430-95520-001 (Rev. 04/05)ABOptions/Vol InfBackPrimeFor use with Lists: 12391, 20679-04,and 11971-04 (-03 and above ONLY)System Operating ManualRfor usewithfor usewithHOSPIRA, INC., LAKE FOREST, IL 60045, USADraft Manual- Not to be usedin a Clinical Situation.
Draft Manual- Not to be usedin a Clinical Situation.Change History430-95520-001 (Rev. 04/05) Change HistoryTitleDescription of Changes Pages Affected430-95520-001 (Rev. 04/05)First Release All
430-95520-001 (Rev. 3/05)ContentsSection 1, Descriptive Information  . . . . . . . . . . . . . . .  1-1PRODUCT DESCRIPTION  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1INDICATIONS FOR USE   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2user qualification   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2CONVENTIONS   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2warnings, cautions, and notes  . . . . . . . . . . . . . . . . . . . . . 1-3PRECAUTIONS   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4healthcare professionals and patient related  . . . . . . . . . . 1-4concurrent flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5epidural administration  . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7battery operation   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8sets and accessories  . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9backpriming  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9general  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9cleaning  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10bolus related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1-10artifacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1-11GUIDENCE ON EMC COMPATIBILITY . . . . . . . . . . . . . . . . . . . . . .  1-13FCC INFORMATION  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1-14Section 2, Principles of Operation   . . . . . . . . . . . . . . .  2-1FEATURES  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1programs   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1line programming options . . . . . . . . . . . . . . . . . . . . . . . . . 2-1plumset capabilities  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2air management  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2battery   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2biomedical   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2options  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2-2other features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3Section 3, Equipment Description   . . . . . . . . . . . . . . .  3-1OPERATING KEYS  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2INDICATORS  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4DISPLAY SYMBOLS  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Contents430-95520-001 (Rev. 3/05) REAR CASE CONTROLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6ADMINISTRATION SETS   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7PREPARING THE ADMINISTRATION SET . . . . . . . . . . . . . . . . . . . . 3-7PRIMING THE ADMINISTRATION SET  . . . . . . . . . . . . . . . . . . . . . . 3-8LOADING THE CASSETTE  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9PREPARING THE SECONDARY LINE   . . . . . . . . . . . . . . . . . . . . . 3-10DISCONTINUING ELECTRONIC FLOW CONTROL & SETTING GRAVITY FLOW   . . . . . . . . . . . . . . . . . . . 3-11DISCONTINUING FLUID ADMINISTRATION . . . . . . . . . . . . . . . . . . 3-12Section 4, Basic Operation   . . . . . . . . . . . . . . . . . . . . .4-1GETTING STARTED  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1unpacking  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1tandem carrier instructions  . . . . . . . . . . . . . . . . . . . . . . . . 4-2system self-tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2data retention  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3POWER ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4SIMPLE DELIVERY  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6TITRATION   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8PIGGYBACK DELIVERY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12CONCURRENT DELIVERY  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15STOP AND START WITH ONLY 1 LINE PUMPING . . . . . . . . . . . . . 4-18STOP AND START WITH BOTH LINES PUMPING  . . . . . . . . . . . . . 4-19BACKPRIMING  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20CLEARING SETTINGS  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21UPDATING DRUG LIBRARY  . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23Section 5, Advanced Programs  . . . . . . . . . . . . . . . . . .5-1DOSE CALC (MCG/KG/MIN ON A)   . . . . . . . . . . . . . . . . . . . . . . . . 5-2DOSE CALC (MG/MIN ON B)  . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9LOADING DOSE  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15MULTISTEP PROGRAMMING  . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22Section 6, Additional Features . . . . . . . . . . . . . . . . . . .6-1SIMPLE DELIVERY USING DELAYED START . . . . . . . . . . . . . . . . . 6-1PIGGYBACK WITH NURSE CALLBACK   . . . . . . . . . . . . . . . . . . . . . 6-2POSSIBLE NON-DELIVERY PROGRAMMED . . . . . . . . . . . . . . . . . . 6-4USING THE STANDBY FEATURE  . . . . . . . . . . . . . . . . . . . . . . . . . 6-4SELECT OPTION- VOLUMES INFUSED, PRESSURE/POST INFUSION
430-95520-001 (Rev. 3/05)RATE, AND LIGHTING/CONTRAST . . . . . . . . . . . . . . . . . . . . . . . . 6-6VARIABLE RATE CAP   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10EXAMPLES OF AUTOMATIC CALCULATION . . . . . . . . . . . . . . . . . 6-10at startup   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10while running (titration) . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10at kvo  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6-11CHANGING CCA WHILE INFUSING  . . . . . . . . . . . . . . . . . . . . . . 6-12AUTO-PROGRAMMING  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13Section 7, Alarms and Troubleshooting  . . . . . . . . . . .  7-1WARNING MESSAGES  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1RESPONSE TO ALARMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2GENERAL ALARMS   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7-3LINE A ALARMS   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7-5LINE B ALARMS   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7-7MALFUNCTIONS   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7-8GENERAL MESSAGES  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9Section 8, Cleaning, Maintenance, and Storage   . . . .  8-1cleaning and sanitizing . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1battery maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3storage  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4service  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4Section 9, Specifications   . . . . . . . . . . . . . . . . . . . . . .  9-1PHYSICAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1ELECTRICAL  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1WIRELESS LAN DEVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9-2VTBI RANGE  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9-2ENVIRONMENT  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3DELIVERY RATE RANGE  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9-3AIR-IN-LINE ALARM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3OCCLUSION ALARM AND LIMITS . . . . . . . . . . . . . . . . . . . . . . . . . 9-4TIME TO DETECT DOWNSTREAM OCCLUSIONS . . . . . . . . . . . . . . 9-5BOLUS VOLUME RELEASED AFTER DOWNSTREAM OCCLUSIONS ARECORRECTED  . . . . . . . . . . . . . . . . . . . . . . . . . 9-5DELIVERY ACCURACY  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6ENTERAL & HIGH VISCOSITY FLUIDS EFFECTS . . . . . . . . . . . . . . 9-6TRUMPET CURVES  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Contents430-95520-001 (Rev. 3/05) example   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7Section 10, Supplies and Accessories . . . . . . . . . . . .10-1ADMINISTRATION FLUIDS  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1administration fluids  . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1enteral and high viscosity fluids  . . . . . . . . . . . . . . . . . . . 10-1containers  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2general  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2networked application . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-2Section 11, Warranty  . . . . . . . . . . . . . . . . . . . . . . . . .11-1Section 12, Default Drug Library (DDL)  . . . . . . . . . . . 12-1
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 1SECTION 1Descriptive InformationThe Plum A+ Volumetric Infusion System is designed to meet the fluid delivery requirements of today’s evolving healthcare environments. The infuser is a cassette based multi-function infusion system. The Plum A+ allows two lines in and one line out. The infuser can be used for standard, piggyback, or concurrent delivery. Programming modes include:• Standard Infusions• Multistep Programming• Loading Dose• Dose CalculationThe Plum A+ is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types.The Plum A+ is designed to be used in most areas of patient care, including, but not limited to:Product DescriptionThe system includes a pumping module (hereafter called the infuser) and an assortment of disposable IV sets (hereafter called a set), optional accessories, and this operator’s manual.The Plum A+ host device contains a Connectivity Engine peripheral module that provides wired Ethernet and wireless • General Floor • Labor/Delivery/Post Partum• Burn Unit• Medical/Surgical • OR/Anesthesia • Hemodialysis• ICU/CCU • Post Op/Recovery • Oncology• Pediatrics • Cardiac Cath Lab • Mobile Intensive Care• Neonatology • Emergency • Nutritional
1- 2 430-95520-001 (Rev. 04/05) 802.11b local area networking capabilities. This allows the Hospira MednetTM networked application software for drug library downloads  and Auto-programming to the infuser.Indications for UseUSER QUALIFICATIONThe Plum A+ is intended for use at the direction or under the supervision of licensed physicians or certified healthcare professionals who are trained in the use of the infuser and the administration of parenteral, enteral, and epidural fluids and drugs and whole blood or red blood cell components. The training should emphasize preventing related IV complications, including appropriate precautions to prevent accidental infusion of air. The epidural route can be used to provide anesthesia or analgesia.WARNINGADMINISTER ONLY ANESTHETICS/ANALGESICSAPPROVED FOR EPIDURAL ADMINISTRATION (ASINDICATED OR ALLOWED BY THE DRUGS’ FDAAPPROVED LABELING). EPIDURAL ADMINISTRATION OFDRUGS OTHER THAN THOSE INDICATED FOR EPIDURALUSE COULD RESULT IN SERIOUS INJURY TO THE PATIENT.ConventionsThis section describes the conventions used throughout this manual, as follows:CONVENTION APPLICATION EXAMPLEItalic Reference to a section, figure, or tableFunction or mode specific instructions(See Figure 3-1, Priming Cassette)Primary Only: Attach an empty container.
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 3WARNINGS, CAUTIONS, AND NOTESAlert messages used throughout this manual are described below. Pay particular attention to these messages.WARNINGA WARNING MESSAGE CONTAINS SPECIAL SAFETYEMPHASIS AND MUST BE OBSERVED AT ALL TIMES.FAILURE TO OBSERVE A WARNING MESSAGE ISPOTENTIALLY LIFE THREATENING.CAUTION: A CAUTION USUALLY APPEARS IN FRONT OF APROCEDURE OR STATEMENT. IT CONTAINS INFORMATION THATCOULD PREVENT IRREVERSIBLE PRODUCT DAMAGE OR HARDWAREFAILURE. FAILURE TO OBSERVE A CAUTION COULD RESULT INSERIOUS PATIENT OR USER INJURY.[BRACKETEDALL CAPS]Keys or buttons on the device are displayed in [BRACKETED ALL CAPS] or with a graphic.[START]or V [Italic] Softkey OptionsV [Choose]Initial Caps lowercaseScreen displays and device labels (as appropriate)ProgramDose CalculationBold Emphasis ...sets are supplied Sterile and are for....CONVENTION APPLICATION EXAMPLE
1- 4 430-95520-001 (Rev. 04/05) NOTE: A Note highlights information that helps explain a concept or procedure.This symbol directs the user to consult accompanying documents.When visible on the display, this symbol informs the user to use CAUTION because the specified drug has NOT been programmed with specified safety limits.NOTE: Figures are rendered as graphic representations to approximate the actual product. Therefore, figures may not exactly reflect the product.PrecautionsThe Plum A+ has been designed and manufactured to be safe, reliable, and easy to use. This section details precautions and possible hazards.For safe operation of the Plum A+, observe the following precautions and hazards.HEALTHCARE PROFESSIONALS AND PATIENT RELATED• In vitro studies have suggested that packed red blood cells with unusually high hematocrit be diluted with blood-compatible fluids, such as 0.9% sodium chloride injection, to decrease hemolysis and increase flow rate.• Setting the primary rate greater than the secondary rate will result in a more rapid infusion of any residual secondary drug remaining in the line and the cassette.• Consult drug labeling to confirm drug compatibility, concentration, delivery rates, and volumes are all suitable for secondary, concurrent and piggyback delivery modes.• Arrange tubing, cords, and cables to minimize the risk of patient strangulation or entanglement.• Before opening the door, close clamp on the primary line or remove the secondary container from the secondary port to prevent mixing of primary and secondary fluids.
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 5• Although unlikely, failure of certain robust mechanical components such as the anti-free flow mechanism or valve control springs could cause fluid delivery limited to the contents of the fluid container. Single fault failure of certain electronic/motor control components would result in no more than 5 mL of unexpected fluid delivery.• A small amount of fluid is expelled from the set (less than 0.05 ml) each time the door is opened or closed with a set installed. If potent drugs are being used, take appropriate action to guard against overmedication of the patient.• Before disconnecting a syringe from the cassette, pull up the plunger slightly to avoid spilling the fluid. For rigid containers, close the upper slide clamp, open the cassette door, then remove and invert the cassette (ports down).• Air bubbles may form distal to the cassette as result of normal outgassing of dissolved air in the fluid.   This may occur if chilled solution is in use, if the infuser is mounted significantly above the patient, or when using certain fluids known to routinely outgas.   In these cases, an air eliminating filter may be used.• Repeated opening and closing of the door may defeat the proximal air-in-line alarm and may cause a distal air-in-line alarm, requiring repriming. • The screen displays the VTBI (volume to be infused) in integers when value is above 99.9. Any fraction of a milliliter delivered is not displayed, but is retained in memory.CONCURRENT FLOWGUIDELINESWhen delivering short half-life critical drugs (see Critical Drugs, this section) using the Plum A+ in the Concurrent mode, the following delivery rate guidelines should be observed:• If the critical drug (with half-life less than 6 minutes) is to be infused at less than 2.0 mL/hr, the other infusion should be no faster than 5 times the critical drug’s rate. Dopamine, for example, delivered at 1.5 mL/hr should not be accompanied by an infusion programmed any faster than 7.5 mL/hr.
1- 6 430-95520-001 (Rev. 04/05) • If the critical drug (with half-life less than 6 minutes) is to be infused at 2.0 to 5.0 mL/hr the other infusion should be no faster than ten times the critical drug’s rate. Dopamine, for example, delivered at 3.5 ml/hr should not be accompanied by an infusion programmed any faster than 35 mL/hr.• If the critical drug (with half-life less than 6 minutes) is to be infused at 5.1 mL/hr or greater, the other infusion can be programmed at any desired rate.NOTE: The total of the primary rate plus the secondary rate cannot exceed 500 mL/hr.These guidelines apply only when infusing short half-life critical drugs in Concurrent mode. Individual patient responses may vary requiring adjustment of delivery rates.DELIVERY RATE GUIDELINESSHORT HALF-LIFE (LESS THAN 6 MINUTES) CRITICAL DRUG INFUSION RATEMAXIMUM RATE OF ACCOMPANYING INFUSION0.5 - 1.9 mL/hr 5 Times the Critical Drug Rate2.0 - 5.0 mL/hr 10 Times the Critical Drug Rate5.1 or Greater Any Desired Ratio
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 7CRITICAL DRUGSExamples of drugs with a short half-life (approximately 6 minutes or less when given IV) include:For these drugs, the concurrent flow guidelines should be followed when the infusion rate of the drug will be 5 mL/hr or less.NOTE: This list of critical drugs is not intended to be all-inclusive of critical drugs or drugs with a short half-life.The clinician should become familiar with the pharmacodynamics of any critical drug before administration.This information is presented to inform clinicians of a rare situation that could be misinterpreted if they are unfamiliar with this phenomenon.EPIDURAL ADMINISTRATION• Recommended use of the epidural route is to provide anesthesia or analgesia for periods up to 96 hours.• This device can be used to administer only those anesthetics/analgesics approved for epidural administration (as indicated or allowed by the drugs’ FDA approved labeling). Epidural administration of drugs other than those indicated for epidural use could result in serious injury to the patient.• For epidural administration, the use of Hospira  catheters, Plum sets without Y-sites, and "epidural" stickers indicating ongoing epidural administration are recommended.Dobutamine LidocaineDopamine NitroglycerinEpinephrine NitroprussideEpoprostenol NorepinephrineEsmolol OxytocinIsoproterenol Procainamide
1- 8 430-95520-001 (Rev. 04/05) • Administration of drugs via the epidural route should be limited to personnel familiar with associated techniques and patient management problems. Proper epidural placement of the catheter is essential since catheter migration could result in intravascular or intrathecal administration. Facilities practicing epidural administration must be equipped with resuscitative equipment, oxygen, naloxone, and other resuscitative drugs. Adequate monitoring equipment (e.g., Oximetry) is recommended for continuous monitoring of the patient during epidural administration. Patients must be observed frequently for side effects in a fully-equipped and staffed environment for at least 24 hours following completion of drug administration by the epidural route. DELAYED RESPIRATORY DEPRESSION FOLLOWING CONTINUOUS EPIDURAL ADMINISTRATION OF PRESERVATIVE-FREE MORPHINE SULFATE HAS BEEN REPORTED.• The epidural space has 58 openings through which fluid can exit. Pressure buildup during administration is transient. However, if a large volume of fluid is administered over a short time period, the pressure will take longer to return to normal. If overdelivery occurs during administration, observe the patient closely for signs of spinal cord compression (disorientation, headache, transient neuralgias) and drug overdose.BATTERY OPERATION• When the battery is removed from  the Plum A+, do not operate on patients. Use of a properly maintained and charged battery helps confirm proper operation. • The battery may not be fully charged upon receipt. Connect the infuser to AC power for at least six hours.• Use AC power whenever possible. Connect to AC power during storage to ensure a fully charged battery for emergencies. If the quality of the earth grounding source is in doubt, use battery power.• If the low-battery alarm sounds, connect the infuser to AC power immediately.
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 9SETS AND ACCESSORIES• Only compatible LifeCare PlumSets can be used with the Plum A+. See individual set instructions for additional information.• Administration sets should be changed per CDC guidelines or healthcare provider policy. Discard after use.• LifeCare IV infusion sets with integral nonblood filters are not for use in the administration of blood, blood products, emulsions, suspensions, or any medications not totally soluble in the solution being administered. These medications may be administered through the lower Y-injection site, below the filter.• When infusing at low delivery rates (5.0 mL/hr or less) the use of thick-walled microbore PlumSets is recommended. This will reduce the amount of the fluid bolus that may be delivered when a distal line occlusion is released.• Syringes must be larger than 3 cc. Use syringe adapter (List 11986-48) when using syringes smaller than 10cc. Some 10cc syringes may require use of a syringe adapter. Syringes larger than 10cc may be attached directly to the secondary port of the cassette. Use of a syringe adapter may decrease the occurrence of proximal occlusion alarms.• Use a 19-gauge or larger needle or catheter at the venipuncture site for viscous fluids if operating at rates greater than 500 ml/hr.See Section 10 for information on sets and accessories.BACKPRIMING• Backpriming is not recommended for reconstituting secondary containers containing dry powders.• To avoid pressurization when backpriming into a syringe, the user must confirm there is sufficient empty space to accept the backprimed fluid.GENERAL• Possible explosion hazard exists if used in the presence of flammable anesthetics.
