check failed. Contact Olympus. To continue reprocessing, choose a program with the ALT Setting disabled. (Program 3 or 4)Message 086 “Leak” has been selected as a manual leak test result.1 Close the lid by pushing until it clicks. The water will automatically be discharged from the basin.2 Perform “Leaking Scope Decontamination”.Message 087 Alcohol cannot be supplied. 1 Confirm the amount of alcohol in the alcohol tank. If no alcohol is remaining, replenish the alcohol.2 Confirm the alcohol tubing is not bent and press “Restart”.3 To cancel the process, press “Stop”.Refer to Section 5.9, “Inspecting and replenishing alcohol”.Message 088 Fluid cannot be drained from the alcohol line.1 Confirm the nozzle of the cap is out of the beaker.2 Confirm the alcohol tubing is not bent and press “Restart”.3 To cancel the process, press “Stop”.Message 088 Fluid cannot be drained from the detergent line.1 Confirm the nozzle of the cap is out of the beaker.2 Confirm the detergent tubing is not bent and press “Restart”.3 To cancel the process, press “Stop”.Message 089 Fluid cannot be supplied from the alcohol line.1 Confirm the nozzle of the cap is in the beaker.2 Confirm the alcohol tubing is not bent and press “Restart”.3 To cancel the process, press “Stop”.Message 089 Fluid cannot be supplied from the detergent line.1 Confirm the nozzle of the cap is in the beaker.2 Confirm the detergent tubing is not bent and press “Restart”.3 To cancel the process, press “Stop”.Message 091 To enable the mode setting, turn the power off and boot with service mode.No additional action is required.Message 093 Detergent cannot be supplied. 1 Confirm the amount of detergent in the detergent tank. If no detergent is remaining, replace a new detergent tank.2 Confirm the detergent tube is not bent and press “Restart”.3 To cancel the process, press “Stop”.Refer to Section 8.3, “Replacing the detergent tank”.Message 095 An invalid time has been entered. Set the time within following range.Range: 1/1/2012 00:00 – 12/31/2087 23:59No. Message Remedial actions
13.2 Troubleshooting guide625OER-Elite OPERATION MANUALCh.13Message 100 Leak test air tube for scope 1 has not been connected within 5 minutes.1 Enter scope ID for scope 1 again.2 Within 5 minutes, connect the leak test air tube for scope 1 to the E1 connector.Message 101 Leak test air tube has been connected to the endoscope before entering scope ID.1 Disconnect the leak test air tube from the connector.2 Enter scope ID for scope 1 again.3 Connect the leak test air tube for scope 1 to the E1 connector.Message 102 Program has been changed and OER-Elite cannot detect the proper connection of leak test air tube.1 Disconnect the leak test air tube from the connector.2 Press “OK”.Message 110 Process interrupted before rinse. Chemical may remain in the basin.1 Press “Rinse” to continue now.2 Press “OK” to perform “Rinse” later.Message 120 Detergent cycle count/usage exceeds the set value.Perform “Replace Detergent”.Message 132 An invalid date has been entered. Set the expiration date properly.Warning 026 This endoscope requires special procedures for reprocessing in the OER-Elite.Before reprocessing the endoscope in the OER-Elite, thoroughly review and follow the procedures described in the respective endoscope reprocessing manual or the OER-Elite Instructions – Operation Manual. Otherwise, reprocessing may be insufficient.Warning 096 “ALT Self-Check” has not been performed.1 Press “ALT Self-Check” to continue now.2 Press “OK” to perform ALT Self-Check later.Warning 121 Detergent has expired based on its shelf life expiration date.Perform “Replace Detergent”.Caution 007 Incompatible combination of endoscope.Refer to the “List of compatible Endoscopes/Connecting Tubes <OER-Elite>” to confirm the applicable combination of endoscopes.Caution 013 This endoscope is incompatible with OER-Elite.1 Reprocess this endoscope using another method.2 For a list of compatible endoscopes, refer to the “List of compatible Endoscopes/Connecting Tubes <OER-Elite>”.Caution 033 LCG has expired based on its shelf life expiration date.Perform “Drain LCG”, then “Load LCG”.Caution 039 Load LCG has not been completed properly.Perform “Drain LCG”, then “Load LCG”.Caution 040 “Fail” has been entered as an MRC check result, cannot perform the process.Perform “Drain LCG”, then “Load LCG”.Caution 041 “Mix LCG” has not been performed. 1 Press “Mix LCG” to continue now.2 Press “OK” to perform Mix LCG later.Caution 046 More than 4 hours has passed since the last cycle. To ensure accurate MRC testing, perform “Mix LCG”.1 Press “Mix LCG” to continue now.2 Press “OK” to perform Mix LCG later.Caution 084 System clock battery has run out. Contact Olympus.No. Message Remedial actions
62613.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13Caution 085 System clock battery low. 1 Contact Olympus.2 To continue using the OER-Elite, make sure the system clock indicates the actual current date and time. If not, perform “Date and Time setting” and set the current date and time. The system clock will function until OER-Elite is powered off.Caution 092 “No” has been selected as the result of manual cleaning and leak test for scope 1.1 Perform leak test and manual cleaning before starting reprocessing.2 If a leak has been detected, perform “Leaking Scope Decontamination”.Caution 092 “No” has been selected as the result of manual cleaning for scope 1.Perform manual cleaning before starting reprocessing.Caution 092 “No” has been selected as the result of leak test for scope 1.1 Perform leak test before starting reprocessing.2 If a leak has been detected, perform “Leaking Scope Decontamination”.Caution 092 “Pass” has been selected as the result of leak test. “No” has been selected as the result of manual cleaning.Perform manual cleaning before starting reprocessing. (Do not use leaking scope decontamination for this endoscope)Caution 092 “Pass” has been selected as the result of leak test.Perform reprocessing. (Do not use leaking scope decontamination for this endoscope.)Caution 094 Water Line Disinfection has not been completed after replacing the water filter.Perform “Water Line Disinfection”.No. Message Remedial actions
13.2 Troubleshooting guide627OER-Elite OPERATION MANUALCh.13When the message screen “Message 087” is displayedIf alcohol has run out and the message screen “Message 087” is displayed, the reprocessor will stop the process. In this case, you can restart the process from the “alcohol flush” by implementing the following procedure.Figure 13.11NOTE• The reprocessing process can be restarted from the “alcohol flush” if the “Restart” button is pressed in Message 087. In that case, the reprocessing record is recorded as normal.• The lid will remain locked until the message screen is cleared, and you cannot access the reprocessing basin even after pressing the foot pedal. If you want to interrupt the reprocessing process, press STOP button and follow the procedure described in Section 13.1, “Emergency stop and automatic processing after stopping”.
62813.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13Figure 13.12WARNING• The alcohol used with the reprocessor must be 70% ethyl alcohol or 70% isopropyl alcohol. Using any other kind of alcohol may result in malfunction of the reprocessor or the endoscope, difficulty drying the endoscope, fire hazard, or a hazard due to toxic vapor emitted from the alcohol.• Alcohol is flammable and should be handled with extra care.• Remove the alcohol in the alcohol tank and replace it with new alcohol at least once a week. Otherwise, the alcohol in the alcohol tank may degrade.• Before handling the alcohol, read the cautions carefully and use it as instructed.When the Message 087 is displayed during the alcohol flush processMessage 087 “Alcohol cannot be supplied” is displayed.Confirm whether alcohol still present in the alcohol tank.1.Replenish alcohol.(Refer to Section 5.9, “Inspecting and replenishing alcohol”.)2.Press the “Restart” button to restart the reprocessing cycle.3.Reprocessing process restarts from alcohol flush process.4.If no alcohol remain in the tank.If alcohol still present in the tank.Refer to “If the “Message 087” is displayed again after replenishing alcohol and restarting the alcohol flush process:” on page 629.Inspect the tube and connector connected to the alcohol tank.Correct irregularities.Press the “Stop” button to cancel the reprocessing process and perform remedial actions of E093 to resolve the problem.If any irregularity is found.If no irregularity is found.If the message screen is displayed again.
