Intromedic INTROMEDIC5 CAPSULE ENDOSCOPE SYSTEM User Manual

Intromedic Co., Ltd. CAPSULE ENDOSCOPE SYSTEM

User Manual

  User Manual
 Copyright© 2011 IntroMedic Co., Ltd.                   MM1100-U-1311 Version 3.00                                     Date: 2013-11-21   User Manual
  Warranty   Page 5  Trademarks MiroCam®,  0LUR9LHZ, IntroMedic, and the associated logos are the registered trademarks RUWUDGHPDUNVof IntroMedic Co., Ltd. © IntroMedic Co., Ltd. July 2007.  Except as required by applicable copyright laws; any use of the IntroMedic trademarks, or any reprinting, reproduction, modification, referencing and translations of the User Manual, without the prior written approval of IntroMedic Co., Ltd. is strictly prohibited.    Warranty Every effort has been made to ensure the information contained in this User Manual is accurate, and is believed to be correct at time of printing. IntroMedic reserves the right to change any content contained with this User Manual without prior notice.     IntroMedic Co., Ltd. warrants the product against defects in material and workmanship for a period of twelve (12) months from the date of sale, unless different local regulations apply. IntroMedic Co., Ltd. will repair or replace products that are ascertained by IntroMedic to have defects during the warranty period. IntroMedic Co., Ltd. is not liable for the defects occurred by misuse, careless handling, unauthorized modifications or erroneous use, or any use that is non-compliant with instructions detailed within this User Manual. This includes use of the product in non-appropriate locations or conditions. Any other warranties are neither represented here nor recognized by implication.    To validate the warranty, please complete product registration with the local authorized IntroMedic distributor.
   Warranty   Page 6 Exclusive warranty service The warranty service provided hereby is applicable exclusively to the purchaser of the product. IntroMedic will only warranty the product for purposes and usage as defined in this User Manual. Any usage not heeding the warnings, cautions and recommended usages as defined in this manual will nullify the warranty.    Support For warranty or repair service please contact the local authorized IntroMedic distributor.  For customer service or support please contact your point of purchase or IntroMedic Co., Ltd. Service agreements are only applicable to products of IntroMedic Co., Ltd.    IntroMedic Customer Service TEL: 82-2-801-9300 FAX: 82-2-801-9330 http://www.intromedic.com E-mail: helpdesk@intromedic.com  Safety Non-compliance with WKH XVHU¶V PDQXDO XQDXWKRUL]HG PRGLILFDWLRQV RI WKHproduct or replacement of parts, and/or opening of the product casing is prohibited and may be hazardous.  Declaration on translation of labeling   Based on buyer's request, IntroMedic will provide the labeling, such as ID label, Instruction for use and other promotional materials, which is translated into the national language(s) of European countries. And it will be evaluated by the local language expert, and will be confirmed by the native speaker of the local language.
  Contents   Page 7  CONTENTS  1. SAFETY INFORMATION  1-3 1.1 Warnings 1-3 1.2 Symbols for Safety  1-5 1.3 Function Symbols  1-6 1.4 Notes for Safe Use  1-11  2. SYSTEM OVERVIEW  2-3 2.1 Intended Purpose  2-3 2.2 Observable Diseases  2-4 2.3 Product Warranty Period  2-6 2.4 Product Specification  2-7  3. PERFORMING CAPSULE ENDOSCOPY  3-3 3.1 General Comments  3-3 3.2 Safety Warnings  3-3 3.3 Examination Preparation  3-4 3.4 Patient Preparation  3-7 3.5 Sensor Placement & Capsule Ingestion  3-9 3.6 Using Magnetic Controller  3-18 3.7 Uploading Image Data  3-22 3.8 MiroCam® Receiver Management ² Post Procedure 3-23
  Contents   Page 8 4. 86,1*0,529,(: SOFTWARE  4-3 4.1 +:6SHFLILFDWLRQVIRU0LUR9LHZ6RIWZDUH 4-6 4.2 Overview ² 0LUR9LHZ6HUYHU 4-10 4.3 Overview - 0LUR9LHZ2SHUDWRU 4-17 4.4 Overview - 0LUR9LHZ&OLHQW 4-45 4.5 Overview - 0LUR9LHZ9LHZHU 4-89 4.6 Performing Capsule Endoscopy  4-90 4.7 How to import 0LUR9LHZSoftware 1.0 Files  4-123  5. 86,1*0,529,(:579 5-3 5.1 Overview 5-3 5.2 Using MiroView579 5-5  6. 86,1*:,5(/(660,529,(:579L 6-3 6.1 Overview 6-3 6.2 8VLQJ:LUHOHVV0LUR9LHZ579L 6-6
  1           Safety Information
  Safety Information                                            Chapter 1   1- 2
  Chapter 1                                            Safety Information  1-3 1. SAFETY INFORMATION  1.1 Warnings  MiroCam® Capsule Endoscope System has been manufactured to conform with the International Standard for Medical Electrical Equipment: IEC 60601-1, together with the Collateral Standard for Electromagnetic Compatibility Requirement and Tests IEC 60601-1-2.    MiroCam® Capsule Endoscope System has been manufactured to conform to the electric shock, fire and mechanical hazard standards as defined in CAN/CSA C22.2 NO.601.1.  Based on request of the buyer, IntroMedic will provide the labeling, such as ID labels, and the User Manual in the national language(s) of European countries. Translated documents will be evaluated by a local language expert, and will be confirmed by a native speaker of the respective national language.
  Safety Information                                            Chapter 1   1- 4 Safety Symbols: The User Manual incorporates various safety symbols to ensure safe and correct use of the product and to prevent any personal injury or property damage. These symbols are defined in the following table:     WARNING WARNING indicates a potential hazard that, if not avoided, could result in serious personal injury or damage to the product.  CAUTION CAUTION indicates a potential hazard that, if not avoided, could result in minor personal injury or damage to the product.  NOTE NOTE does not indicate potential hazards as in Caution or Warning, but contains important information regarding the installation, operation or maintenance of the product.
  Chapter 1                                            Safety Information  1-5 1.2 Symbols for Safety  This section describes a set of symbols that the IEC (International Electrotechnical Commission) has established for medical electronic equipment to classify connections and warnings of any potential hazards. EN980: Attention. See instruction manual for use.  IEC 878-02-03: Indicates that this is classified into Type BF equipment  EN 980: Denotes Date of Manufacture  EN 980: Denotes Address of Manufacture  ,(&$11(;''HQRWHV³21´VWDWXVRIPDLQSRZHUVZLWFK  ,(&$11(;''HQRWHV³2))´VWDWXVRIPDLQSRZHUVZLWFK SN  EN 980: Denotes serial number  IEC 417-5031: Denotes DC (Direct Current)  IEC 417-5032: Denotes AC (Alternating Current) A  Denotes Ampere, the unit of current V  Denotes Volt, the unit of Voltage Hz  Denote Hertz, the unit of Frequency  IEC 417-5021: Denotes potential equalization terminal  Single Use Only  Use by date  EN980: Authorized representative in the European community IPX  Protected against the effects of continuous immersion in water
  Safety Information                                            Chapter 1   1- 6 1.3 Function  Symbols  1.3.1  0LUR9LHZSoftware Function Symbols  The following table describes symbols or icons used in the MiroViewSoftware.   1.3.1.1 0LUR9LHZClient Symbol Description Symbol Description  Open the patient data screen.   Open screen to review a specific patient  Open the report screen to create a patient capsule endoscopy report.   Open screen to export (save externally) selected image data for a specific patient.  Print the selected report on the List.   Delete the selected case on the List.  Add a new review for the selected case on the List.   Delete the selected review on the List.  Open the Recycle Bin dialog.   Find the Item from the list with the keyword typed in the search box.  Show all the items on the List.   Open the Position Window  Show images in the Single View.    Show images in the Dual View.  Show images in the Quad View.   Show images in the Range View.  Show images in the Map View.   Play all images.  Play the selected images only. View images via Express Play function.   SGIB - Play the images captured via Suspected GI Bleeding function.
  Chapter 1                                            Safety Information  1-7   Symbol Description  Symbol Description  Move to the first image in the video.   Move to the previous captured image.  Play images in reverse order.   Stop playback of images.  Play  Play images in sequential order.  Move to the next captured image.   Move to the final image in the video.  Open the Comparison Window   Expand/Shrink Capture Box  Indicating the current Capture Box / Finding Summary Box / Drag Bag status is Capture Box. Click this icon to change WKH VWDWXV WR µ)LQGLQJ6XPPDU\ER[¶  Indicating the current Capture Box / Finding Summary Box / Drag Bag status is Finding Summary Box. Click this icon to change the status WRµ'UDJ%DJ¶  Indicating the current Capture Box / Finding Summary Box / Drag Bag status is Drag Bag. Click this icon to change the VWDWXVWRµ&DSWXUH%R[¶  Show Circumference Markers  Add a Circle on the image  Add a Line on the image  Delete selected Circle / Line   Color Palette  Undo the action  Redo the action
  Safety Information                                            Chapter 1   1- 8   Symbol Description Symbol Description  Check the selected captured images   Uncheck the selected captured images  Print  Preview  Save as PDF  Export Review  Open PACS function (Registered User Only)   Export Captured Images with EXMIF  Export Captured Images as Image / Video / MiroVideo   Export Segments as Image / Video / MiroVideo  Export Case as Video / MiroVideo   Select the area as a Segment  Deselect the area from Segment
  Chapter 1                                            Safety Information  1-9 1.3.1.2 0LUR9LHZOperator  Symbol Description Symbol Description  Open the patient data screen.  Connect to the MiroCam® Receiver and open MiroCam® Receiver control screen.  Print the selected report on the List.  Delete the selected case on the List.  Open the Recycle Bin dialog.  Export the case selected on the List.  Find the Item from the list with the keyword typed in the search box.  Show all the items on the List.  Upload the image data from the selected MiroCam® Receivier  Show / Edit the patient information for the selected MiroCam® Receiver  Initialize the selected MiroCam® Receiver    Unconnected  Unformatted the MiroCam® Receiver.  Empty the MiroCam® Receiver.  Prepared the MiroCam® Receiver.  Recorded the MiroCam® Receiver.  Uploaded the MiroCam® Receiver.
  Safety Information                                            Chapter 1   1- 10 1.3.2  MiroCam® Receiver Function Symbols  Symbol Description SIG  Indicates status of signal from the MiroCam® Capsule.   Green : Signal is being received from the MiroCam® Capsule. Yellow : Signal is not being received from the MiroCam® Capsule. INI  Initialization status of the MiroCam® Receiver   Green : MiroCam® Receiver is initialized Yellow : MiroCam® Receiver is not initialized BAT  Battery Status   Green : Fully charged   Yellow : Not charged Wi-Fi  Wi-Fi Switch Status (LED) On : Enabled (LED) Off : Disabled
  Chapter 1                                            Safety Information  1-11 1.4   Notes for Safe Use      Follow the safety instructions included in this User Manual and clinical precautions advised by the medical professional.     The manufacturer is not liable for harm or damage caused by improper, unauthorized, unprofessional or inexpert use of the device and/or product.   IntroMedic Co., Ltd. is NOT responsible for physical harm or HTXLSPHQW SUREOHPV FDXVHG E\ WKH XVHU¶V FDUHOHVV RSHUDWLRQ RUmismanagement of the device and/or product.   Users MUST have read and understood the User Manual. ONLY trained and qualified medical professionals or authorized representatives of IntroMedic Co., Ltd. may operate the system.   User Manual must ALWAYS be with the equipment. This is the 86(5¶65(63216,%,/,7<.  CAUTION: foreign substances including water, cleaning fluids, disinfecting cleanser and such substances may harm the equipment, and should not enter the equipment.     ONLY authorized personnel may perform repairs. Never attempt to open covers, panels or casings.   DO NOT crease, bend, fold or twist the data cable (or data belt). Take care to guard them against mechanical stress (e.g. wheels or heels)!     The sensor pads, the MiroCam® Receiver, the data cable, the data belt and the capsule must not be exposed to mechanical shock (e.g. by dropping). Any damage caused that way will void the product warranty.  CAUTION: Damage/injuries to the sensor pad, data cable or data belt may cause a safety hazard. Damaged items MUST be repaired IMMEDIATELY.   DO NOT handle fluids in the vicinity of the system.
  Safety Information                                            Chapter 1   1- 12   When using a cart purchased elsewhere, ensure the brake or latch guard is in use to prevent the wheels from rolling.   DO NOT USE in moist or damp places.     DO NOT operate the equipment with wet hands.   Avoid using the equipment in extreme temperatures or humid environments.   DO NOT keep the equipment or carry out the MiroCam® Capsule endoscopy procedure in places such as areas exposed to direct sunlight, vicinity of heaters, vicinity of chemical materials or gases, areas moist/damp or dusty, or poorly ventilated areas.   DO NOT disassemble or open the equipment without permission. This will invalidate the warranty.     DO NOT carry out the MiroCam® Capsule endoscopy procedure in areas with high vibrations or in environments where high electro-magnetic waves are generated.   DO NOT pull out the power cord by grabbing the cable. When disconnecting the power cord, grasp the plug, and pull out.  This prevents short-circuits, disconnection, or cord damage.  CAUTION: Verify that the power rating supplied from the power receptacle matches with the voltage the system requires. Check Voltage and Frequency on the AC/DC adaptor.  CAUTION: Turn off the power switch before connecting the sensor pad.   DO NOT discard sensor pads, cables and connectors with general waste. Discard separately as industrial or medical waste.     DO NOT carry out the MiroCam® Capsule endoscopy procedure simultaneously with other procedures using medical products or equipment.   DO NOT use for purposes other than medical treatment.
  Chapter 1                                            Safety Information  1-13   DO NOT connect the USB cable to the MiroCam® Receiver while the MiroCam® Receiver¶Vdata cable is still connected to sensor pads or data belt.   DO NOT charge the MiroCam® Receiver while the MiroCam® 5HFHLYHU¶V GDWD FDEOH LV VWLOO FRQQHFWHG WR VHQVRU SDGV RU GDWDEHOWwhile still attached to the human body.    Connect USB cable to the MiroCam® Receiver only after mounting it on charger.   DO NOT install any other programs onto the workstation utilized for review and diagnosis of patient image data (i.e computers with the 0LUR9LHZ Software).   The MiroCam® Capsule is disposable and should not be reused.   The sensor pads and data belt are medical waste, and should be disposed of according to local regulations or WEEE directive on waste disposal.  Only use the capsule, the MiroCam® Receiver, the data cables, the sensor pads and the data belt in the medical environment condition.   All products connected with the MiroCam® Endoscope system must be compliant with requirements of IEC60950-1 or UL certifications.   Check the outer surface of the endoscope for rough surfaces, sharp edges, or protrusions which may cause a safety hazard.  3OHDVHGRQ¶WWU\WRFXWWKHODEHO
  Safety Information                                            Chapter 1   1- 14 1.4.1 Environmental Conditions for Storage   Temperature    : -10 ୅ ~ 70 ୅ (Operating 0 ୅ ~ 40 ୅)   Relative humidity    : 10 % ~ 80 % (Operating 45 % ~ 75 %)   Atmospheric pressure  : 700hPa to 1060hPa  WARNING  DO NOT operate the equipment in the vicinity of generators, power stations, X-ray devices, and broadcasting stations where high levels of electro-magnetic waves are generated. The electro-magnetic waves can cause equipment malfunctions.    CAUTION  If the equipment has been brought in from a cold environment (stock room, airfreight) into a warm room, initial activation should take place after a few hours, to allow for temperature adjustment and balance and evaporation of condensed humidity.  WARNING  DO NOT operate the equipment in the vicinity of heat sources, strong electric or magnetic fields (close to a transformer), or near instruments generating   high-frequency signals.    WARNING  DO NOT use MiroCam® alongside or together with medical devices or procedures involving electrical currents.   DO NOT use MiroCam® with h.f. surgical equipment. It may result in burns at the site of the electrodes and possible damage to the MiroCam® Capsule and the MiroCam® Receiver. DO NOT use the unit in close radius (within 1 m) of short wave or microwave therapy equipment. It may produce instability in the captured image.  WARNING  This device is a Class B device according to EN60601-1-2 standards. This equipment can cause radio interference in residential areas. In this case, the owner (or operator) can be held responsible to take appropriate measures or take proper measures for compensation.