1- 10 430-95520-001 (Rev. 04/05) • Do not place Plum A+ in service if it fails the self-test.• Do not operate the Plum A+ with the case opened.• Keep the cassette door securely closed while the infuser is not in use, to avoid cassette door damage.• Values beyond a fields maximum hard limit will be diplayed as dashes (-- -- --). The user must clear these fields using the [CLEAR] key prior to entering new values.CLEANINGFor more information on cleaning the infuser, see Section 8.• To avoid mechanical or electronic damage, do not immerse the Plum A+ in any fluids or cleaning solutions.• Do not spray cleaning solutions toward any opening in the instrument.• Certain cleaning and sanitizing solutions may slowly degrade components made from some plastic materials. Using abrasive cleaners or cleaning solutions not recommended by Hospira may result in product damage. Do not use compounds containing combinations of isopropyl alcohol and dimethyl benzyl ammonium chloride.• Never use sharp objects such as fingernails, paper clips, or needles to clean any part of the infuser.• Do not sterilize by heat, steam, ethylene oxide (ETO), or radiation. • To avoid infuser damage, cleaning solutions should only be used as directed. The disinfecting properties of cleaning solutions vary; consult the manufacturer for specific information.BOLUS RELATEDUse the following procedure to avoid the administration of a bolus following a distal occlusion (i.e., a closed distal clamp):• If a secondary container is in use, clamp proximal tubing before opening cassette door.• Open cassette door and remove the cassette.
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 11• Open the flow regulator briefly to dissipate the pressure and then close it.• Eliminate the source of occlusion (closed clamp).• Reinsert the cassette and close the cassette door.• Open all clamps and resume infusion.NOTE: When troubleshooting an occlusion where all clamps are in the OPEN position, use care to avoid delivery of a bolus by opening the flow regulator to release any built-up pressure. Close the clamp between the cassette and the patient before opening the flow regulator to relieve the pressure. See Section 7, Alarms and Troubleshooting, for more information.ARTIFACTS• Nonhazardous, low-level electrical potentials are commonly observed when fluids are administered using infusion devices. These potentials are well within accepted safety standards, but may create artifacts on voltage-sensing equipment such as ECG, EMG, and EEG machines. These artifacts vary at a rate that is associated with the infusion rate. If the monitoring machine is not operating correctly or has loose or defective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actual physiological signals. To determine if the abnormality in the monitoring equipment is caused by the infusion device instead of some other source in the environment, set the infusion device so that it is temporarily not delivering fluid. Disappearance of the abnormality indicates that it was probably caused by the electronic noise generated by the infusion device. Proper setup and maintenance of the monitoring equipment should eliminate the artifact. Refer to the appropriate monitoring equipment system documentation for setup and maintenance instructions.• The Plum A+ system is designed to operate normally in the presence of most encountered electromagnetic interference (EMI) conditions. In the event of extreme levels of interference, such as encountered next to an electrosurgical generator, it is possible that the normal operation of a sensor
1- 12 430-95520-001 (Rev. 04/05) or microcomputer might be disrupted. Even in this event, the outcome would likely be a false alarm or detected system malfunction and would not result in a hazard to patient or operator.• This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN 55011 Class B and IEC/EN 60601-1-2:2001). These limits are designed to provide reasonable protection against harmful interference in a typical medical installation. The equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices (radio or television reception), which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:• Reorient or relocate the receiving device• Increase the separation between the equipment• Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected• Consult the manufacturer or field service technician for help• Portable and mobile RF communications equipment, such as cellular telephones, 2-way radios, Bluetooth devices, microwave ovens, in close proximity to this device may affect wireless and wired communications with the Infusion pump and/or the operation of the Infusion pump. Special precautions need to be exercised regarding EMC, These include:• Use of a shielded Ethernet cable (CAT5 STP or better) for plugging into the RJ45 Ethernet connector. Using an unshielded Ethernet cable may result in increased emissions.
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 13• Maintaining a minimum separation distance of 2 ½ ft between the Infusion pump system and portable/mobile RF communications equipment• List Numbers 12391 and 11971 are compliant to IEC/EN 60601-1-2 (1993)List Number 20679 is compliant to IEC/EN 60601-1-2 (2001) and have been tested and found to comply with EMC limits for the Medical Device Directive 93/42/EEC (EN 55011 Class B and EN 60601-1-2:2001).Guidance on EMC Compatibility• There is a shared responsibility between manufacturers, customers and users to ensure that Medical Equipment and Systems are designed and operated as intended. Medical electrical equipment needs special precautions regarding electromagnetic compatibility and needs to be installed and used according to the electromagnetic compatibility information provided in this manual.• The device is suitable for use in all establishments, including domestic establishments.  If extended operation during power mains interruption is needed, use battery power.• Always manage the electromagnetic environment. • The guidance included in this manual provides information needed to:• Determine the device’s suitability for use in the intended environment.• Manage the electromagnetic environment to permit the device to perform as intended without disturbing other equipment.• Separate the device from all other electronic equipment. If the device must be used near other electrical equipment, monitor the equipment to ensure there is no electromagnetic interference.• Devices should not be used adjacent to or stacked with other equipment. If the device must be used adjacent to or stacked
1- 14 430-95520-001 (Rev. 04/05) with other equipment, monitor the devices to verify normal operation.• USE ONLY components specifically labeled for use with the Plum A+ Infusion System to help ensure the device operates as intended.• If you suspect external RF sources or other equipment are influencing device operation, contact the biomedical engineering department for additional guidelines concerning electromagnetic immunity.• Contact the biomedical engineering department for additional information in the technical service manual concerning operating devices near RF sources. FCC InformationU.S. Pat. No. 5,681,285 FCC ID: STJ-20677IC: 5627A-20677US FCC (FEDERAL COMMUNICATIONS COMMISSION) STATEMENT• This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause interference, and (2) This device must accept any interference, including that may cause undesired operation of this device.FCC INTERFERENCE STATEMENT• This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy. If not installed and used in accordance with the instructions, it may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 1- 15to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try and correct the interference by one or more of the following measures:• Reorient or relocate the receiving antenna.• Increase the distance between the equipment and the receiver.• Connect the equipment to an outlet on a circuit different from that to which thereceiver is connected. • Consult the dealer or an experienced radio/TV technician for help. • Changes or modifications not expressly approved by Hospira could void the user's authority to operate the equipment.RADIO FREQUENCY EXPOSURE STATEMENT• The Wireless LAN radio device in the Connectivity Engine peripheral board with this infusion device has been evaluated and found compliant to the requirements of the following Radio Frequency exposure standards:• Federal Communications Commission, OET Bulletin 65 (Edition 97-01), Supplement C (Edition 01-01), Evaluating Compliance with FCC Guidelines for Human Exposure to Radio frequency Electromagnetic Fields, July 2001.• Industry Canada, Evaluation Procedure for Mobile and Portable Radio Transmitters with respect to Health Canada's Safety Code 6 for Exposure of Humans to Radio Frequency Fields, Radio Standards Specification RSS-102 Issue 1 (Provisional): September 1999.• Federal Communications Commission, OET Bulletin 65 (Edition 97-01), Supplement C (Edition 01-01), Evaluating Compliance with FCC Guidelines for Human Exposure to Radio frequency Electromagnetic Fields, July 2001.• The radiated output power of this Wireless LAN device is far below the FCC radio frequency exposure limits. The Wireless
1- 16 430-95520-001 (Rev. 04/05) LAN device has been evaluated with zero inches separation of human body from the antenna and found to be compliant with FCC RF exposure limits.
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 2- 1SECTION 2Principles of OperationFeaturesThe Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for hospital-wide, as well as alternate site and home healthcare, standardization. With its primary line, secondary line, and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical and critical care applications. Full compatibility with LifeCare Plum Series administration sets and accessories and the LifeShield® and CLAVE® needleless protection systems, makes the Plum A+ a convenient and cost-effective infusion system.PROGRAMS• Standard Programming• Loading Dose• Multistep ProgrammingLINE PROGRAMMING OPTIONS• Drug List• Nurse Call Back• Delayed Start/Standby Setting• Concurrent delivery•Titration• Micro 0.1-99.9 mL/hr (in 0.1 mL increments) flow rate range for both lines• Macro 100-999 mL/hr (in 1 mL increments) flow rate range for both lines• Automated Secondary drug delivery (Piggyback)
2- 2 430-95520-001 (Rev. 04/05) PLUMSET CAPABILITIES• Anti Free-Flow Protection• Direct Connection for syringe deliveryAIR MANAGEMENT•Air Trap• Air Removal/Backpriming• Air Detection-Proximal• Air Detection-DistalBATTERY• Battery Gauge• Less than 8-Hour Battery Recharge Time• Long battery life (3 hours) for emergency backup and temporary portable operationBIOMEDICAL• Upgradability (Field)• Variable Rate Cap• Alarm History• Nurse Call Relay ConnectorOPTIONS• Volumes Infused (A, B, Total Volume)• KVO at dose end (1 ml/hr or less depending on delivery rate) or Continue Rate (CR) to continue at the current rate• Variable Distal Pressure SettingsOTHER FEATURES• Nonpulsatile volumetric accuracy• Microprocessor control• Large liquid crystal display (LCD) screen
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 2- 3• Panel back illumination on mains power• Lockout switch• Standard syringe use• Parenteral and nonparenteral (enteral) fluid delivery• Blood and blood products delivery• Wide range of Standard and Specialty administration sets• Password protected keypad lockCOMMUNICATION• Networked Communication (Ethernet & Wireless)• Autoprogramming• Drug Library DownloadDefinitionsTERM DEFINITIONAdministration SetThe sterile, disposable assembly with flexible tubing that connects to a source fluid container for input to the Cassette pumping chamber and to some output device for administration to the patient.Alarm A condition requiring operator intervention that invokes both audible and visible alarm indicators. Confirmation of the condition usually is required.Alert While entering a program, a visual warning that informs the clinician of the violation of a hospital-defined rule set, e.g., a dose limit for a specific drug.Alternate Units Dose Rate units that may be selected. Alternate Units are any units other than mL/hr.Alternate Units ParametersDrug Amount, Diluent, Patient Weight (if applicable), and Dose Rate.
2- 4 430-95520-001 (Rev. 04/05) Auto-Program A physician order sent to the infuser.Auto-Program (new)An auto-program is considered “new” if either of the following conditions are met: A) There is no program on the associated line. B) The drug and concentration in the auto-program is different from a currently-programmed drug and concentration.Auto-Program (titration)An auto-program is considered a “titration” if the drug and concentration in the auto-program match the currently-programmed drug and concentration. Non-promotion titration programs include some subset of the following: Rate, Dose, Duration. Promotion titration programs include some subset of the following: Rate, Dose, Duration, Patient Weight, Dosing Units, Drug Amount, Drug Units, Diluent Amount.Awake CE The CE is considered awake when the infuser is powered on with battery power and the battery is above the low battery threshold or is plugged into AC.Backpressure Resistance to fluid flow on the Distal or output portion of the Administration Set, usually expressed as PSIG.Backprime Removal of air or fluid from the proximal tubing and Cassette air trap by rapid pumping of fluid from Line A to an output receptacle on Line B. No fluid delivered distal to the cassette during a Backprime.Basic Delivery A single-step delivery specifying rate, VTBI, and duration only. Drug name may or may not be specified.Biomed Mode The name for the non-delivery or service mode of Infuser operation. TERM DEFINITION
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 2- 5Bolus A single uninterrupted discrete volume of fluid delivered over a discrete period of time.Cassette A Hospira standard Plum Macro Cassette that contains the actual pumping chamber with inlet and outlet valves. It also contains access ports to diaphragms for proximal and distal line pressure sensing and chambers for air trapping and air quantity measurement.Caution Contains information which could prevent hardware failure, irreversible damage to equipment, or loss of data.Channel Distal Line output to the patient.Cleared Program Same as Cleared Settings, but restricted to the program delivery settings for an individual line.Cleared Settings Programmed delivery settings for both lines and the general control parameters are at their default values.Clinical Care Area (CCA)An area of the hospital for which authorized hospital staff has allowed use of up to 149 specific drugs, plus No Drug Selected. The clinician selects a CCA after turning on the Infuser. The hospital may create from one to eighteen CCA’s.Concentration For Dose Calculation based Line A or B delivery programming of premixed drugs, the concentration consists of Drug Amount (in micrograms [mcg], milligrams [mg], grams, milliequivalents [mEq], millimols [mmols], or units [USP Units]) and Diluent (Volume in milliliters [mL]). Abbreviated on screen displays as “Conc”.TERM DEFINITION
2- 6 430-95520-001 (Rev. 04/05) Concurrent DeliverySimultaneous delivery of two fluids at independent flow rates.Concurrent Mode The Line B control selection mode that allows either Line A or Line B (or both) of the Mechanism to be programmed for delivery that completes with KVO or CR. Line A operation is unaffected by Line B delivery.Connectivity Engine (CE)A component of the infuser that controls ethernet and wireless communication between the MMU and Infuser processor.Continue Rate (CR)An option used to provide fluid delivery at the current fluid delivery rate when the VTBI becomes zero.Default Drug Library (DDL)A drug list of 99 drug names, default units and default concentrations, plus ‘No Drug Selected’. Does not contain Soft or Hard Limits. The DDL is a subset of Plum A+ Version 11.3 drug list. The Infuser functions initially with the DDL.Deliver Together A Biomed Delivery option setting, which selects default of Piggyback or Concurrent delivery mode.Delivery Mode The normal operational mode for the Infuser, either Concurrent or Piggyback, during which fluid delivery may occur.Delivery Rate The delivery rate (as Volume/Time) assigned by the operator or calculated from other delivery parameters.Device The Plum A+ IV Infusion Device (Infuser), not including the disposable administration sets.Diluent (Volume) The volume of fluid used with the Diluent Unit (e.g., 250 mL), which dilutes the drug to be infused.TERM DEFINITION
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 2- 7Distal Downstream (output) from the Cassette portion of the Administration Set.Door The lever-operated, bottom-hinged, Cassette holder on the front of the Mechanism.Dose (Rate) Delivery rate times drug concentration.Dose Calculation Allows programming Dose Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep.Drug Amount The mass or quantity of the drug premixed with a Diluent to express the Concentration for the Line being programmed.Drug Library Drug names available for use with the Infuser in the selected CCA, and each drug’s concentration, units of measure, and dose limits, as determined by authorized hospital staff. The Drug Library is based on hospital-defined best practices. The Drug Library is customized and downloaded into the Infuser by authorized hospital personnel using Hospira MedNet Networked software.Duration The time period used to deliver a VTBI at an existing Delivery Rate or as set by the operator.Enteral Delivery via an intestinal route.Error A condition resulting in an audible warning and often an accompanying message, such as an invalid key press, resulting from an operator attempt to make an input that is not valid for the current operational context. Not a Malfunction.TERM DEFINITION
2- 8 430-95520-001 (Rev. 04/05) Filling Head Height (FHH)The gravity induced Proximal Line pressure due to fluid height in source container above the Distal Line output level.Flow Rate The resulting rate of fluid flow. (See Rate and Dose.)Fluid Delivery BeginThe time required, measured from the initial start of a keypad press, for a digit in the third decimal place of a scale measuring grams to change when pumping water, the delivery tube is fluid filled, and the end of the delivery tube is immersed.Hard Limit The dosing amount (upper and/or lower amount) that when assigned to a drug cannot be overridden. Defined by the hospital for each drug in its Drug Library. The hard limits for a particular drug may vary across different CCAs.Infuser A Plum A+ device, not including the disposable administration set.Initialized PositionThe plunger is in the Plunger Park Position, and each valve is in the Valve Home Position.Key Any of the marked locations on the front panels intended for user input via a pressing action.KVO (Keep Vein Open) Internally selected minimal delivery rate, the lesser of 1.0 mL/Hr or actual Rate, intended to provide sufficient fluid flow to decrease the potential for clotting at the IV infusion site.Line A Inlet tubing on the top center of Cassette when viewed after it has been loaded in Mechanism Door.TERM DEFINITION
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 2- 9Line B Inlet connector on the top right of Cassette when viewed after it has been loaded in Mechanism Door.Loading Dose Allows programming of an initial infusion rate for a specific volume, followed automatically by a maintenance rate from the same container (e.g., a fluid challenge). If Dose Calc is used, the Loading Dose and Maintenance Dose are in the same unit of measure, over the same period of time (mcg/min), from the same container. Macro Cassette The Plum Cassette as part of an Administration Set whose attainable delivery performance is for general purpose uses.Malfunction One of a number of conditions that indicate a failure of the Infuser. These conditions usually require a repair action.Mechanism The provider of the physical activities of: 1) valve control for Line selection and inlet to or outlet from the Cassette; 2) Plunger extension and retraction for changing fluid volume of the Cassette pumping chamber.MedNet software Networked application software used to define and download the custom drug library to the Infuser.TERM DEFINITION
2- 10 430-95520-001 (Rev. 04/05) Multistep Allows a sequential program to deliver up to 10 steps; fluid volumes and delivery rates may be programmed for each step. The program may be entered based on Rate and Volume or Volume and Time. If Dose Calc is used, the delivery steps are in the same unit of measure, over the same period of time, from the same container.MMS Medication Management Suite- a product suite that will be used to support safe and efficient caregiver administration of intravenous medications.MMU Medication Management Unit- the MMU is a product of the MMS and provides external communications with infusers  and other systems supporting medication management. It supports drug library  downloads and status/log updates to the infusers. MMU also delivers  a variety of pre-built and custom reports related to the infuser activities.Network-Enabled ModuleThe hardware that facilitates communication between the infuser and the network.Note A Note explains something new, gives tips on performing an activity, or provides a reminder of necessary preparations.Override The clinician acknowledges an Alert, confirms and proceeds with a program containing a parameter that falls outside the hospital-defined Soft Limits.Parenteral Delivery via other than an intestinal route, such as intravenous (IV) injection.TERM DEFINITION
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 2- 11Patient Weight The Alternate Units Parameter whose value entry is required with Dose rates (quantity/time) having the third parameter of weight, such as mcg/kg/min.Piggyback DeliveryInfusion will stop Line A and infuse Line B until VTBI completes.Piggyback Mode The Line B control selection mode that requires Line A delivery suspension while Line B delivers and then its resumption when Line B VTBI completes. If Line A is not programmed and started, Line B delivery just stops when its VTBI completes. Line A operation is delayed for duration of its interruption by Line B.Plunger Park PositionThe position of the pumping plunger that allows the cassette to be withdrawn from the Infuser without restriction, with an opened Door.POST Power On Self Tests. Startup initialization and test of critical hardware modules and software functions, to determine Infuser readiness for its Intended Use.Power On When the power is first applied to the Infuser processor. It occurs when Infuser is in the off condition and the AC power is applied or during battery operation when the ON/OFF key is pressed.Pressure The measured fluid pressure for Proximal Line A, Proximal Line B, or the Distal line when a fully primed administration set is loaded into a Mechanism with the Cassette Door closed.TERM DEFINITION
2- 12 430-95520-001 (Rev. 04/05) Programs A Concentration-based extension to basic Volume/Rate delivery for Dose Rate applications. Softkey-selected types are: 1) Standard Program for single rate delivery; 2) Loading Dose for up to two steps of delivery at differing rates; 3) Multistep for up to ten steps of delivery at differing rates. Loading Dose and Multistep include the basic Delivery Rate option.Protocol In the medical field, hospital-approved procedures. For devices with an electronic communications interface, the format and content of messages being exchanged.Proximal Upstream (input, as Line A and/or B) with respect to the Cassette portion of the Administration Set.Rate See Delivery RateRounding To round-off to the closest whole number (e.g., 25.5  would round to 26, but 25.4 would round to 25).Rule Sets The compilation of Soft and Hard Limits associated with a drug entry.Screen That active portion of the LCD assigned to a delivery or other user interactive display operation.Softkey A front panel key whose function label is on the bottom portion of the display screen just above its location, so that label may be changed to indicate differing key functions within the operational context of the defined screens.TERM DEFINITION
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 2- 13Soft Limit The dosing amount (upper and/or lower amount) that when assigned to a drug can be overridden based on confirmation. Defined by the hospital for each drug in its Drug Library. The soft limits for a particular drug entry may vary across CCAs.Standard Programming(see Basic Delivery)Standby A delivery program state that resembles a delayed start for a line in all respects, but status display of STANDBY instead of DELAYED. A press of the START key (and a Start Softkey if necessary) will begin the programmed delivery.System The Plum A+ Infuser system with its associated accessories and Administration Sets.Tall-Man LetteringTall-Man lettering uses upper-case letters in combination with lower-case letters to help clinicians differentiate among sound-alike or look-alike drug names. Certain Hospira Default Library drug names are displayed in Tall-Man lettering.Truncate To round-down to the last whole number (e.g., 103.7 would be 103).Units A Drug Amount (without a quantity of mass referent) adopted as the United States Pharmacopoeia (USP) standard of measure for use in a Concentration.VI Volume Infused.VTBI Volume To Be Infused – The volume of fluid or IV solution (remaining) for delivery by a program or Therapy step from a Line.TERM DEFINITION
2- 14 430-95520-001 (Rev. 04/05) Warning Audible indication at the defined Warning sound level, often accompanied by a brief message, to advise operator that: 1. Input was attempt to use a function in wrong sequence, with wrong values, or at wrong time (resulting in an Invalid Key Warning); 2. Input was properly related to a function (resulting in a Valid Key Warning). (The Low Battery Warning is an exception, using the Alarm Indication sound level.)Weight The Alternate Units Parameter whose value entry is required with Alternate Units for Dose that include /kg/.TERM DEFINITION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 3- 1SECTION 3Equipment DescriptionFront Panel Display and KeypadBackPrimeOptions/Vol InfABLine A Flow IndicatorStatus RegionWorking RegionMessage RegionStatus RegionWorking RegionMessage RegionStart and Stop KeysCharging/Line IndicatorON/OFF KeyLine B FlowIndicatorLCDBatteryChargeIndicatorNumericKeys(Up/Left)SelectionKey(Down/Right)SilenceKeyNumeric Field Clear Key Decimal PointKeySoftkey LabelSoftkey Selector
Draft Manual- Not to be usedin a Clinical Situation.3- 2 430-95520-001 (Rev. 04/05) Operating KeysThe [ON/OFF] key is used to control the power of the Plum A+. NOTE: All deliveries must be stopped in order to turn the power off.NOTE: To confirm proper shutdown of infuser, do not unplug until “double beep” is heard (approximately 5 seconds). If “double beep” is not heard, the infuser will turn itself back ON sensing the AC power was lost and revert to battery back-up power.The [START] key is used to begin infusion. It brings up the Confirmation screen. If all informtion appears correct, select V [Yes] to start program.The [STOP] key is used to stop the current delivery on one line; used in conjunction with soft keys to stop delivery on both lines.
