EBR Systems 5100 Model 5100 Programmer User Manual Users manual

EBR Systems Inc Model 5100 Programmer Users manual

Users manual

     WiSE™ CRT System    INSTRUCTIONS FOR USE Programmer Model 5100 Software Release 6  For use with: • WiSE CRT Transmitter Model 4000, Software Release 3 • WiSE CRT Transmitter Model 4100, Software Release 1   EBR Systems, Inc.   480 Oakmead Parkway Sunnyvale, CA  94085 USA   www.ebrsystemsinc.com +1 408 720 1906 support@ebrsystemsinc.com     Premier Research Group Limited  1st Floor, Rubra 2, Mulberry Business Park Fishponds Road, WOKINGHAM Berkshire, RG41 2GY, United Kingdom  Australian Sponsor: EBR SYSTEMS, INC. PTY. LIMITED LELEL 21, 133 CASTLEREAGH STREET SYDNEY, NSW, 2000  LBL-03843-EN Rev. A2
     WiSE CRT System Programmer Model 5100 Instructions for Use  1  Table of Contents 1  Precautions ........................................................................................................................................... 2 1.1  Storage and Handling .................................................................................................................... 2 1.2  Co-Implanted Pacemaker and Defibrillator Compatibility ............................................................ 2 1.3  Programming ................................................................................................................................ 2 1.4  Environmental Interference .......................................................................................................... 2 2  WiSE CRT System Programmer ............................................................................................................. 3 2.1  Using the Programmer .................................................................................................................. 3 2.1.1 Configuring Connections to the Radio Module and the Tablet Computer .............................................. 4 2.1.2 Turning the Programmer On/Off ............................................................................................................. 4 2.1.3 Using the Touch Screen for Selection ...................................................................................................... 5 2.1.4 Navigating with the Main Menu .............................................................................................................. 6 2.1.5 Programmer Setup ................................................................................................................................... 7 2.2  Establishing a Communication Session with a Pulse Generator ................................................... 8 2.2.1 Programmer Status Bar............................................................................................................................ 9 2.2.2 Pulse Generator Status Bar ...................................................................................................................... 9 2.3  Reviewing and Printing Device Data ........................................................................................... 10 2.4  Programming Permanent Pulse Generator Parameters ............................................................. 11 2.5  Patient Information and Device Tracking ................................................................................... 12 2.6  Test and Evaluation Tools ........................................................................................................... 13 2.6.1 Initializing Synchronization with the Co-Implant ................................................................................... 13 2.6.2 Set-Up the Alignment of the Transmitter Model 4000 with the Acoustic Window .............................. 15 2.6.3 Set-Up the Alignment of the Transmitter Model 4100 with the Acoustic Window .............................. 18 2.6.4 Transmitter Model 4100 Speaker Test for Battery ................................................................................ 20 2.6.5 Initiating Temporary LVOO Mode .......................................................................................................... 21 3  System Evaluation during Implant and Follow-up .............................................................................. 22 3.1  Initializing the System ................................................................................................................. 22 3.2  Battery Model Selection ............................................................................................................. 22 3.3  Battery Status and End of Service ............................................................................................... 23 3.4  Battery Longevity ........................................................................................................................ 24 3.5  Using Marker Events ................................................................................................................... 25 3.6  Troubleshooting .......................................................................................................................... 26 4  Technical Specifications ...................................................................................................................... 29 4.1  Physical and Electrical Specifications .......................................................................................... 29 4.2  Parameter Specifications ............................................................................................................ 29 4.3  Programable Pulse Generator Values ......................................................................................... 30 4.4  Technical Service and Support .................................................................................................... 30 4.5  Cleaning and Disposal of Devices ................................................................................................ 30 4.6  EMC Information ......................................................................................................................... 31 5  Symbol Glossary .................................................................................................................................. 32
   2  WiSE CRT System Programmer Model 5100 Instructions for Use    1 PRECAUTIONS PRECAUTION  -  For  use  only  by  Qualified  Medical  Personnel  in  professional  healthcare  environments:    The WiSE  CRT  System  Programmer  is  intended  to  be  used  by  qualified  physicians  and  medical  technicians.    The Programmer is only to be used in professional healthcare facilities.  EBR Systems provides hands-on pre-clinical training.  Before using the system, contact EBR Systems to schedule a two-hour training session for Programmer operators.   At least annually,  review  didactic  training materials  supplied  and  presented by support  personnel from EBR Systems. Obtaining this training prior to using the Programmer is a responsibility of the operator.  The information provided in these Instructions for Use and in the WiSE CRT System Instructions for use (LBL-02744) should be considered as a required supplement for qualified and experienced medical professionals.  1.1 STORAGE AND HANDLING • Devices  should  be  stored  in  their  original  packages  in  a  dry,  clean  temperature-monitored environment. 1.2 CO-IMPLANTED PACEMAKER AND DEFIBRILLATOR COMPATIBILITY  • The WiSE CRT System is intended to be used to provide bi-ventricular pacing in conjunction with a co-implanted pacemaker or defibrillator.  Program the co-implanted pacemaker or defibrillator to deliver right ventricular pacing with the appropriate mode and timing interval settings as would be required for CRT; e.g. typically a dual chamber pacing modality that utilizes right atrial pacing unless the patient is in permanent atrial fibrillation.  1.3 PROGRAMMING • Use only the Programmer Model 5100 to attempt to communicate and program the WiSE CRT System. • The Programmer Model 5100 is only intended to be used in a professional healthcare environment.   • Program  the  parameters  of  the  co-implant  device  to  appropriate  settings  to  achieve  bi-ventricular pacing.  Do not attempt to program the co-implant pacemaker or defibrillator and the WiSE CRT System to an LV-only pacing operation.  • Check parameter  settings  and  appropriate system  operation  on  a  regular  basis, performing  a  patient follow-up at least every 3 months. • Use only the AC power adapter provided with the Programmer to maintain electrical safety. • If it is necessary to isolate the Programmer from the mains power, remove the power cord from  the external supply module. 1.4 ENVIRONMENTAL INTERFERENCE • Medical electrical equipment needs special precautions regarding electro-magnetic compatibility (EMC) and  needs  to  be  installed  and  put  into  service  according  to  the  EMC  information  provided  in  these Instructions for Use. • The  use  of  accessories  or  cables  other  than  those  supplied  with  the  Programmer  Model  5100  may result in increased emissions of or decreased immunity to electromagnetic interference
