  Operating Manual  EMGZ310 Digital Microprocessor Controlled Tension Amplifier   EMGZ310.ComACT Digital Microprocessor Controlled Tension Amplifier  with intuitive operation via FMS ComACTTM app   Version 2.10  10/2016  NS   Firmware Version EMGZ310  V1.14   Firmware Version EMGZ310.ComACT  V2.10     This operation manual is also available in German. Please contact your local FMS representative.  Diese Bedienungsanleitung ist auch in Deutsch erhältlich. Bitte kontaktieren Sie Ihren nächstgelegen FMS Vertreter.   © by FMS Force Measuring Systems AG, CH-8154 Oberglatt – All rights reserved.
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   2  Table of contents  1Safety Instructions ..................................................................................... 31.1Description conditions 31.2List of safety instructions 41.3Regulatory Notices 51.4Notes Réglementaires 62Quick Installation Guide ............................................................................ 72.1Preparations for Set-up 72.2Installation Procedure 72.3Mounting the Force Sensors 72.4Installation and wiring 72.5Block diagram 82.6Wiring the Amplifier 82.7Modify the wiring 93Configuring the Amplifier ....................................................................... 103.1Power up the EMGZ 310 103.2Operating the EMGZ310 over the Operating Panel 103.3Operating Panel 103.4Change the Device Mode 113.5Offset Compensation 113.6Calibration 123.7Calibration Procedure in the Force Gauge Mode 133.8Calibration Procedure in Volt Gauge Mode 144Configuration via operating panel ......................................................... 154.1Reset to Default Parameter Set 174.2Parameter List 184.3Description of Parameters 195Configuration via FMS ComACTTM app .................................................. 225.1Screenshots 236Dimensions .............................................................................................. 247Technical Specification ........................................................................... 258Your Settings ............................................................................................  269Notes ......................................................................................................... 27
 Operating Manual EMGZ 310   10.01.2017   3  1 Safety Instructions All safety related regulations, local codes and instructions that appear in the manual or on equipment must be observed to ensure personal safety and to prevent damage to the equipment connected to it. If equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Do not stress the equipment over the specification limits neither during assembly nor operation. To do so can be potentially harmful to persons or equipment in the event of a fault to the equipment.  1.1 Description conditions a) High danger of health injury or loss of life Danger This symbol refers to high risk for persons to get health injury or loss of life. It has to be followed strictly.  b) Risk of damage of machines Caution This symbol refers to information, that, if ignored, could cause heavy mecanical damage. This warning has to be followed absolutely.  c) Note for proper function Note This symbol refers to an important information about proper use. If not followed, malfunction can be the result.
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   4  1.2 List of safety instructions  Proper function of the Tension Measuring Amplifier is only guaranteed with the recommended application of the components. In case of other arrangement, heavy malfunction can be the result. Therefore, the installation instructions on the following pages must be followed strictly.   Local installation regulations are to preserve safety of electric equipment. They are not taken into consideration by this operating manual. However, they have to be followed strictly.   Bad earth connection may cause electric shock to persons, malfunction of the total system or damage of the control unit. It is vital to ensure a proper earth ground connection.   Each change of the Device Mode requires a re-calibration of the system    Changes or modifications made to this equipment not expressly approved by FMS AG may void the FCC authorization to operate this equipment.