13.2 Troubleshooting guide629OER-Elite OPERATION MANUALCh.13If the “Message 087” is displayed again after replenishing alcohol and restarting the alcohol flush process:When message screen “Alcohol cannot be supplied” is displayed again after restarting the reprocessing process, check that there is no irregularity regarding the following points.• The tube connected to the alcohol tank is not kinked.• The connector of alcohol tank is attached firmly.If any irregularity is found by inspection, correct these points and press the “Restart” button on the touch screen to restart the alcohol flush process.If no irregularity is found during an inspection or the message screen is displayed again, internal problem with the reprocessor may be suspected. In this case, follow the procedure below.WARNING• The alcohol used with the reprocessor must be 70% ethyl alcohol or 70% isopropyl alcohol. Using any other kind of alcohol may result in malfunction of the reprocessor or the endoscope, difficulty drying the endoscope, fire hazard, or a hazard due to toxic vapor emitted from the alcohol.• Alcohol is flammable and should be handled with extra care.• Remove the alcohol in the alcohol tank and replace it with new alcohol at least once a week. Otherwise, the alcohol in the alcohol tank may degrade.• Before handling the alcohol, read the cautions carefully and use it as instructed.1Inspect the alcohol tank. If alcohol is still present in the alcohol tank, go to the procedure of “If the “Message 087” is displayed again after replenishing alcohol and restarting the alcohol flush process:” on page 629 described in this section.2Replenish alcohol into the alcohol tank as described in Section 5.9, “Inspecting and replenishing alcohol”.3Press Restart button on the touch screen to restart the reprocessing process.Figure 13.134Reprocessing process restarts from alcohol flush process.
63013.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13(a) Required itemsTable 13.2(b) Remedial actionsCheck Required itemsSyringeTube1Press the “Stop” button on the touch screen to cancel the reprocessing process.Figure 13.142The touch screen displays the error code [E093]. Press the “OK” button. Then, troubleshooting guide is displayed on the touch screen.Figure 13.15
13.2 Troubleshooting guide631OER-Elite OPERATION MANUALCh.133Connect the provided syringe and tube. For detail, refer the detail button on E093 on the touch screen.Figure 13.16Figure 13.174Disconnect the ventilation tube from the guide of the alcohol tank.Figure 13.185Fill the syringe with air and insert the tube connected to the syringe all the way into the far end of the alcohol tank ventilation tube.Figure 13.19Ventilation tubeGuide
63213.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.136Press the “OK” button repeatedly until error screen is closed.Figure 13.207Go to the Function menu. Press the “Alcohol Flush” button on the second page.Figure 13.218Enter the operator's user ID. For entering ID, refer to Section 3.6, “Entering ID” (If applicable).NOTEThe input of the user ID can be omitted by modifying the “User ID Setting”. For details, refer to Section 4.5, “User ID Setting”.
13.2 Troubleshooting guide633OER-Elite OPERATION MANUALCh.139Press the “Start” button on the touch screen. Flush air in the syringe within 30 seconds.Figure 13.2210 Press the “Stop” button to stop the alcohol flush. Then, error code E000 is displayed on the touch screen. After confirming that, disconnect the syringe's tube from the ventilation tube.Figure 13.23Figure 13.24
63413.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.1311 Disconnect the tube from the cap on the alcohol tank, remove the tank from the detergent/alcohol drawer and insert the ventilation tube into the guide.Figure 13.2512 Carefully pour the 70% ethyl alcohol or 70% isopropyl alcohol into the alcohol tank until it is level with the line inside the tank. If any alcohol is spilled from the tank, wipe it with a clean cloth. Replace the cap on the alcohol tank, making sure that it is tight.NOTEWhen the alcohol tank is filled to the line, it will hold about 1 L (33 ounces) of alcohol (enough for about 20 alcohol flushes).13 After confirming that alcohol has not spilled from the alcohol tank and there is enough alcohol remaining in the alcohol tank, place the alcohol tank in the detergent/alcohol drawer and connect the tube that is designed to be connected to the cap of the alcohol tank.Figure 13.2614 Close the detergent/alcohol drawer. Rinse the syringe and tube with running water, dry them completely, and store in a clean place.15 Perform the “alcohol flush” again from the beginning.Alcohol tank Ventilation tube
13.2 Troubleshooting guide635OER-Elite OPERATION MANUALCh.13When the “Message 093” is displayedIf detergent has run out and the message screen “Detergent cannot be supplied” is displayed, the reprocessor will stop the process. In this case, reprocessing can be restarted by implementing the following procedure.Figure 13.2716 If alcohol flush is successfully completed, the problem is resolved and alcohol flush was performed with the endoscopes. If the “Message 087” is displayed again, the reprocessor component malfunction is suspected. In this case, do not use the reprocessor and contact Olympus. Remove endoscope(s) from the basin and perform alcohol flush again by another method.
63613.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13NOTE• The reprocessing process can be restarted only when the message screen “Detergent cannot be supplied during the process” is displayed.• The lid will remain locked until the error is cleared, and you cannot access the reprocessing basin even after pressing the foot pedal. If you want to interrupt the reprocessing process, press the “Stop” button and follow the procedure described in Section 13.1, “Emergency stop and automatic processing after stopping”.Figure 13.28When the message 093 is displayed during the reprocessing processMessage screen is displayed.Confirm whether the detergent still present in the detergent tank.1.Replace the detergent tank.2.Press the “Restart” button to restart the reprocessing process.3.Restarted process is completed.4.If no detergent remain in the tank.If detergent still present in the tank.Refer to “If the message screen “Message 093” is displayed again after replacing the detergent tank and restarting the reprocessing process:” on page 638.Inspect the tube and connector connected to the detergent tank.Correct irregularities.Perform remedial action to resolve the problem.If any irregularity is found.If no irregularity is found.If the same message screen is displayed again.
13.2 Troubleshooting guide637OER-Elite OPERATION MANUALCh.13WARNING• Before handling the detergent, read the cautions carefully and use it as instructed. It is especially important to know what to do if the detergent solution comes in contact with your skin and eyes.• When handling the detergent, always wear appropriate personal protective equipment, such as eyewear, face mask, moisture-resistant clothing, and chemical-resistant gloves that fit properly and are long enough so that your skin and eyes is not exposed. All personal protective equipment should be inspected before use and replaced periodically before it is damaged.NOTE• The input of the user ID can be omitted by modifying the user ID input setting. For details, refer to Section 4.5, “User ID Setting”.• When the “Delete” button is pressed, the entered ID is deleted.• When the detergent lot and shelf-life management setting are activated, the touch screen displays the lot entry screen or shelf-life entry screen. For details, refer to “When entering the lot number of detergent and shelf life:” on page 372.1Inspect the detergent tank. If detergent is still present in the detergent tank, go to the procedure of “If the message screen “Message 093” is displayed again after replacing the detergent tank and restarting the reprocessing process:” on page 638 described in this section.2Replace the detergent tank as described in Section 5.8, “Inspecting the remaining detergent”.3Enter the operator’s user ID. For the detailed procedure, refer Section 3.6, “Entering ID” (If applicable). Then, press the “Restart” button to restart the process.Figure 13.29
63813.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13If the message screen “Message 093” is displayed again after replacing the detergent tank and restarting the reprocessing process:When the message screen “Detergent cannot be supplied” is displayed again after restarting the process, check that there is no irregularity regarding the following points.• The tube connected to the detergent tank is not kinked.• The connector of detergent tank is attached firmly.If any irregularity is found by inspection, correct these points and press start button to restart the process.If no irregularity is found during an inspection or the “Detergent cannot be supplied” is displayed again, internal problem with the reprocessor may be suspected. In this case, follow the procedure below.WARNING• Before handling the detergent, read the cautions carefully and use it as instructed. It is especially important to know what to do if the detergent solution comes in contact with your skin and eyes.• When handling the detergent, always wear appropriate personal protective equipment, such as eyewear, face mask, moisture-resistant clothing, and chemical-resistant gloves that fit properly and are long enough so that your skin and eyes is not exposed. All personal protective equipment should be inspected before use and replaced periodically before it is damaged.NOTEDetergent replacement indicator is turned off when running a reprocessing program after performing 1 to 11 below to fill detergent in the detergent supply piping.