  Chapter 1                                            Safety Information  1-15 1.4.2 Safety Precaution  CAUTION  -  Make sure the environment is without interference from electromagnetic fields. -  Make sure the environment is without noise and vibration. -  DO NOT carry out the MiroCam® Capsule endoscopy procedure while using other equipments, devices or products. -  The instruction for use of the sensor pads MUST be observed. -  DO NOT use on patients with pacemakers or defibrillators. -  DO NOT use the MiroCam® Capsule if the package is unsealed.    DO NOT reuse a used capsule.   To prevent unexpected accidents like fire or explosion, do not use any product near or in the presence of inflammable or ignitable substances.   DO NOT disassemble the equipment case nor open the cover. In case service is required, please contact IntroMedic customer support or local point of sale immediately.   Only the accessories authorized and designed by IntroMedic Co., Ltd. should be used with this equipment. Faults resulting from the usage of unapproved or inappropriate accessories are not guaranteed against.   This equipment may affect on other products or be affected by other products.  )ROORZ \RXU GRFWRU¶V LQVWUXFWLRQV DQG DELGH E\ WKH JXLGHOLQHV LQ WKHUser Manual.   DO NOT try to upload the image data while the data cable or data belt is still connected to the MiroCam® Receiver.
  Safety Information                                            Chapter 1   1- 16   DO NOT change the rechargeable battery in the MiroCam® Receiver while the data cable or data belt is still connected to the MiroCam® Receiver.   Stay away from high frequency radiation sites (such as high voltage, radar, installation power plants, MRI, CT or electric blankets etc.) during your capsule endoscopy procedure. (It may result in serious side effects requiring an emergency operation.)   In case of any symptoms of abdominal pain, vomiting, fever, heart trouble, dizziness or seizure during or after the MiroCam® Capsule endoscopy procedure, please notify your doctor.  Always check whether the data cable or data belt is connected to the MiroCam® Receiver.   Always check that the battery in the MiroCam® Receiver is fully charged before use.   DO NOT use the MiroCam® Capsule if the package is unsealed.     After ingesting the capsule, always confirm whether the MiroCam® Capsule has been excreted. Excretion of the MiroCam® Capsule can be confirmed with an X-ray procedure.  Prior to undergoing the MiroCam® Capsule endoscopy procedure, patients with diabetes must be informed via a medical professional regarding appropriate medication & dosage.   For more accurate data and better analysis, patients can have solid food for lunch, but must have liquid food for dinner on the day prior to the MiroCam® Capsule endoscopy procedure. The patient must fast for 11 hours prior to ingesting the MiroCam® Capsule.     If recommended by the physician, the patient may take a laxative prior to the MiroCam® Capsule endoscopy procedure.     DO NOT smoke for at least 12 hours prior to the MiroCam® Capsule endoscopy procedure.
  Chapter 1                                            Safety Information  1-17   DO NOT apply body lotion before the MiroCam® Capsule endoscopy procedure.   DO NOT bite the capsule.   Avoid excessive physical activity during the MiroCam® Capsule endoscopy procedure.   4 hours after ingesting the capsule, patients may have liquid food.   It is recommended to drink mineral water (non-colored liquid) frequently, so that it may raise the possibility to take much better quality of the image.   When undergoing the MiroCam® Capsule endoscopy procedure, DO NOT make physical contact with another person undergoing the same procedure.   During the MiroCam® Capsule endoscopy procedure, DO NOT touch the MiroCam® Receiver, or get the MiroCam® Receiver wet.     Only use the provided batteries, and never remove the battery from the MiroCam® Receiver during the MiroCam® Capsule endoscopy procedure.   Avoid disconnecting the USB while uploading the image data from the MiroCam® Receiver to the workstation.   Always confirm that the USB is connected by checking the MiroCam® Receiver screen on the 0LUR9LHZ Software.     Always check the AC Power rating prior to using the workstation.   DO NOT touch AC power cable with wet hands.   DO NOT open the MiroCam® Receiver bag or touch the MiroCam® Receiver outside of the hospital.   DO NOT use on patient with electrical implantable device.   Magnetic controller must be stored in special storage box.   If electrical risk is detected from the MiroCam® Receiver, remove the battery.
  Safety Information                                            Chapter 1   1- 18   WARNING The MiroCam® Capsule takes images for 12 hours and naturally excretes in about 24 hours under normal conditions. If the MiroCam® Capsule has not been excreted from the patient within 72 hours, patient should contact the physician. After examining, the physician may need to perform a surgical operation or treatment to remove the capsule.  WARNING  Before doing capsule endoscopy with MC1000-WM, always check whether the previous capsule is retained or not in human body. If the previous capsule is detected in human body, remove it by surgical treatment. When two   MC1000-WM are existed in human body at the same time, those can stick to each other. WARNING Make sure the capsule had been excreted before going through an MRI, CT and etc. It may cause injury to the patient.
  Chapter 1                                            Safety Information  1-19 1.4.3  Cleaning and Maintenance   Capsule The capsule is disposable and should not be reused.   System and accessories -  All products should be cleanly maintained. For cleaning, rub them lightly with a soft cloth wet with warm water at least once a week. DO NOT use organic solvents such as lacquer, thinner, ethylene and oxide because they can damage the equipment. Be careful that foreign substances do not enter the main system when cleaning.  -  ALWAYS operate the equipment under sanitary environmental conditions. DO NOT use heat or gas for disinfection of the capsule.     Service Document If required, or upon request, the local IntroMedic distributor (authorized IntroMedic representative) may provide block diagrams, lists of spare parts, descriptions, adjustment instructions or other related information which may help qualified technical personnel in repairing specified parts of the equipment which have been defined repairable by IntroMedic Co., Ltd. .    Moving the Equipment -  CAUTION when moving equipment. -  WARNING: Excessive impact/shock causes internal damage. -  If wiring is connected/disconnected when moving, check the exact wiring status after moving. -  If damage to the equipment is discovered after moving, immediately contact IntroMedic or local Distributor.
  Safety Information                                            Chapter 1   1- 20
  2                 Overview & Intended Usage
  Overview & Intended Usage                                   Chapter 2   2- 2
  Chapter 2                                   Overview & Intended Usage  2-3 2. SYSTEM OVERVIEW  MiroCam® is an orally ingested capsule endoscope designed to capture and wirelessly transmit images of the lining of the small intestine. Captured images are viewed via the 0LUR9LHZ6RIWZDUH for diagnosis of diseases related to the small intestine.   The MiroCam® Capsule Endoscope System consists of the MiroCam® Capsule, the MiroCam® Receiver, the 0LUR9LHZ 6RIWZDUH, and accessories.  2.1 Intended  Purpose  MiroCam® is intended for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an electrode array, a receiving/recording unit, a data storage device, computer software to process the images (0LUR9LHZ Software), and accessories.
  Overview & Intended Usage                                   Chapter 2   2- 4 2.1.1 Contraindications    Not intended for patients with known or suspected gastrointestinal tract obstructions, perforations, strictures or fistular.   Not intended for patients who have difficulty ingesting food or pills (dysphagia).   Not intended for patients who have difficulty communicating.   Not intended for patients with indigestion, or slow gastric emptying.    Not intended for patients who may be affected by electromagnetic radiation, such as pregnant women, infants, and patients with heart disease or epilepsy.   Not intended for patients with diverticulosis in the Small Bowel.   Not intended for patients who are recommended against having the MiroCam® Capsule endoscopy procedure by a Physician.   Not intended for patients with pacemakers and/or defibrillators.   This device is intended for use in adults and children from two years of age. If a patient is younger than 18, the doctor should decide whether to perform the MiroCam® Capsule endoscopy procedure.  2.2   Observable  Diseases  2.2.1 Obscure gastrointestinal bleeding (OGIB)   Angioectasia  Meckel diverticulum  NSAIDs   2.2.2  &URKQ¶V'LVHDVH
  Chapter 2                                   Overview & Intended Usage  2-5 2.2.3 Small Bowel Tumors    Polyps  Peutz-Jeghers Syndrome  Familial adenomatous polyposis  Lymphoma  Carcinoid   Gastro-Intestinal Stromal Tumor (GIST)  Lipoma  Hemangioma  2.2.4 Inflammation   Intestinal Tuberculosis  Typhoid Fever  Angeitis  Amyloidosis  Eosinophilic enteritis  2.2.5 Celiac Disease  2.2.6  %HKFHW¶V'LVHDVH  2.2.7 Cowden Disease  2.2.8 Cronkhite-Canada Syndrom
  Overview & Intended Usage                                   Chapter 2   2- 6 2.3 Product  Warranty  Period  2.3.1  Maximum Storage Time of MiroCam® Capsule   MC1000  : 1 Year  MC1000-F   : 1 Year  MC1000-G  : 1 Year  MC1000-W  : 1 Year  MC1000-WG  : 1 Year  MC1000-WM  : 1 Year  MC1000-B  : 1 Year  MC1200  : 1 Year  MC1200-B  : 1 Year  MC1200-G  : 1 Year  MC1200-M  : 1 Year  MC1600  : 1 Year  MC1600-B  : 1 Year  MC1600-G  : 1 Year  MC1600-M  : 1 Year   2.3.2  MiroCam® Receiver, MR1100     MiroCam® Receiver:1 Year   MiroCam® Receiver Battery: 6 Months
  Chapter 2                                   Overview & Intended Usage  2-7 2.4 Product  Specification  2.4.1 MiroCam® Capsule   Component Component Model Name  Quantity Capsule   MC1000 MC1000-F MC1000-G MC1000-W MC1000-WG MC1000-WM MC1000-B MC1200 MC1200-B MC1200-G MC1200-M MC1600 MC1600-B MC1600-G MC1600-M 1 Piece (Selection) Sensor Pads  MC1000-S  10 Piece Magnetic Controller MC1000-C 1 Piece (Optional)
  Overview & Intended Usage                                   Chapter 2   2- 8  Specification Item Specification Capsule Unit (MC1000, MC1000-F, MC1000-G) Size  10.8 X 24 mm Weight  3.25 ± 0.05 g (-F) 3.5 ± 0.05 g Frame Rate  3 frame/sec Field of View  150 degrees Battery Life  12 hours Pixels 102,400 Lighting 6 LEDs Communication Method Human Body Communication Magnetic Controller (MC1000-C) Length 27.3 cm
  Chapter 2                                   Overview & Intended Usage  2-9  Item Specification Capsule Unit (MC1000-W, MC1000-WG, MC1000-WM, MC1000-B) Size  10.8 X 24.5 mm   10.8 X 25.5 mm (-WM) Weight 3.25 ± 0.05g 4.75 ± 0.05g (-WM) Frame Rate  3 frame/sec Field of View  170 degrees Battery Life  12 hours 8 hours (-WM) Pixels 102,400 Lighting 6 LEDs Communication Method Human Body Communication
  Overview & Intended Usage                                   Chapter 2   2- 10  Item Specification Capsule Unit (MC1200 MC1200-B MC1200-G, MC1200-M) Size  10.8 X 24.5 mm   Weight  3.2 ± 0.05 g 4.5 ± 0.05 g (-M) Frame Rate  3 frame/sec Field of View  170 degrees   Battery Life  12 hours 8 hours (-M) Pixels 102,400 Lighting 6 LEDs Communication Method Human Body Communication
  Chapter 2                                   Overview & Intended Usage  2-11  Item Specification Capsule Unit (MC1600 MC1600-B MC1600-G, MC1600-M) Size  10.8 X 24.5 mm   Weight  3.2 ± 0.05 g 4.5 ± 0.05 g (-M) Frame Rate  6 frame/sec Field of View  170 degrees   Battery Life  12 hours 8 hours (-M) Pixels 102,400 Lighting 6 LEDs Communication Method Human Body Communication
  Overview & Intended Usage                                   Chapter 2   2- 12 2.4.2 MiroCam® Receiver   Component Component Model Name Quantity Receiver   MR1100  1 piece Battery Pack  MR1100-B  2 piece Data Cable  MR1000-D  1 piece Data Belt  MR1000-D(M), MR1000-D(L) 1 piece Receiver Bag  MR1000-G  1 piece System Carrying Case  MR1000-A  1 piece Battery Charger  MR1000-C  1 piece Adaptor MR1000-T 1 piece Power Cable  MR1000-P  1 piece Measuring Tape  MR1000-M  1 piece USB Cable  MR1100-U  1 piece   Specification Item Specification Receiver  (MR1100) Battery Life  12 Hours Size  85 X 140 X 40 mm Weight 350g (including Battery)
  Chapter 2                                   Overview & Intended Usage  2-13  Item Specification Battery Pack (MR1100-B) Type  Lithium-Ion storage cell Capacity 10400 mA Voltage 3.6 V Recharging Time  Approx. 4 hours Size  73.8 X 84.3 X 21.4 mm Data Cable (MR1000-D) Channel 9 Size  700 x 900 x 110 mm Data Belt (MR1000-D(M), MR1000-D(L)) Channel 9 Size  1000 X 180 mm (M) 1500 X 180 mm (L) Receiver Bag (MR1000-G) Size  100 X 190 X 55 mm System Carrying Case(MR1000-A) Size  480 X 350 X 100 mm Battery Charger (MR1000-C) Power Source  Adaptor Size  135 X 75 X 105 mm Weight 285 g Adaptor (MR1000-T) Weight 231g Power Rating  AC100~240V/  50~60 Hz
  Overview & Intended Usage                                   Chapter 2   2- 14 Item Specification Power Cable (MR1000-P) Length 1.5 m Measuring Tape (MR1000-M) Length 1.5 m   2.4.3 MiroCam® Software   Component Component Model Name Version 0LUR9LHZ:RUNVWDWLRQ  0LUR9LHZ Software  2.5 0LUR9LHZ579 0LUR9LHZ579 2.0 Wireless MiroView೰ RTVi  Wireless  MiroView೰ RTVi 1.4
  3                 Performing Capsule Endoscopy
  Performing Capsule Endoscopy                                Chapter 3   3- 2
  Chapter 3                                Performing Capsule Endoscopy  3-3 3. PERFORMING CAPSULE ENDOSCOPY  3.1 General  Comments      This chapter describes how to operate and control the MiroCam® Capsule Endoscope System.  Installation and initial operation must be done by an Authorized IntroMedic Service Technician.    3.2 Safety  Warnings  WARNING Before turning on the system, check the environment condition for the MiroCam® Capsule Endoscope System.  WARNING This device is intended for use in adults and children from two years of age. If a patient is younger than 18, the doctor should decide whether to perform the MiroCam® Capsule endoscopy procedure.  CAUTION  In case the system has been brought from cold environment (stock room or airfreight) into a warm room, first switch on it after some hours for temperature balance and removing of condensation humidity (Risk of leakage current).  CAUTION  For the electrical safety, make sure that peripheral equipments (external PC, Monitor, printer etc) are connected to a wall socket of independent power supply with a perfect ground. Particularly make sure whether or not those peripheral equipments should be authorized for medical usage.  CAUTION  DO NOT touch the dome of the capsule.   CAUTION  If SIG indicator of the MiroCam® Receiver turns yellow after ingesting the capsule, immediately contact the physician.
  Performing Capsule Endoscopy                                Chapter 3   3- 4  CAUTION  If SIG indictor of the MiroCam® Receiver does not turn green when the patient holds the MiroCam® Capsule by the gold bands, immediately stop the MiroCam® Capsule endoscopy procedure.  NOTE If images of the small bowel are not captured during the MiroCam® Capsule endoscopy procedure, ensure the MiroCam® Capsule has been excreted by the patient and perform procedure again.    3.3   Examination  Preparation  The following steps (3.3.1 to 3.3.4) explain the process to initialize the MiroCam® Receiver and register patient data for the MiroCam® Capsule endoscopy procedure. For more information, see section 4.2.    3.3.1 Start System    Turn on the power of the workstation.    Click the   to start the 0LUR9LHZOperator.   Type the ID and the password in each field, and then click the .  WARNING  Before starting the system, check the power rating for the workstation.
  Chapter 3                                Performing Capsule Endoscopy  3-5 3.3.2 Initialize the MiroCam® Receiver     Connect the smaller side of the USB Cable to the MiroCam® Receiver    Connect the larger side of the USB cable to the workstation.  Turn on the MiroCam® Receiver.   Click the   in  the  0LUR9LHZ Operator to initialize the MiroCam® Receiver.  Click the   in  the  0LUR9LHZ Operator to input patient information.   After completing the required fields, click the   to save the patient information onto the MiroCam® Receiver.  WARNING If the MiroCam® Receiver is initialized, any patient data stored on the device will be deleted permanently.    NOTE During initialization with MR1100, it is not possible to cancel the process.  3.3.3 MiroCam® Receiver Preparation     Connect the battery to the MiroCam® Receiver. - Insert pins and bumps on back side of the MiroCam® Receiver in the slots on the battery.   -  Press battery to the MiroCam® Receiver until clasp locks battery in place.     Turn on the power switch of the MiroCam® Receiver.  Check the battery indicator (BAT) of upper side of the MiroCam® Receiver. If the battery is fully charged, the LED light will be green. If the LED light is not green, charge the battery.