 Operating Manual EMGZ 310   10.01.2017   5  1.3 Regulatory Notices NOTE: This device complies with Part 15 of the FCC Rules and with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions:  this device may not cause harmful interference, and   this device must accept any interference received, including interference that may cause undesired operation.   Fig. 1a: FCC labeling on device      FCC_Label_EMGZ310ComACT.ai  NOTE:  This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   6  NOTE: Changes or modifications made to this equipment not expressly approved by FMS Force Measuring Systems AG may void the FCC authorization to operate this equipment.  Radiofrequency radiation exposure Information: This equipment complies with FCC and IC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance of [50] cm between the radiator and your body. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. 1.4 Notes Réglementaires NOTE: Cet équipement est conforme à la section 15 du règlement de la Commission fédérale américaine des communications (FCC) et au(x) Cahier(s) des charges sur les normes radioélectriques (CNR) d’Industrie Canada exemptes de licence. Son utilisation est soumise aux deux conditions ci-dessous :    cet équipement ne peut pas causer d’interférences nuisibles, et  cet équipement doit accepter toutes les interférences, y compris celles qui pourraient provoquer un fonctionnement indésirable.  NOTE:  Cet équipement a été testé et déclaré conforme aux limites applicables aux appareils numériques de classe A, selon la section 15 du règlement FCC. Ces limites visent à garantir une protection raisonnable contre les interférences nuisibles lorsque l’équipement est utilisé dans un environnement commercial.  Cet équipement produit, utilise et peut émettre une énergie de fréquence radio et, s’il n’est pas installé ni utilisé conformément au manuel d’instruction, il peut générer des interférences affectant les communications radio. L’utilisation de cet équipement dans une zone résidentielle peut causer des interférences nuisibles, auquel cas l’utilisateur sera tenu d’y remédier à ses frais.  NOTE: Des changements ou modifications apportés à cet équipement sans l’autorisation expresse de FMS Force Measuring Systems AG peuvent invalider l’agrément d’utilisation de cet équipement accordé par la FCC.  Informations sur l’exposition aux rayonnements radioélectriques Cet équipement est conforme aux limites d'exposition aux rayonnements IC établies pour un environnement non contrôlé. Cet équipement doit être installé et utilisé avec un minimum de  [50] cm de distance entre la source de rayonnement et votre corps. Ce transmetteur ne doit pas etre place au meme endroit ou utilise simultanement avec un autre transmetteur ou antenne. “
 Operating Manual EMGZ 310   10.01.2017   7  2 Quick Installation Guide In a Plug & Play configuration the set-up of the EMGZ310 and force sensor is limited to only the offset compensation and the calibration.   2.1 Preparations for Set-up 1. Read the Operation Manual of your force measuring sensors 2. Check your requirements such as: − Desired output configuration  ±10V or 0/4… 20mA − Devise Mode (Force Gauge or Volt Gauge) − Cut off frequency of low-pass filter 3. Draw the wiring diagram for your configuration (ref. to 2.6 “Wiring the Amplifier“)  2.2 Installation Procedure  1. Mount your force maesuring sensors to the machine 2. Wire the Force Measuring Sensor(s) (ref. to Fig. 1 and 2) 3. Connect the amplifier to the power supply. 4. The power supply voltage must be in the range 18 to 36V DC.  5. Perform offset compensation and calibration (ref. to chapters 3.4 to 3.7) 6. If required, do additional settings (ref. to 4 “ Parameterization”)  2.3 Mounting the Force Sensors Mounting of the force sensors is done according to the FMS Installation manual which is delivered together with the force sensors.  2.4 Installation and wiring Caution Proper function of the Tension Measuring Amplifier is only guaranteed with the recommended application of the components. In case of other arrangement, heavy malfunction can be the result. Therefore, the installation instructions on the following pages must be followed strictly.  Caution Local installation regulations are to preserve safety of electric equipment. They are not taken into consideration by this operating manual. However, they have to be followed strictly.
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   8  2.5 Block diagram   Fig. 1: Block diagram EMGZ 310      E310001e  2.6 Wiring the Amplifier One or two force sensors can be connected to the measuring amplifier. When using two force sensors, the output signal of the measuring amplifier will then correspond to the average value of the two sensors. The connection between force sensors and measuring amplifier has to be done using a 2x2x0.25mm2 [AWG 23] shielded twisted-pair cable.   Fig. 2: Wiring of EMGZ 310      E310002e
 Operating Manual EMGZ 310   10.01.2017   9    Caution Bad earth connection may cause electric shock to persons, malfunction of the total system or damage of the control unit. It is vital to ensure a proper earth ground connection.   Note The shield should be connected only to the electronic unit. On the force sensor side the shield should stay open.   Note Sensor cables must be installed separate from power lines. 2.7 Modify the wiring For an easy and quick exchange of an amplifier it is possible to exchenge the single terminal blocks as assembly. You neither have to disconnect the wiring, nore do you have to open the housing of the device.  Make sure that the amplifier is disconnected from the power supply.     Insert a screw driver with a small blade in between the terminal block and the housing on the top end of the block Using the screw driver as a lever will loosen the block from the housing. It will slide downwards   The block is now disconnected. You can slide it out the device. Insert the block on the desired position. Push the block fully into the slot. You will hear a “click” sound as the block interlocks
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   10  3  Configuring the Amplifier 3.1 Power up the EMGZ 310 1. Connect the first force sensor (see Fig.2 Wiring Diagram EMGZ310) 2. Check whether applying a force in measuring direction (in the direction of the red point) on the first sensor results in a positive output signal. If not, exchange the two signal wires of this force sensor in the terminal block (terminals 6/7). 3. If applicable, connect the second force sensor. 4. Check whether applying a force in measuring direction on the second sensor results in a positive output signal. If not, exchange the two signal wires of this force sensor in the terminal block (terminals 6/7).  3.2 Operating the EMGZ310 over the Operating Panel The EMGZ310 amplifier can go in 4 different states:  Operation State:    Amplifier operating   Offset Compensation:   Automatic procedure to offset the roller weight  Calibration State:    Procedure to calculate the amplifier gain  Parameterisation State:   Set or change parameters When using the operating panel, the user can toggle between these modes by pressing a key or a combination of keys.   3.3 Operating Panel The keys on the operation panel have the following functions:        Fig. 3: Operating Panel EMGZ 310 E310003 Key Key Function  Change the selections, increase / decrease the values or choose   Enter >0< + PARA Enter Offset Compensation Procedure. Press the two keys at the same time for longer than 3 sec. Cal + PARA Enter Calibration Procedure. Press the two keys at the same time for longer than 3 sec. PARA  Enter the Parameter Setting Mode. Press the  key for longer than 3 sec.