13.2 Troubleshooting guide639OER-Elite OPERATION MANUALCh.13(a) Required itemsTab le 1 3. 3(b) Remedial actionsCheck Required itemsSyringeTube1Press the “Stop” button.Figure 13.302The touch screen displays the error code [E095]. Press the “OK” button. Then, troubleshooting guide is displayed on the touch screen.Figure 13.313Step on the foot pedal to open the lid.4Prepare the provided syringe and the provided tube.
64013.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.135Press the “Next” button.Figure 13.326Connect the provided syringe and tube.Figure 13.337Connect the tube to the detergent nozzle inside the reprocessing basin and suction it with the syringe until detergent comes out.Figure 13.34
13.2 Troubleshooting guide641OER-Elite OPERATION MANUALCh.138Pinch the tube at the closest point to the detergent nozzle with your fingers and disconnect the tube from the detergent nozzle.Figure 13.359Rinse the syringe and tube thoroughly in running water, dry them well, and store in a clean place.10 Press the “OK” button repeatedly until the error screen is closed.Figure 13.3611 Start the reprocessing process from the beginning.12 If the reprocessing process is successfully completed, the problem is resolved and reprocessing process was performed with the endoscopes. If the “Message 093” is displayed again, the reprocessor component malfunction is suspected. In this case, do not use the reprocessor and contact Olympus. Remove endoscope(s) from the basin and perform reprocessing process again by another method.
64213.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13Other problems and remedial actionsProblem Possible causes Remedial actionsThe odor of the disinfectant solution is stronger than expected.• The gas filter has expired.• Disinfectant solution is leaking.• Replace the gas filter as described in Section 8.6, “Replacing the gas filter (MAJ-822)”. If the problem persists after replacement, contact Olympus.• Check if disinfectant solution is leaking. If it is, do not use the reprocessor and contact Olympus.Water leak from the reprocessor.Improper installation of water filter housing.Stop the current process and reattach the water filter housing as described in Section 8.4, “Replacing the water filter (MAJ-824 or MAJ-2318)”.However, the STOP button may not work if stopping the water might cause a hazardous situation, for example during disinfectant solution replacement. In this case, close the water faucet, tighten the water filter housing, then open the water faucet again and continue the process. If an error code is displayed, take the appropriate remedial action for that error.Irregularity in the lid packing Inspect the lid packing.oRefer to Section 5.5, “Inspecting the lid and lid packing”.If the lid packing is abnormal, contact Olympus.Internal problem with the reprocessor.Close the water faucet, set the power switch to OFF, disconnect the power cord plug from the wall mains outlet and contact Olympus.Fluid leak from the disinfectant removal port.Something is clogging the disinfectant removal port.Connect the drain connector as described in to Section 3.7, “Checking the MRC level and entering the check result”, push the valve on the connector several times so that the material stuck in the port is removed. If leakage is still detected, attach the rubber cap to the disinfectant solution drain port and contact Olympus.
13.2 Troubleshooting guide643OER-Elite OPERATION MANUALCh.13Flow of the water in the reprocessing basin is weaker than before.Incomplete opening of the water faucet.Open the water faucet fully.Improper installation of water filter.Reattach the water filter as described in Section 8.4, “Replacing the water filter (MAJ-824 or MAJ-2318)”.Clogging of water filter. Replace the water filter as described in Section 8.4, “Replacing the water filter (MAJ-824 or MAJ-2318)”.Clogging of the mesh filter in the water supply hose connector.Clean the mesh filter as described in Section 9.7, “Cleaning the mesh filter in the water supply hose connector”.The disinfectant bottle drawer cannot be pulled out.The disinfectant bottle drawer is locked.After draining the disinfectant solution from the reprocessor as described in Section 8.2, “Replacing the disinfectant solution”, perform “Load LCG” and unlock the disinfectant bottle drawer.Connecting tube cannot be connected.Not using the appropriate connecting tube.Consult the List Of Compatible Endoscopes/Connecting Tubes <OER-Elite>.The lid cannot be closed. The lid is locked. Step on the foot pedal to unlock the lid.Internal components are pressing against the lid.Check that the lid is not pushed by the fluid level sensor, washing case or endoscope inside the basin. If it is pushed out of position, correct its positioning.Disinfectant solution remains in the reprocessing basin.STOP button was pressed during disinfection process to force it to stop.If an error code is displayed, take the corresponding remedial action. Collect or drain the disinfectant solution and then rinse the basin as described in Section 7.5, “Rinse”. Since the endoscopes may not be properly disinfected, they should be put through the reprocessing process again from the beginning.Cleaning fluid remains in the reprocessing basin.STOP button was pressed during cleaning process to force it to stop.Rinse the basin as described in Section 7.5, “Rinse”. Since the endoscopes may not be properly disinfected, they should be put through the reprocessing process again from the beginning.Problem Possible causes Remedial actions
64413.2 Troubleshooting guideOER-Elite OPERATION MANUALCh.13The touch screen appears to turn off completely during an operation.• The Heat LCG Timer process started.• Power cord is disconnected from the power outlet.• Circuit breaker is activated.• A power failure has occurred.• Check that the Heat LCG Timer indicator on the control panel is lit. If it is lit, pressing the touch screen or a button will cause the touch screen to display a screen.• Perform the checks described in Section 5.3, “Inspecting the power activation”. Error code [E041] will be displayed when the reprocessor is turned ON.After checking, release the error code. Note that, however, any button is disabled during collection of the disinfectant solution from the reprocessing basin to the tank.Power indicator does not light when the reprocessor is turned on.Power switch is set to ON immediately after it was set to OFF.Set the power switch to OFF, wait for a few seconds or more and set the power switch to ON again.Reprocessing operator feels sick during work.The operator may be allergic to the disinfectant, detergent or alcohol.Stop doing any reprocessing, move away from the reprocessor and consult a medical specialist.Disinfectant solution is judged to be ineffective with the test strip.Expiration of the service life of the disinfectant solution.Replace the solution as described in Section 8.2, “Replacing the disinfectant solution”.Bacteria were detected as a result of culture test of a reprocessed endoscope.• Expiration of service life of filters, degradation of disinfectant solution, etc.• Water supply piping is not disinfected.Inspect the reprocessor as described in Chapter 5, “Inspection and Preparation Before Use”, preclean the endoscope and put it through the reprocessing process again from the beginning. If bacteria are detected again in the next culture test, contact Olympus.Bacteria were detected as a result of culture test of rinse water collected from the reprocessor.• Expiration of service life of filters, degradation of disinfectant solution, etc.• Water supply piping is not disinfected.Inspect the reprocessor as described in Chapter 5, “Inspection and Preparation Before Use”. If bacteria are detected again in the next culture test, contact Olympus.Endoscopes were not precleaned before being reprocessed. –Inspect the reprocessor as described in Chapter 5, “Inspection and Preparation Before Use”. Then, preclean the endoscopes and reprocess them again from the beginning.Printed paper is not output from the printer.• Printer paper roll has run out.• Paper jam.Take the remedial action by referring to the MAJ-2144 instruction manual for printer.Abnormal noise from the reprocessor.Internal problem with the reprocessor.Contact Olympus.Non-connected status is displayed (yellow indicator) on the touch screen.The connection with a network is not established.Check the connection of the network cable and the network setting of the reprocessor.Problem Possible causes Remedial actions