    WiSE CRT System Programmer Model 5100 Instructions for Use  3  2 WISE CRT SYSTEM PROGRAMMER The  WiSE  CRT  System  has  programmable  parameters  and  settings  that  are  adjustable  in  the  Pulse  Generator Transmitters Model 4000 and Model 4100 using the Programmer Model 5100 with a compatible software release.   Delivery of Cardiac Resynchronization Therapy relies extensively on the settings of the co-implanted pacemaker or defibrillator for the pacing mode, pacing rate, tracking intervals, etc.  The WiSE CRT System depends on sensing the right ventricular pacing electrical output of the co-implant and immediately triggering an ultrasonic transmission to activate the Electrode in the LV to provide bi-ventricular pacing for CRT.  The WiSE CRT System has a minimal set of programmable parameters.  These parameters are used to ensure adequate sensing of the right ventricular pacing output of the co-implant and to optimize the ultrasound transducer transmission to efficiently use energy from a WiSE CRT Battery.  The Programmer is packaged in a case containing the following: 1. Model 5100 radio module  with an integrated USB cable for connection to the tablet computer (REF: Model 5100 COM). 2. Model 5100 tablet computer (REF: Model 5100 TAB). 3. An AC power adapter, with country-specific power cord. 4. A recording system connection cable for connecting analog output signals for Marker Events. 5. A copy of these Programmer Model 5100 Instructions for Use and the WiSE-CRT System Instructions for Use. 2.1 USING THE PROGRAMMER The Programmer and Pulse Generator communicate using a standard radiofrequency protocol referred to as the Medical Implant Communications Service (MICS) or MedRadio.  This is an ultra-low power radio link that supports transmitting  data  for  diagnostic  or  therapeutic  operation  of  implantable  medical  devices.    This  communication standard has been integrated into the WiSE CRT System. No  magnet  is  required  to  establish  a  communication  connection  between  the  Programmer  and  the  implanted Pulse Generator.  Communication can be established by the Programmer scanning the immediate vicinity for Pulse Generators.    The  Programmer  can  only  establish  a  connection  with  WiSE  CRT  devices.    Additionally,  no  other devices using the MICS protocol can establish communications with the WiSE CRT Programmer.   IMPORTANT!  The Programmer’s radio module should be placed on a flat surface within 2 meters (6 feet) of the patient  in  order  to  establish  sufficient  communication  signal  strength  between  the  Programmer  and  the  Pulse Generator. Move the radio module closer to  the patient  to improve signal strength.   In  some circumstances the radio module may have trouble establishing a link with a Pulse Generator, change position and orientation of the radio module until a link is established.   IMPORTANT! Position the Programmer Tablet such that the power connections are accessible.   IMPORTANT! Do not touch patient and equipment at same time.
   4  WiSE CRT System Programmer Model 5100 Instructions for Use    2.1.1 Configuring Connections to the Radio Module and the Tablet Computer  The radio  module has an integrated USB-based cable for connecting to the tablet computer.   The radio  module also has a connection  for  analog  outputs  for  Marker  Events and  contains  two  USB  ports  for  the  use  of  memory modules (Figure 1).  Printed reports can be sent directly to the USB memory modules connected to this port.   IMPORTANT! Use only USB memory modules with the radio module.  Do not connect any other USB devices to the radio module.  IMPORTANT! Use only the Marker cable supplied by EBR with the Marker Events port. Do not connect any other cables or devices to this port.     Figure 1: Radio Module Connection Panel The radio module connections are located on one side of the device.  The  tablet  computer  has  a  DC  input  power  cable.  The  cable  and  the  DC  converter  are  supplied  with  the Programmer. IMPORTANT! Use only the AC power adapter provided with the Programmer to maintain electrical safety. 2.1.2 Turning the Programmer On/Off The Programmer’s computer tablet is powered from an integrated battery or by connecting the tablet to AC line power using the DC power supply.  The radio module is powered from the tablet computer via the USB connection. The  Programmer  is  powered  ON  from  a  switch  on  the  upper  right  hand  side  of  the    tablet  computer.    The Programmer is powered OFF/shutdown using an on-screen selection button (Figure 2).   To turn  the  Programmer  ON, press and hold  the power  switch  until  the  green light  within  the  power-on  switch remains on. .  The hardware and software will take 75-90 seconds to initialize.  The display indicates that the start-up  process  is  active  until  the  first  application  screen  appears.  When  a  successful  communication  between  the tablet computer and the radio module is established a solid blue indicator light on the radio module will illuminate.
    WiSE CRT System Programmer Model 5100 Instructions for Use  5  The Programmer may also be turned OFF using the power switch.  Press and hold the switch for at least 6 seconds and  then  release  the  switch.    The  Programmer  will  then  turn  off  automatically  and  the  green  light  within  the power-on switch will be extinguished.       Figure 2: Programmer Menu Located on the left side of the display screen To turn the Programmer OFF, first select the [SHUTDOWN] button on the Main Menu and confirm the shutdown selection in the pop-up dialog box by selecting the [OK] button.  The Programmer will then turn off automatically. The power-on indicator will remain illuminated for a short while after the screen darkens.  Wait until the indicator is off before pressing the ON switch again. The  Programmer  will  display  a  Shutdown  Confirmation  or Disconnect  Confirmation  dialog  box  (Figure  3).    If the Pulse  Generator  has  been  programmed  to  an  OFF  or  SENSE  ONLY  mode  setting,  then  this  dialog will  notify  the operator that the operational setting is not programmed to pace.  If the Pulse Generator has been programmed to the RV Synchronous mode, then the Transmit Level and Pulse Width will be shown in the dialog box.  Figure 3: Disconnect Confirmation Dialog Box  2.1.3 Using the Touch Screen for Selection User selections on the Programmer are made on a touch-screen display.   Selectable areas of the screen appear as labeled graphic buttons, icons, scrolling up/down arrow buttons, selectable list items, or text entry areas. Buttons,  icons,  list  items,  and  active  screen  areas  for  text  entry  can  be activated  by  touching  the  screen  with  a finger.      With  each  successful  button  contact  an  audible  sound  is  produced  to  further  discern  the  response.    A “click” sound indicates the button action has been initiated.  A “boop” sound indicates that the button action has not been initiated and an error has occurred in the selection of the button. Buttons that indicate the current active selection are highlighted in blue.  Buttons are selectable (active) if they are highlighted in dark gray.  Buttons are not selectable if they are shown in light gray.  The [PROGRAM] button is an exception when it is active; it is always displayed in an orange color. On-screen values that are displayed in blue or orange are transferred to the Pulse Generator when the [PROGRAM] button is selected.