 Operating Manual EMGZ 310   10.01.2017   11  3.4 Change the Device Mode The EMGZ310 amplifier has two Device Modes: ● Display of Force Gauge:    Tension data are displayed in a force unit  ● Display of Voltage Gauge:  Tension data are displayed in Volt (V)  1. Press the PARA key for longer than 3 seconds, to enter the parameter-selection mode. The display changes to 2. Scroll down the parameter list with the keys   until you see on the LCD   3.  Press the key   to enter the Parameter Chang Mode 4. Select V for Voltage Gauge or kgkNN for Force Gauge 5. Confirm the parameter change with the key   6. Leave the Parametrisation by holding the key   for longer than 3 sec.   Caution Each change of the Device Mode requires a re-calibration of the system   3.5 Offset Compensation With the Offset Compensation the roller weight is compensated. It is always performed before the calibration. The Force Measuring Sensor should not be loaded while the Offset Compensation is performed.  1. Start offset compensation by pressing the two keys >0< + PARA at the same time for longer than 3 seconds.  2. The offset compensation is done automatically.(see state diagram Fig.4)   Fig. 4: State diagram offset compensation EMGZ310008e  Output  Mode
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   12  3.6 Calibration By calibrating (setting the Gain), amplifier and force sensors are adjust so that the display gives the actual tension value. There are two methods of calibrating the system. The first method uses a defined weight attached at a rope (see Fig. 5). The second method is based on a calculation process in conjunction with the FMS Calculator tool. The method with the calibration weight is preferred since it simulates the real material path in the machine and delivers the most accurate results. If the material tension cannot be simulated in the machine by the weight method (e.g. because of space restrictions), the gain can be calculated with the FMS Calculator tool. This Calculator can be down loaded from the FMS web page.    Fig. 5: Calibrating the measuring amplifier C431011e
 Operating Manual EMGZ 310   10.01.2017   13  3.7 Calibration Procedure in the Force Gauge Mode Note If you do not press any key on the operating panel the display will automatically switch to the intitial screen. You have to repeat any previously performed steps. Execute the single steps uninterrupted and prepare any required material in advance.  1. Load a rope with a defined weight corresponding to your calibration force on the roller. The roller configuration must correspond to the real configuration in the machine (wrap angle, distance of the rollers etc.). 2. Enter calibration mode by pressing the two keys Cal and PARA at the same time for longer than 3 seconds. The display changes to  3. The System Force [Sys_F] determines the measuring capability of your measuring roller. E.g. if two 500 N sensors are installed, enter 1000N. If only one 500N sensor is used, enter 500N (see chapter 4.3 “Descriptions of Parameter”, [Sys_F]). 4. Use the keys     to enter the system force. The LCD shows: 5. Press the   key to confirm your input. 6. Enter the force corresponding to your calibration weight with the  keys       7. Press the   key to confirm your input. The display changes t  8. The parameter [F@mOut] determines the force corresponding to the maximum amplifier output (10V or 20mA) (see chapter 4.3)  9. Enter the value of [F@mOut] with the keys    and confirm your input with the   key. The calibration procedure has been completed successfully.   Fig. 6: State diagram calibration EMGZ310009e N 1000 Sys_F 1000 Gain N 1000 F@mOut
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   14  1) The display shows the unit of measure that was previously selected. 2) Enter the force corresponding to your calibration weight 3) Permanent key pressing expedites the changing speed.  3.8 Calibration Procedure in Volt Gauge Mode 1. Load a rope with a defined weight corresponding to your calibration force on the roller. The roller configuration must correspond to the real configuration in the machine (wrap angle, distance of the rollers etc.). 2. Enter calibration mode by pressing the key Cal + PARA for longer than 3 seconds. The display changes to   3. Use      keys to change the output to the correct value, corresponding to the calibration force. Note The force corresponding to the calibration weight must be inputted in Volt (V) in the following calibration procedure (Volt Gauge).  4. Press the   key to confirm the input. The calibration procedure has been completed successfully.   V        7.4 Gain