13.3 OER-Elite return645OER-Elite OPERATION MANUALCh.13Contact Olympus for information about packing and shipping the reprocessor for return. When you return the reprocessor, include a description of the malfunction or damage and how it occurred.Auto leak test conducted independently gave the “Leaked” judgment result.• Water leak of an endoscope.• Irregularity with the leak test air tube.• Endoscope is warm because it was cleaned manually with warm water.• Check if the endoscope has a water leaking point with manual leak test. If a leak point exists on the outer surface, stop it with a piece of tape and perform the leaking scope decontamination as described in Section 7.15, “Leaking scope decontamination”. After it, return the endoscope to Olympus for servicing. If the leak point exists in a position where taping is impossible, do not execute this process but contact Olympus.• Perform the ALT self-check. If an error is detected, contact Olympus.Do not know how to return the leaking endoscope to Olympus for servicing.– Check the leaking point of the endoscope with manual leak test. If the leak point exists on the outer surface, stop it with a piece of tape and perform the leaking scope decontamination as described in Section 7.15, “Leaking scope decontamination”. After it, return the endoscope to Olympus for servicing. If the leak point exists in a position where taping is impossible, do not execute this process but contact Olympus.Leaking scope decontamination is performed without attaching connector jigs.Connection of connector jigs was forgotten.If the leaking scope decontamination is already completed, return the endoscope to Olympus for servicing. Do not retry leaking scope decontamination, as this may cause excessive water leak inside the endoscope.Error code [E115] is generated several times during leaking scope decontamination.Water leak from the endoscope is serious (the leaking hole is big).Contact Olympus.13.3 OER-Elite returnProblem Possible causes Remedial actions
64613.3 OER-Elite returnOER-Elite OPERATION MANUALCh.13
App.System chart647OER-Elite OPERATION MANUALAppendixThe recommended combinations of equipment and accessories that can be used with this reprocessor are listed below. Some items may not be available in some areas. New products released after the introduction of this reprocessor may also be compatible with this equipment. For further details, contact Olympus.WARNINGIf combinations of equipment other than those shown below are used, Olympus cannot guarantee that the device will perform as expected. Nor can Olympus guarantee the safety of patients and operators. Nor can the durability of the device be guaranteed when nondesignated equipment is used. Any damage resulting from improper combinations will not be serviced or repaired free of charge.System chart
App.648System chartOER-Elite OPERATION MANUALReprocessing basinOlympus endoscopes (for compatible models, refer to the separately provided “List of Compatible Endoscopes/Connecting Tubes <OER-Elite>”).Connecting tubes (for the connecting tubes compatible to each endoscope model, also refer to the “List of Compatible Endoscopes/Connecting Tubes <OER-Elite>”).Connector hanger (MAJ-865)Olympus-validated detergentWater filter (MAJ-824 or MAJ-2318)AlcoholGas filter (MAJ-822)Air filter (MAJ-823)Endoscope reprocessor (OER-Elite)Water pre-filtration systemUsing a prefilter can extend the life of the water filter.Olympus-validated concentrated disinfectant solution1  Always use Olympus designated external devices for communication with OER-Elite. If not, Olympus cannot guarantee that the device will perform as expected. For more details on external devices available in your area, contact Olympus.Olympus-designated external devices (for communication with OER-Elite)*1Portable memory (MAJ-1925)Bar code reader (MAJ-2130)Printer set (MAJ-2144)This device is connected to back side.
App.Specifications649OER-Elite OPERATION MANUALShipping environmentOperating environmentNOTE• Purified water is water that has been produced through the methods of either reverse osmosis (RO), deionization (DI), distillation or other methods that meet USP standards to remove impurities.• When the water supply flow is less than 18 l/min, the reprocessing process time is extended.SpecificationsShipping environmentAmbient temperatures –47 to +60qC (–52 to +140qF)Relative humidity 10 – 95Atmospheric pressure 700 – 1060 hPaOperating environmentAmbient temperatures 10 – 40qC (50 – 104qF)Relative humidity 30 – 85Elevation 3000 metersDesigned for use IndoorsWater supply flow • Minimum quantity of water supply (from the water supply/circulation nozzle) to complete a reprocessing process: 6 L/min or more• Recommended quantity of water supply (from the water supply/circulation nozzle) to achieve the minimum reprocessing time: 18 L/minWater supply pressure 0.1 – 0.5 MPa (include water hammer)Water supply temperature 5 – 28qC (41 – 82qF)Water type Potable water, softened water or purified waterWater hardness 0 – 400 ppm0 – 150 ppm (Recommended value) *1*1 The recommended value is in reference to AAMI TIR34: 2014 “Water for the reprocessing of medical devices”.
App.650SpecificationsOER-Elite OPERATION MANUALSpecificationsCompatible endoscopes Olympus flexible endoscopes(Refer to the “List of Compatible Endoscopes/Connecting Tubes <OER-Elite>” for details)Note:The OER-Elite does not sterilize endoscopes. Therefore, after cleaning/disinfecting an endoscope that requires sterilization, always sterilize the endoscope as instructed in its instruction manual.Number of reprocessed endoscopes Max. 2 (1 with certain models)(Refer to the “List of Compatible Endoscopes/Connecting Tubes <OER-Elite>” for combinations)Cleaning method Exterior surfaces Ultrasonic cleaning, turbulent bathChannel interiors Fluid flushingValves Ultrasonic cleaning, fluid flushingDisinfection method Exterior surfaces Disinfectant solution immersionChannel interiors Disinfectant solution flushing and fillingValves Disinfectant solution immersionCleaning time setting 3 minutesDisinfection time setting 7 minutesHeat LCG setting Heating immediately before disinfection process in a reprocessing program.20qC (68qF)(If the temperature of disinfectant solution is below 20qC, it is heated to 20qC (68qF).)Heating before the start of a reprocessing program.22qC (72qF)(If the temperature of disinfectant solution is below 20qC, it is heated to 22qC (72qF).)Heating by the timer.22qC (72qF)(If the temperature of disinfectant solution is below 20qC, it is heated to 22qC (72qF).)Heat LCG method Built-in heater in the disinfectant solution tank.1 Heating before the start of a reprocessing program.2 Heating by the timer.Built-in heater in the reprocessing basin.1 Heating immediately before disinfection process in a reprocessing program.Water discharge method Forced draining using a pump (The top of the drain hose should be 60 cm or less. A floor drain is recommended.)