  Performing Capsule Endoscopy                                Chapter 3   3- 6   Check the initialization indicator (INI) of upper side of the MiroCam® Receiver. Prior to beginning a procedure, this light must be green.     Connect the data cable to the MiroCam® Receiver.  Indicator Description SIG   Indicates status of signal from capsule   Green : Signal is being received from capsule Yellow : Signal is not being received from capsule INI   Initialization status of the MiroCam® Receiver   Green : The MiroCam® Receiver is initialized Yellow : The MiroCam® Receiver is not initialized BAT  Battery Status   Green : Fully charged   Yellow : Not charged Wi-Fi  Wi-Fi Switch Status (LED) On : Enabled (LED) Off : Disabled  WARNING  DO NOT do begin the MiroCam® Capsule endoscopy procedure when the MiroCam® Receiver¶VEDWWHU\LQGLFDWRUis not green. This indicates battery is not fully charged or out of order, and patient data may be lost.   CAUTION Switch the MiroCam® Receiver off prior to connect the sensor pad.
  Chapter 3                                Performing Capsule Endoscopy  3-7 3.4 Patient  Preparation  Provide the patient with the following instructions to prepare for the MiroCam® Capsule endoscopy procedure.  3.4.1 Patient Preparation    One day prior to the MiroCam® Capsule endoscopy procedure  - Lunch: Patient can have a normal mean around noon, followed by a liquid diet as instructed by the physician. -  The patient should fast for at least 12 hours prior to the MiroCam® Capsule endoscopy procedure. Patient may only drink water (no food or other beverages such as milk or coffee). -  The patient should not take any medications two hours before ingesting the MiroCam® Capsule endoscope. -  Male patients may need to shave area, as advised by physician or nurse. -  Patient should stop taking iron supplements 1 week before the MiroCam® Capsule endoscopy procedure, and should not ingest any medicine at least 2 hours prior to the MiroCam® Capsule endoscopy procedure. -  Diabetic patients need to follow any changes to insulin dosage as prescribed by the physician.   - Physician is recommended to prescribe patient a laxative such as PEG or Sodium Phosphate. Patient should ingest laxative 12 hours prior to procedure.
  Performing Capsule Endoscopy                                Chapter 3   3- 8 - Physician is recommended to prescribe patient an anti-foaming agent (such as simethicone), to reduce bubbles in the GI tract. This should be ingested after the laxative.   -  The patient should abstain from smoking 12 hours before the MiroCam® Capsule endoscopy procedure.   -  On the day of the MiroCam® Capsule endoscope test, the patient should wear comfortable and loose cloths. One-piece clothing should not be worn.   -  Prior to the MiroCam® Capsule endoscopy procedure, do not apply any lotions or perfumes.    On the day of capsule endoscopy procedure  -  After arriving in a hospital at the appointed time, the patient should submit an examination permission form and check in.   After ingesting capsule endoscope  -  The patient should refrain from ingesting any food or drink (except water) two hours after swallowing the MiroCam® Capsule endoscope. - Patient is recommended to drink water every half hour during the MiroCam® Capsule endoscopy procedure. -  The patient can start with light food after 4 hours. When the examination is complete, the patient can start having normal meals. -  In case of any abdominal pain, nausea or vomiting after ingesting the MiroCam® capsule, the patient should contact physician or nurse immediately.
  Chapter 3                                Performing Capsule Endoscopy  3-9 3.5 Sensor Placement & Capsule Ingestion  Following is the detailed procedure required to correctly administer the MiroCam® Capsule endoscopy procedure.  3.5.1 Connecting Sensor Pads  After attaching sensor pads to the data cables, attach the sensor pad correctly to each area as shown below.  Xiphoid  process Line Pelvic Line 146785R23
  Performing Capsule Endoscopy                                Chapter 3   3- 10     3.5.2                  Attach sensor pads according to the number on the data cable. - Locate the center between the umbilical and the xiphoid process. From this center point, attach the sensor to the distal SRLQWRQWKHULJKWIODQNRIWKHSDWLHQW¶VERG\ - Place the sensor pad on the xiphoid process line, directly below the right clavicle. - Place the sensor pad on the xiphoid process line, directly below the left clavicle. - Locate the center between the umbilical and the xiphoid process. From this center point, attach the sensor to the distal SRLQWRQWKHOHIWIODQNRIWKHSDWLHQW¶VERG\ - Place the sensor pad on the right pelvic line, 2 cm behind sensor #1. 5 8  Pelvic Line Xiphoid process Line 1  4
  Chapter 3                                Performing Capsule Endoscopy  3-11 - Place the sensor on the right inguinal line, 2 cm to the outside of sensor #2. - Place the sensor on the right inguinal line, 2 cm to the outside of sensor #3. -  Place the sensor pad on the right pelvic line, 2 cm behind sensor #4. -  Place the sensor directly below the right clavicle.  CAUTION  The numbering of the data cables and sensor pad must match the specified area. Please follow the numbering sequence appropriately.    CAUTION  To prevent entanglement of the data cables during procedure, organize cables and place appropriately in cable bag.  WARNING Correct image signal may not be obtained if a sensor pad is not in direct, immediate contact with the skin. Ensure the sensor pads are in direct contact with the skin. Shaving may be required for male patients.      The pictures below show the sensor pads attached in the correct locations.
 Performing Capsule Endoscopy                                Chapter 3   3- 12 3.5.3  Connecting Data Belt    Attach stitched point on the belly button.   Tighten the data belt around the waist with the velcro.   Use the data belt as shown below.      After placing sensor pads on the patient, connect the cable to the MiroCam® Receiver as detailed in the following image.   CAUTION  Do not attached the Data belt above clothing
  Chapter 3                                Performing Capsule Endoscopy  3-13 3.5.4  Removing Capsule from Package    While holding the case upside down, peel off the sterilized package cover.  3.5.5  Removing the Case    Turn the package over, the MiroCam® case will slip out of the package. While holding the case, open the lid.
  Performing Capsule Endoscopy                                Chapter 3   3- 14 3.5.6  Removing the Capsule    Grip capsule with right hand, and remove from cast. DO NOT touch the dome of the capsule.    Make sure that the MiroCam® Capsule packaging is not damaged in any way. When removed capsule from case, the MiroCam® Capsule should be flashing.  CAUTION Do not use a MiroCam® Capsule Endoscope if the packaging or the MiroCam® Capsule is damaged.   CAUTION Do not touch the dome of the MiroCam® Capsule. This may result in poor image quality.    Check if the front of the MiroCam® Capsule is flashing when it is removed from the case
  Chapter 3                                Performing Capsule Endoscopy  3-15   After washing hands, patients should grip the gold section of the MiroCam® Capsule between the thumb and the forefinger of the both hands.   CAUTION Initialize the MiroCam® Receiver prior to the MiroCam® Capsule Endoscopy Procedure. Once the MiroCam® Receiver confirms signal transmissions with a green SIG indicator, the receiver will start recording for 12 hours, even while the power switch of the MiroCam® Receiver is in the off position.  3.5.7  Checking Operation of the MiroCam® Receiver and Signal Reception  After the patient holds the MiroCam® Capsule as detailed above, the signal indicator on the MiroCam® Receiver is green and signal indication melody is generated. If signal indicator is yellow, there is no signal being received from the capsule.    Switch on the MiroCam® Receiver.   Check if the battery indicator (BAT) at the top of the MiroCam® Receiver shows that the battery is fully charged. The battery indicator on the MiroCam® Receiver should be green.   Check if the initialize indicator (INI) at the top of the MiroCam® Receiver shows that initialization is complete. The INI indicator on the MiroCam® Receiver should be green.
  Performing Capsule Endoscopy                                Chapter 3   3- 16   Connect the data cable to the MiroCam® Receiver.   After patient grasps capsule by opposing gold bands (as detailed in picture above, check the SIG indicator on the MiroCam® Receiver is green and signal indication melody is generate.  CAUTION If the battery indicator on the MiroCam® Receiver shows that the battery is not fully charged, do not start the endoscopy procedure. The battery may be depleted during the MiroCam® Capsule endoscopy procedure and images may not be saved properly.  CAUTION  Please be noticed to confirm the shape and the direction of the connector when you connecting the data cable to the MiroCam® Receiver, since it may cause cable damage problem.  3.5.8 Ingesting the Capsule  After checking if the signal indicator on the MiroCam® Receiver is green, ingest the MiroCam® Capsule with water.  CAUTION DO NOT bite the MiroCam® Capsule while ingesting.   NOTE To increase the effectiveness of the MiroCam® Capsule endoscopy procedure, please be sure to drink water on a regular basis.
  Chapter 3                                Performing Capsule Endoscopy  3-17  WARNING  The MiroCam® Capsule should enter the stomach. Make sure to always check that the MiroCam® Capsule enter the stomach by checking the images on the MiroView RTV (or the Wireless MiroView RTVi) and color of SIG indicator on the MiroCam® Receiver turns green right after ingesting the capsule. WARNING  DO NOT USE a MiroCam® Capsule that has been dropped or coughed up by a patient.  New MiroCam® Capsule must be used. Take note of the new MiroCam® Capsule number. The MiroCam® Receiver does not need to be initialized again prior to swallowing the MiroCam® Capsule.  3.5.9  Checking and Wearing the MiroCam® Receiver      After ingesting the capsule, make sure that the SIG indicator is green.   Place the MiroCam® Receiver in the bag.   Organize the contents of the bag so that the cables are not entangled.   Adjust cables, and place excess in cable bag.   Adjust the bag strap as needed.
  Performing Capsule Endoscopy                                Chapter 3   3- 18   The MiroCam® Capsule endoscopy procedure will be complete after 12 hours.    CAUTION Patient should not touch the MiroCam® Receiver during the MiroCam® Capsule endoscopy procedure. This may lead to loss of image data.  CAUTION During the MiroCam® Capsule endoscopy procedure the SIG indicator should be green.   In this case, check data cable to ensure it is securely inserted and check the sensor pads to ensure these are fixed to the body.  3.5.10  During the Procedure     After ingesting the MiroCam® Capsule Endoscope instruct the patient to: -  Immediately report any episodes of abdominal pain, nausea or vomiting, to the Nurse or Physician in charge. -  After ingesting the MiroCam® Capsule you may ingest water, and only water. At least 250ml of water should be ingested per hour.   -  Four hours after swallowing the MiroCam® Capsule, you will be permitted to resume a clear liquid diet. A clear liquid diet includes water, clear broth, Jell-O, Sprite, apple juice, and popsicles. No solid foods, milk or milk products, or dark-colored liquids are allowed. A light diet (e.g. applesauce, cereal, toast) can be resumed six hours after the MiroCam® Capsule was swallowed.   -  Do not touch or disturb the MiroCam® Receiver or get them wet for the duration of the procedure. The MiroCam® Receiver and sensors must remain on the body for a period of 12 hours after ingestion of the MiroCam® Capsule.
  Chapter 3                                Performing Capsule Endoscopy  3-19 -  If a sensor pad or cable is removed during the procedure, the patient can replace the sensor pad or cable and take a note of the time of the occurrence.   -  Patient should not exercise or lift heavy items - Daily activities, aside from strenuous activities are permitted for the patient.   -  Please refrain from bathing or showering while sensor pads and MiroCam® Receiver are attached to your body.  -  Patient must Avoid MRI rooms at the hospital and radio transmission towers.   -  Patient should not use electric blankets or electric heating pads. -  Use of computers, radios, cell phones are permitted -  Keep away from other patients who are undergoing the MiroCam® Capsule endoscopy procedure. -  Once the patient has ingested a MiroCam® Capsule, the patient may leave the clinic. WARNING Patient should not open the MiroCam® Receiver bag. The safety of ESD (electrostatic discharge) when the patient leaves the clinic is not published.  WARNING Patient should not contact any electric device to the patient abdomen directly. This may lead to loss of image data.    CAUTION  Tell the patient to notify the nurse or physician of symptoms of abdominal pain, or vomiting during or after the MiroCam® Capsule endoscopy procedure.  CAUTION  Tell the patient to avoid excessive physical activity during the MiroCam® Capsule endoscopy procedure.  CAUTION  4 hours after ingesting the MiroCam® Capsule patients may resume a liquid diet.
  Performing Capsule Endoscopy                                Chapter 3   3- 20 CAUTION  When undergoing the MiroCam® Capsule endoscopy procedure, DO NOT make physical contact with another person undergoing the same procedure.  3.5.11  Completion of the Procedure    The MiroCam® Receiver set, sensor pad and cables can be removed by the patient 12 hours after ingestion of the MiroCam® Capsule.   Hospital or clinic is to arrange the time for the patient to return the MiroCam® Receiver set.    3.5.12 After the Procedure  - Inform the patient that:     The MiroCam® Capsule will be naturally excreted.    Instruct patient to contact nurse or physician if the patient feels acute or continuing abdominal pain.   Meals: eat as usual.     The day following the MiroCam® Capsule endoscopy procedure the patient may resume all usual activities.    MRI examinations: patient should avoid MRI examinations prior to excretion of the MiroCam® Capsule.    WARNING After ingesting the MiroCam® Capsule, always check whether the MiroCam® Capsule has been excreted. The MiroCam® Capsule captures images for up to 12 hours and gets naturally excreted in about 24 hours under normal conditions. If the MiroCam® Capsule has not been excreted from the patient within 72 hours, patient should contact the physician. If examination confirms retention of the MiroCam® Capsule, surgical or medical treatment to remove the MiroCam® Capsule may be necessary. NOTE  The MiroCam® Capsule and sensor pads are medical waste, and should be disposed of according to local regulations or Waste Electrical and Electronic Equipment (WEEE) directive on waste disposal.
 Chapter 3                                Performing Capsule Endoscopy  3-21  3.6 Using  Magnetic  Controller    After ingesting the capsule, make sure that the SIG indicator is green.  Connect to the 0LUR9LHZ5TV or Wireless 0LUR9LHZRTVi and then confirm current location of capsule.   Control the direction of the MiroCam® Capsule in human body with the magnetic controller like following pictures.       Lead the MiroCam® Capsule into the duodenum with magnetic controller by FKHFNLQJ WKH LPDJHV RQ WKH 0LUR9LHZ 579 or the Wireless MiroView RTVi.  CAUTION If the patient feels pain during doctor controls a magnetic capsule with magnetic controller, carry out conventional endoscopy. CAUTION DO NOT put any steel or magnetic materials within 1m radius during the MiroCam® Capsule endoscopy procedure.
  Performing Capsule Endoscopy                                Chapter 3   3- 22 3.7 Uploading  Image  Data  Once the MiroCam® Capsule endoscopy procedure is complete, upload the image data from the MiroCam® Receiver to the workstation and start diagnosis of the data using 0LUR9LHZSoftware.  3.7.1  Transferring and Checking image data   Switch on the workstation.   Connect the smaller end of the USB cable to the MiroCam® Receiver  Click the    on the desktop to launch 0LUR9LHZOperator.   When the login window appears, enter the user ID and password, and then click the  .   Connect the smaller end of the USB cable to the MiroCam® Receiver.   Connect the other end of the USB cable to the workstation.  Click the   in 0LUR9LHZOperator.   After the image data in the MiroCam® Receiver has been successfully uploaded, execute 0LUR9LHZ Client to review and report on the patient file.  WARNING )RUPRUHLQIRUPDWLRQRQXVLQJ0LUR9LHZ Software please refer to Chapter 4. 8VLQJ0LUR9LHZ Software.
  Chapter 3                                Performing Capsule Endoscopy  3-23 3.8   MiroCam®  Receiver  Management  ± Post Procedure  3.8.1 Cleaning    Remove data cable (or data belt) from the MiroCam® Receiver .   Capsule, data belt and sensor pads should not be reused.   To clean the MiroCam® Receiver and the workstation, dampen a soft cloth with warm water or with a commercial, nonabrasive cleaner (or a mild soap or detergent solution) and wipe the exterior surface lightly. DO NOT allow any liquids to come in contact with the power connector, cable connector or switches. DO NOT allow any liquids to penetrate connectors or openings in the system cover. Cleaning of the system and accessories should be conducted at least once a week.  CAUTION  DO NOT immerse the MiroCam® Capsule Endoscope System or its accessories in liquid or clean with caustic or abrasive cleaners. DO NOT spray or pour any liquid on the system or its accessories. DO NOT use the organic solvents such as a lacquer, thinner, ethylene and oxide because they can damage the equipment. When cleaning the data cable, do not use the cleaning tools such as brush and sandpaper.
  Performing Capsule Endoscopy                                Chapter 3   3- 24 3.8.2 Charging Battery    Battery must be recharged after every procedure.   Battery can be individually recharged using the battery charger or while attached to the MiroCam® Receiver.   Battery must be recharged in the battery charger.   If two batteries are inserted in charger, the batteries will automatically consecutively charge (not simultaneous).      Connect adapter DC cable to battery charger.