   6  WiSE CRT System Programmer Model 5100 Instructions for Use    2.1.4 Navigating with the Main Menu The Main Menu is located along the left side of the display screen (Figure 4).     ← Search for/Connect to Pulse Generators within radio range or disconnect from currently connected Pulse Generator ← Store patient and device information in the Pulse Generator   ← Initialize the Pulse Generator for sensing RV pacing spikes  ← Perform temporary asynchronous pacing in the LV   ← Select and program Pulse Generator pacing parameters ← Align the Transmitter within the acoustic window  ← View and print data and settings of the Pulse Generator   ← Configure the Programmer e.g. Marker Event outputs ← Turn off the Programmer  Figure 4: Main Menu Located on the left side of the display screen
    WiSE CRT System Programmer Model 5100 Instructions for Use  7  2.1.5 Programmer Setup  Figure 5: Programmer Setup Screen Accessible by pressing the [SETUP] button from the main menu (Figure 4) Programmer setup can be accomplished via the Programmer Setup Screen (Figure 5). The time and date are adjusted by selecting the associated buttons. By selecting the associated buttons the local language for the buttons and messages, the brightness of the display, and the volume level of the Programmer’s speaker may be adjusted. The analog outputs for the  Marker  Events are adjusted by selecting the associated buttons.  To test the  Marker Events  output  levels,  select  the  [CALIBRATE  MARKER  OUTPUT]  button.  For  additional  information  about  using Marker Events and the Programmer output signal when the [CALIBRATE MARKER OUTPUT] button is pressed, refer to section 3.5, Using Marker Events Figure 24.
   8  WiSE CRT System Programmer Model 5100 Instructions for Use    2.2 ESTABLISHING A COMMUNICATION SESSION WITH A PULSE GENERATOR  Figure 6: Connect Pulse Generator Screen Accessible by pressing the [CONNECT] button from the main menu (Figure 4) Select the [SEARCH] button to initiate a radio module scan for WiSE CRT Pulse Generators (Figure 6).  The scan may take up to 60 seconds and the progress bar will fill as the radio module scans for devices. The radio module’s indicator light will blink green while scanning for devices and turn on green if a device is identified.  All devices that are found by the scan are listed by Serial Number and by Patient Name, if the Patient Name has been programmed into the Pulse Generator. Select the appropriate device from the list.  Select the [CONNECT] button on the bottom of the window to establish a session; select the [DISCONNECT] button to stop a session and select a different Pulse Generator from the list.  If the radio connection between the Programmer and the Pulse Generator stops communication the No Communication with the Pulse Generator dialog box will be displayed.   To start a new programming session with the same Pulse Generator select the [RECONNECT] button in the dialog box.   Select the [SEARCH] button in the dialog box to return to the Connection Screen.
    WiSE CRT System Programmer Model 5100 Instructions for Use  9   2.2.1 Programmer Status Bar The Programmer Status Bar is continuously displayed at the top of the screen and contains a radio signal strength indicator for the link between the radio module and the Pulse Generator and a section that updates Marker Events in real-time as an indicator of Pulse Generator operations.  Figure 7: Programmer Status Bar Located at the top of the display screen  2.2.2 Pulse Generator Status Bar The Pulse Generator Status Bar is continuously displayed  on  the  right  hand  side  of  the screen so  that the latest settings and interactions with the Pulse Generator may be viewed at any time (Figure 8).  The Status Bar contains a section to report the estimated remaining battery life, and a section to report any status conditions such as device resets or end of service life flags.   Figure 8: Pulse Generator Status Bar Located at the top of the display screen
   10  WiSE CRT System Programmer Model 5100 Instructions for Use    2.3 REVIEWING AND PRINTING DEVICE DATA  Figure 9: Report Data Screen Accessible by pressing the [REPORT] button from the main menu (Figure 4) After connecting to a Pulse Generator,  the  Programmer  will interrogate  the  device  for  its settings and  data  and display the Report Data screen (Figure 9).  Select [EXPAND REPORT] to view and print an expanded set of collected counter and statistic data.  To review the complete settings and data, use the scroll bar to adjust the window view.      To print the settings and data select the [PRINT REPORT]  button. Printed reports can be sent directly to the USB memory modules connected to USB port on the radio module.   Select [CLEAR FAULT] to reset errors that may have been reported.   The  collected  data  in  the  device  may  be  cleared  after  selecting [EXPAND  REPORT].   Select  the  [RESET  HISTORY] button to clear/reset the counter and statistic data.  The following data will be cleared: • Total number of LV Pace Attempts • Total number of RV Pacing Spike Detections  • Query statistics including targeting transmission data  • RV pacing spike detection data • Electrode location and distance statistics
    WiSE CRT System Programmer Model 5100 Instructions for Use  11  2.4 PROGRAMMING PERMANENT PULSE GENERATOR PARAMETERS  Figure 10: Mode/Output Screen Accessible by pressing the [MODE/OUTPUT] button from the main menu (Figure 4) The Pulse Generator may be programmed to one of three operational modes (Figure 10): • RV Synchronous – This is the pacing mode for the Pulse Generator.  In this mode, the Pulse Generator senses RV pacing electrical outputs from the co-implanted pacemaker or defibrillator and immediately triggers an ultrasonic transmission to the Electrode to pace the LV for bi-ventricular pacing. • Sense Only – This is a sensing mode that disables LV pacing but allows the device to track, count, and transmit RV Pacing Spike Detection marker events.  • Off – This  mode  disables  pacing and  sensing  in  the  device.    This  is  the  factory  default  setting  of  the device.  In  RV  Synchronous  mode  the  Pulse  Generator  may  be  programmed  to  fixed  settings  for  the  transmit  level  and pulse width.  The pulse width is analogous to the pacing pulse width in a conventional pacing pulse.  The setting of this pulse width determines the delivered pulse width for the LV pace pulse.  The transmit level does not correlate to  a  specific electrical pacing  voltage  amplitude.   The transmit level is related to the intensity of the ultrasonic output field and is indirectly correlated to the delivered pace pulse amplitude.   The  Pulse  Generator  may  be  programmed to  one  of  three targeting  modes.      Targeting  is  essential  for  efficient transfer of energy from the Pulse Generator to the Electrode.   Targeting sends a series of short ultrasonic pulses toward the last known region or location of the Electrode and senses the Electrode’s electrical output response.  A sufficient electrical response indicates that the Pulse Generator is focused on the Electrode.