App.Specifications651OER-Elite OPERATION MANUALDisinfectant solution discharge method 1 Draining through disinfectant collection hose2 Draining through drain hoseReprocessing basin capacity Approximately 17.9 LDisinfectant solution tank capacity Approximately 21.1 LDisinfectant solution Olympus-validated disinfectant solution (Refer to Section 2.8, “Consumable accessories (Optional)” for Olympus-validated disinfectant solution.)Detergent Olympus-validated detergent (Refer to Section 2.8, “Consumable accessories (Optional)” for Olympus-validated disinfectant solution.)Manual leak test Visual inspection of bubble during immersion.Auto leak test Detects automatically for a pin hole by feeding air into the endoscope and computing a changed air pressure.Alcohol flush Automatic flushing/draining using a pump and compressorDimensions 500 (W) u 945 (H) u 774 (D) mmWeight 133 kg (dry condition)Power supply Voltage 120 V ACFrequency 60 HzRated input 660 VAVoltage fluctuation Within r10Frequency fluctuationWithin r1HzFuse rating 10 A, 250 VFuse size ø5 × 20 mmUltrasonic wave Frequency 36 r2kHzPower 100 WEMC Applied standard IEC 61326-1: 2012• CISPR 11 of emission:Group 1, Class BElectrical safety Applied standard UL 61010-1: 2012CSA C22.2 No. 61010-1: 2012CSA C22.2 No. 61010-2-040: 2016IEC 61010-1: 2010IEC 61010-2-040: 2015• Installation category: IIPollution degree: 2Radio transmitter Compliance ISO/IEC 18000-3 (Mode1)Center frequency 13.56 MHzModulation ASKEffective radiated power 200 mWr20
App.652SpecificationsOER-Elite OPERATION MANUALNOTEThis equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:Reorient or relocate the receiving antenna.Increase the separation between the equipment and receiver.Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.Consult the dealer or an experienced radio/TV technician for help.UDI label Indication A label required by some countries’ regulations regarding identification of medical device also known as Unique Device Identification (UDI).The following information is being coded in the 2-dimensional barcode (GS1 Date Matrix):(01) 14-digit GS1 Global Trade Item Number;(11) 6-digit date of manufacture;(21) 7-digit serial number.FCC and IC informationThis equipment complies with part15 of the FCC rules and the IC RSS210. FCC ID: S8Q-RU2020IC: 4763B-RU2020FCC WARNINGChange or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
App.EMC information653OER-Elite OPERATION MANUALGuidance and manufacturer’s declaration — Electromagnetic emissionsThis model is intended for use in the electromagnetic environment specified below. The customer or the user of this model should assure that it is used in such an environment.EMC informationEmissions test Compliance Electromagnetic environment — GuidanceRF emissionsCISPR 11Group 1 This instrument uses RF (Radio Frequency) energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.Radiated emissionsCISPR 11Class B This instrument’s RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.Main terminal conducted emissionsCISPR 11Harmonic emissionsIEC 61000-3-2Not applicable Power supply specification of this instrument is less than 220 VAC, and this instrument is exempt from requirements of IEC 61000-3-2.Voltage fluctuations/flicker emissionsIEC 61000-3-3Not applicable Power supply specification of this instrument is less than 220 VAC, and this instrument is exempt from requirements of IEC 61000-3-3.
App.654EMC informationOER-Elite OPERATION MANUALGuidance and manufacturer’s declaration — Electromagnetic immunityThis model is intended for use in the electromagnetic environment specified below. The customer or the user of this model should assure that it is used in such an environment.Immunity test IEC 61326-1 test level Compliance level Electromagnetic environment — GuidanceElectrostatic discharge (ESD)IEC 61000-4-2Contact:r2, r4kVAir:r2, r4, r8kVSame as left Floors should be made of wood, concrete, or ceramic tile that hardly produces static. If floors are covered with synthetic material that tends to produce static, the relative humidity should be at least 30%.Electrical fast transient/burstIEC 61000-4-4r2kV for power supply linesr1kV for input/output linesSame as left Mains power quality should be that of a typical commercial (original condition feeding the facilities) or hospital environment.SurgeIEC 61000-4-5Differential mode:r1kVCommon mode:r1, r2kVSame as left Mains power quality should be that of a typical commercial or hospital environment.Voltage dips, short interruptions, and voltage variations on power supply input linesIEC 61000-4-110% UT(100% dip in UT)for 1 cycleSame as left Mains power quality should be that of a typical commercial or hospital environment. If the user of this instrument requires continued operation during power mains interruptions, it is recommended that this instrument be powered from an uninterruptible power supply or a battery.40% UT(60% dip in UT)for 200 ms70% UT(30% dip in UT)for 500 ms0% UT(100% dip in UT)for 5 secondsPower frequency (50/60 Hz) magnetic fieldIEC 61000-4-830 A/m Same as left It is recommended to use this instrument by maintaining enough distance from any equipment that operates with high current.Definition UT is the a.c. mains voltage prior to application of the test level.
App.EMC information655OER-Elite OPERATION MANUALGuidance and manufacturer’s declaration — Electromagnetic immunityThis model is intended for use in the electromagnetic environment specified below. The customer or the user of this model should assure that it is used in such an environment.Portable and mobile RF communications equipment should be used no closer to any part of this model, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.NOTE• At 80 MHz and 800 MHz, the higher frequency range applies.• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.• Electromagnetic interference may occur in the vicinity of high-frequency electrosurgical equipment and/or other equipment marked with the following symbol:Immunity test IEC 61326-1 test levelCompliance level Electromagnetic environment — GuidanceRecommended separation distance(V1=3 according to the compliance level)Conducted RFIEC 61000-4-63 Vrms(150 kHz – 80 MHz)3V (V1)Recommended separation distance(E1=10, E2=3, E3=1 according to the compliance level)Radiated RFIEC 61000-4-310V/m (80MHz – 1GHz)3V/m (1.4 GHz – 2 GHz)1V/m (2 GHz – 2.7 GHz)10 V/m (E1)3 V/m (E2)1 V/m (E3)80 MHz – 800 MHz800 MHz – 1 GHz1.4 GHz – 2 GHz2GHz – 2.7GHzDefinition Where “P” is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and “d” is the recommended separation distance in meters (m).d3.5V1------- P=d3.5E1------- P=d7E1------P=d7E2------P=d7E3------P=
App.656EMC informationOER-Elite OPERATION MANUALNOTE• Field strength from fixed RF transmitters as determined by an electromagnetic site surveya) should be less than the compliance level in each frequency rangeb).a) Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which this model is used exceeds the applicable RF compliance level above, this model should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating this model.b) Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
App.EMC information657OER-Elite OPERATION MANUALRecommended separation distances between portable and mobile RF communications equipment and this modelThis model is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this model can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and this model as recommended below, according to the maximum output power of the communications equipment.NOTE• At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.Rated maximum output power of transmitterP (W)Separation distance according to frequency of transmitter (m)(calculated as V1=3, E1=10, E2=3 and E3=1)150 kHz – 80 MHz 80 MHz – 800 MHz 800MHz – 1GHz0.01 0.12 0.04 0.070.1 0.38 0.12 0.2311.2 0.35 0.7010 3.8 1.2 2.3100 12 3.5 7Rated maximum output power of transmitterP (W)Separation distance according to frequency of transmitter (m)(calculated as V1=3, E1=10, E2=3 and E3=1)1.4 GHz – 2 GHz 2 GHz – 2.7 GHz0.01 0.24 0.70.1 0.76 2.312.4 710 7.6 23100 24 70Others For transmitters rated at a maximum output power not listed above, the recommended separation distance ‘d’ in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where ‘p’ is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.d1.2 P=d0.35 P=d0.7 P=d2.4 P=d7P=
App.658License information of Open Source SoftwareOER-Elite OPERATION MANUALGNU GPLThis product uses software modules licensed under the terms of the GNU GENERAL PUBLIC LICENSE(“GPL”).Please see the following table for the software module made the target.zEnd User License Agreement of Open Source Software modulesThis section must not be translated by any language and must not be changed from original text.Because these license description must be described original description in English according to these license.GNU GENERAL PUBLIC LICENSE Version 2 with ExceptionAbout libgcc and libstdc++, those are licensed under the terms of the GNU GENERAL PUBLIC LICENSE Version 2 with special exception or additional permission to show in the following.a) libstdc++The libstdc++ source code is distributed under the GNU General Public License version 2, with the so-called “Runtime Exception” as follows (or see any header or implementation file). As a special exception, you may use this file as part of a free software library without restriction. Specifically, if other files instantiate templates or use macros or inline functions from this file, or you compile this file and link it with other files to produce an executable, this file does not by itself cause the resulting executable to be covered by the GNU General Public License. This exception does not however invalidate any other reasons why the executable file might be covered by the GNU General Public License.b) libgccGCC is free software; you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation; either version 2, or (at your option) any later version.License information of Open Source SoftwareSoftware modules concerned End User License Agreement of software moduleslibgcc GNU GENERAL PUBLIC LICENSE Version 2 with Exceptionlibstdc++
App.License information of Open Source Software659OER-Elite OPERATION MANUALIn addition to the permissions in the GNU General Public License, the Free Software Foundation gives you unlimited permission to link the compiled version of this file into combinations with other programs, and to distribute those combinations without any restriction coming from the use of this file. (The General Public License restrictions do apply in other respects; for example, they cover modification of the file, and distribution when not linked into a combined executable.) GCC is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General Public License for more details.Others• This instrument contains software modules which are not licensed under the terms of GPL. This section must not be translated by any language and must not be changed from original text. Because these license description must be described original description in English according to these license.• The Licensed Software may include (i) some open source software and (ii) other third party software (collectively, “Third Party Software”). Regarding the use of the Third Party Software, you shall follow the license terms and conditions applied to such Third Party Software. Any part of this License Agreement does not restrict, change or alter any of your rights and obligations imposed by the license terms and conditions of the Third Party Software.newlibThis product uses the open source software program “newlib” which applies the terms and conditions provided by owner of the copyright to the “newlib”.The newlib subdirectory is a collection of software from several sources.Each file may have its own copyright/license that is embedded in the source file.(1) Red Hat IncorporatedCopyright (c) 1994-2007 Red Hat, Inc. All rights reserved.This copyrighted material is made available to anyone wishing to use, modify, copy, or redistribute it subject to the terms and conditions of the BSD License. This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY expressed or implied, including the implied warranties of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. A copy of this license is available at http://www.opensource.org/licenses. Any Red Hat trademarks that are incorporated in the source code or documentation are not subject to the BSD License and may only be used or replicated with the express permission of Red Hat, Inc.(2) University of California, BerkeleyCopyright (c) 1981-2000 The Regents of the University of California.All rights reserved.