 Operating Manual EMGZ 310   10.01.2017   15  4 Configuration via operating panel After having done the offset and calibration procedures the system is configured and ready for operation. Description of parameterization: 1. Press the key PARA for longer than 3 seconds, to enter the parameter-select mode. The display changes to 2. Use     keys to select the parameter to be changed. 3. Press the   key to enter the parameter changing mode. Use     keys to change the parameter value. 4. Press the   key to confirm the input. The display changes back to the parameter-select mode. 5. If needed, press     keys to select another parameter. 6. To leave the parameterization, press the key PARA for longer then 3 seconds or wait for the timeout after 30 seconds, the parameters will be saved then.   The state diagrams (Fig. 6 -7) in the following page help to navigate through the parameter setting menu.  Output
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   16    In the Parameterisation State the EMGZ 310 can be brought into the 2 modes: “Parameter Selection” and “Parameter Change”.   Fig. 6: State Diagram Parameter Setting (1st part)  E310010e 1) The display shows the unit of measure that was previously selected. 2) Enter the force corresponding to your calibration weight 3) Permanent key pressing expedites the changing speed.
 Operating Manual EMGZ 310   10.01.2017   17     Fig. 7: State Diagram Parameter Setting (2nd  part)  E310011e    4.1 Reset to Default Parameter Set   Fig. 8: Reset to Default Parameter Set  E310012e
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   18  4.2 Parameter List Parameter  Unit  Min  Max  Default  Remarks Output  [mA]  0...20 or 4...20  4…20  Current output Filter [Hz] 0.1 999.9 10.0 Noise filter Unit  [-]  N; kN; g; kg; or lb  N  System unit Sys_F  N; kN; g; kg; or lb 1  100’000  1000  System force of the roller F@mOut  N; kN; g; kg; or lb 1  100’000    Force at max. output Offset   [Digit]  –8000  8000  8000  Roller weight Gain  [-]  0.1 20.0 20.0 Amplification Mode  [-]  Volt or Force  Force  Requires new calibration
 Operating Manual EMGZ 310   10.01.2017   19  4.3 Description of Parameters  Current Output Selection  LCD: .Output Use:  This parameter selects the current output signal of the amplifier. The voltage output (+/-10V) is parallel in use Unit  Parameter Range  Selection  Default Min  Max - -  -  4..20mA 0..20mA  4..20mA  Low-pass Filter  LCD: Filter Use:  The amplifier provides a low-pass filter to prevent noise overlaying the output signal. The parameter Filter stores the cut off frequency of the filter. The lower the cut off frequency, the more sluggish the output and the display will be. The filter stabilizes the output signal in case of highly fluctuating force values. A correctly tuned filter will steady the value shown in the display.  Unit  Parameter Range  Selection  Default Min  Max Hz 0.1  999.9  -  10.0
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   20   Unit of Measure  LCD: Unit Use:  This parameter determines the unit used in the system. The label on the force sensor shows always the nominal force in N. By changing the units to lb (pounds) the whole unit system will   change from metric to imperial units.  If the Device Mode [VoltGauge] is set, this parameter is deactivated  Unit  Parameter Range  Selection  Default Min  Max - -  - N kN lb g kg N  System Force  LCD: Sys_F Use:  The System Force [Sys_F] determines the measuring capability of your measuring roller. E.g. If two 500 N sensors are installed in your measuring roller, enter 1000N. If only one 500N sensor is used, enter 500N. If sensors with sheaves are used (e.g. RMGZ-Series), the nominal force 500N must be entered.  If the Device Mode Volt Gauge is set, this parameter is deactivated Unit  Parameter Range  Selection  Default Min  Max N, kN, g, kg, lb 1) 1 100’000  -  1000 1) The display shows the unit of measure that was previously selected  Force at max. Amplifier Output  LCD: F@mOut Use:  This parameter defines what force value (N, kN, lb, g, kg) corresponds to the maximum output of the amplifier (10V or 20mA).  If the Device Mode  Volt Gauge is set, this parameter is deactivated Unit  Parameter Range  Selection  Default Min  Max N, kN, g, kg, lb 1) 1 100’000  -  1000 1) The display shows the unit of measure that was previously selected