   12  WiSE CRT System Programmer Model 5100 Instructions for Use    • Nominal – This is the default targeting mode in the Pulse Generator.  This mode uses the recent history of  electrical  responses  to  the  targeting  pulses  to  gauge  the  systems  effectiveness  on  targeting  the Electrode. • Reduced  Global  Search –  This  targeting  mode  is  used  in  circumstances  where  the  system  would otherwise  use  a  significant  number  of  targeting  pulses  to  unsuccessfully  locate  the  Electrode.    This mode is used to conserve energy usage in the system.   • Alternate –  The  Alternate  targeting  mode  is  used  in  circumstances  where  the  electrical  output  is somewhat  variable.    This  algorithm  does  not  rely  on  prior  history  related  to  the  amplitude  output signals.   To program the Pulse Generator parameters: 1. Select the [RV SYNCHRONOUS], [SENSE_ONLY], or [OFF] button. 2. If [RV SYNCHRONOUS] is selected then select the Transmit Level and the Pulse Width settings for the device. 3. If  [RV  SYNCHRONOUS]  is  selected  then  select  the  [NOMINAL],  [REDUCED  GLOBAL  SEARCH],  or [ALTERNATE] button. 4. Select the [PROGRAM] button. There are  3 set capture threshold buttons to the right of the Transmit Level and Pulse Width selections.  These are used to record values onto  the  report that  have been determined to be the capture threshold  for  the  Transmit Level and Pulse Width as tested  in various postures, i.e. supine, sitting, and/or standing.   Select the icon or the value display area to set these values to those that are the currently active in the Transmitter.  These values are not  stored  in  the  Transmitter  but  do  appear  on  the  print  report  as  an  indication  that  a  capture  threshold  was determined during the follow-up session. 2.5 PATIENT INFORMATION AND DEVICE TRACKING  Figure 11: Enter Patient Information Screen Accessible by pressing the [PATIENT] button from the main menu (Figure 4) The  tracking  of  patients  and  devices  is  an  important  safety  and  regulatory  requirement.    Information  must  be entered into the Pulse Generator using the Programmer (Figure 11), and an Implanted Device Tracking Registration Form must be completed and returned to EBR Systems, Inc.
    WiSE CRT System Programmer Model 5100 Instructions for Use  13  To enter a Patient Information data field, select the data field to highlight it and then use the on-screen keyboard to  enter alphanumeric  information.    Obtain  the  serial  number from the packaging,  registration cards, or patient records and check for its correct entry.   Check other data entry information against patient records.   Selecting the Battery Model data field will display the Select Battery Model dialog box.  Select [PROGRAM] to communicate and confirm the changes to the Pulse Generator.   An  Implanted  Device  Tracking  Registration  Form  and  a  Patient  Device  Identification  Card  are  included  in  each device  package  and  should  be  completed  promptly  at  the  time  of  the  implant  procedure.  The  tracking  form  is generic and allows for 1-4 devices to be registered on a single form.  One copy of the Registration Form should be kept  with  the  procedure/patient  record  and  one  copy  must  be  returned  to  EBR  Systems.    The  Patient  Device Identification Card is provided to the patient.  Self-adhesive serial number labels are included in each package and can be conveniently applied to the Registration Form and to patient records as may be required. Implanted Device Tracking Registration Forms should be returned or faxed to EBR Systems, Inc. Please refer to the cover page of this document for EBR Systems’ address and fax number.  2.6 TEST AND EVALUATION TOOLS 2.6.1 Initializing Synchronization with the Co-Implant The WiSE CRT System relies on detecting RV pacing electrical outputs (spikes) from the co-implanted pacemaker or defibrillator.  The WiSE CRT Pulse Generator must be initialized with specific right ventricular pacing pulse width from the co-implanted device to reliably discriminate right ventricular pacing from atrial pacing.  This must be done at implant and may be re-initialized during follow-up.  Figure 12: Co-implant Synchronization Screen Accessible by pressing the [CO-IMPLANT SYNCHRONIZATION] button from the main menu (Figure 4) The process requires the operator to interact with both the co-implant device and with the Pulse Generator.  More specifically, the co-implant device must be programmed to specific values for the RV pacing pulse width and the atrial pacing pulse width.
   14  WiSE CRT System Programmer Model 5100 Instructions for Use    Select the [INITIALIZATION] button after the process has started to start over. On-screen directions are provided in the text area to set up the co-implant parameters and verify the initialization (Figure 12): 1. Set the co-implant to provide RV pacing.   2. Set the co-implant ventricular and atrial pacing pulse widths within the required bounds.  The WiSE CRT Transmitter  must  be  initialized  with  the  pacing  pulse  widths being  used  by  the  co-implanted  device.  The right ventricular pacing pulse width of the co-implant pacemaker must be programmed at or above  0.35 ms.   In dual chamber modes, where atrial pacing may be expected, the right atrial pacing pulse width  of  the  co-implant  pacemaker  must  be  programmed  at  least  0.3  ms  above  or  below  the  right ventricular  pulse  width.      A  setting  for  right  atrial  pulse  width  above  1.0ms  is  most  effective  for  the Pulse Generator to distinguish atrial pulses from ventricular pulses.   3. Select  the  [PROCEED]  button  to  confirm  that  the  co-implant  is  providing  RV  pacing.    The  Enter  Co-Implant Pacing Pulse Widths dialog box (Figure 13) will appear for the operator to specify the RV pacing pulse  widths  being  used.    Select  the  scroll-up  and  scroll-down  buttons  to  enter  the  RV  pacing  pulse width.  RV pulse width value selections may not exactly match the resolution value of the co-implant.  In this case select the value that is the closest to the actual programmed value in the co-implant.  Select [ACCEPT] to use the pulse width value and begin the initialization.  Select [CANCEL] to start the process from the beginning.  Figure 13: Enter Co-Implant Pacing Pulse Widths dialog box 4. Use the Marker Events to confirm that the Pulse Generator is sensing RV pacing electrical outputs.  The RV Pace Detect icon should be flashing in the Marker Event section of the Pulse Generator Status Bar and analog output marker events should be visible on the recording system.  Confirm that the marker events are synchronized to RV pacing with the ECG on the recording system.   5. The progress bar will reach 100% when synchronizing with the co-implanted device completes.  If the Pulse Generator is  unable to synchronize with the co-implant, the user will be shown a message “RV Pacing Spike Initialization unsuccessful” and can press the [PROCEED] button to start over.   Once co-implant  synchronization  has  succeeded  the  user  can  select  the  [PROCEED]  button  to  confirm  to  the Pulse Generator that it is tracking RV pacing electrical outputs.  Select the  [GO BACK] button to start the process over.