App.660License information of Open Source SoftwareOER-Elite OPERATION MANUALRedistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:* Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.* Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.* Neither the name of the University nor the names of its contributors may be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED.IN NO EVENT SHALL THE COPYRIGHT OWNER OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITYOF SUCH DAMAGE.(3) David M. Gay at AT&TThe author of this software is David M. Gay.Copyright (c) 1991 by AT&T.Permission to use, copy, modify, and distribute this software for any purpose without fee is hereby granted, provided that this entire notice is included in all copies of any software which is or includes a copy or modification of this software and in all copies of the supporting documentation for such software.THIS SOFTWARE IS BEING PROVIDED “AS IS”, WITHOUT ANY EXPRESS OR IMPLIED WARRANTY. IN PARTICULAR, NEITHER THE AUTHOR NOR AT&T MAKES ANY REPRESENTATION OR WARRANTY OF ANY KIND CONCERNING THE MERCHANTABILITY OF THIS SOFTWARE OR ITS FITNESS FOR ANY PARTICULAR PURPOSE.-------------------------------------------------------------------The author of this software is David M. Gay.Copyright (C) 1998-2001 by Lucent TechnologiesAll Rights Reserved
App.License information of Open Source Software661OER-Elite OPERATION MANUALPermission to use, copy, modify, and distribute this software and its documentation for any purpose and without fee is hereby granted, provided that the above copyright notice appear in all copies and that both that the copyright notice and this permission notice and warranty disclaimer appear in supporting documentation, and that the name of Lucent or any of its entities not be used in advertising or publicity pertaining to distribution of the software without specific, written prior permission.LUCENT DISCLAIMS ALL WARRANTIES WITH REGARD TO THIS SOFTWARE, INCLUDING ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS.IN NO EVENT SHALL LUCENT OR ANY OF ITS ENTITIES BE LIABLE FOR ANY SPECIAL, INDIRECT OR CONSEQUENTIAL DAMAGES OR ANY DAMAGES WHATSOEVER RESULTING FROM LOSS OF USE, DATA OR PROFITS, WHETHER IN AN ACTION OF CONTRACT, NEGLIGENCE OR OTHER TORTIOUS ACTION, ARISING OUT OF OR IN CONNECTION WITH THE USE OR PERFORMANCE OF THIS SOFTWARE.(4) Advanced Micro DevicesCopyright 1989, 1990 Advanced Micro Devices, Inc.This software is the property of Advanced Micro Devices, Inc (AMD) which specifically grants the user the right to modify, use and distribute this software provided this notice is not removed or altered. All other rights are reserved by AMD.AMD MAKES NO WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, WITH REGARD TO THIS SOFTWARE. IN NO EVENT SHALL AMD BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THE FURNISHING, PERFORMANCE, OR USE OF THIS SOFTWARE.So that all may benefit from your experience, please report any problems or suggestions about this software to the 29K Technical Support Center at 800-29-29-AMD (800-292-9263) in the USA, or 0800-89-1131 in the UK, or 0031-11-1129 in Japan, toll free. The direct dial number is 512-462-4118.Advanced Micro Devices, Inc.29K Support ProductsMail Stop 5735900 E. Ben White Blvd.Austin, TX 78741800-292-9263(5) C.W. SandmannCopyright (C) 1993 C.W. SandmannThis file may be freely distributed as long as the author's name remains.
App.662License information of Open Source SoftwareOER-Elite OPERATION MANUAL(6) Eric Backus(C) Copyright 1992 Eric BackusThis software may be used freely so long as this copyright notice is left intact.There is no warrantee on this software.(7) Sun MicrosystemsCopyright (C) 1993 by Sun Microsystems, Inc. All rights reserved.Developed at SunPro, a Sun Microsystems, Inc. business.Permission to use, copy, modify, and distribute this software is freely granted, provided that this notice is preserved.(8) Hewlett Packard(c) Copyright 1986 HEWLETT-PACKARD COMPANYTo anyone who acknowledges that this file is provided “AS IS” without any express or implied warranty: permission to use, copy, modify, and distribute this file for any purpose is hereby granted without fee, provided that the above copyright notice and this notice appears in all copies, and that the name of Hewlett-Packard Company not be used in advertising or publicity pertaining to distribution of the software without specific, written prior permission.Hewlett-Packard Company makes no representations about the suitability of this software for any purpose.(9) Hans-Peter NilssonCopyright (C) 2001 Hans-Peter NilssonPermission to use, copy, modify, and distribute this software is freely granted, provided that the above copyright notice, this notice and the following disclaimer are preserved with no changes.THIS SOFTWARE IS PROVIDED “AS IS” AND WITHOUT ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, WITHOUT LIMITATION, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.(11) Christopher G. DemetriouCopyright (c) 2001 Christopher G. DemetriouAll rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:
App.License information of Open Source Software663OER-Elite OPERATION MANUAL1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.3. The name of the author may not be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED BY THE AUTHOR “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED.IN NO EVENT SHALL THE AUTHOR BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(12) SuperH, Inc.Copyright 2002 SuperH, Inc. All rights reservedThis software is the property of SuperH, Inc (SuperH) which specifically grants the user the right to modify, use and distribute this software provided this notice is not removed or altered. All other rights are reserved by SuperH.SUPERH MAKES NO WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, WITH REGARD TO THIS SOFTWARE. IN NO EVENT SHALL SUPERH BE LIABLE FOR INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THE FURNISHING, PERFORMANCE, OR USE OF THIS SOFTWARE.So that all may benefit from your experience, please report any problems or suggestions about this software to the SuperH Support Center via e-mail at softwaresupport@superh.com .SuperH, Inc.405 River Oaks ParkwaySan JoseCA 95134USA(13) Royal Institute of TechnologyCopyright (c) 1999 Kungliga Tekniska H skolan (Royal Institute of Technology, Stockholm, Sweden).All rights reserved.