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 3- 3The [SILENCE] key is used to silence an audible alarm during actions to correct its cause.The [SELECT] key (or  ) is used to move the highlighting cursor between the programming fields.The numeric keys are used to enter values for any highlighted field requiring numeric data.The [CLEAR] key will cause a highlighted numeric field to be cleared in preparation for data entry. It is also used to clear the displayed dashes (-- -- --) that occur when entering values beyond the pre-programmed hard limits. It WILL NOT clear the entire program.The [DECIMAL POINT] key is used to manually enter numbers other than whole numbers (i.e., 1.2 mL).
Draft Manual- Not to be usedin a Clinical Situation.3- 4 430-95520-001 (Rev. 04/05) Softkeys are located at the bottom of the main display and serve a variety of functions. What each key does is indicated by the text in the display above the actual softkey.IndicatorsThe Charge/Line Indicator is illuminated to indicate the battery is currently charging.NOTE: If the device is plugged into AC power, with a battery installed, and the Charge Indicator is not illuminated, contact technical support.The fluid drop symbol, when flashing, indicates an active infusion on Line A and/or Line B.Cancel/BackStandby Enter
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 3- 5Display SymbolsBattery Capacity Symbol (located on right side of message region on display) indicates relative charge with number of white boxes from left to right.Cautionary Symbol is displayed to alert the user that the infuser is being operated without rule sets (soft or hard limits). Read the drug package insert for information on proper usage.Soft Limit Override Symbol indicates the programmed soft limits upper (up arrow) or lower (down arrow) have been overridden.NOTE: The infuser will display the rule set icon from the first step in a situation where a 2-step process is completed and the infuser is stopped after going to KVO.The "Wireless Connection Available" icon is displayed when the device is receiving a wireless signal. The infuser will connect to the network if a wireless network access point is recognized. !
Draft Manual- Not to be usedin a Clinical Situation.3- 6 430-95520-001 (Rev. 04/05) It may appear on any Main Delivery Screen whether pumping or stopped. It also may appear regardless of Auto-Programming. However, in a wireless network environment, Auto-Programming will not be successful if the icon is not visible during the Auto-Program transfer. If wireless Auto-Programming is desired and the icon is not displayed on the Main Delivery Screen, try relocating the pump towards the wireless access point until the icon is displayed. The icon will not be displayed if the infuser is communicating via an ethernet connection.Rear Case ControlsThe audio level rotary knob, located on the back, adjusts the alarm audio level.The toggle switch, located below the audio level control, activates the lockout function when the lever is placed in the “up” position, disabling all front panel keys except [STOP].Password Protected Keypad Lock may also be used. It both locks and unlocks the keypad by pressing the decimal key, followed by 9, 6, & 3.AABBSTOPPED PUMPINGRatemL/hrVol InfmLOptions/Vol Inf26.310.130Dose 2mg/minmcg/kg/minMedSurgLidocaineDose CalculationDOPamineDose Calculation
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 3- 7NOTE: Pressing of a key while lockout is active results in the display of the lockout enabled message. This action is recommended to confirm each lockout activation.NOTE: If the lockout switch is put in the “Locked” position while the infuser is off, the [ON/OFF] key allows the unit to turn on.The connector located below the lockout switch is the Nurse Call Relay Connector, for providing remote notification of a nurse call event.The Ethernet port is used to connect to a local area connection.Administration SetsThe Plum A+ is compatible with a wide range of LifeCare Plum series administration sets. Become familiar with the set components before preparing the administration set.Primary LineFlow RegulatorPumping ChamberSecondaryInlet Port(Capped)Finger GripPatient LineAir Trap
Draft Manual- Not to be usedin a Clinical Situation.3- 8 430-95520-001 (Rev. 04/05) Preparing the Administration SetNOTE: For detailed instructions, see administration set packaging.WARNINGARRANGE TUBING, CORDS, AND CABLES TO MINIMIZETHE RISK OF PATIENT STRANGULATION ORENTANGLEMENT.Use the aseptic technique to prepare the administration set for priming, then proceed as follows:1Make sure flow regulator is closed.2Insert piercing pin into container outlet with a twisting motion.3Fill Drip chamber to about 1/2 full or to the score mark. Do not completely fill.NOTE: To avoid getting Distal Air Alarms, confirm the fluid bag is positioned higher than the infuser. This should be set up prior to priming the cassette.
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 3- 9Priming the Administration Set1Invert the cassette.2Turn the flow regulator counter-clockwise until a drop of fluid is seen in the pumping chamber.3Turn the cassette upright, then prime the remainder of the administration set.4Push in the flow regulator to close. Confirm there is no flow.Loading the cassetteTo load the primed cassette into the plum A+, proceed as follows.1Open the cassette door by lifting up the handle.
Draft Manual- Not to be usedin a Clinical Situation.3- 10 430-95520-001 (Rev. 04/05) 2Insert cassette into door guides. Close the cassette door. Confirm there is no flow and no kinks appear in tubing.NOTE: If flow is observed, close the tubing clamps and replace the administration set.NOTE: The administration set should be changed per CDC guidelines or healthcare provider policy. Discard after use.Preparing the Secondary LineCAUTION: CONSULT THE DRUG LABELING TO CONFIRM DRUGCOMPATIBILITY, CONCENTRATION, DELIVERY RATES, AND VOLUMESARE ALL SUITABLE FOR CONCURRENT DELIVERY OR PIGGYBACKDELIVERY (SECONDARY FOLLOWED AUTOMATICALLY BY PRIMARY)MODES.The Plum A+ features concurrent or piggyback delivery modes when program requires administering more than one fluid. In addition to standard containers, the Plum A+ can use syringes on the secondary port. The secondary line can be prepared without removing or repriming the cassette.Before preparing the secondary line, observe the following guidelines:• Review the backpriming function.• Attach the secondary container using an 18- or 19-gauge, 31 mm (or shorter) needle, if the cassette secondary inlet port has a piercing reseal.• Syringes must be larger than 3 cc. Use syringe adapter (List 11986-48) when using syringes smaller than 10cc. Some 10cc syringes may require use of a syringe adapter. Syringes larger than 10cc may be attached directly to the secondary port of the
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 3- 11cassette. Use of a syringe adapter may decrease the occurrence of proximal occlusion alarms.• When using a syringe adapter, retract the plunger to draw approximately 1 mL of fluid into the syringe to clear air from the adapter filter.To prepare the secondary line, use aseptic technique and proceed as follows:1With cassette door closed: Loosen and remove white cap, then discard. Add secondary tubing or syringe.2Open all proximal clamps before pressing [START].3Set up per operating instructions. (See Sections 4 - 6 of this manual.)4Syringe: Secure the container to the cassette door using the optional container support arm (List 12095-03).Discontinuing Electronic Flow Control & Setting Gravity FlowCAUTION: BEFORE OPENING DOOR, CLOSE CLAMP ON PRIMARYOR SECONDARY SET OR REMOVE SECONDARY CONTAINER FROMSECONDARY PORT TO PREVENT MIXING OF PRIMARY AND SECONDARYFLUIDS.If electronic flow needs to be stopped, the precision flow regulator on the cassette can be used to manually set and
Draft Manual- Not to be usedin a Clinical Situation.3- 12 430-95520-001 (Rev. 04/05) maintain fluid flow in a gravity mode (see the following graphics)..Discontinuing Fluid AdministrationPRESS STOPOPEN DOOR & REMOVE CASSETTESET GRAVITY RATE  BYTURNING FLOW REGULATORCOUNTER-CLOCKWISENOTE: FLOW REGULATOR CLOSESAUTOMATICALLY WHEN DOOR IS OPENEDTO PREVENT ACCIDENTAL FREE FLOWNOTE: CASSETTE SHOULD BEIN UPRIGHT POSITIONNOTE: IF DOOR OPEN ALARM OCCURS, TURN PUMP OFFSTOP123PRESS STOPCLOSE DOORDISCARD SET AND FLUIDCONTAINER PER HOSPITALPROCEDUREOPEN DOOR, REMOVE CASSETTESTOP4321
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 1SECTION 4Basic OperationGetting StartedUNPACKINGCAUTION: PRODUCT DAMAGE MAY OCCUR UNLESS PROPERCARE IS EXERCISED DURING UNPACKING AND INSTALLATION. DO NOTUSE THE PLUM A+ IF IT APPEARS DAMAGED IN ANY WAY. THEBATTERY MAY NOT BE CHARGED UPON RECEIPT.Inspect the Plum A+ packaging for possible shipping damage. If damage is found, contact the delivery company immediately.Use care when unpacking the infuser. Retain the packing slip and save all packing material in case the infuser is damaged or fails the infuser self-test and has to be returned to Hospira.Inspect thoroughly for damage. CAUTION: IF THE PLUM A+ APPEARS TO BE DAMAGED, CONTACTHOSPIRA, INC.To set up, connect the AC (mains) cord to a properly grounded receptacle, unless temporary battery operation is desired.NOTE: Use AC (mains) power whenever possible. Store the infuser connected to AC (mains) to confirm a fully charged battery for emergencies.NOTE: If the quality of the earth grounding source is in doubt, use battery power.CAUTION: THE PLUM A+ SYSTEM IS DESIGNED TO OPERATENORMALLY IN THE PRESENCE OF MOST ENCOUNTEREDELECTROMAGNETIC INTERFERENCE (EMI) CONDITIONS. IN THEEVENT OF EXTREME LEVELS OF INTERFERENCE,  SUCH ASENCOUNTERED NEXT TO AN ELECTROSURGICAL GENERATOR,CELLULAR TELEPHONES,  OR 2-WAY RADIOS, IT IS POSSIBLE THATTHE NORMAL OPERATION OF A SENSOR OR MICROCOMPUTER MIGHTsec04.fm  Page 1  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 2 430-95520-001 (Rev. 04/05) BE DISRUPTED. OPERATION OF THE INFUSER UNDER THESECONDITIONS SHOULD BE AVOIDED.TANDEM CARRIER INSTRUCTIONSThe Plum A+ may be safely and conveniently mounted on an IV stand or on a tandem carrier mounted on an IV stand.Always mount the tandem carrier (List# 12270-01) to the pole before attaching the infuser(s) to the tandem carrier.Make sure tandem carrier clamp is attached firmly to pole.Make sure infusers are firmly attached to tandem carrier.This section details the Plum A+ instrument setup procedures.SYSTEM SELF-TESTSThe infuser performs a suite of System Self-Tests when power is applied to confirm readiness for use.Failure during the Self-Tests is reported as a Malfunction Condition.CAUTION: DO NOT PLACE THE PLUM A+ IN SERVICE IF IT FAILSTHE SYSTEM SELF-TESTS.Connect the AC power (mains) cord to an AC power receptacle, then confirm that the Charging/Line indicator illuminates (above the [ON/OFF] key on the front panel).A systematic self-testing of the processing, delivery, and safety systems is performed whenever the Plum A+ is turned on, to verify readiness for operation.NOTE: If the quality of the earth grounding source is in doubt, use battery power.sec04.fm  Page 2  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 3Press the [ON/OFF] key to turn the power on. Check screen display and listen for a beep to indicate the audio is working. Wait for the self-tests to complete. If successful, put a fully primed macro cassette into the cassette door and close the door. See Loading Cassette in Section 3 for details of correct procedure.When the Charging/Line indicator is off and the battery symbol on the display is flashing, this indicates that the Plum A+ is operating on low battery power and should be recharged.CAUTION: WHEN THE BATTERY IS REMOVED FROM  THE PLUMA+,  DO NOT OPERATE ON PATIENTS. USE OF A PROPERLYMAINTAINED AND CHARGED BATTERY HELPS CONFIRM PROPEROPERATION. To confirm battery is fully charged, connect the Plum A+ to AC (mains) power for a minimum of six hours while in the OFF mode. If an alarm occurs during the power on self-test, identify the alarm message, then take corrective action (see Section 7, Alarms and Troubleshooting). Turn infuser OFF, then back ON. If the self-test alarm recurs, remove the infuser from service and contact the local Hospira representative.The infuser performs a suite of System Self-Tests when power is applied to confirm readiness for use.Failure during the Self-Tests is reported as a Malfunction Condition.DATA RETENTIONDelivery program settings and programming option selections are retained in non-volatile memory. If the Plum A+ has been turned off for longer than four hours, all delivery settings are cleared and programming option selections are restored to their default selections for next use.sec04.fm  Page 3  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 4 430-95520-001 (Rev. 04/05) Power On1Press   to Power On the Plum A+ and the System Self-Tests Screen is displayed.After self-testing successfully completes, the display will proceed to the next screen.Programming the infuser may begin during the cassette check phase of the start-up testing.NOTE: If battery is missing or defective, a “Battery not installed” message with a “Continue without battery” prompt will appear (AC power only).CAUTION: WHEN THE BATTERY IS REMOVED FROM  THE PLUMA+,  DO NOT OPERATE ON PATIENTS. USE OF A PROPERLYMAINTAINED AND CHARGED BATTERY HELPS CONFIRM PROPEROPERATION. 2To set up the Plum A+, insert selected Plum Set and close cassette holder door lever. NOTE: If cassette is already inserted, this screen is skipped.If Infuser ID Entry screen appears, consult your hospital’s biomed personnel.HOSPIRA Plum A+13.X   Jan 31 2005 12:30:30Copyright Hospira, Inc.2005System Self TestIn Progress**SCREENS SHOWN IN THIS MANUAL ARE EXAMPLES ONLY AND DON’T NECESSARILY REFLECT ACTUAL CLINICAL PRACTICE**SETUPInsert PLUM SetClose Leversec04.fm  Page 4  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 5The Area Selection screen will appear if your system has been programmed with the Hospira MedNet accessory. Use   to choose. If only one CCA has been entered, it becomes the default and this screen is skipped.NOTE: To change the CCA while infusing, see Section 6.8, Changing CCA While Infusing.If not using MedNet, the display will go straight to the “Clear Settings ?” option.Example of Clear Settings Screen3The on-screen message “Clear Settings?” may appear. Select  [Yes] to clear all settings.Clear ALL settings when changing patients or multiple programming parameters.NOTE: This clears all delivery parameters, such as Rate and VTBI on both lines, as well as setting items, such as the KVO and Nurse Callback options, to their default values.NOTE: If no settings currently exist or all settings are zero, this screen will be bypassed.Area SelectionEnterSelect, then EnterAny Hospital, USAMM/DD/YY XXYYZZ- 123456ICUMedSurgORCardioUrologyPsychRecovL&D5 WestRenalNeo ICUOB/GYNOncolAd ICUER7 South3 North4 NorthSETUPA BClear Settings?Yes clears ALL settingsYes Nosec04.fm  Page 5  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 6 430-95520-001 (Rev. 04/05) Simple DeliveryExample of MedNet Main Delivery Screen (with cleared settings). The selected area is displayed at the bottom.NOTE: For clarity, whole number font sizes will appear larger than fractional number font sizes (numbers on the right-side of the decimal).1To Program a delivery rate, select  [A].2To enter delivery rate, use numeric keypad.3Use   to move down to the VTBI field.4To enter VTBI amount, use keypad.NOTE: While the VTBI is being entered, the duration is automatically computed. If the computed value is greater than 99:59, dashes (- -) appear in place of hours.AABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol InfBackPrime0000MedSurg**SCREENS SHOWN IN THIS MANUAL ARE EXAMPLES ONLY AND DON’T NECESSARILY REFLECT CURRENT ACTUAL