    WiSE CRT System Programmer Model 5100 Instructions for Use  15  2.6.2 Set-Up the Alignment of the Transmitter Model 4000 with the Acoustic Window To pace the LV with the Electrode an ultrasonic pulse from the Transmitter is emitted and targeted to the location of the Electrode.  For this to be done efficiently, the Transmitter Model 4000 must be implanted above an acoustic window (a lung-free, bone-free path) to the Electrode.  The Transmitter is composed of transducers arranged as an array with active dimensions of 2.4 cm by 3.2 cm on one side of the enclosure.  The acceptability and optimization of alignment of the Transmitter’s transducer array can be done at implant and at follow-up. PRECAUTION – Pacing in the LV may be inhibited while set up of the acoustic window is operating.  Figure 14: Transmitter Model 4000 Setup Acoustic Window Screen Accessible by pressing the [ACOUSTIC WINDOW] button from the main menu (Figure 4) The setup of the acoustic window evaluates the position of the Transmitter array relative to the Electrode position.  The  on-screen  display  shows  a  graphic  of  the  measured  sensing  and  efficiency  values  in  real  time  as  the Transmitter locates the Electrode (Figure 14).  This is useful for monitoring the effect of the current selection being tested.   The following characteristics are visible in the graphic display: • The array elements and position currently programmed in the Transmitter are illustrated in green. • The array elements and position that are currently  being tested are outlined with  a tan border.   The number of elements and position to be tested is adjustable. • Information  related  to  the general  location  and  distance  to  the Electrode, relative to the face of the Transmitter, is continuously displayed as a blue dot  on the graphic and as an ACTUAL distance  value above the graphic. • The  graphic  depicts  a  circular  representation  of  the  full  180-degree  hemisphere  beneath  the  Pulse Generator, with tick marks along the x- and y-axes indicating +/- 30 and 60 degrees. The targeting zone for locating the Electrode is shown by a blue circle, the center of which is shown by a blue “x”. This “x” and  the  corresponding  target  zone  may  be positioned  anywhere  within  the 180-degree  hemisphere. The blue dot is expected to be within the target zone and generally near its center. • A data area below the graphic display indicates either sensed signal levels or efficiency measurements related  to  the  alignment  of  the  Transmitter  with  the  Electrode.    The  signal  data  are  the  electrical amplitudes  output  by  the  Electrode  and  sensed  by  the Transmitter  in  response  to  a  targeting  pulse.  They  are  used  to  determine  whether  the  Transmitter  is  sufficiently  focused  on  the  Electrode.    The Efficiency  is  a  relative  measure  of  improvements  in  energy  transfer  when  compared  to  other  tested
   16  WiSE CRT System Programmer Model 5100 Instructions for Use    values.  Both the signal and efficiency data are displayed as a moving histogram.  Select [EFFICIENCY] button or select [SIGNAL] button to the left of the graphic to alternate the data display. • One or more location sensing channels being used by the system to locate the Electrode are highlighted by  yellow  dots  in  the  graphic  display.    These  are  in  the  general  location  for  the  sense  electrodes arranged on the image of the Transmitter.  If the system is using the battery - battery location sensing channel this will be shown as two yellow dots on the cable graphic.  The sensing channel configuration being  used  is also  displayed  to  the right  of  the  graphic  and  as  described  below  can  be  selected  and programmed.    Select  the  [LOCATION  SENSING]  button  to  open  a  dialog  box  for  selecting  sensing channel configurations.  • Data values related to the sensing signals used by the system to locate the Electrode are displayed to the  left  of  the  graphic.    EFFICIENCY  indicates  a  relative  measure  of  the  energy  transfer  efficiency.  SIGNAL indicates the relative strength of the sensing signal used to locate the Electrode. NOISE is the background electrical signal noise present in the sensing signal. THRESHOLD is a function of the noise and  a  statistical  history  of  the  noise.    The  system  uses  the  sensed  signals  above  the  threshold  to determine whether the ultrasound pulse is targeted on the Electrode.  The higher the signal is above the noise, the more statistically likely it is that the targeting is accurate.  The function can be modified by selecting a THRESHOLD TUNING setting in the pull-down menu. At  implant  or  at  discharge, set up  of acoustic window  alignment  with  the  Transmitter’s transducer array  should always be completed. At follow-up, set up of acoustic window alignment with the Transmitter’s transducer array may be used to optimize the service life of the battery. To use the Programmer for this assessment: 1. Adjust  the  Acoustic  Window  Size,  i.e.  the  number  of  the  array  elements,  to  test  by  selecting  the up/down  buttons  for  the  Width  and  Length  of  active  elements  to  be  used.    As  this  is  adjusted  the graphic display will outline the elements to be tested with a white/tan border. 2. Adjust  the  position  of  the  array  section  being  tested  by  selecting  the  Position  arrow  buttons.    The position of the array will adjust Up, Down, Left, Right within the graphic display. 3. Position  the  “x”  representing  the  center  of  the  target  zone  at  the  expected  Electrode  location,  by touching  the  desired  center  of  the  target  zone  with  a  finger  or  stylus.  If  the  Electrode  has  been previously  found  and  tracked,  a  red  ellipse  will  be  displayed  indicating  the  average  location  of  the electrode.  The  “x”  may  be  placed within  that  ellipse manually  by  tapping  at the new  location,  or  by pressing [CENTER GSC] which will automatically adjust it to the center of the ellipse. 4. Select a TARGET range from the Transmitter to the Electrode’s location by selecting a DISTANCE using the pull-down menu.  The DISTANCE is used by the system to focus the Pulse Generator’s ultrasound output.  The  measured  range  is  displayed  as  the ACTUAL  in  this screen area.      The  selections  for  the DISTANCE are in centimeters of separation between the Electrode and the Transmitter face and include [Close (< 4)], [Near (4 – 5)], [Typical (5 – 8)], and [Far (> 8)]. The DISTANCE is normally selected based on the ACTUAL distance reported.  If the Electrode is at a large angle, performance may be improved by selecting  a  DISTANCE  that  is  greater  than  the  ACTUAL  distance  displayed.  The  DISTANCE  pull-down menu  also  contains  a  setting  labeled  OFF.  Selecting  OFF  may  be  preferable  for  especially  large Transmitter-to-Electrode angles (total angle > 45˚). The TOTAL ANGLE is displayed to the upper right of the graphic display area. 5. Select the sensing distance [LIMIT] button to open the Select Distance Limit dialog box (Figure 15) to select the sensing range. When the CLOSE distance is set to AUTO, the system automatically calculates the minimum sense blanking distance. In many cases, targeting can be improved by manually selecting a CLOSE sensing distance. Typically choose a CLOSE distance that is at least 2cm less than the expected minimum Electrode distance. The FAR distance is typically set to at least 2cm farther than the expected maximum Electrode distance, up to a  maximum of 16.9cm. Select [ACCEPT] or [CANCEL] to close the dialog box.