  Chapter 3                                Performing Capsule Endoscopy  3-25   Connect AC power cable to adaptor.   Connect AC power cable to AC outlet or multi-tap.   Battery charger indicator lights: -  Battery is charging when green light is blinking -  Battery is fully charged when light is solid blue -  Battery is not-charging if light is yellow    NOTE  Second Battery in charger will automatically begin charging once first battery is fully charged.    Waiting Charging Charging Charged
  Performing Capsule Endoscopy                                Chapter 3   3- 26  WARNING  If the indicator on the battery charger is blinking yellow there is a bad connection. Please disconnect the battery from the battery charger and try again.   WARNING  Only use batteries provided by IntroMedic. Usage of inappropriate batteries can cause serious damage to the MiroCam® Receiver.     WARNING  Recharge the batteries with the adaptor and the battery charger only from IntroMedic. Improper adaptors and battery chargers cause serious damage.     WARNING  Try not to contact the metallic terminal board of batteries with conductive objects or human body. It can cause an electric shock or damage and/or breakage of the batteries.   WARNING  DO NOT use the batteries with visible surface damage on. Usage of damaged batteries can result in leakage of contents of the batteries.       WARNING  To avoid the risk of explosion, keep batteries away from fire.   WARNING  Treat the batteries carefully from serious impact or shock. This can cause serious damage to the batteries.     WARNING  DO NOT expose the batteries to liquids. This can cause serious damage.
  4           8VLQJ0LUR9LHZSoftware
  Using MiroView Software                                   Chapter 4   4- 2
  Chapter 4                                    Using 0LUR9LHZ Software  4-3 4. USING 0,529,(: SOFTWARE    The  0LUR9LHZ Software is a computer software program that aids in diagnosis of disease of the small bowel via displaying images obtained from the MiroCam® Capsule.    Following is a list of new & updated features: - User Interface  The  0LUR9LHZ Software enables adjustable resolution, to appropriately fit to the screen size.    The 0LUR9LHZ Software has been designed with a dark background to reduce eye strain.  The 0LUR9LHZSoftware provides intuitive and unified icons so that the user can adapt quickly.  - Reporting Tool - Express Play - SGIB Play - Range View - Map View - Network Support - Account Management - Multi-Review / Report Generation   - Multi-Receiver Management   - Multi-Data Path - Customizing Print Dialog - Data Multi-Language - ODF Report - Drag Bag - Finding Setting (Custom Finding)
  Using MiroView Software                                   Chapter 4   4- 4 The workstation where the 0LUR9LHZ Software is installed properly has icons for four applications on the Desktop folder: 0LUR9LHZ Client, 0LUR9LHZOperator, 0LUR9LHZServer, and 0LUR9LHZViewer. However this manual will not explain details on how to use the 0LUR9LHZViewer. For additional information for 0LUR9LHZViewer, please refer to the 0LUR9LHZ9LHZHU,QVWUXFWLRQ 0DQXDO 7RJHW WKH LQVWDOODWLRQ SURFHGXUH IRU0LUR9LHZ Software, refer to the MiroCam® Service Manual.
  Chapter 4                                    Using 0LUR9LHZ Software  4-5   0LUR9LHZClient  0LUR9LHZOperator  0LUR9LHZServer  0LUR9LHZViewer This program is used for reviewing the images from the MiroCam® Capsule and generating a patient report. The 0LUR9LHZServer should be started prior to starting the 0LUR9LHZClient. This program is used for managing the MiroCam® Receivers (initialize, upload etc.) and managing user accounts. The 0LUR9LHZServer should be started prior WRVWDUWLQJ0LUR9LHZOperator. The server stores the patient data files. Start the 0LUR9LHZServer prior to starting 0LUR9LHZSoftware or 0LUR9LHZOperator. The  0LUR9LHZViewer is used for reviewing, editing and creating the images exported IURPWKH0LUR9LHZSoftware; EXMIF or JPEG images. Functions: - List Mode - Review Mode - Report Mode - Export Mode - Image Atlas - SGIB     (Suspected     GI Bleeding) - Express Play - Position Window - Network connectivity - Multi-Review - Range View - Map View - GI Dictionary   - Finding / Diagnosis    Information - Drag Bag Functions: - Input Patient    Information - Receiver    Initialization - Upload Image Data from Receiver - Data Backup - Account    Management - Patient Manager -Network connectivity Functions: - DB, Mass storage -Network connectivity - Multiple Data Path Functions: - Study Information - Image Information - Show Findings /    Diagnosis
  Using MiroView Software                                   Chapter 4   4- 6 4.1   H/W  Specifications  for  0LUR9LHZ Software  4.1.1  0LUR9LHZServer   Minimum Hardware Requirements -  OS: Windows XP (32 bit, 64 bit), Windows Vista Business (32 bit, 64 bit), Windows 7 (32 bit, 64 bit) -  CPU: > Core 2 Duo E6300 -  Memory: > DDR II 4GByte -  Video Card: Windows Compatible -  HDD: SATA 500GB - ODD: DVD+-RW - Monitor Resolution: 1024x768   Recommended Hardware Requirements  -  OS: Windows XP (32 bit, 64 bit), Windows Vista Business (32 bit, 64 bit),   Windows 7 (32 bit, 64 bit) -  CPU: > Core 2 Duo E6300 -  Memory: > DDR II 4GByte -  Video Card: Windows Compatible -  HDD: SATAII RAID 1TB - ODD: DVD+-RW -  Monitor Resolution: 1680x1050, 1280 x 1024
  Chapter 4                                    Using 0LUR9LHZ Software  4-7 4.1.2  0LUR9LHZClient   Minimum Hardware Requirements  -  OS: Windows XP (32 bit, 64 bit),   Vista or Windows7 (32 bit, 64 bit) - CPU: Pentium4 2GHz - Memory: 1GB -  Video Card: Geforce 7600GT 128MB -  HDD: SATA 160GB - ODD: DVD+-RW - Monitor Resolution: 1024x768 -  Sound Card / Mike / Speaker   Recommended Hardware Requirements  -  OS: Windows XP (32 bit, 64 bit),   Vista or Windows7 (32 bit, 64 bit) -  CPU: > Core 2 Duo E6300 -  Memory: > DDR II 2GByte -  Video Card: > Geforce 7600GT 128MB -  HDD: SATA II 500 GB - ODD: DVD+-RW - Monitor Resolution: 1680x1050 -  Sound Card / Mike / Speaker
  Using MiroView Software                                   Chapter 4   4- 8 4.1.3  0LUR9LHZOperator   Minimum Hardware Requirements  -  OS: Windows XP (32 bit, 64 bit),   Vista or Windows7 (32 bit, 64 bit) - CPU: Pentium4 2GHz - Memory: 1GB -  Video Card: Windows Compatible -  HDD: SATA 160GB - ODD: DVD+-RW - Monitor Resolution: 1024x768   Recommended Hardware Requirements  -  OS: Windows XP (32 bit, 64 bit),   Vista or Windows7 (32 bit, 64 bit) -  CPU: > Core 2 Duo E6300 - Memory: 1GB -  Video Card: Windows Compatible -  HDD: SATA 160GB - ODD: DVD+-RW - Monitor Resolution: 1680x1050
  Chapter 4                                    Using 0LUR9LHZ Software  4-9 4.1.4  0LUR9LHZViewer   Hardware Requirements  -  OS: Windows XP (32 bit, 64 bit),   Vista or Windows7 (32 bit, 64 bit) - CPU: Pentium4 2GHz - Memory: 1GB -  Video Card: Windows Compatible -  HDD: SATA 160GB - ODD: DVD+-RW - Monitor Resolution: 1024x768  CAUTION  The  0LUR9LHZ Software cannot be viewed on a monitor with a resolution under 1024x768.  NOTE  7KH*8,LVEHVWYLHZHGRQDZLGHPRQLWRURI´RUODUJHU
  Using MiroView Software                                   Chapter 4   4- 10 4.2   Overview  ± 0LUR9LHZServer  The  0LUR9LHZ Server software is intended to be used as a system administrative managing tool. The 0LUR9LHZ Server manages file systems and database systems. When switching the workstation on, the 0LUR9LHZ Server will start automatically. When the following window appears, set up the configuration and press the   to continue.  ཛ  Set the port number for the Network. If unknown, please leave the ILHOG DV µ¶ 7KLV ILHOG VKRXOG EH PDQDJHG E\ D 1HWZRUNAdministrator at the hospital. c d e f
  Chapter 4                                    Using 0LUR9LHZ Software  4-11 ཛྷ 6HW WKH SDVVZRUG IRU µ$GPLQLVWUDWRU¶ DFFRXQW $GGLWLRQDO DFFRXQWVcan be made via Account Manager function after logging in the 0LUR9LHZ2SHUDWRUDVµ$GPLQLVWUDWRU¶ ཝ  Confirm the password. ཞ  Set the Active Path; Video Path and Data Path as the desirable path. Select one of multiple paths from the paths list for the active path. The paths list can be modified via Server Menu Æ Multi Data Path on the 0LUR9LHZServer running.  § ,QSXWWKH+RVSLWDO¶V1DPHLQWKLVILHOG འ ,QSXWWKH+RVSLWDO¶VAddress in this field. ཡ ,QSXWWKH+RVSLWDO¶V7HOHSKRQHQXPEHULQWKLVILHOG g h i
  Using MiroView Software                                   Chapter 4   4- 12   After setting up the Configuration Windows, the 0LUR9LHZServer window above will appear. Click the   to hide the window.
  Chapter 4                                    Using 0LUR9LHZ Software  4-13   The Video Path list and the Data Path list for the Active Paths can be modified via Server Menu Æ Multi Data Path on the 0LUR9LHZServer running.  -  Login window will appear. Input the password for the Administrator account.
  Using MiroView Software                                   Chapter 4   4- 14 -  Case Path Setting will appear. Click the   to show the Active Path Setting.
  Chapter 4                                    Using 0LUR9LHZ Software  4-15 - Click the  to set the path for the Video Path and the Data Path respectively.
  Using MiroView Software                                   Chapter 4   4- 16  -  The user can modify the Video Path and Menu Path using the  (add path) and  (remove path) as well. -  ,I WKH $FWLYH 3DWK¶ PHPRU\ EHFRPHV LQVXIILFLHQW WR XVH(below 10GBytes), set the path as new one larger than 30 GB.   NOTE The  0LUR9LHZ Server provides the task manager to show the activating status on the server. The 0LUR9LHZ Server offers the case list and the analysis function as well.
  Chapter 4                                    Using 0LUR9LHZ Software  4-17 4.3 Overview  -  0LUR9LHZOperator  The  0LUR9LHZ Operator software is intended to be used for administrative purposes, including the Account Manager, the Patient Manager and the MiroCam® Receiver management functionality.  4.3.1 Start 0LUR9LHZOperator  Double-click the   to run the 0LUR9LHZ Operator.   When executed for the first time, the following window will appear.   ,QSXW³´IRUWKH,3$GGUHVVDQG³´IRUWKH3RUW ଖ If the 0LUR9LHZ Server is installed on the network, input the workstation¶V,3RQZKLFKWKH0LUR9LHZServer is installed for this field, and the port which the 0LUR9LHZ Server is using, referring to the IT engineer in the hospital to point the operator to the 0LUR9LHZServer.
  Using MiroView Software                                   Chapter 4   4- 18   Set the Active Path; Video Path and Data Path as the desirable path. Select one of multiple paths from the paths list for the active path.      Direct the Default Export Data Folder.   Choose the preferred Language settings and press   to access the 0LUR9LHZOperator.  After the 0LUR9LHZOperator has launched, the Log-in window will appear.    ,QSXWµ$GPLQLVWUDWRU¶IRU,'DQGGHVLJQDWHGSDVVZRUG7KHQSUHVVthe  .  Press the    to change the IP address the 0LUR9LHZServer is using.
  Chapter 4                                    Using 0LUR9LHZ Software  4-19   After clicking the   the window will appear like the above. Type the 0LUR9LHZServer IP address and port. Then press the   after inputting ID and password.  CAUTION  DO NOT forget the ID and password after they are encrypted.  CAUTION  Before the 0LUR9LHZ Operator is launched, a password for the administrator account should be created. When running the 0LUR9LHZ Server for the very first time, a password for the administrator account should be created.  NOTE  Create new and edit existing accounts by using the µAccount Manager¶ menu in 0LUR9LHZ Operator.    The following subsections describe the main interfaces and functions of the 0LUR9LHZOperator.
  Using MiroView Software                                   Chapter 4   4- 20 4.3.2 List Mode                            £ List Mode  Opens List Mode ¤ Receiver Mode  Opens Receiver Mode ¥ Print Report  Print out the selected Report ¦ Delete Case  Delete the selected Cases § Recycle Bin  Open Recycle Bin Window ¨ Export  Export the selected Report © Search  Search Cases and/or Reviews by keyword ª Show all  Refresh and Show all the Cases and Review on the List  NOTE  Right click the mouse at the list item to use the context menu functionality, for example the Backup function.  e f g ij  c d h
  Chapter 4                                    Using 0LUR9LHZ Software  4-21 4.3.3 Receiver Mode  Screen enables administration and status display for up to four MiroCam® Receivers.     £ Upload  Upload image data in the MiroCam® Receiver to the server ¤ ,QSXW 3DWLHQW¶VInformation ,QSXW 3DWLHQW¶V ,QIRUPDWLRQ WR WKH VHOHFWHGReceiver ¥ Initialization  Delete all data from the MiroCam® Receiver and/or prepare the MiroCam® Receiver for the next patient ¦ Receiver Port  Shows the status of the connected the MiroCam® Receiver for each of the four ports § Detail Information  6KRZV WKH SDWLHQW¶V LQIRUPDWLRQ IRU WKHselected  Receiver  ¥ ¦ c d e
  Using MiroView Software                                   Chapter 4   4- 22  CAUTION  When running MiroView Operator on a 64bits OS, Do not turn off or disconnect the MiroCam® Receiver while uploading a patient file.  4.3.4 Receiver Initialization  Click the    to initialize the selected Receiver.
  Chapter 4                                    Using 0LUR9LHZ Software  4-23  4.3.5 Inputting Patient Information  Click the    to enter the patient information. After completing the required fields, click the   to save the patient information onto the MiroCam® Receiver. The fields in the patient information are editable and additional fields can be added.  This can be accessed via Tools Æ Preferences Æ Receiver Æ Set Æ Add / 0RGLI\ 7KH ILHOG ³5HDVRQ IRU UHIHUUDO´ LV QHZO\ DGGHG RQ WKH0LUR9LHZ Software version.)          * indicates required fields.
  Using MiroView Software                                   Chapter 4   4- 24  4.3.6 Image Data Upload  After the MiroCam® Capsule endoscopy procedure, upload the image data from the MiroCam® Receiver to the MiroCam® system via the 0LUR9LHZ Operator. Click the   to upload the image data from the MiroCam® Receiver.  After the image data in the MiroCam® Receiver has been successfully uploaded, enter the List Mode and check if the new Case has been created. - Double-click the   to run the 0LUR9LHZClient. - Select the newly uploaded image data and enter the Review Mode. - Play the case to check the uploaded image data.  CAUTION  If uploading of the image data failed, disconnect the USB cable from the workstation, turn off the MiroCam® Receiver and back on.    Reconnect the USB cable to the workstation and try uploading once more. If the problem continues, please contact the local IntroMedic distributor.  NOTE  0DNH VXUH WKH SDWLHQW¶V QDPH KDV EHHQ LQSXWWHGproperly prior to uploading the image data. If the image data is uploaded with the wrong patient name, it could result in misdiagnosis or delay in diagnosis.   NOTE  If the user uploads an image data which has been XSORDGHGEHIRUHDQHUURUPHVVDJHµ'DWDKDVEHHQpreviously uploaded. Would you like to upload agDLQ"¶ZLOODSSHDU  NOTE  The user can stop the uploading of the image data while uploading process file by clicking the  .
  Chapter 4                                    Using 0LUR9LHZ Software  4-25 4.3.7 User Account Management  The 0LUR9LHZOperator enables management of multiple users & Groups.  Users can be organized into Groups to readily control access.    The following screen can be accessed by Tools Æ Account Manager.