PRACTICE**APROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackDrug ListEnter Value using keypad0000:00MedSurgAPROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackDrug ListEnter value using keypad5.5000:00MedSurgsec04.fm  Page 6  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 75Use   to move down to the DURATION field.If auto-calculated Duration is acceptable, skip to step 9.6Use keypad to change/enter the hours part of duration if desired.NOTE: When all parameters are displayed, changing any one parameter will result in other parameters automatically being computed. To understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.7Use   to move from hours to minutes.NOTE: The maximum hours allowed is 99 and the maximum minutes is 59.8Use keypad to enter minutes.9Press .CAUTION: VERIFY ALL VALUESBEFORE STARTING INFUSION.Program Confirmation screen will appear. If correct, press  [Yes] to begin infusion.APROGRAMRateVTBIDurationProgramOptionsCancel/BackDrug ListEnter value using keypad5.50.500:05mL/hrmLhr : minMedSurgAPROGRAMDrugRateVTBIDelivery will be:5.50.5mL/hrmLConfirm Program?No Drug Selected!Yes N oMedSurgsec04.fm  Page 7  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 8 430-95520-001 (Rev. 04/05) Example of Main Delivery Screen just after being programmed and started.NOTE: The Line A Fluid Flow Indicator also will flash to indicate pumping.NOTE: VTBI (volume to be infused) values over 100 mL are displayed truncated. This means that the number displayed rounds-down to the last whole number (e.g., 105.7 would be displayed as 105). An example would be the Programming or Pumping screens.NOTE: VI (volume infused) values over 100 mL are displayed rounded. This means that the number displayed rounds-off to the nearest number (e.g., 105.7 would be displayed as 106, but 105.3 would be displayed as 105). An example would be the Program Options screen.Titration1Select Line  [A] (or  [B]) at the Main Delivery Screen. AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.50.100MedSurg!AABBPUMPING PUMPINGRatemL/hrVol InfmLOptions/Vol Inf2541 2050!!MedSurgsec04.fm  Page 8  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 92Enter the desired Rate using the keypad. NOTE: When Rate is changed, the duration is automatically recomputed using the current VTBI. To understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.3Press  .CAUTION: VERIFY ALL VALUESBEFORE STARTING INFUSION.Program Confirmation screen will appear. If correct, press  [Yes] to begin infusion.NOTE: During Titration, infuser continues to deliver at previous parameters until new program is confirmed.Example of Main Delivery Screen after being titrated.NOTE: Volume Infused is an incrementing value which is cleared in  [Options/Vol Inf]. The following 3 screens are examples of changed values.APROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackDrug ListEnter Value using keypad155003:20MedSurgAPROGRAMDrugRateVTBIDelivery will be:1550mL/hrmLConfirm Program?No Drug Selected!Yes N oMedSurgAABBPUMPING PUMPINGRatemL/hrVol InfmLOptions/Vol Inf1541.62050!!MedSurgsec04.fm  Page 9  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 10 430-95520-001 (Rev. 04/05) Example of changed rate.4From the Main Delivery Screen, select  [A].5Use keypad to change the RATE value.NOTE: The display time is only accurate to ± one minute. As shown in this example, the delivery shows 3 hours and 20 minutes (not 3 minutes and 20 seconds).6Press  .Program Confirmation screen will appear. If correct, press  [Yes] to begin infusion.NOTE: [START] accepts the change. Delivery is changed to the new RATE and DURATION and display returns to Main Delivery Screen.Example of Changed VTBI value.7Use   to highlight the VTBI field.8Use the keypad to change the VTBI value.NOTE: When the VTBI is changed, the Duration is recomputed with respect to the current Rate. To understand the Automatic APROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackDrug ListEnter Value using keypad155003:20MedSurgAPROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackEnter Value using keypad1510006:40MedSurgDrug Listsec04.fm  Page 10  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 11Calculation feature, see Section 6.7 Examples of Automatic Calculation.9Press  .Program Confirmation screen will appear. If correct, press  [Yes] to begin infusion.NOTE: [START] accepts the change. Delivery is changed to the new VTBI and DURATION and the display brings up the Confirmation screen before returning to the Main Delivery Screen.Example of changed DURATION field.10 Use   to highlight the Duration field, then use the keypad to change its value.NOTE: When DURATION is changed, RATE is automatically recomputed using the current VTBI value. To understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.11 Press  .CAUTION: VERIFY ALL VALUES BEFORE STARTING INFUSION.Program Confirmation screen will appear. If correct, press  [Yes] to begin infusion.APROGRAMRateVTBIDurationProgramOptionsCancel/BackEnter Value using keypad15010000:40mL/hrmLhr : minMedSurgDrug Listsec04.fm  Page 11  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 12 430-95520-001 (Rev. 04/05) NOTE: [START] accepts the change. Delivery is changed to the new DURATION and RATE and the display brings up the Confirmation screen before returning to the Main Delivery Screen.Piggyback DeliveryPIGGYBACK DELIVERY- Infusion will stop Line A and infuse Line B until VTBI completes. Line A will automatically restart.1With Line A pumping, to program a Piggyback delivery, select  [B].The Delivery Mode field is highlighted. NOTE: If an entry was previously programmed, RATE is highlighted. If no previous entries, MODE is highlighted.2If field reads “Concurrent”, select  [Change Mode] to switch to “Piggyback”.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.520 05!!MedSurgBPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsChangeModeCancel/BackChange using Change ModePiggyback0000:00MedSurgDrug Listsec04.fm  Page 12  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 133Select   to highlight the RATE field.4To enter delivery rate, use numeric keypad.5Use   to move down to the VTBI field.6To enter VTBI amount, use keypad.NOTE: While the VTBI is being entered, the duration is automatically computed. If the computed value is greater than 99:59, dashes (- -) appear in place of hours. To understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.BPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackEnter value using keypadPiggyback200000:00MedSurgDrug ListBPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackEnter value using keypadPiggyback20010000:30MedSurgDrug Listsec04.fm  Page 13  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 14 430-95520-001 (Rev. 04/05) 7Use   to move down to the DURATION field.(Optional) Use keypad to change duration.When all parameters are displayed, changing any one parameter will result in other parameters automatically being computed.8Press  .CAUTION: VERIFY ALL VALUES BEFORE STARTING INFUSION.Program Confirmation screen will appear. If correct, press  [Yes] to begin infusion.NOTE: If Line A has a Delay Start remaining or is stopped, a message will appear warning of a possible non-delivery (see section 6.4).Example of Main Delivery Screen just after being programmed and started.BPROGRAMModeRateVTBIDurationProgramOptionsCancel/BackEnter value using keypadPiggyback13310000:45mL/hrmLhr : minMedSurgDrug ListAABBDELAYED PUMPINGRatemL/hrVol InfmLOptions/Vol Inf5.520 0.1133!!MedSurgsec04.fm  Page 14  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 15Concurrent DeliveryCONCURRENT DELIVERY- Simultaneous delivery of two fluids at independent flow rates.CAUTION: CONSULT DRUGLABELING TO CONFIRM DRUGCOMPATIBILITY,  CONCENTRATION,DELIVERY RATES, AND VOLUME ARE ALLSUITABLE FOR CONCURRENT ANDPIGGYBACK DELIVERY MODES.  SEESECTION 8 FOR INFORMATION ONCONCURRENT FLOW,  SEE SECTION 9FOR DELIVERY RATE RANGES.NOTE: In concurrent mode with certain alarm conditions, one line may stop while the other line may continue to deliver.1With pumping started on line A, to program a Concurrent delivery rate, select  [B].The Delivery Mode field is highlighted. NOTE: If an entry was previously programmed, RATE is highlighted. If no previous entries, MODE is highlighted.If field reads “Piggyback”, select  [Change Mode] to switch to “Concurrent”.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.50.100!MedSurgBPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsChangeModeCancel/BackChange using Change ModePiggyback0000:00MedSurgDrug Listsec04.fm  Page 15  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 16 430-95520-001 (Rev. 04/05) Verify change to “Concurrent” mode.2Use   to highlight Rate field.3To enter delivery rate, use numeric keypad.NOTE: When Rate for Line A plus Rate for Line B is greater than valid rate ranges, a “Concurrency violation” message will explain why second line did not start. See Section 9.3 for valid delivery rate ranges.NOTE: “Concurrency violation” may not be displayed if unit is displaying “VTBI Complete”. The device will not start and an invalid keypress sound will occur when [START] is pressed.4Use   to move down to the VTBI field.5To enter VTBI amount, use keypad.NOTE: While the VTBI is being entered, the duration is automatically computed. If the computed value is greater than 99:59, dashes (- -) appear in place of hours. To BPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsChangeModeCancel/BackChange using Change ModeConcurrent0000:00MedSurgDrug ListBPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackEnter value using keypadConcurrent10000:00MedSurgDrug ListBPROGRAMModeRateVTBIDurationProgramOptionsCancel/BackEnter value using keypadConcurrent1010010:00mL/hrmLhr : minMedSurgDrug Listsec04.fm  Page 16  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 17understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.6Use   to move down to the DURATION field.(Optional) Enter values for desired hours and minutes.NOTE: When Duration is entered, Rate is automatically computed.7Press  .CAUTION: VERIFY ALL VALUES BEFORE STARTING INFUSION.Program Confirmation screen will appear. If correct, press  [Yes] to begin infusion.Example of a concurrent delivery.AABBPUMPING PUMPINGRatemL/hrVol InfmLOptions/Vol Inf5.520.4110!!MedSurgsec04.fm  Page 17  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 18 430-95520-001 (Rev. 04/05) Stop and Start with only 1 Line Pumping1Press   at the Main Delivery Screen.NOTE: If  [A] (or  [B]) is selected prior to pressing [STOP], the display will return to the programming screen. Select  [Cancel/Back] to return to the Main Delivery screen.Example of STOPPED delivery.2To discontinue infusion, press .3Clamp secondary tubing if connected.4Open door, remove set and close door. Or, press  to resume infusion.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol InfBackPrime1008.500!MedSurgAABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol InfBackPrime1008.500!MedSurgsec04.fm  Page 18  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 19Stop and Start with Both Lines pumping1Press   at the Main Delivery Screen.2Select  [Stop A],  [Stop B], or  [Stop All].NOTE: If  [A] (or  [B]) is selected prior to [STOP], the display will show the programming screen. Select  [Cancel/Back] to return to the Main Delivery screen.Example of stopped infusion.NOTE: This is an example using  [Stop All].CAUTION: VERIFY ALL VALUESBEFORE STARTING INFUSION.AABBPUMPING PUMPINGRatemL/hrVol InfmLOptions/Vol Inf253 20.450!!MedSurgABPUMPING PUMPINGRatemL/hrVol InfmLCancel253 20.450StopAllStopAStopBChoose line(s) to stop:!!MedSurgAABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol Inf253 20.450BackPrime!!MedSurgsec04.fm  Page 19  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 20 430-95520-001 (Rev. 04/05) 3To resume infusion, press .4Select  [Start A],  [Start B], or  [Start All]. Or, to discontinue infusion, press .5Clamp secondary tubing if connected.6Open door, remove set and close door.BackprimingBACKPRIMING- Removal of air or fluid from the proximal line and cassette airtrap.1To initiate a backprime, select  [Back Prime].NOTE: Backpriming is available only when all delivery is stopped. Secondary container is necessary to receive the air.ABSTOPPED STOPPEDRatemL/hrVol InfmLCancel253 20.450StartAllStartAStartBChoose line(s) to start:!!MedSurgAABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.58.9 0 0BackPrime!MedSurgsec04.fm  Page 20  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 212Select and hold  [Back Prime] until fluid pumped from line A to line B has expelled the air from cassette air trap into Secondary container.NOTE: Fluid flow is at maximum rate possible.3Upon release of  [Back Prime], the cassette test is performed to confirm readiness for pumping. Pressing   will resume programmed delivery.Clearing Settings1Press   at the Main Delivery Screen.Clear ALL settings when changing patients or multiple programming parameters.AABBBACKPRIMINGRatemL/hrVol InfmLOptions/Vol Inf5.58.9 0 0BackPrimeRelease Backprime to Stop!MedSurgAABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.50.100!MedSurgsec04.fm  Page 21  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 22 430-95520-001 (Rev. 04/05) 2Select Line  [A] (or B if active) from the stopped Main Delivery Screen.NOTE: If Line A or B is programmed with a Therapy (i.e., Dose Calculation), Rate, VTBI and Duration will not display values. The values to be cleared may be viewed in applicable programming screen under  [Drug List].3Select  [Yes].Clear ALL settings when changing patients or multiple programming parameters.Example screen of “Cleared Settings”.4Select  [Cancel/Back] to return to the Main Delivery screen or program new settings.AABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.58.9 0 0BackPrime!MedSurgPROGRAMAClear Line A Settings?Yes NoAPROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackEnter Value using keypad0000:00MedSurgDrug Listsec04.fm  Page 22  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 4- 23Updating Drug LibraryIf a new Drug Library is available, the New Drug Library Install screen will be displayed.1Select  [Yes] to begin the installation (It may take up to an hour). Select  [No] to skip the installation and continue using the current library.Example screen of Library Installing.Library InstallYes N oNew Drug LibraryAvailableThe Installation maytake up to an hour.Ensure pump isconnected to AC power.Install Now?Library InstallInstallingNew Drug LibraryThe Installation maytake up to an hour.Ensure pump isconnected to AC power.sec04.fm  Page 23  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.4- 24 430-95520-001 (Rev. 04/05) NOTESsec04.fm  Page 24  Monday, April 11, 2005  8:34 AM
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 1SECTION 5Advanced ProgramsNOTE: During programming, user must identify desired drug with softkey Drug List prior to setting program parameters.Use drug manufacturer recommendations for IV administration when using the Plum A+ infuser.StandardProgram(Dose Calculation) Allows programming Dose Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep.Loading DoseAllows programming of an initial infusion rate for a specific volume, followed automatically by a maintenance rate from the same container (e.g., a fluid challenge).If Dose Calc is used, the Loading Dose and Maintenance Dose are in the same unit of measure, over the same period of time(mcg/min), from the same container. MultistepAllows a sequential program to deliver up to 10 steps; fluid volumes and delivery rates may be programmed for each step. The program may be entered based on Rate and Volume or Volume and Time. If Dose Calc is used, the delivery steps are in the same unit of measure, over the same period of time, from the same container.
