2−11Press the   ID   key.Enter the ID number using the numeric keypad.3. Enter the patient’s birth date.Press the   Age   key.Enter the birth date using the numeric keypad.4. Select the patient type, and   Used   /    Not used   of pacemaker use.Select the patient type from   Adult  ,   Child ,or   Neonate .Select the pacemaker use from   Used , Not used .WARNING The patient type and pacemaker use must be selected.    The patient typeselection influences the precision of the QRS detection and NIBPmeasurement.  Make sure the correct selection is made.2To Start Monitoring
2−12− Basic Operation  −Scale, Lead, Baseline Position Setup (Parameter Key)1. Select the parameter to set the scale / lead / baseline position. (Ex.: ECG)Press the parameter key where heart rate is displayed.2. Press the  Lead・Size  key.  The menu to adjust size / lead / baseline position will bedisplayed.3. Adjust the waveform size.Select an appropriate waveform size for the monitoring waveform.CAUTION The arrhythmia detection level corresponds with the displayed waveformsize.  Select an appropriate size for monitoring.
2−134. Select the lead.Select an appropriate lead according to the monitoring purpose.5. Adjust the baseline position.Use the   ? ,   ?  keys to adjust the waveform baseline position.ReferenceUse the same procedure for each parameter.Refer to section for each parameter in “6. Parameter Setup”.2Basic Operation
2−14Scale, Lead, Baseline Position Setup (User Key)Pressing the   Lead・Size  key will display thearrow keys on the home display to adjustwaveform size, scale, lead, baseline position.1. Select the waveform size, lead, baseline position for ECG waveform.Adjust the waveform suitable for monitoring.Pressing the   Lead   key will sequentially changethe lead.3-electrode:Ⅰ→Ⅱ→Ⅲ→Ⅰ4-electrode:Ⅰ→Ⅱ→Ⅲ→aVR→aVL→aVF→Ⅰ5-electrode:Ⅰ→Ⅱ→Ⅲ→aVR→aVL→aVF→V→ⅠPressing the   Size  key will sequentially change thesize.×1/4→×1/2→×1→×2→×4→×1/4Use the  ↑,  ↓ keys to adjust the baseline positionup or down.2. Select the scale for BP, CO2 waveform.Pressing the   Scale  key will sequentially switch thescale.3. Select the waveform size for impedance respiration waveform, SpO2 waveform.Pressing the   Size  key will sequentially change thesize.×1/4→×1/2→×1→×2→×4→×1/4
2−15Alarm Setup for Each ParameterThe alarm can be set for each parameter.  By pressing the selected parameter key, upper andlower alarm limit and ON/OFF of alarm can be set.1. Select the parameter to set the alarm. (Ex.: HR alarm)2. Press the   HR Alarm   key.  The menu to adjust the alarm limit will be displayed.3. Set the upper and lower alarm limit.Use the   ? ,   ?  keys to adjust the lower limit.Use the   ? ,   ?  keys to adjust the upper limit.ReferenceUse the same procedure for the setup of each parameter.Refer to section for each alarm in “6. Parameter Setup”Lower LimitUpper Limit2Basic Operation
2−16ON/OFF of Parameter Display Waveform/Numeric Data DisplayThe waveform and numeric data display for each parameter can be turned ON or OFF withoutchanging the display configuration.  If not performing the ECG or SpO2 measurement while theECG cable or SpO2 sensor is connected to the monitor, the equipment status alarm such as “LeadOff” will generate.  Removing ECG or SpO2 from the display configuration will not generate suchalarm, but this function may be more useful as it allows to turn off the measurement withoutchanging the display configuration.This function is not available for NIBP monitoring.1. Select the parameter to turn off the display. (Ex.: ECG)             2. Press the  Display ON/OFF   key.  The confirmation display for ON/OFF of ECG displaywill appear.             3. Select   Display ON  or    Display OFF .Pressing the   Display ON   key will display thewaveform and numeric data.Pressing the   Display OFF   key will not display thewaveform and numeric data.4. Automatic resetFor ECG, impedance RESP, SpO2, CO2, properly connecting the sensor will automatically set thedisplay ON/OFF function to “Display ON”.ReferenceFor automatic reset condition, refer to Display ON/OFF section for each parameter in“6. Parameter Setup”.The Display OFF message will bedisplayed inside the parameter key.
2−17Recording●Start / Stop of Waveform RecordingPressing the   Rec. START/STOP   key on thehome display will start the waveform recording.Up to 3 waveforms can be recorded.【Example of Recording】ReferenceFor the manual recording, number of recording waveforms and recording duration can be set.Refer to “4. Monitoring Setup  Manual Recording” for details.For the alarm recording, number of recording waveforms, recording duration, alarm factor can beset.Refer to “4. Monitoring Setup  Alarm Recording” for detailsFor the periodic recording, number of recording waveforms, recording duration, recordingintervals can be set.Refer to “4. Monitoring Setup  Periodic Recording” for details.The monitoring data of the patient such as graphic trend and tabular trend can be recorded.Refer to sections on graphic trend and tabular trend in “7. Function”.Bed IDPatient NameWaveform Type, Lead, SizeRecording Speed       Rec. Mode      Delay Time       Telemetry Channel      Filter ModeRecorded Time   Sex    Age   Patient ID   Patient Type    Measured Data2Basic Operation
2−18To Install the Paper1. Press the button located at the right side of the recorder magazine.2. The magazine will come out.3. Pull out the magazine from the monitor.4. Set the recording paper so that the mark printed on each paper comes to the right.5. Place the magazine back into the monitor.  Push in until it locks into place with a clicksound.Mark
2−19To Suspend MonitoringWhen not monitoring for a while, turning OFF the power will erase the recall data, STmeasurement, OCRG data.However, using the cease monitoring function allows suspension of data measurement, alarmgeneration, automatic measurement, automatic recording without erasing the data or setupdetails.1. Press the   Cease   key on the admit / discharge menu.Press the   Menu   →    Admit / Discharge   → Cease   key.2.  Suspend monitoring.Pressing the   OK   key on the confirmationdisplay will interrupt monitoring.Pressing the   Cancel   key will return to theprevious display.3. Verify that the monitoring is suspended.The   Resume  key will be displayed on thehome display.  On the home display, numericdata and waveform display will be ceased, andall the key operation except the   Resumekey will become ineffective.NOTEThe telemetry transmission will cease when monitoring is suspended.(for DS-7141, DS-7101LT)Note that the square wave will be displayed on the central monitor indicating thetoo far condition of the telemetry.2Basic Operation
2−20Discharging ProcedureThis procedure will erase the past data, such as patient name, ID, age, tabular / graphic trend, andrecall.1. Press the   Discharge  key on the admit / discharge menu.Press the   Menu   →    Admit / Discharge   → Discharge   key.2. Perform discharge procedure.Pressing the   OK   key on the confirmationdisplay will discharge the patient.Pressing the   Cancel   key will return to theprevious display.
2−21− Operation Flow  −The operation flow of this system is as follows.【Menu】Admit /DischargePatient Name / ID / Sex Discharge Room / Bed IDPatient Type (Adult / Child / Neonate) Mode Selection PacemakerAge Cease Monitor Filter ModeParameter Setup Zero BalanceGraphic Trend OCRGTabular Trend Alarm SetupNIBP List Volume SetupRecall Display ConfigurationST Display Night ModeOther Bed【System Configuration】Night Mode SetupRecord SetupSweep SpeedBrightness SetupIC CardColor SetupTelemetry Waveform Setup【Preset】Alarm Mode SetupHospital SetupWard SetupMonitor SetupDisplay Mode SetupTest Menu2Operation Flow