  Using MiroView Software                                   Chapter 4   4- 26  Click the    to show the window as displayed below to create a new user ID.  ID: User ID Password: Password Name: User Name Type: User type. - User: can access own reviews and all the reviews in the *URXSRIWKH8VHU¶VRQO\ - Super User: can access all the reviews in the System - Operator: can log in via the 0LUR9LHZ Operator S/W only. ୔ Administrator account can log in via both the 0LUR9LHZ Client and the 0LUR9LHZOperator. Group: Set of users in the same Clinic or Laboratory.  Click the    to delete the selected account
  Chapter 4                                    Using 0LUR9LHZ Software  4-27 4.3.8 Patient Manager  The  0LUR9LHZ Operator enables management of Patient. The following screen can be accessed by Tools Æ Patient Manager   Click the    to show the window as displayed below WRFUHDWHDSDWLHQW¶V,' ID: ID for the new patient. Name3DWLHQW¶VQDPH Date of birth: The date of birth for the patient which will appear on the report Gender7KHSDWLHQW¶VJHQGHU
  Using MiroView Software                                   Chapter 4   4- 28 4.3.9 Preferences  To set and/or change the settings for the 0LUR9LHZ Operator, access Preferences via Tools Æ Preferences. Preferences for the 0LUR9LHZ Operator falls into two categories: Global and Local. Global preference is saved on the 0LUR9LHZ Server and has in effect for all the users using the 0LUR9LHZServer. Local preference is saved on the workstation on which the 0LUR9LHZOperator is installed and has in effect for the workstation only.   ic d e f ghj
  Chapter 4                                    Using 0LUR9LHZ Software  4-29  Global  £ Settings  - Hospital Information - Measurement - Agent & Distributor Information ¤ PACS  - PACS Registration - Technical Settings Local  ¥ Settings  - Language - GI Tract Color Reference - Free Hard Disk - Display - Default Export Data Folder - Finding Setting ¦ User  - 8VHU¶V,QIRUPDWLRQ - Change Password § Network  - 0LUR9LHZServer - Server Manager ¨ List  - Display ± Show/Hide Columns © Receiver  - Patient Information Configuration - Multiple Upload ª Print  - Page Setup
  Using MiroView Software                                   Chapter 4   4- 30 1) Global Settings  i. Hospital Information ཛ Country ± Choose the Country where the hospital is located. ཛྷ Hospital Name ± ,QSXWWKH+RVSLWDO¶V1DPHLQWKLVILHOG ཝ Address ± ,QSXWWKH+RVSLWDO¶V$GGUHVVLQWKLVILHOG ཞ Telephone ± ,QSXWWKH+RVSLWDO¶V7HOHSKRQHQXPEHULQWKLVfield. ཟ Logo Image ± Add the Logo Image ୔ These information will be applied and shown on the final      report. § ¨ © ¢ £ ¤ ¥ ¦
  Chapter 4                                    Using 0LUR9LHZ Software  4-31 ii. Measurement འ Height ± &KRRVHWKHPHDVXUHPHQWXQLWIRUSDWLHQW¶VKHLJKW ཡ Weight - &KRRVHWKHPHDVXUHPHQWXQLWIRUSDWLHQW¶VZHLJKW ར Waist - &KRRVHWKHPHDVXUHPHQWXQLWIRUSDWLHQW¶VZDLVW ୔ These information will be applied and shown on the case   information and patient information   iii.    Agent & Distributor Information ལ $JHQF\,QSXWWKH$JHQF\¶VLQIRUPDWLRQLQWKLVILHOG ཤ  Contact: Input the contact information for the agency.  ª «
  Using MiroView Software                                   Chapter 4   4- 32 2) Global PACS   i.  Registration: Process to enable the PACS related functionalities. - Click the  . PACS registration window will appear like the below. -  Carry out in alphabetical order like following picture.  NOTE  Generation key is a key that unlocks the PACS function on the MiroCam® workstation.  ¢ £ ¤ § ¦ ¥ ÚÛÜ
  Chapter 4                                    Using 0LUR9LHZ Software  4-33 -  Generation key will be generated automatically. Click the    to send the Generation key to IntroMedic. Mail Box window will appear like below. -  Input the e-mail address of the hospital administrator in WKH µ)URP¶ ILHOG ,QSXW DOO WKH UHTXLUHG LQIRUPDWLRQ WRsend to IntroMedic: Hospital name, Hospital address, Main contact, Telephone, E-mail address and Generation key. - Click the   to send the information to IntroMedic. IntroMedic will reply with the Registration Key to the
  Using MiroView Software                                   Chapter 4   4- 34 e-PDLO DGGUHVV W\SHG LQ µ)URP¶ ILHOG ,W ZLOO WDNH WLPH WRreceive the reply. - Click the    to close the Mail Box window. -  After receiving the Registration Key, input the Key in the Registration Key field on PACS registration window.   Click the    to finalize the registration process.  ii.  Caller: Technical properties for Caller. ཛ  Title: Input the Title for Caller. ཛྷ  Modality: Input the Modality for Caller. ཝ  Port: Input the port number to use for Caller.  iii.  Image Storage Server: Technical properties for Image Storage. ཞ  Title: Input the title for Image Storage Server. ཟ  IP Address: Input the IP Address for Image Storage Server. འ  Port: Input the port number to connect to Image Storage Server.
  Chapter 4                                    Using 0LUR9LHZ Software  4-35  iv.  Work list Server: Technical property for Work list Server. ཡ  Title: Input the Title for Work list Server. ར  IP Address: Input the IP Address for Work list Server. ལ  Port: Input the port number to connect to Work list Server.  v.  Other Settings: Technical properties for PACS setting. ཤ  PACS Module: Choose the appropriate PACS Module, DreamWell or Vieworks. ཥ  Log Level: Choose the appropriate Log Level for the PACS functionality: Level 0, Level 1 and Level 2. ས Time Delay Choose the Time delay from 0 to 9 seconds. ཧ Station Name ཨ Wild-card character ¯ ®  ¬ ª © ¨ «
  Using MiroView Software                                   Chapter 4   4- 36  3) Local Settings  i. Language ཛ  Display: Choose the display language. ཛྷ  CEST: Choose the CEST language. ii.  GI Tract Color Reference: Show the color references for the GI Tract. Free Hard Disk: Decide the number of cases on the local drive.   NOTE  If the number of cases saved from the Server is more than the number of Keeping Case, it will automatically delete the oldest cases from the local drive.  ¢ £
  Chapter 4                                    Using 0LUR9LHZ Software  4-37  iv. Display ཝ  Display: Font setting to display. v.  Default Export Data Folder: Direct the folder for exporting data. Click the    to open the browser. vi.  Finding Setting: Configure the finding setting. Click the   to open the configuration dialogue. ¤
  Using MiroView Software                                   Chapter 4   4- 38 4) Local User  i.  8VHU¶V,QIRUPDWLRQ6HWPRGLI\WKHLQIRUPDWLRQIRUWKHXVHU ཛ  User Name: Input User Name which will be used for log-in ID. ཛྷ  Email Address: Type in the e-mail address here. ཝ  Telephone: Type in the telephone number here. ii. Change Password ཞ  Current: Input the current password ཟ  New: Input the desired password to use. འ  Confirm: Input the desired password again.  ¢ £ ¤ ¥ ¦ §
  Chapter 4                                    Using 0LUR9LHZ Software  4-39 5) Local Network  i.  0LUR9LHZServer ཛ  IP Address: Input the IP Address where the 0LUR9LHZServer is installed. ཛྷ Port: Input the port number to connect with 0LUR9LHZServer. ii. Server Manager ཝ  Port: Input the port number for connection between the Server Manager and the 0LUR9LHZServer. ¢ £ ¤
  Using MiroView Software                                   Chapter 4   4- 40  6) Local List  i. Display Columns: Configure the columns which will appear on List.
  Chapter 4                                    Using 0LUR9LHZ Software  4-41  7) Local Receiver  i. Patient Information Configuration: Configure the items shown on Patient Information after initializing the MiroCam® Receiver.
  Using MiroView Software                                   Chapter 4   4- 42 - Click the   to  show  the  Patient  Information Configuration window.  -  Add/Delete/Modify the items. After completing the tasks, click the   to close the window.  ii. Multiple Upload Count: Set the number of MiroCam® Receivers to upload to the 0LUR9LHZOperator at once. Choose a number, 1 to 4.
  Chapter 4                                    Using 0LUR9LHZ Software  4-43  8) Local Print  i. Page Setup ཛ  Footer: Type in the string for footer which will appear on the final report. ¤  Page Setup: Edit the layout of the report. ¢ £
  Using MiroView Software                                   Chapter 4   4- 44 - Click the    to show the Page Setup window.  - Choose the paper type; A4 or Letter and adjust the margins for Header, Footer, Right and Left.
  Chapter 4                                    Using 0LUR9LHZ Software  4-45 4.4 Overview  -  0LUR9LHZClient  4.4.1 Start 0LUR9LHZClient   The 0LUR9LHZ Client is used to review images from the MiroCam® Capsule procedure and generating a patient report. 2WKHU IXQFWLRQV LQFOXGHG ,Q 0LUR9LHZ &OLHQW DUH Network Connectivity, List Mode, Review Mode, Report Mode, Export Mode, Image Atlas, Multi-Review, SGIB(Suspected GI Bleeding)Play, Express View, Range View, Map View, Positioning Window, GI Dictionary, and Finding / Diagnosis Information.  Double-click the   to  run  the  0LUR9LHZ Client. When executed for the first time, the following window will appear.
  Using MiroView Software                                   Chapter 4   4- 46  ,QSXW³´IRUWKH,3$GGUHVVDQG³´IRUWKH3RUW If the 0LUR9LHZ Server is installed on the network, input the workstation¶V ,3 RQ ZKLFK WKH 0LUR9LHZ Server is installed for this field, and the port which the 0LUR9LHZ Server is using, referring to the IT engineer in the hospital.  'LUHFWWKH³Active Path´WRWKHFDFKLQJIROGer.    Direct the Default Export Data Folder.   Choose the preferred Language settings and press   to access the 0LUR9LHZClient.  After the 0LUR9LHZClient has launched, the Log-in window will appear.    ,QSXWµ$GPLQLVWUDWRU¶IRU,'DQGSDVVZRUGDOUHDG\W\SHGLQ7KHQpress the  .  Press the    to change the IP address the 0LUR9LHZServer is using.
  Chapter 4                                    Using 0LUR9LHZ Software  4-47   After clicking the    the window will appear as the above. Type the  0LUR9LHZ Server IP address and port. Then press the   after inputting ID and password. (Use the   to change the port number for the Server Manager, which is used to manage the server availability)  CAUTION  ID and password are encrypted. DO NOT forget the ID and password.  CAUTION  Before the 0LUR9LHZClient has launched, a password for the administrator account should be created. When running the 0LUR9LHZServer for the first time, a password for the administrator account should be created.  NOTE  Create new and edit new or existing accounts by using the Account Management tool on the 0LUR9LHZOperator.  NOTE  Access to the 0LUR9LHZ Client at Local Mode by clicking the   to use the 0LUR9LHZClient without connecting to the 0LUR9LHZServer.
  Using MiroView Software                                   Chapter 4   4- 48 4.4.2 List Mode   List  £ List Mode  Open List Mode ¤ Review Mode  Open Review Mode ¥ Report Mode  Open Report Mode ¦ Export Mode  Open Export Mode § Print Report  Print selected Report ¨ Delete Case  Delete Selected Case © Add a Review  Add a Review for the Selected Case ª Delete Review  Delete selected Review « Recycle Bin  Open Recycle Bin ¬ Search  Search Cases and/or Reviews by key word  Show all  Refresh and Show all the Cases/Reviews on the List ® Selection Bar  A rectangular area to show selected item ¯ Detail View  Show Detail Information for selected Case or Review c d e f g h i j k l ̺11  ̺12  ̺13
  Chapter 4                                    Using 0LUR9LHZ Software  4-49 4.4.3 Review Mode    Select a case to review on the List Mode and click the   to enter the Review Mode.    Capture Box Mode   £ List Mode  Open List Mode ¤ Review Mode  Open Review Mode ¥ Report Mode  Open Report Mode ¦ Export Mode  Open Export Mode § Position Window  Current position of the   MiroCam® Capsule in 2D ¨ Single View  Change to Single View © Dual View  Change to Dual View ª Quad View  Change to Quad View c d e f ̺5  ̺6  ̺7  ̺8  ̺9  ̺10   ̺11   ̺12   ̺13   ̺14   ̺15   ̺16   ̺17   ̺18   ̺19   ̺20   ̺21   ̺22   ̺23   ̺24   ̺26   ̺27   ̺25   ̺28   ̺29
  Using MiroView Software                                   Chapter 4   4- 50 « Range View  Change to Range View ¬ Map View  Change to Map View  Normal Play  Show all images.     ® Express Play  This mode actively removes images based upon similarity; effectively reducing the number of images for review. This mode is recommended to review the patient file but not to diagnose the patient symptoms. ¯ SGIB Play  Shows the images that have a high possibility of bleeding and / or blood. This mode may assist in finding areas of suspected bleeding; however, it is NOT 100% reliable, and should only be used as an adjunctive tool to assist in reviewing the images.   ° First Frame  Go to the First Image Frame ± Previous Captured Image Go to Previous Captured Image Frame ԧ Play in reverse  Play in reverse Ԩ Play / Pause  Play / Pause ԩ Play  Play Forward Ԫ Next Captured Image  Go to the Next Captured Image ԫ Last Frame  Go to the final Image Frame  Capture Image  Image capture button.  Set landmark  Landmark button. Use this to set landmarks for transition points in the GI tract.   Image Enhancement  Show the Image Enhancement window  Full Screen Mode  Fit image to full screen  Time Display  Display the time duration of the MiroCam® Capsule from procedure start.  Annotation Box Handle Expand / Shrink the control panel (Show / Hide the Capture Box) The video image area may widen by Annotation Box Handle.
  Chapter 4                                    Using 0LUR9LHZ Software  4-51  Comparison Window  Display the image selected from the Image Atlas    Thumbnail Extension  Extend the Capture Box to display all captured images.   (Hide the Video area)  Capture Box /   Finding Summary Box / Drag Bag Toggle Icon  Replace the Capture Box with Finding Summary Box or Drag Bag.    Finding Summary Box     Finding Summary Box  Displays Capture Box, Thumbnail and Finding Summary Box.   This is changed by clicking the  . When new findings are added in Report mode, new information for captured image will be shown in Finding Summary Box.   
  Using MiroView Software                                   Chapter 4   4- 52 -  Choose a section of Summary Box to show the thumbnails. When a section of Summary Box is selected, the thumbnails will be shown in the screen. Click the   to  close  the thumbnail mode.
  Chapter 4                                    Using 0LUR9LHZ Software  4-53  Drag Bag -  Choose a section of Drag Bag to show the captured images.  -  There are two ways to set the captured image, the drag bag: set the main image to a drag directly and capture the main image first and set the image to a drag bag via the pop-up menu.
  Using MiroView Software                                   Chapter 4   4- 54 i.  Left click the mouse and drag the mouse downwards. Fan shaped Drag Bags will appear like below. Release the mouse in one of them.  ii. Left click the mouse and drag to upper direction. Fan shapes for Landmarks will appear like below. Release the mouse on one of them.
  Chapter 4                                    Using 0LUR9LHZ Software  4-55 iii.  Right click the mouse on a captured image and then set the image to a Drag Bag.
  Using MiroView Software                                   Chapter 4   4- 56  Mode Control - Full Screen Mode Click the    to change to Zoom View mode   NOTE  Play Slide Control is available on the Full Screen Mode to set playing direction, playing speed and playing status: Play and Pause. NOTE  Move the cursor away from the bottom of the screen and the Play Control will appear.
  Chapter 4                                    Using 0LUR9LHZ Software  4-57  - Viewing Modes   Select one of the   for preferred Viewing     Mode.  Single View                   Dual View Quad View                   Range View  NOTE  Range View has been designed to assist in readily finding transition points in the GI tract. The display will show 9 images before and 9 images after the current position. The 9 images are defaulted to 10 minutes apart, resulting in an overview of 90 minutes prior to, and 90 minutes after the current position. The number of images displayed is set at 9, and is not configurable. The 90 minute duration is configurable and can be adjusted by dragging the side line in blue on the Time Bar and dragging the mouse icon on the time boundary.    It can also be adjusted by going to Tools -> Preference -> Review.
  Using MiroView Software                                   Chapter 4   4- 58 Map View  -  Use the wheel of the mouse to zoom in/out the map. -  Double-click the desired image to capture. Right click the mouse for the desirable image to capture or set landmark.   - Click the desirable image and press <Space> to play the adjacent image frames.  NOTE  Map view has been designed to assist in reviewing the patient file by showing all the images on the same page to assist in readily identifying landmakers and potential bleeding areas.   Move the cursor on the designated image to see the original image in detail.
  Chapter 4                                    Using 0LUR9LHZ Software  4-59 - The    will change to playing Mode, as below.              Normal Play                  Express Play SGIB Play  NOTE  The Express Play shows the meaningful images only so that the user can review the whole video for the patient quickly. When entering the Express Play, the 0LUR9LHZ Software will show the message, The Express Play is not intended to detect abnormalities and should not be used as a substitute for reviewing the entire video.  NOTE  The SGIB Play shows the images in red or the obscure bleeding images. IntroMedic does not guarantee the performance and/or accuracy. Thus, use this function for reference only.