Draft Manual- Not to be usedin a Clinical Situation.5- 2 430-95520-001 (Rev. 04/05) Dose Calc (mcg/kg/min on A)DOSE CALCULATION- Allows programming Dose Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep.Clear ALL settings when changing patients or multiple programming parameters.The Area Selection screen will appear if your system has been programmed with the MedNet accessory. Use   to select CCA and bring you to the Main Delivery Screen.NOTE: To change the CCA while infusing, see Section 6.8, Changing CCA While Infusing.1Select  [A] at the Main Delivery Screen.Area SelectionEnterSelect, then EnterAny Hospital, USAMM/DD/YY XXYYZZ- 123456ICUMedSurgORCardioUrologyPsychRecovL&D5 WestRenalNeo ICUOB/GYNOncolAd ICUER7 South3 North4 NorthAABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol InfBackPrime0000MedSurg**SCREENS SHOWN IN THIS MANUAL ARE EXAMPLES ONLY AND DON’T NECESSARILY REFLECT CURRENT ACTUAL PRACTICE**
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 3The Line A Programming screen appears.2Select  [Drug List] for Dose Calculation programming.The Drug List Screen appears, in mainly alphabetical order.NOTE: Default entry is “No Drug Selected”.3Use   to highlight desired drug name.When list does not contain desired generic drug name or group name select  [Page Down] (or  [Page Up]) until it shows, then use  to highlight choice.4Select  [Enter] to indicate selection and advance to the Therapy selection screen.Based on this selection, assume the hospital has set the following rules for the Med/Surg Area:APROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackDrug ListEnter Value using keypad0000:00MedSurgAMedSurgPageUpPageDownCancel/BackEnterSelect, then EnterNo Drug SelectedAlbumin 5%AldesleukinAlfentanilAlteplaseAlteplaseAmikacinAmiodarone25 grams/500 mL2.6 mg/50 mL10 mg/250 mL50 mg/50 mL100 mg/100 mL100 mL150 mg/100 mLAMedSurgPageUpPageDownCancel/BackEnterSelect, then EnterAntibioticDOBUTamineDOBUTamineDocetaxelDOPamineDOPamineDOPamineHeparin250 mg/500 mL500 mg/500 mL100 mg/250 mL___ ___/___ mL400 mg/250 mL800 mg/250 mL25,000 units/500 mL
Draft Manual- Not to be usedin a Clinical Situation.5- 4 430-95520-001 (Rev. 04/05) Soft upper limit of 8 mcg/kg/minHard upper limit of 12 mcg/kg/minNOTE: The SOFT and HARD limits are not displayed. However, the limit is provided if a value is programmed outside the acceptable range. Typically, the pharmacy will set the limit.The Select Therapy screen for Line A appears.5Select  [Choose] to select Standard Program (Dose Calc) and continue.If selected drug has been programmed with established dosing units and concentration, skip to step 8.The Program Dose Calc screen appears.NOTE: Cursor defaults to package insert (or hospital) recommended dose units associated with the selected drug.6Select  [Choose] to continue. Or, use   to change the units to use in the delivery. If desired, select  [Page Down] to see more units of measure.APROGRAMChoose BackSelect, then ChooseStandard ProgramLoading DoseMultistepDOPamineAProgram Dose CalcChoose BackSelect, then ChoosemL/hrmcg/kg/minmcg/kg/hrmcg/minmcg/hrmg/kg/hrmg/minmg/hrgrams/hrng/kg/minmmol/minmmol/hrmEq/hrDOPaminePageUpPageDown
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 5The Container Concentration Selection screen appears whenever the selected Dose Units are gram-based (e.g., grams, mg, mcg, or ng), otherwise this screen is skipped.NOTE: Cursor defaults to package insert (or hospital) recommended concentration units associated with the selected drug.7Select  [Choose] to continue. Or, use   to change the Container Concentration units to use in the delivery.The next page of the Program Dose Calc screen appears to provide for entry of delivery program parameters.The Concentration and Volume fields are filled in per drug selection.8Use keypad to enter desired values. Once each value is entered, use   to highlight the next programming field.NOTE: The Concentration elements and weight value must be entered before the [SELECT] key allows access to remaining fields.AProgram Dose CalcChoose BackSelect, then ChooseDrug Conc in ContainermcgmggramsDOPamineAProgram Dose CalcProgramOptionsCancel/BackEnter value using keypadConcWeight40070mgkg250 mLDoseVTBIDurationRate0000: 000mcg/kg/minmLhr : minmL/hrDOPamine
Draft Manual- Not to be usedin a Clinical Situation.5- 6 430-95520-001 (Rev. 04/05) 9Continue to enter values using the keypad, and advancing to the next field using  .NOTE: As the Dose value is entered, the Rate is automatically calculated. To understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.10 Use   to scroll down to the VTBI field.11 Enter a value for VTBI using the keypad.NOTE: As the VTBI value is entered or changes, Duration is automatically updated.NOTE:  At this point, programming is complete unless [Program Options] are desired. See Section 6, Additional Features, for more information.12 Press .CAUTION: VERIFY ALL VALUES BEFORE STARTING INFUSION.AProgram Dose CalcProgramOptionsCancel/BackEnter value using keypadConcWeight40070mgkg250 mLDoseVTBIDurationRate10000: 0026.3mcg/kg/minmLhr : minmL/hrDOPamineAProgram Dose CalcProgramOptionsCancel/BackEnter value using keypadConcWeight40070mgkg250 mLDoseVTBIDurationRate1025009: 3026.3mcg/kg/minmLhr : minmL/hrDOPamine
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 7Because the Dose falls outside the soft limits, the infuser beeps three times (Invalid Key Warning), and generates the error message:“Dose > 8. Override?”13 Select  [Yes].NOTE: A SOFT limit override will be recorded in the History Events Log.This will bring up the PROGRAM CONFIRMATION screen.14 If Dose, Drug, Conc., Weight & Rate appear correct, select  [Yes] to begin program.NOTE: If the selected drug has been pre-programmed with dosing units and concentration, the concentration drug amount or diluent volume cannot be changed whether delivering or stopped. The only way to change to a different concentration is to clear the program.However, if selected drug has not been pre-programmed (or still operating with the DDL), the concentration can be changed for a program.AProgram Dose CalcYes NoDose > 8. Override?ConcWeight40070mgkg250 mLDoseVTBIDurationRate1025009: 3026.3mcg/kg/minmLhr : minmL/hrDOPamineAProgram Dose CalcYes N oDoseDrugConcWeightRateDelivery will be:104007026.3mcg/kg/minmgkgmL/hrConfirm Program?DOPamine250 mLMedSurg
Draft Manual- Not to be usedin a Clinical Situation.5- 8 430-95520-001 (Rev. 04/05) The Main Delivery screen appears displaying the information just entered.NOTE: The Program Type and Drug are displayed as well as the Dose Rate and Dose Units.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf26.300010mcg/kg/minDOPamineMedSurgDoseDose Calculation
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 9Dose Calc (mg/min on B)DOSE CALCULATION- Allows programming Dose Rates in alternative units of measure. Dose Calculation can be used in Simple Delivery, Loading Dose, and Multistep.Clear ALL settings when changing patients or multiple programming parameters.1Select  [B] at the Main Delivery Screen.The Line B Programming screen appears.2Select  [Drug List].AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf26.300010mcg/kg/minDOPamineMedSurgDoseDose CalculationBPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsChangeModeCancel/BackChange using Change ModePiggyback0000:00MedSurgDrug List
Draft Manual- Not to be usedin a Clinical Situation.5- 10 430-95520-001 (Rev. 04/05) The Drug List Screen appears, in mainly alphabetical order.NOTE: Default entry is “No Drug Selected”.3Use   to highlight desired drug name.When list does not contain desired generic drug name or group name select  [Page Down] (or  [Page Up]) until it shows, then use   to highlight choice.4Select  [Enter] to indicate selection and advance to the Program Selection screen.The Select Program screen for Line B appears.5Select  [Choose] to continue.BMedSurgPageUpPageDownCancel/BackEnterSelect, then EnterNo Drug SelectedAlbumin 5%AldesleukinAlfentanilAlteplaseAlteplaseAmikacinAmiodarone25 grams/500 mL2.6 mg/50 mL10 mg/250 mL50 mg/50 mL100 mg/100 mL100 mL150 mg/100 mLBMedSurgPageUpPageDownCancel/BackEnterSelect, then EnterLabetalolLepirudinLidocaineLidocaineLidocaineLiposyn 10%Liposyn 20%Milrinone200 mg/200 mL100 mg/250 mL1 grams/250 mL2 grams/500 mL4 grams/500 mL100 mL500 mL200 mcg/100 mLBPROGRAMChoose BackSelect, then ChooseStandard ProgramLoading DoseMultistepLidocaine
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 11If selected drug has been programmed with established dosing units and concentration, skip to step 8.The Program Dose Calc screen appears.NOTE: Cursor defaults to package insert (or hospital) recommended dose units associated with the selected drug.6Select  [Choose] to continue. Or, use   to change the units to use in the delivery.If desired, select  [Page Down] to see more units of measure.The Container Concentration Selection screen appears whenever the selected Dose Units are gram-based (e.g., grams, mg, mcg, or ng) otherwise this screen is skipped.NOTE: Cursor defaults to package insert recommended concentration units associated with the selected drug.7Select  [Choose] to continue. Or, use   to change the Container Concentration units to use in the delivery.BProgram Dose CalcChoose BackSelect, then ChooseLidocainemL/hrmcg/kg/minmcg/kg/hrmcg/minmcg/hrmg/kg/hrmg/minmg/hrgrams/hrng/kg/minmmol/minmmol/hrmEq/hrPageUpPageDownBProgram Dose CalcChoose BackSelect, then ChooseDrug Conc in ContainermcgmggramsLidocaine
Draft Manual- Not to be usedin a Clinical Situation.5- 12 430-95520-001 (Rev. 04/05) The next page of the Program Dose Calc screen appears to provide for entry of delivery program parameters.8Use keypad to enter desired values. Once entered, use   to highlight the next field.NOTE: If selected drug has established Rule Sets (e.g., non-blank concentration), the Mode field will be highlighted upon entry to this screen.9To change the Piggyback Delivery Mode default, scroll down to Mode using , then select  [Change Mode].NOTE: Delivery Mode Option is available on Line B, its default is Piggyback, and the Weight field is gone.10 Continue entering desired values using the keypad. Use  to advance to the next field.NOTE: When entering the RATE, with a VTBI value entered, the DOSE and DURATION fields are automatically updated. To understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.BProgram Dose CalcProgramOptionsCancel/BackEnter value using keypadConc2500gramsmLDoseVTBIDurationRate0000: 000mg/minmLhr : minmL/hrMode PiggybackLidocaineBProgram Dose CalcProgramOptionsCancel/BackEnter value using keypadConc 2500gramsmLDoseVTBIDurationRate250016 : 4030mg/minmLhr : minmL/hrMode ConcurrentLidocaine
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 13NOTE: The Duration field temporarily changes to dashes when the computed value is larger than what can be displayed.11 Continue entering the RATE value, if desired.NOTE: As Rate is entered, the Duration and Dose are automatically adjusted as individual numbers are entered.NOTE: If the Duration value is changed, the Rate and Dose are automatically adjusted.NOTE: At this point, programming is complete unless [Program Options] are desired. See Section 6, Additional Features, for more information.CAUTION: VERIFY ALL VALUES BEFORE STARTING INFUSION.12 To start the delivery, press .This will bring up the PROGRAM CONFIRMATION screen.13 If Dose, Drug, Conc., Weight, & Rate appear correct, select  [Yes] to begin program.NOTE: If the selected drug has been pre-programmed with dosing units and concentration, the concentration drug amount or diluent volume cannot be changed whether delivering or stopped. The only way to change to a different concentration is to clear the program.However, if selected drug has not been pre-programmed (or still operating with the DDL), the concentration can be changed for a program.BProgram Dose CalcYes N oDoseDrugConcWeightRateDelivery will be:227030mg/mingramskgmL/hrConfirm Program?Lidocaine500 mLMedSurg
Draft Manual- Not to be usedin a Clinical Situation.5- 14 430-95520-001 (Rev. 04/05) Upon  , the Main Delivery Screen shows delivery status for both A and B lines.NOTE: See Delivery Rate Range in Section 9. The ratio of concurrent rates has recommendations specified in Section 1.4.AABBSTOPPED PUMPINGRatemL/hrVol InfmLOptions/Vol Inf26.310.130Dose 2mg/minmcg/kg/minMedSurgLidocaineDose CalculationDOPamineDose Calculation
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 15Loading DoseLOADING DOSE- Allows programming of an initial infusion rate for a specific volume, followed automatically by a maintenance rate from the same container (e.g., a fluid challenge). If Dose Calc is used, the Loading Dose and Maintenance Dose are in the same unit of measure, over the same period of time (mcg/min), from the same container.Clear ALL settings when changing patients or multiple programming parameters.The Area Selection screen will appear if your system has been programmed with the MedNet accessory. Use   to select CCA and bring you to the Main Delivery Screen.NOTE: To change CCA while infusing, see Section 6.8, Changing CCA While Infusing.1Select  [A] (or  [B]) at the Main Delivery screen.Area SelectionEnterSelect, then EnterAny Hospital, USAMM/DD/YY XXYYZZ- 123456ICUMedSurgORCardioUrologyPsychRecovL&D5 WestRenalNeo ICUOB/GYNOncolAd ICUER7 South3 North4 NorthAABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol InfBackPrime0000MedSurg
Draft Manual- Not to be usedin a Clinical Situation.5- 16 430-95520-001 (Rev. 04/05) The selected Line (A or B) Programming screen appears.2Select  [Drug List]. The Drug List Screen appears, in mainly alphabetical order.NOTE: Default entry is “No Drug Selected”. This process will guide you through a “fluid challenge”, so “No Drug Selected” will be entered.3Select  [Enter] to indicate selection and advance to the Program selection screen.The Select Program screen for Line A (or B) appears.NOTE: This example shows line A being programmed.APROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackDrug ListEnter Value using keypad0000:00MedSurgAMedSurgPageUpPageDownCancel/BackEnterSelect, then EnterNo Drug SelectedAlbumin 5%AldesleukinAlfentanilAlteplaseAlteplaseAmikacinAmiodarone25 grams/500 mL2.6 mg/50 mL10 mg/250 mL50 mg/50 mL100 mg/100 mL100 mL150 mg/100 mLAPROGRAMChoose BackSelect, then ChooseStandard ProgramLoading DoseMultistep
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 174Use   to scroll down to Loading Dose.5Select  [Choose] to continue.The next page of the Program Loading Dose screen appears.6Select  [Choose] to continue. Or, use   to change the units to use in the delivery.If desired, select  [Page Down] to see more units of measure.The Loading Dose Programming screen appears.7The default field is Rate. Since a “Fluid Challenge” is being performed in this example, press   to advance to the VTBI field.APROGRAMChoose BackSelect, then ChooseStandard ProgramLoading DoseMultistepAProgram Loading DoseChoose BackSelect, then ChoosemL/hrmcg/kg/minmcg/kg/hrmcg/minmcg/hrmg/kg/hrmg/minmg/hrgrams/hrng/kg/minmmol/minmmol/hrmEq/hrPageUpPageDownAProgram Loading DoseCancel/BackProgramOptionsEnter value using keypadDurationhr : min00 : 0000 : 00VTBImL00mL/hr00Rate12
Draft Manual- Not to be usedin a Clinical Situation.5- 18 430-95520-001 (Rev. 04/05) 8Enter value using keypad.9Use   to highlight the next field.10 Enter Duration using keypad. The Rate value is automatically calculated. To understand the Automatic Calculation feature, see Section 6.7 Examples of Automatic Calculation.11 After the values for loading dose Step-1 are entered, use  to highlight the maintenance dose, Step-2.NOTE: The Loading Dose program has two steps. The first is the loading dose followed by the maintenance dose. These words will appear on the main screen as each step is delivered.AProgram Loading DoseCancel/BackProgramOptionsEnter value using keypadDurationhr : min00 : 0000 : 00mL/hr00Rate12VTBImL2500AProgram Loading DoseCancel/BackProgramOptionsEnter value using keypadmL/hr5000Rate12VTBImL2500Durationhr : min00 : 3000 : 00AProgram Loading DoseCancel/BackProgramOptionsEnter value using keypadDurationhr : min00 : 3000 : 00mL/hr50075Rate12VTBImL2500
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 1912 Enter the set of desired values for Step-2 of the program.13 Select  [Program Options] to select a Delayed Start or Standby.NOTE:  At this point, programming is complete unless  [Program Options] are desired. See Section 6, Additional Features, for more information.NOTE: Nurse Callback enables an alarm when a change in the delivery on a line occurs such as completing Step-1 and starting Step-2. Or on line B, a piggyback delivery is completed and line A is starting its delivery.14 Enter a delay time (less than 24 hours) using the keypad.NOTE: Not available if a delivery is taking place on the line.NOTE: Nurse Callback cannot be changed while a delivery is taking place on the line.15 Select  [Enter] to accept the changes and return to the Program Loading Dose screen.AProgram Loading DoseCancel/BackProgramOptionsEnter value using keypadDurationhr : min00 : 3010 : 00mL/hr50075Rate12VTBImL250750AProgram OptionsCancel/BackEnterStandbyEnter Value using keypadDelay Start for 00:00 hr:minCallback    NoLoading DoseAProgram OptionsCancel/BackEnterYes/NoChange using Yes/NoDelay Start for 00:05 hr:minCallback    YesLoading Dose
Draft Manual- Not to be usedin a Clinical Situation.5- 20 430-95520-001 (Rev. 04/05) NOTE:  [Cancel/Back] returns the fields to their original value and displays the previous screen.16 To start the delivery, press  .This will bring up the PROGRAM CONFIRMATION screen.CAUTION: VERIFY ALL VALUESBEFORE STARTING INFUSION.17 If Drug and Rate appear correct, select  [Yes] to begin program.The Main Delivery Screen appears.NOTE: The words “DELAYED” and “Loading Dose” appear in the upper-left corner. These indicate the options which have been selected.After the programmed delay period has expired, the Plum A+ starts “Pumping”. This is indicated where “Delayed” previously was displayed.AProgram Loading DoseDrugRateVTBIDelivery will be:500250mL/hrmLConfirm Program?No Drug Selected!Yes N oMedSurgAABBDELAYED STOPPEDRatemL/hrVol InfmLOptions/Vol Inf500000Loading DoseMedSurg!AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf500000Loading DoseMedSurg
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 21After the Step-1 has completed, the Main Delivery Screen changes to the second step.NOTE: A flashing “Callback Line A” message and the audio alarm sound indicates activation of that option.18 The user should acknowledge the Callback message and stop the audio alarm by pressing . Upon completion of delivery, the screen shows a flashing “Line A VTBI complete” message and the audible alarm sounds.NOTE: The KVO rate will change to the rate of 1.0 mL/hr or less depending on delivery rate and “Pumping” mode changes to “KVO”. If Continue Rate option was selected, pumping continues at original rate.19 Press   to stop alarm sound.NOTE: Warning message continues to flash and audio alarm will return after a two minute period of silence, unless delivery is stopped or VTBI of last line is changed.AABBSTOPPEDOptions/Vol InfN105 Callback Line APUMPINGRatemL/hrVol InfmL75250 00Maintenance DoseMedSurgAABBSTOPPEDOptions/Vol InfN161 Line A VTBI completeKVORatemL/hrVol InfmL11000 00Maintenance DoseMedSurg
Draft Manual- Not to be usedin a Clinical Situation.5- 22 430-95520-001 (Rev. 04/05) Multistep ProgrammingMULTISTEP- Allows a sequential program to deliver up to 10 steps; fluid volumes and delivery rates may be programmed for each step. The program may be entered based on Rate and Volume or Volume and Time. If Dose Calc is used, the delivery steps are in the same unit of measure, over the same period of time, from the same container.Clear ALL settings when changing patients or multiple programming parameters.The Area Selection screen will appear if your system has been programmed with the MedNet accessory. Use   to select CCA and bring you to the Main Delivery Screen.NOTE: To change CCA while infusing, see Section 6.8, Changing CCA While Infusing.1Select  [A] at the Main Delivery screen.Area SelectionEnterSelect, then EnterAny Hospital, USAMM/DD/YY XXYYZZ- 123456ICUMedSurgORCardioUrologyPsychRecovL&D5 WestRenalNeo ICUOB/GYNOncolAd ICUER7 South3 North4 NorthAABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol InfBackPrime0000MedSurg