    WiSE CRT System Programmer Model 5100 Instructions for Use  17   Figure 15: Select Distance Limit dialog box  6. Select the system’s sensing configuration by selecting a LOCATION SENSING CHANNEL from the Select Location Sensing dialog box (Figure 16). The configuration may be selected from a single sensor or from dual  sensor  pairs.    These  pairs  represent  combinations  of  sensing  electrodes  on  the  exterior  of  the Transmitter and Battery. Up to four selections can be made and can be mixed between single and dual sensors.    The  system  will  dynamically  rotate  the  use  of  these  sensors.    The  rotation  occurs  during operation at times when the system determines that sensing is inconsistently targeting the Electrode.  Select one to four buttons between the Single and Dual columns. Alternatively, the [OPTIMAL SINGLE] or [OPTIMAL DUAL] selections are an automated feature which the system uses to self-select a sensing configuration.  Select [ACCEPT] or [CANCEL] to close the dialog box.  Figure 16: Select Location Sensing dialog box 7. Select the [ENABLE] button to initiate the test process.   8. Monitor the graphic and the data display areas to assess the test performance. A beat-by- beat update of  the  Signal  or  Efficiency  data  will  be  displayed  in  the  bar  graph  area  scaled  between  0  and  the maximum value measured. The average of all values will be displayed as a number.  Also, beat by beat, the data values related to the sensing signals will be updated.  The LOCATION SENSING CHANNEL with the best efficiency should be used to improve system performance.  Monitor the Marker Events during the testing.  The display will show the counts of RV Pace Detect events and Transmit Events next to the flashing icons. 9. During  the  test  the  [PROGRAM]  button  will  be  enabled.    Select  [PROGRAM]  to  permanently  set  the Pulse Generator to the array size, position, distance limits, and sensing configuration being tested. You can only program the settings while the test is in operation. 10. The  test  process  will  continue  until  cancelled,  until  the  communication  with  the  Pulse  Generator  is interrupted, or until a Main Menu button is selected to change to another screen. Select the [CANCEL] button to end the test.  The Pulse Generator will return to its programmed mode of operation.
   18  WiSE CRT System Programmer Model 5100 Instructions for Use    2.6.3 Set-Up the Alignment of the Transmitter Model 4100 with the Acoustic Window To pace the LV with the Electrode an ultrasonic pulse from the Transmitter is emitted and targeted to the location of the Electrode.  For this to be done efficiently, the Transmitter must be implanted above an acoustic window (a lung-free, bone-free path) to the Electrode.  The Transmitter Model 4100 is composed of transducers arranged as an array with active dimensions of 0.8 cm by 2.4 cm on one side of the enclosure.  The testing and optimization of alignment of the Transmitter’s transducer array can be done at implant and at follow-up. PRECAUTION – Pacing in the LV may be inhibited while set up of the acoustic window is operating.   Figure 17: Transmitter Model 4100 Setup Acoustic Window Screen Accessible by pressing the [ACOUSTIC WINDOW] button from the main menu (Figure 4) The  set-up  of  the  acoustic  window  evaluates  the  position  of  the  Transmitter  array  relative  to  the  Electrode position.  The array elements that are currently being tested are not adjustable.  The entire array area of 0.8 cm by 2.4  cm  is  active  during  the  testing.      The  on-screen  display  shows  the  location  and  sensing  signal  values  as measured in real time as the Transmitter locates the Electrode (Figure 17).  This is useful for monitoring the effect of the current selections being tested.   The following characteristics are visible in the graphic display: • Information  related  to  the general  location  and  distance  to  the Electrode, relative to the face of the Transmitter,  is  continuously  displayed  as  a  blue  dot  on the graphic  and  as  actual distance and  angle values in the Results next to the graphic. • The  graphic  depicts  a  circular  representation  of  the  full  180-degree  hemisphere  beneath  the  Pulse Generator, with tick marks along the x- and y-axes indicating +/- 30 and 60 degrees. The targeting zone for locating the Electrode is shown by a blue circle, the center of which is shown by a blue “x”. This “x” and  the  corresponding  target  zone  may  be positioned  anywhere  within  the 180-degree  hemisphere. The blue dot is expected to be within the target zone and generally near its center. • One or more location sensing channels being used by the system to locate the Electrode are highlighted by  yellow  dots  in  the  graphic  display.    These  are  in  the  general  location  for  the  sense  electrodes arranged  on  the  image  of the  Transmitter.  The location  sensing  channel  configuration being used is also displayed below the graphic and as described below can be selected and programmed.
    WiSE CRT System Programmer Model 5100 Instructions for Use  19  • Data areas to the left of the graphic display indicates sensed signal levels related to the alignment of the Transmitter with the Electrode.  The Signal (μV) data  are  the  electrical amplitudes  output  by the Electrode as sensed by the Transmitter in response to a targeting pulse and are used to indicate the relative strength of the sensing signal used to locate the Electrode. Noise is the background electrical signal noise present in the sensing signal. Threshold is a function of the noise and a statistical history of the  noise.    They  are  used  to  determine  whether  the  Transmitter  is  sufficiently  focused  on  the Electrode.  The signal data itself is displayed as an absolute value and as a moving average.  The system uses the sensed signals above the threshold to determine whether the ultrasound pulse is targeted on the  Electrode.    The  higher  the  signal  is  above  the  noise,  the  more  statistically  likely  it  is  that  the targeting is accurate.  The function can be modified by selecting a Threshold setting in the  pull-down menu. At  implant  or  at  discharge, set up  of acoustic window alignment  with  the  Transmitter’s  transducer array should always be completed.  To use the Programmer for this assessment: 1. Position  the  “x”  representing  the  center  of  the  target  zone  at  the  expected  Electrode  location,  by touching  the  desired  center  of  the  target  zone  with  a  finger  or  stylus.  If  the  Electrode  has  been previously  found  and  tracked,  a  red  ellipse  will  be  displayed  indicating  the  average  location  of  the electrode.  The  “x”  may  be  placed within  that  ellipse manually  by  tapping  at the  new  location,  or  by pressing [CENTER GSC] which will automatically adjust it to the center of the ellipse. 2. Select a targeting range from the Transmitter to the Electrode’s location by selecting a  Distance (cm) using  the  pull-down  menu.    