  Using MiroView Software                                   Chapter 4   4- 60   Position Window              NOTE  User is able to choose the option of selecting the type of Position Window between line and region. NOTE  User is able to change the Position Window Setting via Tools -> Preference    Divides the body into 8 areas and displays the location of where the MiroCam® Capsule took the current image.  Organ Bar
  Chapter 4                                    Using 0LUR9LHZ Software  4-61  NOTE  User can now choose the Organ bar option. This organ bar also displays the entire trajectory of the MiroCam® Capsule in the GI track.
  Using MiroView Software                                   Chapter 4   4- 62   Comparison Window   Displays the image shown from the Image Atlas lastly.   Image Enhancement Window  Alter images characteristics (ie. color, saturation etc.) to assist in identifying abnormalities: - Color Weakness : Image adjustment for the reviewer who has color blindness. - ALICE : Image adjustment as NBI(Narrow Band Image)
  Chapter 4                                    Using 0LUR9LHZ Software  4-63 4.4.4 Report Mode  Select the desired review for the Case from the List Mode and click the   to move to the Report Mode.  Report Mode ̺1 ̺2 ̺3 ̺4 ̺5 ¨ ̺7 ྣ ̺12  ̺13  ̺14  ̺15  ̺8 « ¬
  Using MiroView Software                                   Chapter 4   4- 64  £ List Mode  Open List Mode ¤ Review Mode  Open Review Mode ¥ Report Mode  Open Report Mode ¦ Export Mode  Open Export Mode § Circumference markers Opens graduated scale for display on the image ¨ Drawing Tools  Enables users to include annotations for the image. © Image Enhancement  Alter images characteristics (ie. color, saturation etc.) to assist in identifying abnormalities. ª Thumbnail Area  Show the adjacent captured images so that the user can select/deselect the captured images to export. « Image Comments  Write comments for the Capture on the main display and the report. (Image Comments, Reason for Referral, Procedure info & findings, Summary & Recommendation) ¬ Diagnosis  Provides the Level of certainty of diagnosis.  Audio Comment  Record voice comment (The user can record the voice, from 3 seconds to 180 seconds for each captured image) ® GI Dictionary  Provides the frequently-used-words to search. ¯ Input Selected Word  Input the selected word into the text editor. ° Add Word to Dictionary Add the typed word into the Dictionary. ± Remove word from Dictionary Delete selected word from the Dictionary.
  Chapter 4                                    Using 0LUR9LHZ Software  4-65 NOTE  User is able to change the Panel Layout like following pictures via Tools -> Preference -> Local -> Report NOTE Comments can be completed automatically by pressing <Ctrl+ Space Bar> or <Ctrl + Enter> on the keyboard. Refer to 4.6.4  5HYLHZ 0LUR9LHZ &OLHQW for more information. NOTE  Additional tabs in the report can be created or modified via Tools    -> Preferences -> Report NOTE  Right click the mouse on Text Comments area to use Annotations. Refer to 4.6.4  5HYLHZ 0LUR9LHZ &OLHQW for more information.  NOTE  Patient information, including the reason for referral, is registered in the MiroCam® Receiver via 0LUR9LHZ2SHUDWRU
  Using MiroView Software                                   Chapter 4   4- 66 4.4.5 Export Mode    ̺1 ̺2 ̺3 ̺4 g h ̺7 ̺8 ̺9 ̺10  ̺11  ̺12  ̺13  ̺14  ̺15  ̺16  ̺17
  Chapter 4                                    Using 0LUR9LHZ Software  4-67  £ List Mode  Open List Mode ¤ Review Mode  Open Review Mode ¥ Report Mode  Open Report Mode ¦ Export Mode  Open Export Mode § Check Captured Image Select Captured image and add to the list to print/export. ¨ Uncheck Captured Image Deselect Captured image and remove from the candidate list to print/export. © Print  Print out the report for the Review. ª Preview  Preview the report « Save as PDF  Save the report as a PDF file. ¬ Export Review  Export Review data only without the Video. Review data is the own format of the   0LUR9LHZ Software, which contains Captured Images and Comments by the user.  Send to PACS  Send selected captured images to PACS system via DICOM Sender ® Export EXMIF  Export Captured Images as EXMIF   EXchangeable Medical Information Format) ¯ Export Images  Export Captured Images as Image, Video, or MiroVideo. MiroVideo is the own video format for the 0LUR9LHZ Software. °  Export Segment Export segment as Video, MiroVideo ± Export Entire    Export entire file as Video, MiroVideo ԧ Add segments  Add the selected range as segments Ԩ  Remove segments Remove the selected range from segments.
  Using MiroView Software                                   Chapter 4   4- 68 4.4.6 Preferences  Preferences for the 0LUR9LHZ Client falls into two categories: Global and Local. Global preference is saved on the 0LUR9LHZ Server and has in effect for all the users using the 0LUR9LHZServer. Local preference is saved on the workstation on which the 0LUR9LHZ Client is installed and has in effect for the workstation only.     ¢ £ ¤ ¥ ¦ § ¨ © ª « ¬
  Chapter 4                                    Using 0LUR9LHZ Software  4-69  Global  £ Settings  - Hospital Information - Measurement - Agent & Distributor Information ¤ PACS  - PACS Registration - Technical Settings ¥ Report  - User added comment tab Local  ¦ Settings  - Language - GI Tract Color Reference - Captured Image Box - Free Hard Disk - Display - Default Export Data Folder - Auto Recover Information - Finding Setting § User  - 8VHU¶V,QIRUPDWLRQ - Change Password ¨ Network  - 0LUR9LHZServer - Server Manager © List  - Display ± Show/Hide Columns ª Review  - Range View - Summary Control - Capture - Captured Image Box - Drag Bag - SGIB Play - Express Play - Position Window « Report  - Panel Layout ¬ Export  - Export AVI (captured images)  Print  - Page Setup
  Using MiroView Software                                   Chapter 4   4- 70  1) Global Settings  i. Hospital Information ཛ Country ± Choose the Country where the hospital is located. ཛྷ Hospital Name ± ,QSXWWKH+RVSLWDO¶V1DPHLQ this field. ཝ Address ± ,QSXWWKH+RVSLWDO¶V$GGUHVVLQWKLVILHOG ཞ Telephone ± ,QSXWWKH+RVSLWDO¶V7HOHSKRQHQXPEHULQWKLVfield. ཟ Logo Image ± Add the Logo Image ୔ These information will be applied and shown on the final    report. § ¨ © ¢ £ ¤ ¥ ¦
  Chapter 4                                    Using 0LUR9LHZ Software  4-71 ii.  Measurement འ Height ± &KRRVHWKHPHDVXUHPHQWXQLWIRUSDWLHQW¶VKHLJKW ཡ Weight - &KRRVHWKHPHDVXUHPHQWXQLWIRUSDWLHQW¶VZHLJKW ར Waist - &KRRVHWKHPHDVXUHPHQWXQLWIRUSDWLHQW¶VZDLVW ୔ These information will be applied and shown on the case   information and patient information   iii.  Agent & Distributor Information ལ $JHQF\,QSXWWKH$JHQF\¶VLQIRUPDWLRQLQWKLVILHOG ཤ  Contact: Input the contact information for the agency.  ª «
  Using MiroView Software                                   Chapter 4   4- 72 2) Global PACS  i.  Registration: Process to enable the PACS related functionalities. - Click the  . PACS registration window will appear like below.  ¢ ¤ £ ¥ ¦ § § ¦ ¥ ¢ £ ¤ ÚÛÜ
  Chapter 4                                    Using 0LUR9LHZ Software  4-73 -  Generation key will be generated automatically.   Click the  to send the generation key to IntroMedic. Mail Box window will appear like below. -  Input the e-PDLODGGUHVVLQWKHILHOGµ)URP¶for receiving the Registration Key. Fill in the information for the items in the text box: Hospital name, Hospital address, Main contact, Telephone, E-mail address and Generation key. - Click the   to send the information to IntroMedic. IntroMedic will reply with the Registration Key to the e-PDLO DGGUHVV W\SHG LQ µ)URP¶ ILHOG ,W ZLOOtake time to receive the reply.
  Using MiroView Software                                   Chapter 4   4- 74 - Click the    to close the Mail Box window. -  After receiving the Registration Key, input the Key in the Registration Key field on PACS registration window.   Click the    to finalize the registration process.  ii.  Caller: Technical properties for Caller. ཛ  Title: Input the Title for Caller. ཛྷ  Modality: Input the Modality for Caller. ཝ  Port: Input the port number to use for Caller.  iii. Image Storage Server: Technical properties for Image Storage. ཞ  Title: Input the title for Image Storage Server. ཟ  IP Address: Input the IP Address for Image Storage Server. འ  Port: Input the port number to connect to Image Storage Server.
  Chapter 4                                    Using 0LUR9LHZ Software  4-75   iv.  Worklist Server: Technical property for Worklist Server. ཡ  Title: Input the Title for Worklist Server. ར  IP Address: Input the IP Address for Worklist Server. ལ  Port: Input the port number to connect to Worklist Server.  v.  Other Settings: Technical properties for PACS setting. ཤ  PACS Module: Choose the appropriate PACS Module of DreamWell or Vieworks. ཥ  Log Level: Choose the appropriate Log Level for the PACS functionality: Level 0, Level 1 and Level 2. ས  Time Delay: Choose the Time delay from 0 to 9 seconds. ཧ Station Name ཨ Wild-card character ¯ ®  ¬ ª © ¨ «
  Using MiroView Software                                   Chapter 4   4- 76  3) Global Report  i.  User added comment tab: If checking on the checkbox and typing the desired string into the textbox, a comment tab will be added on the Report Mode. To apply the change, click  .
  Chapter 4                                    Using 0LUR9LHZ Software  4-77 4) Local Settings  i. Language ཛ  Display: Choose the display language. ཛྷ  CEST: Choose the CEST language. ཝ  GI Tract Color Reference: Show the color references for the GI Tract. ii.  Captured Image Box: Check the check box to show GI tract reference colors in the background of Captured Image Box.  ¢ £
  Using MiroView Software                                   Chapter 4   4- 78  iii.  Free Hard Disk Keep the number of cases on the local drive. If the cases are cached from the Server more than the configured number, the oldest case on the local drive will be freed up. iv. Display Display: Font setting to display. v.  Default Export Data Folder: Direct the folder for exporting data. Click the    to open the browser. vi. AutoRecover Information Save Frequency: The 0LUR9LHZ Client saves the auto recovery information while the user is reviewing the patient file, periodically. Choose the period to save the auto recovery information from 0 to 60 minutes. ¤
  Chapter 4                                    Using 0LUR9LHZ Software  4-79 vii. Finding Setting: Configure the finding setting. Click the   to open the configuration dialogue.  5)  Local User  i.  8VHU¶V,QIRUPDWLRQ6HWPRGLI\WKHLQIRUPDWLRQIRUWKHXVHU ཛ  User Name: Input User Name which will be used for log-in ID. ཛྷ  Email Address: Type in the e-mail address here. ཝ  Telephone: Type in the telephone number here. ii. Change Password ཞ  Current: Input the current password ཟ  New: Input the desired password to use. འ  Confirm: Input the desired password again.  ¢ £ ¤ ¥ ¦ §
  Using MiroView Software                                   Chapter 4   4- 80 6) Local Network  i.  0LUR9LHZServer ཛ  IP Address: Input the IP Address where the 0LUR9LHZServer is installed. ཛྷ  Port: Input the port number to connect with the 0LUR9LHZServer. ii. Server Manager: Utility to manage the 0LUR9LHZ Server remotely. ཝ Port: Input the port number for connection between Server Manager and the 0LUR9LHZServer. ¤ ¢ £ ¢ £ ¤
  Chapter 4                                    Using 0LUR9LHZ Software  4-81  7) Local List  i. Display Columns: Configure the columns which will appear on List.
  Using MiroView Software                                   Chapter 4   4- 82  8) Review  i. Range View ཛ  Range View Previous Offset: Set the range for range view. ཛྷ  Range View Next Offset: Set the range for range view. ii. Summary Control ཝ  Summary Item Range: Set the period of one item on the Finding Summary Box. iii. Capture Add to report: If the checkbox is checked, the captured images will be shown on Report Mode and added to the final report. The added captured image can be removed from the Report Mode. ¢ £ ¤
  Chapter 4                                    Using 0LUR9LHZ Software  4-83 iv.  Captured Image Box Captured Image Option: If the checkbox is checked, the video will stop when capturing an image on Review Mode.  v.  Drag Bag: Drag Bag related setting. Add/Remove the Drag Bags and set a Bag to Default Bag by choosing one from the combo list. The Bags will be applied for the reviews added newly. If the user makes a new review for a case or open a case not reviewed and choose finding box for the captured image box, the drag bags which are modified on this preference will be shown. ¥ ¦ § ¨ ©
  Using MiroView Software                                   Chapter 4   4- 84 ཞ  Default Bag: Set a Bag to Default Bag by choosing one from the combo list on Drag Bag mode.  vi.  SGIB Play ཟ  Bleeding probability: Choose the Bleeding probability to play on SGIB Play. Set this value as Low to show many images on SGIB Play or set high to show little images.   vii. Express Play འ  Quality: Choose the Quality to play on Express Play. Set this value as Low to skip more images on Express Play than High. viii.  Position Window ཡ  Type: Choose the type for the Position Window. If Line is chosen, the trace of the MiroCam® Capsule will be displayed in line. If Region is chosen, the position of the MiroCam® Capsule will be displayed on region segmented. ར  Organ Bar: Choose the Organ Bar status. If Show is chosen, the Organ Bar will appear on the Position Window.
  Chapter 4                                    Using 0LUR9LHZ Software  4-85 9) Local Report                                                                  i.  Panel Layout: Choose the Panel Layout. The Panel Layout will be shown as per the choice.
  Using MiroView Software                                   Chapter 4   4- 86 10) Export  i. Export AVI ཛ  Save as AVI: Set the length of the AVI to export. ཛྷ  FPS (AVI): Set the frame rate of the AVI to export.  ¢ £
  Chapter 4                                    Using 0LUR9LHZ Software  4-87 11)   Local Print  ii. Page Setup ཛ  Footer: Type in the string for footer which will appear on the final report. ¢ £
  Using MiroView Software                                   Chapter 4   4- 88 ¤  Page Setup: Edit the layout of the report. - Click the    to show the Page Setup window.  - Choose the paper type; A4 or Letter and adjust the margins for Header, Footer, Right and Left.
  Chapter 4                                    Using 0LUR9LHZ Software  4-89 4.5 Overview  -  0LUR9LHZViewer  4.5.1 Start 0LUR9LHZViewer   Double-click the   to run the 0LUR9LHZ Viewer. Open an Image File (EXMIF-Exchangeable Medical Image Format data or -3(*YLDµ)LOH-!2SHQ¶IXQFWLRQ  £ Image Display  Image Display Area (EXMIF or JPEG) ¤ Title  Title for the image in the Image Display ¥ Description  Description for the image in the Image Display ¦  Study Information General Information (Software / Manufacturer / Hospital Information), Patient Information (Doctor, Patient Body Shape) and Study Information (Doctor, Receiver, Capsule and Procedure Information) for the image in the Image Display   § Image Information Creator Information and/or Finding Comments for the image in the Image Display ¨ Findings  Findings information the creator captured and commented before for the image in the Image Display. © Diagnosis  Diagnosis information the creator captured and commented on for the image in the Image Display. d e f g h c i
  Using MiroView Software                                   Chapter 4   4- 90   4.6 Performing  Capsule  Endoscopy  4.6.1 Initializing Receiver    Connect the MiroCam® Receiver to the workstation via USB port.   Run the 0LUR9LHZOperator by double-clicking the   on the desktop or from the Windows Start-up menu. Log in.   Carry out in alphabetical order like following picture.  Ü Ú Û
  Chapter 4                                    Using 0LUR9LHZ Software  4-91   Enter the MiroCam® Receiver mode and select the MiroCam® Receiver port to initialize and click the  .   When following message will appear, click the   to start the initializing process.  Ú Û
  Using MiroView Software                                   Chapter 4   4- 92   After the initializing process is complete, the following message will appear.   Click the   and the Patient Information window will appear. Input the patient information for each field.   Click the    to input the patient name and then click the   to input the doctor name. 7\SHLQWKHSDWLHQW¶VQDPHDQGWKH,'LQWRWKHDSSURSULDWHILHOGVdirectly.)   Carry out in alphabetical order like following picture.  (The fields marked with * is required.) ß à á Ú Û ÜÝ Þ â
  Chapter 4                                    Using 0LUR9LHZ Software  4-93   Click the    to save the Case Information to the MiroCam® Receiver. The MiroCam® Receiver is now ready for the MiroCam® Capsule endoscopy procedure.  NOTE  The fields in the Case Information are editable, and additional fields can be added. This can be accessed via Tools  Æ Preferences Æ Receiver Æ Set Æ Add / Modify.  4.6.2 Upload Image Data (0LUR9LHZOperator)    Connect the MiroCam® Receiver to the workstation via USB port.   Run the 0LUR9LHZOperator by double-clicking the   on the desktop or from the Windows Start-up menu. Log in.   Carry out in alphabetical order like following picture.  Ü Ú Û
  Using MiroView Software                                   Chapter 4   4- 94   After switching the MiroCam® Receiver on, enter the MiroCam® Receiver mode and select the proper the MiroCam® Receiver port to upload the image data.   Carry out in alphabetical order like following picture.   Click the    to start the Upload.  Ú Û
  Chapter 4                                    Using 0LUR9LHZ Software  4-95     After completing the Upload, the following message will appear.  CAUTION  When the MiroView˞ Operator is running on a 64bits OS, Do not turn off or disconnect any MiroCam® Receivers connected to the workstation while uploading.  4.6.3  Change Patient Information After uploading the Image Data (Optional)   From the 0LUR9LHZ 2SHUDWRU XQGHU µ7RROV¶ -! 6HOHFW µ&DVH,QIRUPDWLRQ¶
  Using MiroView Software                                   Chapter 4   4- 96  7KHQ µ&DVH ,QIRUPDWLRQ¶ GLDORJ ER[ ZLOO SRSV XS &KDQJH WKHpatient information and click the  .