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 23The Line A Programming screen appears.2Select  [Drug List].The Drug List Screen appears, in mainly alphabetical order.NOTE: Default entry is “No Drug Selected”.3Use   to highlight desired drug name.When list does not contain desired drug name or drug class name, select  [Page Down] (or  [Page Up]) until it shows, then use  to highlight choice.4Select  [Enter] to indicate selection and advance to the Program Selection screen.APROGRAMRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackDrug ListEnter Value using keypad0000:00MedSurgAMedSurgPageUpPageDownCancel/BackEnterSelect, then EnterNo Drug SelectedAlbumin 5%AldesleukinAlfentanilAlteplaseAlteplaseAmikacinAmiodarone25 grams/500 mL2.6 mg/50 mL10 mg/250 mL50 mg/50 mL100 mg/100 mL100 mL150 mg/100 mLAMedSurgPageUpPageDownCancel/BackEnterSelect, then EnterTirofibanTobramycinTobramycinVancomycinVinCRIStineVinCRIStine5 mg/100 mL40 mg/50 mL80 mg/100 mL500 mg/100mL100 mL50 mL
Draft Manual- Not to be usedin a Clinical Situation.5- 24 430-95520-001 (Rev. 04/05) The Select Program screen for Line A (or B) appears.NOTE: This example shows line A being programmed.5Use   to scroll down to Multistep.6Select  [Choose] to continue.If selected drug has been programmed with established dosing units and concentration, skip to step 9.The Program Multistep screen appears.NOTE: Cursor defaults to package insert recommended dose units associated with the selected drug.7Use   key to highlight the units to use in the delivery.8Select  [Choose] to continue.If desired, select  [Page Down] to see more units of APROGRAMChoose BackSelect, then ChooseStandard ProgramLoading DoseMultistepVinCRIStineAPROGRAMChoose BackSelect, then ChooseStandard ProgramLoading DoseMultistepVinCRIStineAProgram MultistepChoose BackSelect, then ChooseVinCRIStinemL/hrmcg/kg/minmcg/kg/hrmcg/minmcg/hrmg/kg/hrmg/minmg/hrgrams/hrng/kg/minmmol/minmmol/hrmEq/hrPageUpPageDown
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 25measure.The Multistep Programming screen appears.9Rate is the default field. Enter value using keypad.10 Use   to highlight the Duration or VTBI field.NOTE: The VTBI value is automatically computed when the Duration value is entered. If VTBI is entered, the Duration value is automatically computed.11 Use   to drop down to Step-2, Rate field.12 Enter the set of desired values for Step-2 of the program. Continue process for each additional Step.AProgram MultistepCancel/BackEnter value using keypadProgramOptionsMoreStepsRatemL/hrDurationhr : min00 :  0000 :  0000 :  0000 :  00VTBImL0000500001234VinCRIStineAProgram MultistepCancel/BackEnter value using keypadProgramOptionsMoreStepsRatemL/hrDurationhr : min01 : 0000 :  0000 :  0000 :  00VTBImL50000500001234VinCRIStineAProgram MultistepCancel/BackEnter value using keypadProgramOptionsMoreStepsRatemL/hrDurationhr : min01 :  0000 :  0000 :  0000 :  00VTBImL5000050100001234VinCRIStine
Draft Manual- Not to be usedin a Clinical Situation.5- 26 430-95520-001 (Rev. 04/05) NOTE: When Step-4 is programmed, using [SELECT] to advance will highlight another item on this screen.13 When Step-4 is programmed, select  [More Steps] to program up to 10 Steps.NOTE: Using [SELECT] will not advance you to the additional steps.14 To set up Steps 5 through 10, enter desired values as performed for Steps 1 through 4. To see the previous steps, select  [Prev Steps].15 To start the delivery, press .This will bring up the Confirmation screen.CAUTION: VERIFY ALL VALUESBEFORE STARTING INFUSION.AProgram MultistepCancel/BackEnter value using keypadProgramOptionsMoreStepsRatemL/hrDurationhr : min01 :  0008 :  1501 :  0001 : 00VTBImL5082575505010075501234VinCRIStineAProgram MultistepCancel/BackEnter value using keypadProgramOptionsPrevStepsRatemL/hrDurationhr : min00 :  0000 :  0000 :  0000 :  0000 :  0000 :  00VTBImL0000000000005678910VinCRIStineAProgram MultistepDrugRateVTBIDelivery will be:5050mL/hrmLConfirm Program?VinCRIStine!Yes N oMedSurg
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 5- 27The Main Delivery Screen displays Step-1 in the upper-left corner.NOTE: In units other than mL/hr, the Dose rate and units also will be displayed.Example of Multistep Screen Pumping on Step-2.16 To change a program parameter while pumping, select  [A].NOTE: The “*” in a Step number field indicates the step is delivering or “d” indicates each completed step.17 Enter a value using the keypad.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5027.500 VinCRIStineStep 1AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf100652 00 VinCRIStineStep 2AProgram MultistepCancel/BackEnter value using keypadProgramOptionsMoreStepsRatemL/hrDurationhr : min01 :  0008 :  1501 :  0001 :  00VTBImL508257550501007550d34VinCRIStine
Draft Manual- Not to be usedin a Clinical Situation.5- 28 430-95520-001 (Rev. 04/05) If Rate value is changed, the Duration will automatically change.NOTE: Only the current delivering step and future steps can be changed.18 To start the delivery, press .This will bring up the Confirmation screen.CAUTION: VERIFY ALL VALUESBEFORE STARTING INFUSION.NOTE: The Values are now updated on the Main Delivery Screen.The infuser reverts to KVO when last Step completes.The last step can be restarted by entering a new VTBI value.AProgram MultistepCancel/BackEnter value using keypadProgramOptionsMoreStepsRatemL/hrDurationhr : min01 :  0006 :  3601 :  0001 :  00VTBImL508257550501257550d34VinCRIStineAProgram MultistepDrugRateVTBIDelivery will be:125825mL/hrmLConfirm Program?VinCRIStine!Yes N oMedSurgAABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf125658 00 VinCRIStineStep 2
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 1SECTION 6Additional FeaturesSimple Delivery Using Delayed Start1Before or after entering the simple delivery program, select  [Program Options].NOTE: To program a Delayed Start using an advanced program (i.e., Dose Calc) you must first select which line you want (A or B), next select   [Drug List], and then  [Program Options]. If you program the Delayed Start first, it will be cleared upon entry into Select Program mode.The Default screen for Program Options appears.BPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsChangeModeCancel/BackDrugListChange using Change ModeConcurrent525050:00MedSurgBProgram OptionsCancel/BackEnterStandbyEnter Value using keypadDelay Start for 00:00 hr:minCallback    No
Draft Manual- Not to be usedin a Clinical Situation.6- 2 430-95520-001 (Rev. 04/05) 2Enter values for desired hours and/or minutes.3Select  [Enter] to confirm delayed start.CAUTION: VERIFY ALL VALUESBEFORE STARTING INFUSION.4Press  to confirm programming.Line B is delayed for the time entered.NOTE: Delayed line drip indicator is solid, not flashing. It will begin flashing when actually pumping.Piggyback with Nurse Callback1After programming in desired values for both lines, select  [Program Options].NOTE: To set a nurse callback, the infuser must be stopped. This function can be selected either before or after entering the Piggyback delivery program.BProgram OptionsCancel/BackEnterStandbyEnter Value using keypadDelay Start for 01:00 hr:minCallback    NoAABBPUMPING DELAYEDRatemL/hrVol InfmLOptions/Vol Inf125850 05!!MedSurgBPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackEnter value using keypadPiggyback20010000:30MedSurgDrug List
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 3The default screen for Program Options appears.Use   to highlight “Callback” field and select  [Yes/No], if necessary to obtain “Yes”, then select   [Enter].NOTE: Callback default of Yes or No is a biomedical setting and must be configured using the Hospira MedNet accessory.CAUTION: VERIFY ALL VALUES BEFORE STARTING INFUSION.2Press  to confirm program setting and return to Main Delivery screen.At end of Piggyback delivery, with Callback enabled for the line, Callback alarm is issued.3Press   to clear alarm.BProgram OptionsCancel/BackEnterYes/NoChange using Yes/NoDelay Start for 00:00 hr:minCallback         YesBPROGRAMModeRateVTBIDurationmL/hrmLhr : minProgramOptionsCancel/BackEnter value using keypadPiggyback20010000:30MedSurgDrug ListAABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.50.3100200N104 Callback Line BMedSurg!!
Draft Manual- Not to be usedin a Clinical Situation.6- 4 430-95520-001 (Rev. 04/05) Possible Non-Delivery ProgrammedNOTE: Whenever a valid [START] key is pressed, the device checks the delivery program(s) for non-delivery conditions (such as Line B Piggyback without Line A or delayed start) that could permit a period of non-delivery. If so, the display presents this warning screen.1Select  [Yes] to continue with delivery. Or, select  [No] to cancel.Using the Standby FeatureStandby is a feature that enables the clinician to program the infuser up to 23: 59 minutes in advance of Starting.1If Line A is infusing, select  [A] or  [B] to access program screen of line to be placed on Standby.2Select  [Program Options].3Select  [Standby].BPROGRAMYes N oAttention: Line A is stopped ordelayed. Line B in piggyback modegives no KVO or alarm at dose end.Continue anyway?AProgram Dose CalcProgramOptionsCancel/BackEnter value using keypadConcWeight40070mgkg250 mLDoseVTBIDurationRate1025009: 3026.3mcg/kg/minmLhr : minmL/hrDOPamine
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 54Confirm the information displayed is correct, then select  [Yes] to place program on Standby mode.NOTE: This example screen shows a Dose Calculation Program with a Standby. It looks similar to the Program Confirmation screen with the substitution of “Put delivery in Standby” replacing “Confirm Program”. Selecting  [Yes] confirms both the Program and the Standby.5Press  to remove Standby and start the infusion.AProgram Dose CalcYes N oDoseDrugConcWeightRateDelivery will be:104007026.3mcg/kg/minmgkgmL/hrPut delivery in Standby?DOPamine250 mLMedSurg!AABBSTANDBY STOPPEDRatemL/hrVol InfmLOptions/Vol Inf26.300010mcg/kg/min MedSurgDOPamineDose Calculation
Draft Manual- Not to be usedin a Clinical Situation.6- 6 430-95520-001 (Rev. 04/05) Select Option- Volumes Infused, Pressure/Post Infusion Rate, and Lighting/Contrast1To select Options Screen, select  [Options/Vol Inf].2Use   to highlight an option.3Select  [Choose] to select the Volumes Infused screen.This will bring up the Volumes Infused screen which defaults to Total Volume. 4Use   to scroll to Line A or B.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf5.5393 0 0MedSurgDOPamineDose CalculationABOPTIONSChooseChange CCA BackSelect, then ChooseRelease 13.x - xx/xx/xxVolumes InfusedPressure/Post Infusion RateLighting/ContrastAny Hospital, USAMM/DD/YY XXYYZZ-123456ABOPTIONSEnter Cancel/BackPress CLEAR to clear valueVolumes InfusedLine ALine BTotal VolmLmLmL00393DOPamineDose Calculation
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 75To reset Line A (or B) Volume Infused,use  to highlight and press  .6To restore the cleared quantity, select  [Cancel/Back].7To reset all volumes, use   to highlight Total Vol, then press .8To restore all quantities, select   [Cancel/Back] softkey.ABOPTIONSEnter Cancel/BackPress CLEAR to clear valueVolumes InfusedLine ALine BTotal VolmLmLmL00393DOPamineDose CalculationABOPTIONSEnter Cancel/BackPress CLEAR to clear valueVolumes InfusedLine ALine BTotal VolmLmLmL000DOPamineDose Calculation
Draft Manual- Not to be usedin a Clinical Situation.6- 8 430-95520-001 (Rev. 04/05) 9When clearing of quantities is completed, select  [Enter] to return to the Main Delivery screen.10 Select  [Options/Vol Inf] from the Main Delivery screen to return to the Options screen.11 Use   to highlight the Pressure/Post Infusion Rate option.12 Select  [Choose] to continue.NOTE: Prior to changing Pressure/Post Infusion Rate, infuser must be stopped.13 Use   to highlight Continue Rate or Distal Pressure Limit field.14 Continue Rate allows a choice when VTBI is completed. Select  [KVO/Rate].• KVO- The infuser will revert to a fixed KVO of 1 mL/hr or the last programmed Rate, whichever is less.• Rate- The infuser will continue at the Rate programmed.NOTE: Option will be for both line A and line B selections.15 To change Distal Pressure Limit when highlighted, ABOPTIONSChooseChange CCA BackSelect, then ChooseRelease 13.x - xx/xx/xxVolumes InfusedPressure/Post Infusion RateLighting/ContrastAny Hospital, USAMM/DD/YY XXYYZZ-123456ABOPTIONSEnter Cancel/BackChange using KVO/RatePressure/Post Infusion RateContinue RateDistal Pressure LimitDistal PressureKVO60psipsiKVO/RateDOPamineDose Calculation
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 9enter value between 1.0 and 15.0 psi.16 Select  [Enter] to keep changes and return to the Main Delivery screen, or  [Cancel/Back] to restore original values and return to the Options screen.NOTE: Continue Rate default of KVO or Rate must be configured using the Hospira MedNet software.17 Select  [Options/Vol Inf] from the Main Delivery screen to return to the Options screen.18 Use   to highlight Lighting/Contrast Option.19 Select  [Choose] to continue.20 Press   to highlight Backlight Intensity or Display Contrast.21 While viewing display for desired effect, select either  [Increase Setting] or  [Decrease Setting] to change level.22 Select  [Enter] to keep change(s) and return to the Main Delivery screen, or  [Cancel/Back] to restore previous settings and return to the Options screen.ABOPTIONSChooseChange CCA BackSelect, then ChooseRelease 13.x - xx/xx/xxVolumes InfusedPressure/Post Infusion RateLighting/ContrastAny Hospital, USAMM/DD/YY XXYYZZ-123456ABOPTIONSEnter Cancel/BackUse Inc/Decrease SettingsLighting/ContrastBacklight IntensityDisplay ContrastDecreaseSettingIncreaseSetting
Draft Manual- Not to be usedin a Clinical Situation.6- 10 430-95520-001 (Rev. 04/05) Variable Rate CapNOTE: Variable Rate Cap is used to set the maximum rate values to be accepted by the device. The clinician cannot change it once it has been programmed. It must be configured using the Hospira MedNet software.Examples of Automatic CalculationAT STARTUPInitial programming allows the clinician to enter two of the three programming parameters (Rate, VTBI, or Duration) and the third is automatically calculated. (Refer to table below)WHILE RUNNING (TITRATION)Changing two parameters after startup of the infusion will allow for the re-calculation of the third parameter.1st Action 2nd Action [AUTOCALC]enter RATE enter VTBI [DURATION]enter VTBI enter DURATION [RATE]enter RATE enter DURATION [VTBI]
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 11NOTE: VTBI will not be re-calculated if the Rate and Duration are changed. VTBI must be cleared and then the new VTBI re-entered. The new VTBI will automatically calculate a new Rate.AT KVOAudible KVO alarm will sound, clinician will need to re-set the appropriate settings.1st Action 2nd Action [AUTOCALC]change RATE keep VTBI [DURATION]change RATE change VTBI [DURATION]change VTBI keep RATE [DURATION]change DURATION keep VTBI [RATE]change VTBI change DURATION [RATE]change RATE change DURATION Recalculates [RATE] based on previous VTBI and new DURATION1st Action 2nd Action [AUTOCALC]enter VTBI keep RATE [DURATION]enter DURATION keep RATE [VTBI]change RATE enter VTBI [DURATION]enter DURATION enter VTBI [DURATION]enter VTBI enter DURATION [RATE]change RATE change DURATION [VTBI]
Draft Manual- Not to be usedin a Clinical Situation.6- 12 430-95520-001 (Rev. 04/05) Changing CCA While Infusing1Select  [Options/Vol Inf] from the Main Delivery screen.2Select  [Change CCA].3Press  to highlight desired CCA, then select  [Enter] to confirm the change.The delivery continues until VTBI equals zero, or you stop the program and clear the settings.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf26.30.20010mcg/kg/minDOPamineDose CalculationERDoseABOPTIONSChooseChange CCA BackSelect, then ChooseRelease 13.x - xx/xx/xxVolumes InfusedPressure/Post Infusion RateLighting/ContrastAny Hospital, USAMM/DD/YY XXYYZZ-123456Area SelectionEnterSelect, then EnterAny Hospital, USAMM/DD/YY XXYYZZ- 123456ICUMedSurgORCardioUrologyPsychRecovL&D5 WestRenalNeo ICUOB/GYNOncolAd ICUER7 South3 North4 North
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 13The message bar informs the user of the changed CCA.When VTBI reaches zero (infusion is competed), VTBI may no longer be incremented. To program, the user must first select  [Back], then press the  key.Auto-ProgrammingAuto-Programming is a new feature for the Hospira Plum A+ line of infusion pumps. It can only be used if the feature has been installed in conjunction with your Hospital Information System (HIS) and Barcode Point-Of-Care (BPOC) Unit. It works by scanning barcode labels for the patient, pump, and IV bag using a BPOC Unit, such as a PDA. This information is then transferred to the pump utilizing either it’s wireless antenna or RJ-45 Ethernet connector. The feature reduces the number of steps needed to program an infusion.NOTE: If the physician’s order for an auto-program therapy exceeds the capabilities of the pump or is above a hospital defined hard drug limit, the infusion parameters for auto-programming will be rejected and the order will need to be re-checked.AABBPUMPING STOPPEDRatemL/hrVol InfmLOptions/Vol Inf26.30.30010mcg/kg/minDOPamineDose CalculationICUDoseLine A using prior CCAAPROGRAMBackPress Back,then press STOP key.This program was started under aprevious CCA.
Draft Manual- Not to be usedin a Clinical Situation.6- 14 430-95520-001 (Rev. 04/05) NOTE: It is important that biomed personnel ensure that the barcode identification label affixed to each infuser corresponds to the correct Connectivity Engine Identification Number. This way the correct order will be sent to the appropriate infuser.1Power On the pump as described in Section 4.2.2After the CCA has been selected, the pump is ready for auto-programming.NOTE: The Auto-Program feature must begin on the Main Delivery Screen (with the desired line STOPPED).3Following your hospital’s procedure, activate your BPOC Unit.4Scan the patient wristband to pull up the patient’s task list on the BPOC Unit.5Scan the IV bag. IV Task and documentation with Order Details is displayed on the BPOC.NOTE: If the scanned medication does not exist in the Hospira MedNet Rx Rules medication library, that medication will display on the pump as “No Drug Selected”.6Scan the barcode of the available pump channel that is not currently infusing (A or B).AABBSTOPPED STOPPEDRatemL/hrVol InfmLOptions/Vol InfBackPrime0000MedSurg
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 6- 157Programming screen is auto-filled.8Verify ALL parameters. If changes are desired, the infusion parameters can be manually changed using the keypad on the pump.9Press   to display the Confirmation screen.10 Select  [Yes] to confirm programming and begin infusion.The BPOC will reconfirm that the pump parameters match the order.11 Complete the transaction on the BPOC Unit or document the process per hospital procedure.AProgram Dose CalcProgramOptionsCancel/BackEnter value using keypadConcWeight40070mgkg250 mLDoseVTBIDurationRate1025009: 3026.3mcg/kg/minmLhr : minmL/hrDOPamineAProgram Dose CalcYes N oDoseDrugConcWeightRateDelivery will be:104007026.3mcg/kg/minmgkgmL/hrConfirm Program?DOPamine250 mLMedSurg
Draft Manual- Not to be usedin a Clinical Situation.6- 16 430-95520-001 (Rev. 04/05) NOTES
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 7- 1SECTION 7Alarms and TroubleshootingWarning MessagesNOTE: Other error and warning messages will occur when appropriate for the current selection context. For instance, when a change to a required parameter would result in clearing some delivery settings, the warning message will appear when that field is selected. When an input error has just been made, the three quick beeps of an Invalid Key Warning will sound and an appropriate warning message will be shown to explain the MESSAGE POSSIBLE CAUSECORRECTIVE ACTIONStop delivery then turn offAttempting to turn the infuser off while a delivery is in progress.Stop all lines, then turn infuser off.Warning: Replace BatteryThe battery current charge is greater than 200 mA/hr.The battery voltage is less than the depleted threshold and charge level is higher than the low charge threshold.Replace battery.Warning: Charger ServiceThe charging circuitry is not behaving as expected (charge voltage is not changing) and can reduce battery life.Silence key is pressed.Warning: Low Battery Pump not plugged in to AC power.Plug into AC power, only about 30 minutes of battery life remains.