App.664License information of Open Source SoftwareOER-Elite OPERATION MANUALRedistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.3. Neither the name of KTH nor the names of its contributors may be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED BY KTH AND ITS CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL KTH OR ITS CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(14) Alexey ZelkinCopyright (c) 2000, 2001 Alexey Zelkin <phantom@FreeBSD.org>All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
App.License information of Open Source Software665OER-Elite OPERATION MANUAL(15) Andrey A. ChernovCopyright (C) 1997 by Andrey A. Chernov, Moscow, Russia.All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE REGENTS OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(16) FreeBSDCopyright (c) 1997-2002 FreeBSD Project.All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
App.666License information of Open Source SoftwareOER-Elite OPERATION MANUAL(17) S. L. MoshierAuthor: S. L. Moshier.Copyright (c) 1984,2000 S.L. MoshierPermission to use, copy, modify, and distribute this software for any purpose without fee is hereby granted, provided that this entire notice is included in all copies of any software which is or includes a copy or modification of this software and in all copies of the supporting documentation for such software.THIS SOFTWARE IS BEING PROVIDED “AS IS”, WITHOUT ANY EXPRESS OR IMPLIED WARRANTY. IN PARTICULAR, THE AUTHOR MAKES NO REPRESENTATION OR WARRANTY OF ANY KIND CONCERNING THE MERCHANTABILITY OF THIS SOFTWARE OR ITS FITNESS FOR ANY PARTICULAR PURPOSE.(18) Citrus ProjectCopyright (c)1999 Citrus Project, All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(19) Todd C. MillerCopyright (c) 1998 Todd C. Miller <Todd.Miller@courtesan.com>All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:
App.License information of Open Source Software667OER-Elite OPERATION MANUAL1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.3. The name of the author may not be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(26) Mike BarcroftCopyright (c) 2001 Mike Barcroft <mike@FreeBSD.org>All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
App.668License information of Open Source SoftwareOER-Elite OPERATION MANUAL(27) Konstantin Chuguev (--enable-newlib-iconv)Copyright (c) 1999, 2000 Konstantin Chuguev.All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.iconv (Charset Conversion Library) v2.0(28) Artem Bityuckiy (--enable-newlib-iconv)Copyright (c) 2003, Artem B. Bityuckiy, SoftMine Corporation.Rights transferred to Franklin Electronic Publishers.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.
App.License information of Open Source Software669OER-Elite OPERATION MANUALTHIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.libhal_bspThis product uses the open source software program “libhal_bsp” which applies the terms and conditions provided by owner of the copyright to the “libhal_bsp”.Copyright (c) 2003 Altera Corporation, San Jose, California, USA.All rights reserved.Permission is hereby granted, free of charge, to any person obtaining a copy of this software and associated documentation files (the “Software”), to deal in the Software without restriction, including without limitation the rights to use, copy, modify, merge, publish, distribute, sublicense, and/or sell copies of the Software, and to permit persons to whom the Software is furnished to do so, subject to the following conditions:The above copyright notice and this permission notice shall be included in all copies or substantial portions of the Software.THE SOFTWARE IS PROVIDED “AS IS”, WITHOUT WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT. IN NO EVENT SHALL THE AUTHORS OR COPYRIGHT HOLDERS BE LIABLE FOR ANY CLAIM, DAMAGES OR OTHER LIABILITY, WHETHER IN AN ACTION OF CONTRACT, TORT OR OTHERWISE, ARISING FROM, OUT OF OR IN CONNECTION WITH THE SOFTWARE OR THE USE OR OTHER DEALINGS IN THE SOFTWARE.
App.670License information of Open Source SoftwareOER-Elite OPERATION MANUALzEnd User License Agreement of Open Source Software modules• This instrument contains software modules which are not licensed under the terms of GPL. This section must not be translated by any language and must not be changed from original text. Because these license description must be described original description in English according to these license.• The Licensed Software may include (i) some open source software and (ii) other third party software (collectively, “Third Party Software”). Regarding the use of the Third Party Software, you shall follow the license terms and conditions applied to such Third Party Software. Any part of this License Agreement does not restrict, change or alter any of your rights and obligations imposed by the license terms and conditions of the Third Party Software.newlibMAJ-2130, which is “NLV-3101” manufactured by OPTO ELECTORONICS CO.,LTD, uses the open source software program “newlib” which applies the terms and conditions provided by owner of the copyright to the “newlib”.The newlib subdirectory is a collection of software from several sources.Each file may have its own copyright/license that is embedded in the source file(s), the following copyright notices will apply to the contents of the newlib subdirectory:(1) Red Hat IncorporatedCopyright (c) 1994-2009 Red Hat, Inc. All rights reserved.This copyrighted material is made available to anyone wishing to use, modify, copy, or redistribute it subject to the terms and conditions of the BSD License. This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY expressed or implied, including the implied warranties of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. A copy of this license is available at http://www.opensource.org/licenses. Any Red Hat trademarks that are incorporated in the source code or documentation are not subject to the BSD License and may only be used or replicated with the express permission of Red Hat, Inc.(2) University of California, BerkeleyCopyright (c) 1981-2000 The Regents of the University of California.All rights reserved.License information of Open Source Software
App.License information of Open Source Software671OER-Elite OPERATION MANUALRedistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:* Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.* Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.* Neither the name of the University nor the names of its contributors may be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT OWNER OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITYOF SUCH DAMAGE.(3) David M. Gay at AT&TThe author of this software is David M. Gay.Copyright (c) 1991 by AT&T.Permission to use, copy, modify, and distribute this software for any purpose without fee is hereby granted, provided that this entire notice is included in all copies of any software which is or includes a copy or modification of this software and in all copies of the supporting documentation for such software.THIS SOFTWARE IS BEING PROVIDED “AS IS”, WITHOUT ANY EXPRESS OR IMPLIED WARRANTY. IN PARTICULAR, NEITHER THE AUTHOR NOR AT&T MAKES ANY REPRESENTATION OR WARRANTY OF ANY KIND CONCERNING THE MERCHANTABILITY OF THIS SOFTWARE OR ITS FITNESS FOR ANY PARTICULAR PURPOSE.-------------------------------------------------------------------The author of this software is David M. Gay.Copyright (C) 1998-2001 by Lucent TechnologiesAll Rights Reserved
App.672License information of Open Source SoftwareOER-Elite OPERATION MANUALPermission to use, copy, modify, and distribute this software and its documentation for any purpose and without fee is hereby granted, provided that the above copyright notice appear in all copies and that both that the copyright notice and this permission notice and warranty disclaimer appear in supporting documentation, and that the name of Lucent or any of its entities not be used in advertising or publicity pertaining to distribution of the software without specific, written prior permission.LUCENT DISCLAIMS ALL WARRANTIES WITH REGARD TO THIS SOFTWARE, INCLUDING ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS.IN NO EVENT SHALL LUCENT OR ANY OF ITS ENTITIES BE LIABLE FOR ANY SPECIAL, INDIRECT OR CONSEQUENTIAL DAMAGES OR ANY DAMAGES WHATSOEVER RESULTING FROM LOSS OF USE, DATA OR PROFITS, WHETHER IN AN ACTION OF CONTRACT, NEGLIGENCE OR OTHER TORTIOUS ACTION, ARISING OUT OF OR IN CONNECTION WITH THE USE OR PERFORMANCE OF THIS SOFTWARE.(4) Advanced Micro DevicesCopyright 1989, 1990 Advanced Micro Devices, Inc.This software is the property of Advanced Micro Devices, Inc (AMD) which specifically grants the user the right to modify, use and distribute this software provided this notice is not removed or altered. All other rights are reserved by AMD.AMD MAKES NO WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, WITH REGARD TO THIS SOFTWARE. IN NO EVENT SHALL AMD BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THE FURNISHING, PERFORMANCE, OR USE OF THIS SOFTWARE.So that all may benefit from your experience, please report any problems or suggestions about this software to the 29K Technical Support Center at 800-29-29-AMD (800-292-9263) in the USA, or 0800-89-1131 in the UK, or 0031-11-1129 in Japan, toll free. The direct dial number is 512-462-4118.Advanced Micro Devices, Inc.29K Support ProductsMail Stop 5735900 E. Ben White Blvd.Austin, TX 78741800-292-9263
App.License information of Open Source Software673OER-Elite OPERATION MANUAL(7) Sun MicrosystemsCopyright (C) 1993 by Sun Microsystems, Inc. All rights reserved.Developed at SunPro, a Sun Microsystems, Inc. business.Permission to use, copy, modify, and distribute this software is freely granted, provided that this notice is preserved.(8) Hewlett Packard(c) Copyright 1986 HEWLETT-PACKARD COMPANYTo anyone who acknowledges that this file is provided “AS IS” without any express or implied warranty: permission to use, copy, modify, and distribute this file for any purpose is hereby granted without fee, provided that the above copyright notice and this notice appears in all copies, and that the name of Hewlett-Packard Company not be used in advertising or publicity pertaining to distribution of the software without specific, written prior permission.Hewlett-Packard Company makes no representations about the suitability of this software for any purpose.(9) Hans-Peter NilssonCopyright (C) 2001 Hans-Peter NilssonPermission to use, copy, modify, and distribute this software is freely granted, provided that the above copyright notice, this notice and the following disclaimer are preserved with no changes.THIS SOFTWARE IS PROVIDED “AS IS” AND WITHOUT ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, WITHOUT LIMITATION, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.(10) Stephane Carrez (m68hc11-elf/m68hc12-elf targets only)Copyright (C) 1999, 2000, 2001, 2002 Stephane Carrez (stcarrez@nerim.fr)The authors hereby grant permission to use, copy, modify, distribute, and license this software and its documentation for any purpose, provided that existing copyright notices are retained in all copies and that this notice is included verbatim in any distributions. No written agreement, license, or royalty fee is required for any of the authorized uses.Modifications to this software may be copyrighted by their authors and need not follow the licensing terms described here, provided that the new terms are clearly indicated on the first page of each file where they apply.