  Chapter 4                                    Using 0LUR9LHZ Software  4-97 4.6.4 Review (0LUR9LHZClient)   Run the 0LUR9LHZ Client by double-clicking the   on the desktop or from the Windows Start-up menu. Log in.   Carry out in alphabetical order like following picture.   Click the   on the Log-in screen to run the 0LUR9LHZclient as Local Mode. (See  8VLQJ 0LUR9LHZ &OLHQW DWLocal PC without FRQQHFWLQJ WR 0LUR9LHZ 6HUYHU for details.) -  $W WKH ILUVW UXQQLQJ RI µ/RFDO 0RGH¶ WKH ZLQGRZ ZLOO DSSHDUlike below. Input the appropriate values in to the empty fields and click the   to proceed. -  Carry out in alphabetical order like following picture. Ü Ú Û ÚÛÜÝ
  Using MiroView Software                                   Chapter 4   4- 98 -  Open the Case/Review via Menu Æ Home Æ Open &DVH5HYLHZILOH«.  -  Choose a file to open and click the  .  -  After completely reviewing the file, save the review via MenuÆ Home Æ Save Active Review so that the user can import the edited review file on the MLUR9LHZ6HUYHULQWKHfuture.
  Chapter 4                                    Using 0LUR9LHZ Software  4-99   Select the Patient Video Data on the List Mode - Select a patient video data to review from the List Mode and click the    to enter the Review Mode. - Check to see if another user is reviewing the patient file on the Current User column in the List Mode.  - If the patient file is being reviewed by another user, a message will appear.
  Using MiroView Software                                   Chapter 4   4- 100   Review the Patient Video on the Review Mode  NOTE  When using network connection between the 0LUR9LHZ Server and the 0LUR9LHZ Client, for efficient video playback the video is cached locally in the server. During this buffering process, the video may be reviewed.  ± Click the   to readily set landmarks
  Chapter 4                                    Using 0LUR9LHZ Software  4-101 - Use the Range View and Map View to find all the landmarks.  NOTE  Click the    and select landmark type: Esophagus, Gastric, Duodenum, Ileo-cecal, Cecal and User landmark. - After marking the necessary landmarks, play the video via the Express Play and capture abnormalities and / or suspected disease pathologies in the small bowel.
  Using MiroView Software                                   Chapter 4   4- 102  NOTE  7KH XVHU FDQ SUHVV µ(QWHU¶ NH\ WR FDSWXUH WKHimage. -  To identify abnormalities and/or disease pathology more easily, use the Image Enhancement function to adjust preferred image settings. Click the    to open the Image Enhancement window.
  Chapter 4                                    Using 0LUR9LHZ Software  4-103  Converting AVI File -  Click two images (the first image and the last image) while holding the < Ctrl > key among the to convert the segment to an AVI file.  -  Right mouse click and select µSave¶ from dialog box.
  Using MiroView Software                                   Chapter 4   4- 104 -  2QO\FKHFNµ6DYHVHOHFWHGUDQJHDV9LGHR -  Carry out in alphabetical order like following picture.   Choose the location you want to save this file and click the .  Click the    to save selected range as Video. Û Ú
  Chapter 4                                    Using 0LUR9LHZ Software  4-105   Composing the Report on the Report Mode - After capturing images of suspected disease in the Review Mode pathology, click the    to enter the Report Mode.  -  Use the auto complete dictionary slide and choose available disease pathology from the list. -  7KH XVHU FDQXVH KRW NH\ µ&WUO  6SDFH¶ RU µ&WUO  (QWHU¶ WR XVHauto complete function while composing comments. Select a word from the list after pressing the hot key.
  Using MiroView Software                                   Chapter 4   4- 106 -  The user can add the word shown on the GI Dictionary into the text comment by double-clicking the appropriate word or clicking the  . -  The user can prepare Annotations for use by right clicking the mouse in the comments area. x Click µEdit Annotation¶. x Click the  . x  Input annotation and then click the  .
  Chapter 4                                    Using 0LUR9LHZ Software  4-107 -  Completing the report: x  Complete image comments for captured images (as required) x  Complete the following tabs: Reason for referral; Procedure Info & Findings; and Summary & Recommendations x  After completing the report, enter the Export Mode to print out the complete report and/or export the 0LUR9LHZSoftware data.  NOTE  The GI Dictionary tab and findings tab allow the user to adjust the findings and the level of severity. The user can use the GUI easily.  CAUTION  DO NOT detach and/or attach the Microphone from/to the workstation while recording the audio comment.  CAUTION  Sound Card Driver must be installed prior to using the Recording function.  CAUTION  Prior to recording the comments, close all the applications using the Sound Card such as music/video player.
  Using MiroView Software                                   Chapter 4   4- 108  Print report, create file and export the Review on the Export Mode. -  Select the images to select and click the   and add the selected captured images to the candidate list to export. (Remove the selected images from the candidate list using the  ) - The µ6DYHDV3')¶IXQFWLRQLVXSGDWHG   NOTE  User is able to change default settings on whether or not to add the images on the report via Tools -> Preference -> Review.
  Chapter 4                                    Using 0LUR9LHZ Software  4-109 -  Check to include on the report when capturing the image. -  Uncheck to not include the images on the report.   - Click the    to use export function. A window will appear like below.
  Using MiroView Software                                   Chapter 4   4- 110  -  Select the File Type to export and change the desired file location on the Selected Folder and click the  . A window will appear with a progress bar.
  Chapter 4                                    Using 0LUR9LHZ Software  4-111 -  The user can use the   for EXMIF. When clicking the  , the window will appear like below. The user can configure the EXMIF properties: title and description. Findings will be saved as the JPEG file format(.jpg) and diagnosis images will be saved as the multiple image file format(.tiff) -  The user can use the   for Segment Export (Video / Miro Video) and the    for Export Case function.  NOTE  Select the segments on the Timebar to export by selecting two captured images to set as the start point and the end point respectively, or by clicking the landmark flag. If no segment is selected, the   to export segments would be disabled.  NOTE  MiroVideo includes selected Segments so that the XVHUFDQFRQILUPWKHFRQWHQWVDQGHGLWIURPµ2SHQ&DVH5HYLHZILOH«¶µ,PSRUW&DVH5HYLHZILOH«¶main menu.  CAUTION  It is highly recommended to constantly back up patient data through export function in order to prevent the data loss.
  Using MiroView Software                                   Chapter 4   4- 112   Print Report and save the report as PDF on the Report Preview - Click the   to print reports and/or click the   to save the report as PDF. - Show report preview on the screen. The user can include or exclude findings/diagnosis data on the final report via Menu Æ Preview Æ Including findings & diagnosis data. - The user can print out the report and save the report to PDF file in this screen. - IIWKHXVHUVHOHFWVWKHLPDJHVSHUSDJHPRGHWKHPHVVDJHµ,Iyou proceed with the current Layout mode, image comments may be omitted and not printed. Please select 3 images per SDJH PRGH WR LQFOXGH DOO FRPPHQWV LQ WKH UHSRUW¶ ZLOO DSSHDUThis will be defaulted during the next use in that the workstation for all users. - If the user selects the Draft report, no captured images will be included in the final report.
  Chapter 4                                    Using 0LUR9LHZ Software  4-113 -   Click the    to print the report, the Print window will appear. -  Choose the desirable Printer, Print Ranage, Copies/format, Print Option and Layout. -  Carry out in alphabetical order like following picture. - The 0LUR9LHZSoftware privides the ODF for the soft copy of the report so that the user can open the ODF report in other word processors; MS Word and Haansoft; ODF is available for 3 images layout only.  NOTE  The previous software version (0LUR9LHZSoftware 1.0) did not support the draft report style. Ü ß Þ Ý Û Ú
  Using MiroView Software                                   Chapter 4   4- 114   Back up the segment -  If a hospital wants only back up small bowel section, they can XVHWKLVIHDWXUH )URPµ([SRUW¶PRGH VHOHFWWKHUDQJH DQG FOLFNthe    to add segment first.  - Click the  ., Export Segments as Video/MiroVideo.
  Chapter 4                                    Using 0LUR9LHZ Software  4-115 -  6HOHFWµ5HYLHZ'DWD0LUR9LGHRVPHFDQGFOLFNthe  . -  Carry out in alphabetical order like following picture.  4.6.5 Image Atlas   Atlas The following picture show via Help Æ Atlas c d e h i j k f g ̺13  l ̺11  ̺12  ̺14  ̺15  ̺16  Û Ú
  Using MiroView Software                                   Chapter 4   4- 116  £ Atlas  Default Atlas Images ¤ My Atlas  Atlas Images configured by the user ¥ Atlas + My Atlas  Show Atlas Images and My Atlas Images at once. ¦ Import Atlas Images from file Import Atlas Images from file; BMP, JPEG, GIF, EMF, WMF, TIFF and PNG file format. § Export Atlas Images to file(s) Export Atlas Images to file; BMP, JPEG, GIF, EMF, WMF, TIFF and PNG file format. ¨ Edit Information  Edit the Information of the selected Atlas Image. © Export Atlas Image(s) to E-mail Export selected Atlas Image(s) and send it / them to another doctor via E-mail. ª Delete Atlas   Image(s) Delete the selected Atlas Image(s) « Print Atlas   Image(s) Print the selected Atlas image(s) ¬ Save as PDF  Save the selected Atlas Image(s) as PDF  Show/Hide Marks Show / Hide Marks on the Atlas Image(s) ® Search  Search the Atlas Image ¯ Comparison Window Show the Comparison Window ° Browser Tab(Findings /   Diagnosis) Area to select the Category of the Atlas Image and to see the selected image. ± Recent Images  Images the user saw on the Comparison Window ԧ Atlas Thumbnail Area Area to select the Atlas Image. Double-click the Atlas Image to see the Image Information.
  Chapter 4                                    Using 0LUR9LHZ Software  4-117  Ԩ  Study Information  /  Image  Information Show the Information of the Atlas Image ԩ Findings /   Diagnosis Show the Findings / Diagnosis Information Ԫ Atlas Thumbnail Area(2) Search the Atlas Image  ̺18  ̺19  ̺17
  Using MiroView Software                                   Chapter 4   4- 118   ԫ Comparison Window  Show the Current Image and Comparison Image at once.  ̺20
  Chapter 4                                    Using 0LUR9LHZ Software  4-119   To compare a captured image with ATLAS image, click the   (Optional).   6HOHFW DQ LPDJH IURP 0LUR9LHZ Atlas to compare the atlas image with the captured image for a side by side comparison. (Option)
  Using MiroView Software                                   Chapter 4   4- 120  Double-click an ATLAS image for selection.    To show the ATLAS image on the comparison window, click the  on the ATLAS mode.
  Chapter 4                                    Using 0LUR9LHZ Software  4-121   The comparison windows will appear. Figure out the current image and comparison image.    Import Atlas Images from file(s) The 0LUR9LHZ Software allows the user to import image files for personal purpose. - Click the   WRRSHQWKHµ2SHQ¶%URZVHU WR ORRNIRUWKHILOHVWRimport.
  Using MiroView Software                                   Chapter 4   4- 122   Import Atlas Images from file(s) The 0LUR9LHZSoftware allows the user to import image files for personal purpose. - Click the   WRRSHQWKHµ2SHQ¶%URZVHU WR ORRNIRUWKHILOHVWRimport.  -  Select the file to import and click the . -  &RPSOHWH WKH µ7LWOH¶ µ'HVFULSWLRQ¶ DQG µ,PDJH ,QIRUPDWLRQ¶ ILHOGVand click the  .
  Chapter 4                                    Using 0LUR9LHZ Software  4-123 4.7 How  to  import  0LUR9LHZSoftware 1.0 Files  4.7.1  Preparing files for import   5XQWKH0LUR9LHZSoftware 1.0  ([SRUW ILOHV E\ FOLFNLQJ µ([SRUW IRU the  0LUR9LHZ ([SUHVV¶function or back up files by using Backup Mode on the 0LUR9LHZ Software 1.0. Before using the Export to the 0LUR9LHZ Express function, insert DVD media to the DVD writer or connect the external hard disk or USB flash memory to the workstation via USB port.  NOTE  The  0LUR9LHZ Software supports  the 0LUR9LHZ Software Case / Review files, the 0LUR9LHZ  H[SUHVV ILOHV (the MiroCam® Capsule Endoscope System companying with .MMV) and the 0LUR9LHZ Software 1.0 backup files(.ifo companying with .ict and .frg files) to import.    Select a patient file from the list to export to the 0LUR9LHZExpress.  ([HFXWHµ([SRUWWRthe 0LUR9LHZ([SUHVV«¶RQWKHPDLQPHQX
  Using MiroView Software                                   Chapter 4   4- 124  ,I WKH IRUP µ6DYH $V¶ DSSHDUV OLNH EHORZ VHOHFW WKH ORFDWLRQ WRH[SRUWDQGILOOLQWKHµ)LOHQDPH¶ILHOGDQGFOLFNµ6DYH¶   Carry out in alphabetical order like following picture.    A window will appear like below and if the function ends VXFFHVVIXOO\ $ PHVVDJH ZLOO DSSHDU ³([SRUWLQJ KDV EHHQFRPSOHWHG´ Ú Û
  Chapter 4                                    Using 0LUR9LHZ Software  4-125 4.7.2  Import 0LUR9LHZSoftware 1.0 files   Run the 0LUR9LHZOperator or 0LUR9LHZClient.  ([HFXWHWKH¶,PSRUW&DVH5HYLHZ)LOH«-!,PSRUW&DVH¶LQWKHµ/LVW¶menu
  Using MiroView Software                                   Chapter 4   4- 126  Select the 0LUR9LHZ Software 1.0 files from the file extension drop-down box and look for the file to upgrade   /RRNIRUWKH)LOHZLWKH[WHQVLRQµ0LUR&DP¶DQGFOLFNµImport¶.   Carry out in alphabetical order like following picture.   Ú Û
  Chapter 4                                    Using 0LUR9LHZ Software  4-127   Task Manager will appear and will proceed with importation of the file.   Progress bar will appear.   Upgrading the 0LUR9LHZ Software 1.0 file to the 0LUR9LHZSoftware file will be completed.
  Using MiroView Software                                   Chapter 4   4- 128 4.7.3  0LUR9LHZSoftware 1.0 files Batch Import   If the 0LUR9LHZSoftware 1.0 is already installed on the workstation, user can use the 0LUR9LHZ Software 1.0 files Batch Import function via <Menu -> Tools -> Import from the 0LUR9LHZSoftware 1.0> in 0LUR9LHZOperator (or <Menu -> Home -> Import from the 0LUR9LHZ Software 1.0> in the 0LUR9LHZ Client) so that 0LUR9LHZcan use the imported cases. Choose the cases to import and click the  .    To import cases without video file, uncheck the check box on the Video column.
  Chapter 4                                    Using 0LUR9LHZ Software  4-129 4.7.4 Using 0LUR9LHZ Client at Local PC without connecting to 0LUR9LHZServer   Run the 0LUR9LHZ Client by double-clicking the   on the desktop or from the Windows Start-up menu. Log in.   Carry out in alphabetical order like following picture. *If running the 0LUR9LHZ Client as Local Mode, click the   on the Log-in screen.  $W WKH ILUVW UXQQLQJ RI µ/RFDO 0RGH¶ WKH ZLQGRZ ZLOO DSSHDU OLNHbelow. Input the appropriate values in to the empty fields and click the   to proceed.   Carry out in alphabetical order like following picture. Ú Û Ü Ý Ú Û Ü
  Using MiroView Software                                   Chapter 4   4- 130    Open the Case/Review via Menu Æ Home Æ Open Case/Review ILOH«     Choose a file to open and click the  .    After completely reviewing the file, save the review via MenuÆ Home Æ Save Active Review so that the user can import the edited review file on the 0LUR9LHZServer in the future.
