GE Ultrasound Korea ARC Attachable Remote Controller User Manual ARC manual english

GE Ultrasound Korea Attachable Remote Controller ARC manual english

Users Manual

ARC(Attachable Remote Controller) User Manual  Section 1 Overview Section 1-1 Purpose This manual describes the operating procedure for ARC(Attachable  Remote  Controller)  in system.  Section 1-2 Simple Block Diagram   ARC is consist of Bluetooth module and Recharge module. Bluetooth module can transfer data with 2.4GHz . Recharge module is supported by USB 5V.    Operating battery is Li-polymer 3.7V. ARC board can control 2 kind of Keys which are P1 and Freeze.   Fig.1 ARC Block Diagram  Section 1-3 Parts Required   Description  GPN  Remark 1  ARC  5318111  Remote Controller 2  USB type Bluetooth Dongle  5316870  Bluetooth dongle 3  Recharge cable  5324554  Mini A-type USB Cable
Section 2    Operating procedure Section 2-1 Recharging procedure 1) Connect recharge cable to ARC module 2) Connect recharge cable to USB port of rear panel. 3) Confirm LED status.  Table1. LED Status   Color  Operating LED1  Red  Charge in Progress LED2  Green  Full Charge LED3  Blue  Pairing and Button response LED4  Red  Low Battery   Section 2-2 Pairing 1) Turn on System and Press ‘Utility’ button in Keyboard. 2) Click Connectivity and select ‘ADM’ in Title Log In Window.(Fig.2, Fig.3) Operator : ADM, / Password : empty Finaly click OK. 3) Click Wireless and select Enable. 4) Press P1+ Freeze buttons of ARC at the same time more than 3 seconds. 5) Confirm blinking blue LED 6) Click Find Device.(Fig.4) 7) Check notification of Address and then click ‘Save’(Fig.5) 8) Check that ‘RC’ icon is to be green low of the left.(Fig.5)
 Fig.2 Connectivity in Utility menu   Fig.3 Title Login
 Fig.4 Wireless and Find Device   Fig.5 Checking Pairing
Section 2-3 Key Operation 1) Freeze Button is already assigned as Freeze function. When pressed Freeze button of ARC, Live scan image is stopped. 2) P1 is same function which is defined in utility of Keyboard Button. For example, if P1 button is assigned BW printer. BW printer will operate when    pressed P1 button of ARC.   3) If reset is needed, press Reset button by suitable equipment.   Fig.6 Reset procedure
FCC Requirements  Cautions    Modifications  not  expressly  approved  by  the  party  responsible  for  compliance  could  void  the user’s authority to operate the equipment.  FCC compliance Information   This  device  complies  with  part  15 of FCC Rules. Operation is subject to the following two conditions: 1. This device may not cause harmful interference, and 2. This device must accept any interference received. Including interference that may cause undesired operation.  Information to User   This equipment has been tested and found to comply with the limits for a Class B digital device, Pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio Frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However,  there  is  no  guarantee  that interference  will  not  occur  in  a  particular  installation.  If  this equipment does cause harmful interference to radio or television reception, which  can  be  determined by turning the equipment off  and  on,  the  user  is  encouraged  to  try  to  correct  the  interference  by  one  or  more  of  the following measures:   - Reorient or relocate the receiving antenna.   - Increase the separation between the equipment and receiver   - Connect the equipment into an outlet on a circuit different from that  to  which  the  receiver  is connected.   - Consult the dealer or an experienced radio/TV technician for help.