The  Distance  is  used  by  the  system  to  focus  the  Pulse  Generator’s ultrasound output. The measured range is displayed as an actual value in the Results screen area.   The selections for the Distance is in centimeters of separation between the Electrode and the Transmitter face and include [Close (< 4)], [Near (4 – 5)], [Typical (5 – 8)], and [Far (> 8)]. The Distance is normally selected based on the measure value reported.  If the Electrode is at a large angle, performance may be improved by selecting a Distance that is greater than the measured distance displayed. The Distance pull-down menu also contains a setting labeled “Off”. Selecting “Off” may be preferable for especially large Transmitter-to-Electrode  angles (total angle >  45˚).  The Total Angle is displayed  to  the Distance measured value. 3. Select the Distance Limit display area to open the Select Distance Limit dialog box (Figure 18) to select the  sensing range. When the CLOSE distance is set to AUTO,  the system automatically  calculates the minimum sense blanking distance. In many cases, targeting can be improved  by manually selecting a CLOSE sensing distance. Typically choose a CLOSE distance that is at least 2cm less than the expected minimum Electrode distance. The FAR distance is typically set to at least 2cm farther than the expected maximum Electrode distance,  up to a maximum of 16.9cm. Select [ACCEPT] or [CANCEL] to close the dialog box.  Figure 18: Select Distance Limit dialog box
   20  WiSE CRT System Programmer Model 5100 Instructions for Use    4. Select the Location  Sensing configuration display area to open the Select Location Sensing dialog box (Figure 19). The configuration may be selected from a single sensor or from dual sensor pairs.  These pairs represent combinations of sensing electrodes on the exterior of the Transmitter and Battery. Up to  four  selections can  be  made and  can be mixed between single and dual sensors.  The  system  will dynamically rotate the use of these sensors.  The rotation occurs during operation at times when the system  determines  that  sensing  is  inconsistently  targeting  the  Electrode.    Select  one  to  four  buttons between the Single and Dual columns. Alternatively, the [OPTIMAL SINGLE] selection is an automated feature which the  system uses to  self-select a sensing configuration.   Select [ACCEPT]  or [CANCEL] to close the dialog box.  Figure 19: Select Location Sensing dialog box 5. Select the [ENABLE] button to initiate the test process.   6. Monitor  the  data  display  areas  to  assess the test performance.  A  beat-by- beat update  of the Signal level  value  will  be  displayed.  The  average  of  all  values  will  be  displayed  as  a  number.    Monitor  the Marker  Events  during  the  testing.    The  display  will  show  the  counts  of  RV  Pace  Detect  events  and Transmit Events next to the flashing icons. 7. During  the  test  the  [PROGRAM]  button  will  be  enabled.    Select  [PROGRAM]  to  permanently  set  the Pulse Generator to the distance limits and sensing configuration  being  tested. You can only program the settings while the test is in operation. 8. The  test  process  will  continue  until  cancelled,  until  the  communication  with  the  Pulse  Generator  is interrupted, or until a Main Menu button is selected to change to another screen. Select the [CANCEL] button to end the test.  The Pulse Generator will return to its programmed mode of operation. 2.6.4 Transmitter Model 4100 Speaker Test for Battery  An  audible  notification  exists  in  the  Transmitter Model  4100  to  indicate  the  progression of the depletion  of the implanted battery.  The audible notification may be tested from the Setup Acoustic Window screen (Figure 17) to determine whether the notification can be perceived by the patient: 1. Select the Speaker Test button.  The Speaker Test dialog box (Figure 20) will be displayed for 8 seconds while the speaker in the Transmitter emits the audible notification sequence (five half-second audible tones, separated by one second silence between tones).  Figure 20: Transmitter Model 4100 Speaker Test dialog box
    WiSE CRT System Programmer Model 5100 Instructions for Use  21  2.6.5 Initiating Temporary LVOO Mode The Pulse Generator may be programmed to temporarily deliver asynchronous pacing to the LV (Figure 21). PRECAUTION – In Temporary LVOO mode the Pulse Generator delivers pacing without synchronizing its output with  the  co-implanted  device.    Ensure  that  the  co-implanted  rate  is  lower  than  the  LV  rate  to  avoid competitively  pacing  the  ventricles.    Ensure  that  any  co-implant  anti-tachy  therapies  are  disabled  or reprogrammed such that Temporary LVOO pacing is not sensed by the co-implant’s detection algorithms.   Figure 21: Temporary LVOO Mode Screen Accessible by pressing the [TEMPORARY LVOO MODE] button from the main menu ((Figure 4) To program the Pulse Generator to temporary LVOO mode: 1. Select the Pacing Rate, Transmit Level and Pulse Width. 2. Select the [ENABLE] button. Select [CANCEL] button to terminate the temporary pacing and return to the programmed  mode.  If at any time the communication link between the Programmer and Pulse Generator is lost, the temporary  LVOO mode will be automatically cancelled by the Pulse Generator.
   22  WiSE CRT System Programmer Model 5100 Instructions for Use    3 SYSTEM EVALUATION DURING IMPLANT AND FOLLOW-UP 3.1 INITIALIZING THE SYSTEM The Battery and Transmitter  must be within their pockets, and the pockets flushed with sterile saline to remove air,  in  order  to  initialize  the  system.    The  Electrode  and  the  Pulse  Generator  are  not  active  until  the  Pulse Generator  has  been  initialized  using  the  Programmer.      Once  the  Electrode  has  been  implanted,  initializing  the Transmitter is important for optimal performance of the system.   To complete the initialization of the System:   1. Establish a Programming session (a communication link) with the Pulse Generator. 2. Program patient information into the Pulse Generator. 3. Ensure that the co-implant is set to a mode to promote RV stimulation, and that the RV pacing pulse width is set to the value entered during co-implant synchronization. 4. Set up sensing for Co-Implant Synchronization.     5. Set up the acoustic window so that it is optimized for the alignment of the transmitter.  6. Assess  the  EKG  for  bi-ventricular  pacing.    Observe  changes  in  the EKG  to  confirm  that  the  system  is tracking the co-implant and LV pacing is being achieved with the system. 3.2 BATTERY MODEL SELECTION For the Pulse Generator to estimate the service life of the Battery module it must know which Battery Model has been connected.  This is done  at the time of Battery  implant.  At the first connection of the Programmer  to the Pulse Generator after Battery connection, the Programmer will display a Select Battery Model dialog box.  Select the Battery Model that is connected to the Pulse Generator and then select the [PROGRAM] button.