 Operating Manual EMGZ 310   10.01.2017   21    Offset Value  LCD: Offset Use:  The offset value is used to compensate the roller weight (see chapter 3.4 “Offset Compensation”).  Unit  Parameter Range  Selection  Default Min  Max Digit -8000  +8000  -  0  Gain of Amplifier  LCD: Gain Use:  This parameter stores the value determined with the Procedure Calibration (see chapter 3.6).  With a gain value ≥10.000, the display format changes to 10.00. Unit  Parameter Range  Selection  Default Min  Max - 0.001 20.000  -  1.000  Device Mode  LCD: Mode Use:  The Device Mode determines the functionality of the amplifier. If Volt Gauge is set, force values will be displayed only in Volt (V) or mA. Force units won’t be available. Each change of the Device Mode requires a re-calibration of the system (see chapter 3.6). Unit  Parameter Range  Selection  Default Min  Max - -  -  kgkNN V  kgkNN
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   22  5 Configuration via FMS ComACTTM app If you have purchased an EMGZ310.ComACT measuring amplifier you are able to display actual readings and configure the amplifier via the FMS ComACTTM app.  You can identify the type of amp on the type label. The EMGZ310.ComACT has also the last 4 digits of its serial number printed on the front housing.    Only a single mobile device at a time can connect to an amplifier.  As soon as the connection between amp and mobile device is setup, the display of the amp will start flashing. The FMS ComACTTM app is meant as an operation aid to display actual readings and to help configure the amplifier. The settings are only stored in the amplifier and not in the mobile device.
 Operating Manual EMGZ 310   10.01.2017   23  5.1 Screenshots        Download FMS ComACTTM app  Google Play Store (Android); iTunes (IOS) Overview with all amplifiers within the transmitting range. Display of actual readings with the selected settings (unit, gain, output, etc.) “Settings” requires a password. The password is preset to “3231” and it cannot be modified.           „Calibration“ leads you to the offset and gain menu. You can perform an offset without any further input. Make sure that the web does not  material in the  Successful input and operations are confirmed Before you peform the calibration, you have to enter the applied weight.
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   24  6 Dimensions   Fig. 8: Drawing EMGZ 310 rail mount housing E310004
 Operating Manual EMGZ 310   10.01.2017   25  7 Technical Specification  Amplifier type  Digital, microprocessor controlled # of channel  1 channel for 2 sensors @ 350 Ω Sensor excitation  5VDC max. 30mA (high voltage stability) Linearity error  < 0.1% Processor cycle time  1ms Operation / parameter setting  3 keys, 4-digit LCD display (alpha-num) Resolution A/D converter  ±8192 digit (14 bit) Volt output  ± 10V min. 1.0 kΩ (12 bit) Current output  0/4...20mA max. 500 Ω (12 bit) Power supply  24VDC (18...36VDC); max. 0.2A Power consumption  <  3.0W Temperature range  0...+50°C [32...122°F] Protection class  IP20 Special functions  Buttons for calibration and offset compensation
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   26  8 Your Settings Parameter  Your Settings  Remarks Output   Current output Filter   Noise filter Unit   System unit Sys_F    System force of the roller F@mOut    Force at max. output Offset     Roller weight Gain   Amplification Mode    Requires new calibration
 Operating Manual EMGZ 310   10.01.2017   27  9 Notes
Operating Manual EMGZ310 and EMGZ310.ComACT   10.01.2017   28                       FMS Force Measuring Systems AG Aspstrasse 6 8154 Oberglatt (Switzerland) Tel. +41 44 852 80 80 Fax +41 44 850 60 06 info@fms-technology.com www.fms-technology.com FMS Italy Via Baranzate 67 I-20026 Novate Milanese Tel: +39 02 39487035 Fax: +39 02 39487035 fmsit@fms-technology.com FMS USA, Inc. 2155 Stonington Ave. Suite 119 Hoffman Estates, IL 60169 USA Tel. +1 847 519 4400 Fax +1 847 519 4401 fmsusa@fms-technology.com FMS UK Highfield, Atch Lench Road Church Lench Evesham WR11 4UG, Great Britain Tel. +44 1386 871023 Fax +44 1386 871021 fmsuk@fms-technology.com