©2017 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved.No part of this publication may be reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP.OLYMPUS is a registered trademark of OLYMPUS CORPORATION.Trademarks, product names, logos, or trade names used in this document are generally registered trademarks or trademarks of each company.
 Printed in Japan 20170706 *0000GT9882  01Manufactured by2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, JapanFax: (042)646-2429  Telephone: (042)642-2111(Premises/Goods delivery) Wendenstrasse 14-18, 20097 Hamburg, Germany(Letters) Postfach 10 49 08, 20034 Hamburg, Germany3500 Corporate Parkway, P.O. Box 610, Center Valley, PA18034-0610, U.S.A.Fax: (484)896-7128  Telephone: (484)896-5000KeyMed House, Stock Road, Southend-on-Sea, Essex SS2 5QH, United KingdomFax: (01702)465677  Telephone: (01702)616333491B, River Valley Road #12-01/04, Valley Point Office Tower, Singapore 248373Fax: 6834-2438  Telephone: 6834-0010A8F, Ping An International Financial Center, No. 1-3, Xinyuan South Road,Chaoyang District, Beijing, 100027 P.R.C.Fax: (86)10-5976-1299  Telephone: (86)10-5819-9000Elektrozavodskaya str. 27 bld.8, 107023 Moscow, RussiaFax: (7)495-663-8486  Telephone: (7)495-926-70773 Acacia Place, Notting Hill, VIC 3168, AustraliaFax: (03)9543-1350  Telephone: (03)9265-54005301 Blue Lagoon Drive, Suite 290 Miami, FL 33126-2097, U.S.A.Fax: (305)261-4421  Telephone: (305)266-2332Distributed byOlympus Tower 9F, 446, Bongeunsa-ro, Gangnam-gu, Seoul, Korea 135-509Fax: (02)6255-3494  Telephone: (02)6255-3210Fax: (040)23773-4656  Telephone: (040)23773-0
App.674License information of Open Source SoftwareOER-Elite OPERATION MANUAL(11) Christopher G. DemetriouCopyright (c) 2001 Christopher G. DemetriouAll rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.3. The name of the author may not be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED BY THE AUTHOR “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED.IN NO EVENT SHALL THE AUTHOR BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(12) SuperH, Inc.Copyright 2002 SuperH, Inc. All rights reservedThis software is the property of SuperH, Inc (SuperH) which specifically grants the user the right to modify, use and distribute this software provided this notice is not removed or altered. All other rights are reserved by SuperH.SUPERH MAKES NO WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, WITH REGARD TO THIS SOFTWARE. IN NO EVENT SHALL SUPERH BE LIABLE FOR INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THE FURNISHING, PERFORMANCE, OR USE OF THIS SOFTWARE.So that all may benefit from your experience, please report any problems or suggestions about this software to the SuperH Support Center via e-mail at softwaresupport@superh.com .
App.License information of Open Source Software675OER-Elite OPERATION MANUALSuperH, Inc.405 River Oaks ParkwaySan JoseCA 95134USA(13) Royal Institute of TechnologyCopyright (c) 1999 Kungliga Tekniska H skolan (Royal Institute of Technology, Stockholm, Sweden).All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.3. Neither the name of KTH nor the names of its contributors may be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED BY KTH AND ITS CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL KTH OR ITS CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(14) Alexey ZelkinCopyright (c) 2000, 2001 Alexey Zelkin <phantom@FreeBSD.org>All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.
App.676License information of Open Source SoftwareOER-Elite OPERATION MANUALTHIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(15) Andrey A. ChernovCopyright (C) 1997 by Andrey A. Chernov, Moscow, Russia.All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE REGENTS OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(16) FreeBSDCopyright (c) 1997-2002 FreeBSD Project.All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:
App.License information of Open Source Software677OER-Elite OPERATION MANUAL1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(17) S. L. MoshierAuthor: S. L. Moshier.Copyright (c) 1984,2000 S.L. MoshierPermission to use, copy, modify, and distribute this software for any purpose without fee is hereby granted, provided that this entire notice is included in all copies of any software which is or includes a copy or modification of this software and in all copies of the supporting documentation for such software.THIS SOFTWARE IS BEING PROVIDED “AS IS”, WITHOUT ANY EXPRESS OR IMPLIED WARRANTY. IN PARTICULAR, THE AUTHOR MAKES NO REPRESENTATION OR WARRANTY OF ANY KIND CONCERNING THE MERCHANTABILITY OF THIS SOFTWARE OR ITS FITNESS FOR ANY PARTICULAR PURPOSE.(18) Citrus ProjectCopyright (c)1999 Citrus Project, All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.
App.678License information of Open Source SoftwareOER-Elite OPERATION MANUALTHIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(19) Todd C. MillerCopyright (c) 1998 Todd C. Miller <Todd.Miller@courtesan.com>All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.3. The name of the author may not be used to endorse or promote products derived from this software without specific prior written permission.THIS SOFTWARE IS PROVIDED “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(26) Mike BarcroftCopyright (c) 2001 Mike Barcroft <mike@FreeBSD.org>All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:
App.License information of Open Source Software679OER-Elite OPERATION MANUAL1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.(35) - ARM Ltd (arm and thumb variant targets only) Copyright (c) 2009 ARM Ltd All rights reserved. Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.3. The name of the company may not be used to endorse or promote products derived from this software without specific prior written permission. THIS SOFTWARE IS PROVIDED BY ARM LTD “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL ARM LTD BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
App.680License information of Open Source SoftwareOER-Elite OPERATION MANUAL(41) Ed Schouten - Free BSDCopyright (c) 2008 Ed Schouten <ed@FreeBSD.org>All rights reserved.Redistribution and use in source and binary forms, with or without modification, are permitted provided that the following conditions are met:1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.THIS SOFTWARE IS PROVIDED BY THE AUTHOR AND CONTRIBUTORS “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.

Navigation menu