Draft Manual- Not to be usedin a Clinical Situation.7- 2 430-95520-001 (Rev. 04/05) problem. All such messages will be cleared or replaced by another message upon the next valid input key press.Response to Alarms1Disable Lockout by confirming back switch is in the down position or press the decimal key followed by 9, 6, & 3 on the keypad.2Press [SILENCE] key.3Identify/Observe Alarm condition.4Correct Alarm condition (see following tables).5Press [START] to resume infusion.NOTE: Alarm messages begin with an alphanumeric code for tracking purposes only. If troubleshooting does not correct the problem, record the code number and contact technical support.
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 7- 3General AlarmsCODE MESSAGE POSSIBLE CAUSECORRECTIVE ACTIONN100 Unrecognized cassetteAttempting to use unrecognized cassette.Turn off or insert recognized cassette.N101 No Action AlarmPump programming incomplete.Press [START] or continue programming.N102 Infuser Idle 2 MinutesPump is ON without operator input.Standby time has elapsed.Begin programming or turn infuser OFF.Re-enter all programmed data.N103  Memory Failure The infuser did not accept the previous programmed data.Re-enter all programmed data.N230 Prox Air, BackprimeWhen the total proximal air detected exceeds the limit.Check for clamps, or empty containers on line A or B. Correct as necessary. Backprime into line B tubing or syringe.N180, N181, N186 Distal OcclusionDistal line kinked, IV site clotted or positional, pressure limit set too low.Check distal tubing, check IV site, reset pressure limit.See page 1-14 for detailed instructions for avoiding a bolus following a distal occlusion.N187 Neg Distal OcclusionPump too high above patient or defective set.Lower infuser or replace set.N233, N234Distal Air When air is detected at the distal sensor and exceeds the limit.Remove and reprime cassette using standard technique.
Draft Manual- Not to be usedin a Clinical Situation.7- 4 430-95520-001 (Rev. 04/05) N250 Door open while pumpingPump door opened while operating.Close door with cassette inserted.N251 Cassette Test FailureFaulty cassette, proximal or distal occlusion or air in cassette detected at start up.Check clamps, manually reprime set, close door. Backprime into line B tubing or syringe if appropriate. If alarm recurs on retest, replace set. If alarm repeats, replace infuser.N252 Depleted BatteryToo much time on battery power.Plug into AC (mains) power.N253 Hard lockout ViolationThe use of the [STOP] key during delivery, or opening the door, while lockout is enabled.Unlock lockout switch.N254 Hard lockout EnabledPressing of any key except [STOP] during delivery while any alarm is active and lockout is enabled.Unlock lockout switch.N255 Soft lockout ViolationThe use of the [STOP] key during delivery, or opening the door, while lockout is enabled.Unlock lockout by pressing the decimal key, followed by 9, 6, & 3 on the keypad.N256 Soft lockout EnabledPressing of any key except [STOP] during delivery while any alarm is active and lockout is enabled.Unlock lockout by pressing the decimal key, followed by 9, 6, & 3 on the keypad.CODE MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 7- 5Line A AlarmsCODE MESSAGE POSSIBLE CAUSECORRECTIVE ACTIONN105  Callback Line A Alarm programmed by user. Change in delivery for Loading Dose to Maintenance Step or for any step in “Multistep” program provided the step is not the last step.Press [SILENCE].N161  Line A VTBI completeVTBI complete on line A.Program new VTBI.N185  Prox Occl A At StartupProximal line obstructed, line A container (or syringe) disconnected or obstructed during non-delivery (startup).Examine line A for kinks or closed clamps. If there are closed clamps, open the clamps. Then, open and close door. Continue programming. Replace set if problem persists.If clamps are not opened prior to completion of cassette test, peak proximal occlusion may recur.For syringe, select  [Backprime].
Draft Manual- Not to be usedin a Clinical Situation.7- 6 430-95520-001 (Rev. 04/05) N184, N190, N191Prox Occl A / AirProximal line obstructed, line A container disconnect or obstruction.Examine line A for kinks, closed clamps, or air in cassette.N232 Prox Air A, BackprimeWhen proximal air detected exceeds the limit for the line.Check for clamps, or empty container. Correct as necessary. Backprime into line B tubing or syringe.CODE MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 7- 7Line B AlarmsCODE MESSAGE POSSIBLE CAUSECORRECTIVE ACTIONN104 Callback Line BAlarm programmed by user. Change in delivery for piggyback delivery, Loading Dose to Maintenance Step, or for any step in “Multistep” program provided the step is not the last step.Press [SILENCE].N160  Line B VTBI completeVTBI complete on line B.Program new VTBI.N183 Prox Occl B At StartupProximal line obstructed, line B container (or syringe) disconnected or obstructed during non-delivery (startup).Examine line B for kinks or closed clamps. If there are closed clamps, open the clamps. Then, open and close door. Continue programming. Replace set if problem persists.If clamps are not opened prior to completion of cassette test, peak proximal occlusion may recur.For syringe, select [Backprime].
Draft Manual- Not to be usedin a Clinical Situation.7- 8 430-95520-001 (Rev. 04/05) MalfunctionsN182, N188, N189 Prox Occl B / AirProximal line obstructed, line B container disconnected or obstructed.Examine line B for kinks, closed clamps, or air in cassette.Occlusion caused by empty container, backprime into line B tubing or syringe.N231  Prox Air B, BackprimeWhen proximal air detected exceeds the limit for the line.Check for clamps, or empty container. Correct as necessary. Backprime into line B tubing or syringe.CODE MESSAGE POSSIBLE CAUSECORRECTIVE ACTIONE300 or higherMalfunction A failure has occurred in the infuser’s internal systems.Record Malfunction error code. Pump must be turned OFF to clear malfunction. If alarm malfunction repeats, replace infuser.CODE MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 7- 9General MessagesMESSAGE POSSIBLE CAUSECORRECTIVE ACTIONCannot Standby in PiggybackSelection of  [Standby] when Line B is in Piggyback Mode.Select  [Cancel/Back].Cannot Standby with DelaySelection of  [Standby] when Delay time was entered for program.  [Cancel/Back].Changing CCA clears programDisplays when other than last-used Care Area is highlighted.N/AChange Conc clears entriesA program delivery which has been stopped and drug amount value is selected. This is displayed before any screen or field message.User can enter new Concentration using keypad, or press [CLEAR] then make new entry.Change units clears entriesA program delivery which has been stopped and user highlights another concentration unit for the same line.User can select  [Back] to exit, or use [SELECT] arrows to highlight a new value.Change Vol clears entriesA program delivery which has been stopped and diluent value is selected. This is displayed before any screen or field message.User can enter new Concentration volume using keypad, or press [CLEAR] then make another entry.
Draft Manual- Not to be usedin a Clinical Situation.7- 10 430-95520-001 (Rev. 04/05) Concurrency alert Attempt to start a multistep or loading dose delivery whose later steps would exceed concurrent delivery limit if delivery proceeds.User can enter new values using number keypad, or use [SELECT] arrows to move to another field to change the entry. User can also press [CLEAR].Concurrency violation Attempt to start a second line with a delivery rate that would exceed concurrent delivery upper limit immediately. Or, concurrent delivery attempt and rate for the second line is less than concurrent delivery lower limit.  Takes precedence over “Rate must be > n mL/hr” message.  The combined delivery rate must fall within the range of 1.0 mL/hr – 500 mL/hr or CCA-specific Maximum Volumetric Rate, with 0.5 mL/hr minimum per line.User can enter new values using number keypad, or use [SELECT] arrows to move to another field to change the entry. User can also press [CLEAR].Decimal point not allowedAttempt to input Decimal point when either not used in field or a three-digit whole number already input.Press [CLEAR] Hardkey.MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 7- 11Dose > [X.xx]. Override? Attempt to start program with parameter that has caused the Dose (or Rate, if mL/hr program) to be greater than the upper soft limit [uuuuuuuu] for the selected drug and CCA.  Also includes placing program in Standby, and re-starting a stopped program.May be overridden by selecting   [Yes]. Cleared by pressing of [NUMERIC KEYS], [.], [SELECT], [CLEAR] or [STOP] hard keys, or  [Yes/No] softkeys. Also cleared by the [ON/OFF] key when the line is stopped.Dose < [X.xx]. Override? Attempt to start program with parameter that has caused the Dose (or Rate, if mL/hr program) to be less than the lower soft limit [llllllll] for the selected drug and CCA.  Also includes placing program in Standby, and re-starting a stopped program.May be overridden by selecting   [Yes]. Cleared by pressing of [NUMERIC KEYS], [.], [SELECT], [CLEAR] or [STOP] hard keys, or  [Yes/No] softkeys. Also cleared by the [ON/OFF] key when the line is stopped.Hard Lockout Enabled The Lockout Switch has been enabled, no Alarm is active, and any key besides Stop is pressed. Disable the Lockout switch located on the back of the device. (See Section 3)Illegal when Line DeliveringAttempt to select program other than Dose Calculation while a Basic program is delivering on this line.Select  [Cancel/Back] or press [CLEAR]. Pressing [CLEAR] invokes another message.MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.7- 12 430-95520-001 (Rev. 04/05) Input digits > max allowedAttempt to input digit that would exceed the number of decimal places allowed for defined value except Dose and Rate.Press [CLEAR] hardkey.Input value > max allowedAttempt to input number that would cause field to exceed its defined limit.  In the Drug amount, Diluent amount, VTBI, Duration, Patient Weight, Distal pressure, and Delayed Start time fields, the Software shall display ‘---‘, and the user must press the Clear hard key and input a valid value.Press [CLEAR] hardkey.Maximum value reached The display contrast/intensity has reached its maximum value.User can select  [Enter] or  [Cancel/Back].Minimum value reached The display contrast/intensity has reached its minimum value.User can select  [Enter] or  [Cancel/Back].No delivery programmed Delivery attempt at Main Delivery Screen when A and B have both Rate and VTBI equal to zero (no delivery programmed).Select a line to program, or  [Options] or  [Backprime].Rate must be > 0 mL/hr Delivery attempt when VTBI is non-zero but rate is zero.Use number keypad to enter a value.MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 7- 13Soft Lockout Enabled While the Soft Lockout is enabled and no Alarm is active, either 1) any key besides Stop is pressed during delivery, or 2) any key is pressed while stopped.Disable the Soft Lockout.(See Section 3)VTBI must be > 0 Delivery attempt when rate is non-zero but VTBI is zero.Use number keypad to enter a value.Value is too high Attempt to move cursor from field whose input value is too high.Use number keypad to enter a value.Value is too low Attempt to move cursor from field requiring a non-default value.Use number keypad to enter a value.Weight change affects Line A/BSelecting the weight field on line while a weight-based program is entered on other line.Use number keypad to enter a value.Minimum Dose X.xx Attempt to start program with Dose less than lower hard Dose limit n.nn.Use number keypad to enter a value.Maximum Dose X.xx Attempt to start program with Dose greater than upper hard Dose limit n.nn. Attempt to input a Dose greater than the upper Hard Limit.The Software will display ‘---‘, and the user must press the [CLEAR] hard key and input a valid value.MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.7- 14 430-95520-001 (Rev. 04/05) Minimum Rate X mL/hr Attempt to start mL/hr program with Rate less than lower hard Dose limit nn.n.Use number keypad to enter a value.Maximum Rate X mL/hr Attempt to start mL/hr program with Rate greater than the upper Hard Limit or CCA Maximum Volumetric Rate n.Attempt to input a Rate greater than the upper Hard Limit or CCA Maximum Volumetric Rate n.The Software will display ‘---‘, and the user must press the [CLEAR] hard key and input a valid value.MESSAGE POSSIBLE CAUSECORRECTIVE ACTION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 8- 1SECTION 8Cleaning, Maintenance, and StorageThe cleaning, maintenance, and storage of the Plum A+ are described in this section.CLEANING AND SANITIZINGFor proper maintenance of the Plum A+, observe the following cleaning and sanitizing guidelines.Establish a routine weekly schedule for cleaning the infuser. To clean, proceed as follows:• Turn the Plum A+ off using the [ON/OFF] switch• Disconnect from AC (mains) powerThe exposed surfaces of the Plum A+ may be cleaned with a lint-free cloth dampened by one of the recommended cleaning solutions in the following list or mild, nonabrasive soapy water.Clean the cassette door with a soft, lint-free cloth, dampened with one of the cleaning agents listed in the following table, or a mild solution of soapy water. Use a small non-abrasive brush to aid in cleaning the infusion system housing and subsystem chassis components. To thoroughly clean the cassette receptacle, disengage the cassette door from the door latch by pressing the door release tab.CAUTION: DO NOT ALLOW CLEANING SOLUTIONS TO SATURATETHE AIR-IN-LINE DETECTORS OR ENTER THE DEVICE WHEN CLEANINGTHE AIR-IN-LINE DETECTORS.
Draft Manual- Not to be usedin a Clinical Situation.8- 2 430-95520-001 (Rev. 04/05) On a routine basis, clean all of the elements behind the cassette door using cotton-tipped swabs saturated with cleaning solution. The cassette door may be unlatched from the door handle to facilitate cleaning. To unlatch the cassette door from its handle, tilt the device back, open the cassette door, then push the door release tab to open the door fully.CLEANING SOLUTION MANUFACTURER PREPARATIONCoverageTMHBV Steris Corporation, A division of Calgon Vestal LaboratoriesPer manufacturer's recommendationFormula C TM Diversey Corporation Per manufacturer's recommendationDispatch®Caltech Industries Per manufacturer's recommendationPrecise®Caltech Industries Per manufacturer's recommendationHousehold bleach Various Per hospital procedures; do not exceed one part bleach in ten parts waterManu-Klenz®Calgon Vestal LaboratoriesPer manufacturer's recommendationSporicidin®Sporicidin International Per manufacturer's recommendationSuper Edisonite®S. M. Edison Co. Per manufacturer's recommendation
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 8- 3CAUTION: TO AVOID MECHANICAL OR ELECTRONIC DAMAGE, DONOT IMMERSE THE PLUM A+ IN ANY FLUIDS OR CLEANINGSOLUTIONS.CAUTION: DO NOT SPRAY CLEANING SOLUTIONS TOWARD ANYOPENING IN THE INSTRUMENT.CAUTION: CERTAIN CLEANING AND SANITIZING SOLUTIONS MAYSLOWLY DEGRADE COMPONENTS MADE FROM SOME PLASTICMATERIALS. USING ABRASIVE CLEANERS OR CLEANING SOLUTIONSNOT RECOMMENDED BY HOSPIRA MAY RESULT IN PRODUCT DAMAGE.DO NOT USE COMPOUNDS CONTAINING COMBINATIONS OFISOPROPYL ALCOHOL AND DIMETHYL BENZYL AMMONIUM CHLORIDE.CAUTION: NEVER USE SHARP OBJECTS SUCH AS FINGERNAILS,PAPER CLIPS, OR NEEDLES TO CLEAN ANY PART OF THE INFUSER.CAUTION: DO NOT STERILIZE BY HEAT, STEAM, ETHYLENE OXIDE(ETO), OR RADIATION. CAUTION: TO AVOID INFUSER DAMAGE,  CLEANING SOLUTIONSSHOULD ONLY BE USED AS DIRECTED. THE DISINFECTINGPROPERTIES OF CLEANING SOLUTIONS VARY;  CONSULT THEMANUFACTURER FOR SPECIFIC INFORMATION.BATTERY MAINTENANCECAUTION: DO NOT OPERATE THE INFUSER ON PATIENTS WITH THEBATTERY REMOVED. USE OF A PROPERLY MAINTAINED ANDCHARGED BATTERY HELPS CONFIRM PROPER OPERATION. CAUTION:  IF THE LOW-BATTERY ALARM SOUNDS, CONNECT TOAC (MAINS) POWER IMMEDIATELY.The Plum A+ is battery powered for emergency backup and temporary portable operation. A fully charged, new battery provides 3 hours of operation at 125 mL/hr, or delivers 250 mL total volume if pumping at a rate of 126 mL/hr or greater, whichever occurs first. NOTE: For optimum battery life, the Plum A+ should be operated on battery power for three continuous hours at least once every six months and then charged for a minimum of six hours.
Draft Manual- Not to be usedin a Clinical Situation.8- 4 430-95520-001 (Rev. 04/05) The battery charges whenever connected to AC (mains) power. If the infuser is switched OFF, recharge takes approximately six hours. Recharge takes longer if the infuser is turned ON.As a general rule, the more often the battery is partially discharged and recharged, the sooner it will need to be replaced. Consult a qualified biomedical technician for battery replacement if necessary.To maintain maximum battery charge and to prolong battery life, connect the infuser to AC (mains) power whenever possible.When the infuser is used without wireless, the operation times may be proportionally longer.STORAGETo prolong the life of the Plum A+, observe the following storage precautions:• Store the away from excessive heat, cold, and humidity• Store connected to AC (mains) power• Switch the infuser OFF using the [ON/OFF] keySERVICEAll servicing or adjustments to the Plum A+ should be referred to qualified technical personnel. A technical service manual may be ordered from the local Hospira sales office.
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 9- 1SECTION 9SpecificationsPhysicalElectricalDimensions: Approximately 8” X 8” X 6”, excluding pole clamp protrusion and power cord storage.Weight: Approximately 9.5 lbs. with battery.Casing: High-impact plastic.Power Requirements:120 V~, 50-60 Hz, 35 VA. Meets UL 60601-1.Fuses: F1, F2, 250V~, 0.5 A. (internal)Power Cord: Hospital-grade AC cord. 10 ft long, with transparent plug and retainer plate.Battery: One sealed, lead-acid, rechargeable 6 V battery, internal to device.Battery Life: With a new fully charged battery, the infuser operates for a minimum of three hours at 125 mL/hr or less, or delivers 250 mL if > 126 mL/hr. (Time is measured from initial pumping to Depleted Battery Alarm)Recharge: The battery charges whenever the infuser is connected to AC power. The recharge time is approximately six hours with the device operating at 125 mL/hr on one line.Electrical Leakage:Meets IEC 60601-1 standard: Medical Electronic Equipment, Part 1: General Requirements for Safety.