  5                 8VLQJ0LUR9LHZ579
  8VLQJ0LUR9LHZ579                                       Chapter 5   5- 2
  Chapter 8VLQJ0LUR9LHZRTV  5-3 5.  86,1*0,529,(:579  The  0LUR9LHZ 579 D software application of the MiroCam® System Real-time Viewer, consist of a data transmission module that directly receives the image data from the MiroCam® Receiver and an output module for image display.    5.1 Overview  The MiroView579VRIWZDUHFDQEHLQVWDOOHGRQDQ\PRELOHSODWIRUP&OLFN WKH FRUUHVSRQGLQJ LFRQ RQ WKH GHVNWRS WR ODXQFK WKH 0LUR9LHZRTV.  Basic Screen Layout   c Patient Information Display patient name, ID, gender and age d Operation Button Change the 0LUR9LHZ 579 RSHUDWLRQ VWDWXVbetween start, pause and resume. e Main Display  Display captured image. f Time Display  Indicates the MiroCam® Capsule endoscopy procedure time. £ ¢ ¤ ¥
  8VLQJ0LUR9LHZ579                                       Chapter 5   5- 4 Recommended Hardware Specification    OS: Windows Vista (32 bit), Windows XP (32 bit), Windows7 (32 bit)   CPU: Atom  N450(1.66GHz) or higher  Memory: 1GB or higher  Display Adaptor: Windows Compatible   Hard Disk: 80GB memory space available   Monitor Resolution: 1024 X 600 or higher   Monitor Contrast : Windows Compatible   In compliance with IEC60950-1 or IEC60601-1
  Chapter 8VLQJ0LUR9LHZRTV  5-5 5.2 8VLQJ0LUR9LHZ579   5.2.1  %HIRUH8VLQJ0LUR9LHZ579    Installing tKH0LUR9LHZ579VRIWZDUHon a mobile PC.    WARNING  Only recharge the Mobile PC batteries when MiroCam® Receiver does not connect to the Mobile PC.    WARNING  For safe use of the Mobile PC, please refer to the Mobile PC User Manual.  WARNING  DO NOT use the 0LUR9LHZ579VRIWZDUHDQGthe Mobile PC before upgrading MR1000.  WARNING  DO NOT upgrade MiroCam® Receiver during the MiroCam® Capsule endoscopy procedure, it may result loss of image data.  WARNING UVH WKH 0LUR9LHZ RTV Software to check if the MiroCam® Capsule entered to the duodenum, small bowl; during a capsule endoscopy procedure. Thus, prior to UXQQLQJ WKH 0LUR9LHZ RTV software, please make sure that the MiroCam® Capsule endoscopy procedure has been undergone.    Connect the data cable of the Real-time Viewer to the MiroCam® Receiver.  NOTE  The data cable for the MiroView579KDVQRFRYHUIRUWKHUSB port on the MiroCam® Receiver.      Follow the MiroCam® Capsule endoscopy procedure.
  8VLQJ0LUR9LHZ579                                       Chapter 5   5- 6  NOTE  For more information on the MiroCam® Capsule endoscopy procedure, please refer to Chapter 3, Performing Capsule Endoscopy.  5.2.2 Starting 0LUR9LHZ RTV   Switch on the Mobile PC.  Click the    on the screen to launch the 0LUR9LHZ579   Connect the small side of the USB cable to the MiroCam® Receiver.   Connect the large side of the USB cable to the Mobile PC.   Click the    to start image display  Click the   to stop image display  Click the   to restart image display  Click the    on the top-ULJKWWRFORVHWKH0LUR9LHZ579  NOTE  To Start, Pause, and Resume use the same button. The function (Start, Pause or Resume) changes based upon the status of the 0LUR9LHZ579VRIWZDUH  CAUTION  DO NOT disconnect the MiroCam® Receiver while the RTV software is running. If the MiroCam® Receiver is disconnected during use, the software may need to be restarted.
  Chapter 8VLQJ0LUR9LHZRTV  5-7  NOTE  If an error message appears while using the 0LUR9LHZRTV, reconnect the USB cable between the MiroCam® Receiver and the Mobile PC  5.2.3 After Using 0LUR9LHZ RTV    Disconnect USB cable between the MiroCam® Receiver and the Mobile PC.  Continue MiroCam® Capsule endoscopy procedure.
  8VLQJ0LUR9LHZ579                                       Chapter 5   5- 8
  6         Using :LUHOHVV0LUR9LHZ579L
   Using Wireless MiroView RTVi                              Chapter 6   6- 2
  Chapter 6                              Using Wireless MiroView RTVi  6-3 6. USING :,5(/(660,529,(:579L  The  :LUHOHVV 0LUR9LHZ 579L, a software application of MiroCam® System Real-time Viewer, consist of a data transmission module that directly receives the image data from the MiroCam® Receiver and an output module for image display. The :LUHOHVV 0LUR9LHZ 579L receives image data via Wi-Fi channel.    6.1 Overview  The :LUHOHVV0LUR9LHZ579L software can be installed on iOS running on iPad / iPhone. Click the corresponding on the desktop to launch the :LUHOHVV0LUR9LHZ579L.
   Using Wireless MiroView RTVi                              Chapter 6   6- 4 6.1.1  Basic Screen Layout ± iPhone      c Patient Information Display patient name, ID, gender and age d HDR Button  High Dynamic Range Image Button e Main Display  Display captured image. f Wi-Fi Operation Time Indicates the MiroCam® Wi-Fi operation time ྜྷ Recording time Indicates the MiroCam® Capsule recording time.  £ ¢ ¤ ¥ ¦
  Chapter 6                              Using Wireless MiroView RTVi  6-5 6.1.2  Basic Screen Layout ± iPad  c Patient Information Display patient name, ID, gender and age d HDR Button  High Dynamic Range Image Button e Main Display  Display captured image. f Wi-Fi Operation Time Indicates the MiroCam® Wi-Fi operation time ྜྷ Recording time Indicates the MiroCam® Capsule recording time.  NOTE  If turning the iPad upright as Portrait View, the MiroCam® Receiver List will disappear from the Main Display to be enlarged.   6.1.3  Recommended Hardware Platform    iPhone 3G, 3GS, 4, 4S, 5  iPad I, II (with only Wi-Fi) ¢ £ ¤ ¥ ¦
   Using Wireless MiroView RTVi                              Chapter 6   6- 6 6.2 Using  :LUHOHVV0LUR9LHZ579L  6.2.1 Before Using :LUHOHVV0LUR9LHZ579L   Installing tKH :LUHOHVV 0LUR9LHZ 579L VRIWZDUH on iOS on iPhone or iPad.     The effective distance between the MiroCam® Receiver and iPhone or iPad is within 3m.  WARNING  Use the :LUHOHVV0LUR9LHZ579L Software to check if the MiroCam® Capsule entered to the duodenum and small bowl during a capsule endoscopy procedure. Thus, prior to running the :LUHOHVV 0LUR9LHZ 579L software, please make sure that the MiroCam® Capsule endoscopy procedure has been undergone.   NOTE  For more information on the MiroCam® Capsule endoscopy procedure, please refer to Chapter 3, Performing Capsule Endoscopy.  NOTE  When the :LUHOHVV 0LUR9LHZ 579L run at first, network configuration for iPhone or iPad is required.  6.2.2  Turn on the Wi-Fi function of the MiroCam® Receiver.    Press the Wi-Fi switch of MiroCam® Receiver.   Keep pressing it until the buzzer sounds.
  Chapter 6                              Using Wireless MiroView RTVi  6-7 WARNING  DO NOT press the Wi-Fi switch when the buzzer sounds. It can cause the lost of captured images.    WARNING The Wi-Fi switch of MiroCam® Receiver can be turned on involuntarily. Therefore, be careful it is not turned on.    WARNING  The MiroView RTVi application can only be used for a total of 30 minutes during the entire MiroCam® Capsule Endoscopy Procedure.   6.2.3 Wireless MiroView579L.    Run the Settings for the iOS.
   Using Wireless MiroView RTVi                              Chapter 6   6- 8   Turn on the Wi-Fi. If Wi-Fi is already turned on, please turn it off and turn it back on. Carry out in alphabetical order like following picture.   Û Ú
  Chapter 6                              Using Wireless MiroView RTVi  6-9 6.2.4  TCP/IP Configuration (for the first time to connect a MiroCam® Receiver access point: AP)    Check a network with SSID of MiroCam_xxx in the network list.  Click the    on the SSID (MiroCam_xxx).  NOTE  TCP/IP configuration is needed once at first.
   Using Wireless MiroView RTVi                              Chapter 6   6- 10  Click the  .  Click the    on the top-left after entering the IP Address and Subnet Mask. (Setup as Static; IP Address: 192.168.3.200, Subnet Mask:255.255.255.0)   Carry out in alphabetical order like following picture.  Ú Û
  Chapter 6                              Using Wireless MiroView RTVi  6-11   Click the SSID.     Click the  .
   Using Wireless MiroView RTVi                              Chapter 6   6- 12   Enter the password '1234567890' in the field with instruction of system. Afterward, click the  .Carry out in alphabetical order like following picture.   Press home button when network connection finished.    Ú Û
  Chapter 6                              Using Wireless MiroView RTVi  6-13 6.2.5 Running :LUHOHVV0LUR9LHZ579L   Click the    on the desktop to launch the :LUHOHVV0LUR9LHZRTVi.    The Splash window will appear for a while and then will disappear.
   Using Wireless MiroView RTVi                              Chapter 6   6- 14   Enter the password '1234567890' in the field with instruction of system. Afterward, click the  . Carry out in alphabetical order like following picture.     Ú Û
  Chapter 6                              Using Wireless MiroView RTVi  6-15    Check whether Patient Information, MiroCam® Receiver Scan Button, Main Display, Wi-Fi Operation Time and Recording time are displayed normally.  c Patient Information Display patient name, ID, gender and age d HDR Button  High Dynamic Range Image Button e Main Display  Display captured image. f Wi-Fi Operation Time Indicates the MiroCam® Wi-Fi operation time ྜྷ Recording time Indicates the MiroCam® Capsule recording time.  ¢ £ ¤ ¥ ¦
   Using Wireless MiroView RTVi                              Chapter 6   6- 16 6.2.6 After Using :LUHOHVV0LUR9LHZ579L   7R FORVH WKH :LUHOHVV 0LUR9LHZ 579L press the home button and then check whether the Wi-Fi light of the MiroCam® Receiver is off.  NOTE  If MiroCam® Receiver is turned off, following window will appear after about 10 seconds. To Close this window, click the  .     Run the Settings for the iOS.
  Chapter 6                              Using Wireless MiroView RTVi  6-17   Turn Wi-Fi off and turn it back on.
   Using Wireless MiroView RTVi                              Chapter 6   6- 18
  A                 Appendix
  Case Form Report   A- 2
  Case Form Report  A-3  Trial ID :      Version No. : 1.0 : 30/07/2007 CASE REPORT FORM  Subject No.:    Subject Initials :    Center No.:       Principal Investigator  :   Center Name  :   Address :       Telephone No.  :   Fax No.  :   E-mail :                       CONFIDENTIAL
  Case Form Report   A- 4 Trial ID :     Version No. : 1.0 : 30/07/2007  Screening Subject No.:    Subject Initials :    Center No.:   Date of Informed consent signed           D M M Y Y Y Y  Inclusion Criteria  Yes No N/A 1. Informed consent form voluntarily signed        2. Male / Female subject of age 19 years~70 years.                       If there is one or more PDUNRIµ1R¶\RXVKDOOEHFRQWUDGLFWHGIRUWKLVFOLQLFDOWHVW  CONFIDENTIAL
  Case Form Report  A-5  Trial ID :      Version No. : 1.0 : 30/07/2007 Screening Subject No.:    Subject Initials :    Center No.:   Date of visit :             D M M Y Y Y Y  Exclusion Criteria  Yes No N/A 1. Do you have a gastrointestinal disorder (suspected malignantlesions or gastrointestinal obstructive disorders, perforation, stenosis or fistula)?       2.  Do you have a cardiovascular disorder (dysrhythemia, trouble with cardiac conduction system, hypertension(high blood pressure) or ischemic heart disease)?       3.  Have you been diagnosed of cardiac insufficiency, angina pectoris or cardiac arrest (heart attack) in recent 6 months?        4.  Are you with (or implanted) medical devices such as cardiac pacemaker or defibrillator?        5.    Do you have a swallowing disorder?        6.  Do you have delayed gastric emptying or serious gastric disorder?        7.  Are you a pregnant or nursing woman, or a woman of child bearing potential?        8.  Are you an infant or epileptic? Are you easily affected by electronic waves?        9.    Do you have several small bowel diverticulums ?        10.    Did you fail to obtain a voluntary agreement?        11.    Are you judged unqualified for this test by doctors or nurses?        ,IWKHUHLVRQHRUPRUHPDUNRIµ<HV¶\RXVKDOOEHFRQWUDGLFWHGIRUWKLVFOLQLFDOWHVW CONFIDENTIAL
  Case Form Report   A- 6
               EC Representative SeDa S.p.A 20090 Trezzano sul Naviglio(MI),   Via Toistoi, 7 Italy   Office: +39-02-484241 Mobile: +39-02-48424290     IntroMedic Co.,Ltd. Suite 1104, E&C Venture Dream Tower 6-Cha 197-28 Guro-Dong, Guro-Gu, Seoul, KOREA 152-719 Tel : +82-2-801-9300 Fax : +82-2-801-9330 http://www.intromedic.com e-mail: helpdesk@intromedic.com  0120 Disposal of Old Electrical & Electronic Equipment Applicable in the European Union and other European countries with separate collection systems CAUTION: Federal (U.S.A) law restricts this device to sale by or on the order of a physician.
)&& 127,&(7KHGHYLFHFRPSOLHVZLWK3DUWRIWKH)&&5XOHV2SHUDWLRQLVVXEMHFWWRWKHFRQGLWLRQ WKDW WKLV GHYLFH GRHV QRW FDXVH KDUPIXO LQWHUIHUHQFH127(  7KLV HTXLSPHQW KDV EHHQ WHVWHG DQG IRXQG WR FRPSO\ ZLWK WKH OLPLWV IRU D&ODVV % 'LJLWDO 'HYLFH SXUVXDQW WR 3DUW  RI WKH )&& 5XOHV 7KHVH OLPLWV DUHGHVLJQHG WR SURYLGH UHDVRQDEOH SURWHFWLRQ DJDLQVW KDUPIXO LQWHUIHUHQFH LQ DUHVLGHQWLDO LQVWDOODWLRQ 7KLV HTXLSPHQW JHQHUDWHV XVHV DQG FDQ UDGLDWH UDGLRIUHTXHQF\ HQHUJ\ DQG LI QRW LQVWDOOHG DQG XVHG LQ DFFRUGDQFH ZLWK WKH LQVWUXFWLRQPD\ FDXVH KDUPIXO LQWHUIHUHQFH WR UDGLR FRPPXQLFDWLRQ +RZHYHU WKHUH LV QRJXDUDQWHH WKDW LQWHUIHUHQFH ZLOO QRW RFFXU LQ D SDUWLFXODU LQVWDOODWLRQ ,I WKLVHTXLSPHQW GRVH FDXVH KDUPIXO LQWHUIHUHQFH WR UDGLR RU WHOHYLVLRQ UHFHSWLRQ ZKLFKFDQ EH GHWHUPLQHG E\ WXUQLQJ WKH HTXLSPHQW RII DQG RQ WKH XVHU LV HQFRXUDJHG WRWU\ WR FRUUHFW WKH LQWHUIHUHQFH E\ RQH RU PRUH RI WKH IROORZLQJ PHDVXUHV  5HRULHQW RU UHORFDWH WKH UHFHLYLQJ DQWHQQD  ,QFUHDVH WKH VHSDUDWLRQ EHWZHHQ WKHHTXLSPHQW DQG UHFHLYHU  &RQQHFW WKH HTXLSPHQW LQWR DQ RXWOHW RQ D FLUFXLWGLIIHUHQW IURP WKDW WR ZKLFK WKH UHFHLYHU LV FRQQHFWHG  &RQVXOWWKHGHDOHURUDQH[SHULHQFHG UDGLR79 WHFKQLFLDQ IRU KHOS&DXWLRQ$Q\ FKDQJHV RU PRGLILFDWLRQV LQFOXGLQJ WKH DQWHQQD PDGH WR WKLV GHYLFH WKDW DUHQRW H[SUHVVO\ DSSURYHG E\ WKH PDQXIDFWXUHU PD\ YRLG WKH XVHUឧV DXWKRULW\ WRRSHUDWH WKH HTXLSPHQ