9- 2 430-95520-001 (Rev. 04/05) Wireless Lan DeviceVTBI RangeNURSE-CALL System:Circuitry Ratings:NURSE-CALL alarm is factory set for Normally- Open (NO)Contact the Technical Services Center to make an internal adjustment to change the device from Normally-Open (NO) to Normally Closed (NC) system.Voltage-30 VDC MaxCurrent- 0.25 Amps MaxContact Rating- 3 Watts MaxStandards: IEEE802.11bRadio Technology: IEEE802.11b Direct SequenceSpread SpectrumData Transfer  Rate:11, 5.5, 2, 1 Mbps (auto sense)FREQUENCY RANGEIEEE802.11b Channels:Transmit Power:Antenna:Certifications:2412 - 2462 MHz1-11 (FCC)100mW (+20dBm) maxIntegrated surface mount antennaFCC Part 15.247 for US, ICVTBI Range: 0.1 to 99.9 mL (in 0.1 mL increments)100 to 9999 mL (in 1 mL increments)
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 9- 3EnvironmentDelivery Rate RangeAir-in-Line AlarmOperating Temperature:5° to 40° CStorage Temperature:-20° to 60° CAtmospheric Pressure:0 - 10,000 feet (0 - 3,000m) or equivalent pressureRelative Humidity: 10 - 90% (40° C Max)Lines A and B: 0.1 - 99.9 mL/hr (in 0.1 mL increments)100 - 999 mL/hr (in 1 mL increments)Concurrent Delivery:PlumSet:0.5 mL/hr minimum for each line500 mL/hr cumulative (A+B) maximumKVO: 1.0 mL/hr or the last primary delivery rate, whichever is lessPlumSet (Distal): Bolus 0.1 mL (100 microliters) or larger for delivery rates less than 500 mL/hr.Bolus 0.5 mL (500 microliters) or larger for delivery rates equal to or greater than 500 mL/hr.Cumulative 0.25 mL out of 2.6 mL for delivery rates less than 500 mL/hr.Cumulative 0.50 mL out of 5.3 mL for delivery rates equal to or greater than 500 mL/hr.PlumSet (Proximal):Bolus at 0.5 mL, Total 1.0 mL
9- 4 430-95520-001 (Rev. 04/05) Occlusion Alarm and LimitsDistal Occlusion: The DISTAL OCCLUSION alarm sounds after the distal set tubing or set outlet fitting becomes occluded.Proximal Occlusion:The PROXIMAL OCCLUSION alarm sounds if the tubing proximal to the cassette becomes occluded. Distal Pressure Limit           (without alarm):1 to 15 psig. The maximum pressure is user-selectable. Factory default setting is 6 psig.Maximum Infusion Pressure:20 psig
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 9- 5Time To Detect DownstreamOcclusionsBolus Volume Released After Downstream Occlusions AreCorrectedRATEOCCLUSION ALARM PRESSURE SETTINGTYPICAL TIME TO ALARM   (MACRO)MAXIMUM TIME TO ALARM (MACRO)25 mL/hr 6 psig15 psig38 seconds98 seconds47 seconds106 seconds1 mL/hr 6 psig15 psig17 minutes45 minutes22 minutes49 minutes* Baseline backpressure is 0 psig *RATEOCCLUSION ALARM PRESSURE SETTINGTYPICAL BOLUS VOLUME   (MACRO)MAXIMUM BOLUS VOLUME (MACRO)25 mL/hr 6 psig15 psig0.23 mL0.55 mL0.28 mL0.63 mL1 mL/hr 6 psig15 psig0.24 mL0.59 mL0.30 mL0.70 mL* Baseline backpressure is 0 psig *
9- 6 430-95520-001 (Rev. 04/05) Delivery AccuracyThe Plum A+ Infuser was designed and manufactured to maintain a volumetric delivery rate error of the total fluid delivered of less than or equal to ± 5% over the course of 48 hours at a programmed rate of 1 to 999 mL/hr during normal operating conditions. For use of the device at rates below 1 mL/hr, the delivery rate error is less than or equal to ± 10%.Backpressure Effect- Positive backpressure on the distal line may affect delivery accuracy by no more than 0.5% per psig (2% per psig for rates between 0.1 and 0.9 mL/hr) for backpressures up to 15 psig. A typical deviation under these conditions is 0.3%. Negative backpressure may affect delivery accuracy no more than 2% for pressures up to -2 psig (56 inches of water). A typical deviation under these conditions is 0.3%. Filling Head Effect- Variations in filling head (such as container height) may affect delivery accuracy by no more than 2% for variations in the range -15 to +28 inches. A typical deviation under these conditions is 0.5%.Concurrent Delivery Effect- When both lines (A & B) are delivering, the ratio of delivery for the fluid with the lowest rate may be affected by as much as 5% for ratios up to 9 to 1.   For higher ratios, the absolute percentage of delivery for the lowest rate may be affected by no more than 0.5%. When air is present in the bubble trap, the absolute percentage of delivery for the lowest rate may be affected by up to 2.0%. When variations in container height are present, the absolute percentage of delivery for the lowest rate may be affected by up to 2.5% for up to 24 inches of container height differences.Enteral & High Viscosity Fluids EffectsSystem delivery accuracy limits for each Enteral and High Viscosity Fluid, such as those listed in the following table, can be degraded by an additional maximum of 5%. System accuracy for Enteral Fluids is defined only for rates of 1 to 200 mL/hr, with no
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 9- 7suspended air in the solution, and using an Hospira Plum enteral set.Trumpet CurvesThe Trumpet Curve Graphs following the Example show representative maximum and minimum percent flow rate deviation from the programmed rate over time. This information was developed in accordance with IEC 60601-2-24: 1998, Sub-Clause 50.102. Refer to this standard for detailed information.How to read a Trumpet Curve Graph (Refer to example on the following page): The graphs following the Example plot flow rates at 30 second intervals for the first 2 hours and for the 96th hour of delivery. The graph plots mean delivery rate error (Average of 3 infusers) for the 2nd hour and the 96th hour as a straight line. The graph also presents maximum and minimum average delivery rate error for this interval plotted by averaging delivery errors over intervals of 2, 5, 11, 19 and 31 minutes (“Trumpet Curve”).EXAMPLEFrom the Trumpet Curve Graph sample that follows, find the 5 minute interval (A) at the horizontal axis and read the corresponding points (B) and (C) on the vertical axis. The values are approximately +2.8% and -0.5%.This means that at the rate of 25 mL/hr the average maximum flow rate fluctuation for any 5 minute time interval during the 2nd hour of operation was within the limits of +2.8% and -0.5% from ENTERAL/HIGH VISCOSITY FLUIDSDextrose 20% to 50% concentrationIsomil PowderSimilac PowderEnsure Plus HNTwocal HNJevity
9- 8 430-95520-001 (Rev. 04/05) the nominal rate. The average delivery rate error over the entire 2nd hour was +1.6% (D).For other time intervals look at other points at the horizontal axis and determine corresponding limits as above.A trained professional can use the resulting graphs to select a infuserwith the appropriate startup and flow characteristics to suit the clinicalapplication.NOTE: As an example of how the trumpet curves can be used, consider the maximum and minimum deviations at the 5 minute average interval.   The upper curve provides the maximum expected delivery rate error over a 5 minute interval, the lower curve provides the minimum expected delivery rate error over a 5 minute interval. An example would be Dopamine administered at 5 µgm/kg/min. At 5 minutes, the average drug delivery error would be within the range of +2.8% and -0.5% of the expected nominal rate.0510 15 20 25 3015.0%10.0%5.0%0.0%-5.0%-10.0%-15.0%25 ML/HR, 0 mm Hg DELIVERY HEAD, START UPPERCENT ERROR (%)Observation Interval (minutes)+++++Max Mean+MinAverage DeliveryRate Error = +1.6%ADMax Negative Error over 5Minute Interval = -0.5%CMax Positive Error over 5Minute Interval = +2.8%B
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 9- 925 ML/HR, 0 mm Hg DELIVERY HEAD, START UP(Typical)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean25 ML/HR, 0 mm Hg DELIVERY HEAD, START UP(Typical)-10-5051015202530354045500 10 20 30 40 50 60 70 80 90 100 110 120Time (minutes)Rate (ml/hr)Actual RateProgrammed Rate25 ML/HR, 0 mm Hg DELIVERY HEAD, START UP(Composite of 3 Pumps)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean
9- 10 430-95520-001 (Rev. 04/05) 25 ML/HR, 0 mm Hg DELIVERY HEAD , 96 HOURS(Typical)051015202530354045500 1020304050Time (minutes)Rate (ml/hr)Actual RateProgrammed Rate25 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS(Typical)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean25 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS(Composite of 3 Pumps)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 9- 111 ML/HR, 0 mm Hg DELIVERY HEAD, START UP(Typical)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean1 ML/HR, 0 mm Hg DELIVERY HEAD, START UP(Typical)-0.4-0.20.00.20.40.60.81.01.21.41.61.82.00 10 20 30 40 50 60 70 80 90 100 110 120Time (minutes)Rate (ml/hr)Actual Rate Programmed Rate1 ML/HR, 0 mm Hg DELIVERY HEAD, STARTUP(Composite of 3 Pumps)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean
9- 12 430-95520-001 (Rev. 04/05) 1 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS(Typical)0.00.20.40.60.81.01.21.41.61.82.00 1020304050Time (minutes)Rate (ml/hr)Actual Rate Programmed Rate1 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS(Typical)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean1 ML/HR, 0 mm Hg DELIVERY HEAD, 96 HOURS(Composite of 3 Pumps)-15.0%-10.0%-5.0%0.0%5.0%10.0%15.0%0 5 10 15 20 25 30Observation Interval (minutes)Percent Error (%)Max Min Mean
Plum A+ Infusion System430-95520-001 (Rev. 04/05) 10- 1SECTION 10Supplies and AccessoriesAdministration FluidsADMINISTRATION FLUIDS• All parenteral fluids• Whole Blood• Blood fractionsENTERAL AND HIGH VISCOSITY FLUIDS• Dextrose 20% to 50% concentration• Isomil Powder• Similac Powder• Ensure plus HN• TwoCal HN• Packed red cellsCONTAINERS• Dual Chamber Parenteral flexible container (Nutrimix)• Large Volume Parenteral flexible plastic containers, including premixed• Large Volume Parenteral glass containers, including premixed and nutritional• Part Fill Parenteral flexible plastic containers, including ADD-Vantage• Part Fill Parenteral glass containers• Small Volume Parenterals• Syringes (may require special adapters)• Top Filled Enteral bags• Ready to hang Enteral solution containers
10- 2 430-95520-001 (Rev. 04/05) AccessoriesGENERAL• Syringe Adapter- List# 11986-48• Mini Pole- List# 12096-04• Secondary Container Support- List# 12095-03• Tandem Carrier- List# 12270-01NETWORKED APPLICATION• Hospira MedNetTM Medication Management Server Suite (MMSS)List# 20676-04• *Hospira MedNetTM Plug ‘n Play ModuleList# 20677* = For use with Hospira Plum A+ Infusers Lists 11971-04-03 & above, 12391-04, & 20679-04NOTE: Contact Technical Support Operations at 1-800-241-4002 for current listing of accessories.
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 11- 1SECTION 11WarrantySubject to the terms and conditions herein, Hospira, Inc., herein referred to as Hospira, warrants that (a) the product shall conform to Hospira's standard specifications and be free from defects in material and workmanship under normal use and service for a period of one year after purchase, and (b) the replaceable battery shall be free from defects in material and workmanship under normal use and service for a period of 90 days after purchase. Hospira makes no other warranties, express or implied, as to merchantability, fitness for a particular purpose, or any other matter.Purchaser's exclusive remedy shall be, at Hospira's option, the repair or replacement of the product. In no event shall Hospira's liability arising out of any cause whatsoever (whether such cause be based in contract, negligence, strict liability, other tort, or otherwise) exceed the price of such product, and in no event shall Hospira be liable for incidental, consequential, or special damages or losses or for lost business, revenues, or profits. Warranty product returned to Hospira must be properly packaged and sent freight prepaid.The foregoing warranty shall be void in the event the product has been misused, damaged, altered, or used other than in accordance with product manuals so as, in Hospira's judgment, to affect its stability or reliability, or in the event the serial number or lot number has been altered, effaced, or removed.The foregoing warranty shall also be void in the event any person, including the Purchaser, performs or attempts to perform any major repair or other service on the product without having been trained by an authorized representative of Hospira and using Hospira documentation and approved spare parts. For purposes of the preceding sentence, "major repair or other service" means any repair or service other than the
Draft Manual- Not to be usedin a Clinical Situation.11- 2 430-95520-001 (Rev. 04/05) replacement of accessory items such as batteries and detachable mains power cords.In providing any parts for repair or service of the product, Hospira shall have no responsibility or liability for the actions or inactions of the person performing such repair or service, regardless of whether such person has been trained to perform such repair or service. It is understood and acknowledged that any person other than an Hospira representative performing repair or service is not an authorized agent of Hospira.
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 12- 1SECTION 12Default Drug Library (DDL)NOTE: This list is only accurate if no custom drug library has been developed using the Hospira MedNetTM accessory. Once a custom drug library has been developed, this list will no longer be accessable. Only Hospira personnel can re-install the DDL.The information contained in this manual is current as of 11/01/02. User must assure themselves that drug prescribing information has not changed since the publication of this manual.Use drug manufacturer recommendations for IV administration when using the Plum A+ infusion infuser.DRUG NAMEDEFAULT DOSE RATEDEFAULT CONCENTRATIONNo Drug Selected mL/hr N/AAbciximab mcg/kg/min mg/mLAcyclovir mL/hr N/AAlbumin mL/hr N/AAldesleukin mL/hr N/AAlfentanil mcg/kg/min mcg/mLAlteplase (rt-PA) mg/kg/hr mg/mLAmikacin mL/hr N/AAminophylline mg/hr mg/mLAmiodarone mg/min mg/mLAmphotericin B mL/hr N/AAmpicillin mL/hr N/AAtracurium mL/hr N/AAzithromycin mL/hr N/A
Draft Manual- Not to be usedin a Clinical Situation.12- 2 430-95520-001 (Rev. 04/05) Bleomycin mL/hr N/ABlood Products mL/hr N/ABretylium mg/min mg/mLCarboplatin mL/hr N/ACefazolin mL/hr N/ACefepime mL/hr N/ACefoPERAZONE mL/hr N/ACefoTAXIME mL/hr N/ACefoTETAN mL/hr N/ACefoXITIN mL/hr N/ACeftAZIDime mL/hr N/ACeftIZOXime mL/hr N/ACeftriaxone mL/hr N/ACefuroxime mL/hr N/ACimetidine mL/hr N/ACiprofloxacin mL/hr N/ACisplatin mL/hr N/AClindamycin mL/hr N/ACotrimoxazole mL/hr N/ACytarabine mL/hr N/ADiltiazem mg/hr mg/mLDOBUTamine mcg/kg/min mg/mLDocetaxel mL/hr N/ADOPamine mcg/kg/min mg/mLDOXOrubicin mL/hr N/AEpinephrine mcg/min mg/mLEpoprostenol ng/kg/min mg/mLEptifibatide mcg/kg/min mg/mLDRUG NAMEDEFAULT DOSE RATEDEFAULT CONCENTRATION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 12- 3Erythromycin mL/hr N/AEsmolol mcg/kg/min mg/mLEtoposide mL/hr N/AFamotidine mL/hr N/AFentanyl mL/hr N/AFluconazole mL/hr N/AFlumazenil mL/hr N/AFluorouracil mL/hr N/AFurosemide mL/hr N/AGatifloxacin mL/hr N/AGentamicin mL/hr N/AHeparin units/kg/hr units/mLHydromorphone mL/hr N/AIfosfamide mL/hr N/AInsulin units/hr units/mLLabetalol mg/min mg/mLLepirudin mg/kg/hr mg/mLLeucovorin mL/hr N/ALevofloxacin mL/hr N/ALidocaine mg/min grams/mLLorazepam mL/hr N/AMagnesium grams/hr grams/mLMannitol grams/hr grams/mLMesna mL/hr N/AMethoHEXITAL mL/hr N/AMethoTREXATE mL/hr N/AMetoclopramide mL/hr N/AMetronidazole mL/hr N/ADRUG NAMEDEFAULT DOSE RATEDEFAULT CONCENTRATION
Draft Manual- Not to be usedin a Clinical Situation.12- 4 430-95520-001 (Rev. 04/05) Midazolam mL/hr N/AMilrinone mcg/kg/min mg/mLMorphine mg/hr mg/mLNafcillin mL/hr N/ANitroGLYCERIN mcg/min mg/mLNitroPRUSSIDE mcg/kg/min mg/mLNorepinephrine mcg/min mg/mLOfloxacin mL/hr N/AOndansetron mL/hr N/AOxacillin mL/hr N/AOxytocin mUn/min units/mLPaclitaxel mL/hr N/APantoprazole mg/min mg/mLPenicillin mL/hr N/APentobarbital mL/hr N/APhenylephrine mcg/min mg/mLPiperacillin mL/hr N/APotassium mEq/hr mEq/mLProcainamide mg/min grams/mLPropofol mcg/kg/min mg/mLRanitidine mL/hr N/ASTREPTOkinase units/hr units/mLTheophylline mg/kg/hr mg/mLTicar-clavulanate mL/hr N/ATirofiban mcg/kg/min mg/mLTobramycin mL/hr N/AUROkinase mL/hr N/AVancomycin mL/hr N/ADRUG NAMEDEFAULT DOSE RATEDEFAULT CONCENTRATION
Draft Manual- Not to be usedin a Clinical Situation.Plum A+ Infusion System430-95520-001 (Rev. 04/05) 12- 5Verapamil mL/hr N/AVinCRIStine mL/hr N/ADRUG NAMEDEFAULT DOSE RATEDEFAULT CONCENTRATION
Draft Manual- Not to be usedin a Clinical Situation.12- 6 430-95520-001 (Rev. 04/05) NOTES
Plum A+ Infusion System430-95520-001 (Rev. 04/05)For customer service within the United States, contact:1-877-9-HOSPIRA (1-877-946-7747)For technical assistance, product return authorization, and to order parts, accessories, or manuals within the United States, contact Hospira, Inc. Technical Support Operations:1-800-241-4002To order parts using the online eCatalog, download technical publications, technical training courses, and additional services, visit the website at:www.hospiraparts.comAfter authorization, ship prepaid product returns to the following address:Hospira, Inc.Technical Support Operations755 Jarvis DriveMorgan Hill, CA 95037Outside the U.S., contact your local Hospira, Inc. sales office.
430-95520-001 (Rev. 04/05) CAUTION: FEDERAL (USA) LAW RESTRICTS THIS INFUSER TOSALE BY OR ON THE ORDER OF A PHYSICIAN OR OTHER LICENSEDPRACTITIONER.WARNINGPOSSIBLE EXPLOSION HAZARD EXISTS IF THE INFUSERIS USED IN THE PRESENCE OF FLAMMABLEANESTHETICS.Plum A+ Infusion Pump is a registered trademark of Hospira, Inc.Hospira MedNet, LifeShield, and E.L.I. are trademarks ofHospira Incorporated. CLAVE, Teflon, Formula C, Manu-Klenz, SuperEdisonite, Coverage, Sporicidin, Dispatch, and Precise are nottrademarks of Hospira.®C             USClass 1 Mains supply equipment using protective earthIPX1Drip ProofMedical Equipment Protected against dripping waterType CFEquipment providing adequate degree of protection againstelectrical shock and suitable for application to patientUL 60601-1CSA 601.1MCN 160992