    WiSE CRT System Programmer Model 5100 Instructions for Use  23  3.3 BATTERY STATUS AND END OF SERVICE The energy usage is monitored by the Pulse Generator to estimate  and  report  on  the service life of  the Battery module.  Energy usage is based on the individual settings and specific power usage of the device. The Programmer provides an end of service (EOS) date on the Programmer screen and on the  printout as highlighted in Figure 22. Regular device checks are a critical component of care and the EOS date must be reviewed at each device check. At EOS, the transmitter will no longer provide ultrasound transmissions. Replace the Battery module immediately if the Programmer indicates EOS has been reached.  The EOS date is used for determining the Recommended Replacement Time (RRT). The RRT is 3 months before the EOS date reported on the screen and printout. If the EOS date is 3 months or less on the day of the device check, then  the  patient  should  be  scheduled  for  a  procedure  to  replace  the  Battery.    The  Battery  is  replaced  without removing  or  replacing  the  Transmitter.  In  addition  to  the  RRT,  the  battery  voltage  is  measured  by  the  Pulse Generator and the remaining battery capacity is estimated. These are also reported as highlighted in Figure 22. When the battery voltage for the implanted battery reaches 2.6V, the implanted Pulse Generator will generate a patient  notification  consisting  of  twenty  (20)  half-second  audible  tones  separated  by  one  second  of  silence between tones.  This patient alert notification repeats every eight (8) hours, until either the battery is replaced or the  system  is  programed  with settings that extend  battery longevity (e.g.  lower  transmit levels and  wider  pulse widths  could  be  used  to  increase  battery  life).  The  audible  notification  may  be  tested  from  the  Setup  Acoustic Window  screen  (Figure  17)  to  determine  whether  the  audible  notification  can  be  perceived  by  the patient  (see section 2.6.4) IMPORTANT!  Audible  patient  notification  for  Battery  voltage  level  is  only  available  with  the  Transmitter  Model 4100. The  Pulse  Generator  Status  bar  and  the  Report  Data  area  will  contain  indications  of  the  battery  status.    The measured battery voltage is reported, an estimate of the remaining battery capacity is provided, and an estimated date of end of service is reported.    Figure 22: Report Screen with illustrative highlights included showing Battery status Screen is accessible by pressing the [REPORT] button from the main menu (Figure 4)
   24  WiSE CRT System Programmer Model 5100 Instructions for Use    3.4 BATTERY LONGEVITY WARNING – Early Battery Depletion: Higher than expected energy requirements for transmitting ultrasound to activate the Electrode may result in early depletion of the Battery.  Patient follow-up every 3 months is required to assess Battery capacity and projected longevity (i.e. service life). Discuss with each patient the importance of their  regular  device  check  as  a  critical  component  of  their  care.  Advise  the  patients  that  if  they  become symptomatic again, similar to their condition before the WiSE CRT System was implanted, they should contact their physician at the implanting center. Advise the patient that their WiSE CRT System can only be assessed at their  implanting  center.  Prior  to  obtaining  the patient’s  consent  for  the  procedure,  discuss  with  each  patient that Battery replacements will be required.  Discuss the following important topics: 1. The surgical process of Battery removal and replacement. 2. In clinical investigations, the battery was replaced at between 8 months and 3 years.   3. The patient may need a Battery replaced each year. 4. The Battery use and prediction for how long it will last can only be determined at a follow-up visit at their implanting center.  The  longevity  of  the  Battery  is  affected  by  several  important  operating  conditions  that  vary  for  each  implanted system.   The following recommendations should be considered to improve the efficiency of the system:    RECOMMENDATIONS TO IMPROVE PROJECTED BATTERY LONGEVITY During implant procedures • The distance between the Electrode and the face of Transmitter array is 10 cm or less. • The Electrode and the face of the Transmitter array is 30° or less from array-normal. • Implant  the  Transmitter  securely  to  the  inter-costal  muscle  over  an  acoustic  window (lung-free and rib-free transmission path to potential LV implant sites) that is at least 1cm x 2.5cm. • Implant  the  Electrode  in  excitable  tissue  with  an  acute  electrical  capture  threshold  less than 2V at 0.5 ms pulse width. During  patient follow-up visits • Optimize  programmed  transmit  level  and  pulse  width  setting  to  achieve  consistent  left ventricular  pacing  while  maximizing  battery  longevity.  When  possible,  lower  transmit levels and wider pulse widths should be used to increase battery life. • Refer to the WiSE CRT report for battery end of service (EOS) estimate that is generated by the Programmer and based on the system’s current settings. For more information on the WiSE CRT report, see section 3.3, Battery Status and End of Service.  The projected battery longevity in years is shown in the following tables for different WiSE CRT Transmitter model and  Battery  model  combinations.  As  shown,  battery  longevity  is  affected  by  pacing  rate  and  the  programmed settings, such as the transmit level and pulse width.   IMPORTANT! Projected longevity estimates are based on accelerated battery discharge data, device modeling and typical shelf life storage time. Do not interpret these values as precise numbers. TRANSMITTER MODEL 4000 ASSUMING 100% TRACKING AND PACING  Battery Usage Pacing Rate Transmit Level Pulse Width Aperture Projected Longevity (years)* Model 3000 Nominal 70 ppm 4 0.3 ms 8x24 2.2 Model 3100 Nominal 70 ppm 4 0.3 ms 8x24 3.3  IMPORTANT!  Under  extreme  usage  scenarios  (i.e.,  pacing  rate:  90  pmm,  transmit  level:  6,  pulse  width:  1.5ms), projected battery longevity with the Transmitter Model 4000 when using the Battery Model 3000 is 0.3 years, and when using the Battery Model 3100 is 0.4 years.
    WiSE CRT System Programmer Model 5100 Instructions for Use  25  TRANSMITTER MODEL 4100 ASSUMING 100% TRACKING AND PACING  Battery Usage Pacing Rate Transmit Level Pulse Width Aperture Projected Longevity (years) Model 3000 Nominal 70 ppm 4 0.3 ms 8x24 3.0 Model 3100 Nominal 70 ppm 4 0.3 ms 8x24 4.5  IMPORTANT! Under extreme usage scenarios (i.e., pacing rate: 90 pmm, transmit level: 6, pulse width: 1.5ms), projected battery longevity with the Transmitter Model 4100 when using the Battery Model 3000 is 0.7 years and when using the Battery Model 3100 is 1.0 year. 3.5 USING MARKER EVENTS The WiSE CRT Pulse Generator communicates the occurrence of real-time operational events to the Programmer using the RF communication link. The Programmer provides two types of outputs to signal the occurrence of these events: 1) via flashing indicator icons on the display and 2) via an analog output channel that when connected to typical recording equipment aligns the event signal output with ECG signals.  Two real-time operational events are communicated by the Pulse Generator with flashing icons and audible tones: 1. The detection of an RV pacing electrical output spike originating from the co-implanted device, referred to as an RV Pace Detect event.   2. The transmission of an ultrasonic pulse directed at the Electrode intended to pace the LV, referred to as a Transmit Event.   During Acoustic Window testing counts of events are displayed next to the icons.  Audible tones are enabled by selecting or deselecting the speaker icon (Figure 23).    Figure 23: Marker event audible tones are enabled by selecting or deselecting the speaker icon Marker  events  are  dependent  on  the  programmed  settings  and  the  operational  mode  of  the  Pulse  Generator.  In RV Synchronous mode: • RV  Pace  Detect  and  Transmit  Event  occur  with  a  pacing  spike  detection  that  triggers  an  ultrasound transmit.  • RV Pace Detect events occur with pacing spike detections that does not trigger an ultrasound transmit, for  example  if  the  detection  occurs  within  the  High  Rate  Limit  Interval  (430  ms)  or  the  ultrasound transmission does not occur due to an inability to target the Electrode. Marker Events are discernible on an analog output channel by interpretation of the polarity and the amplitude of square  wave  signals  (Figure  24).    The  signal  amplitude  and  pulse  width,  as  well  as  a  calibration  output,  for  the signals may be established in the Programmer Set-up screen previously described (Figure 5).   Inputting the analog signal into recording and display instrumentation that is monitoring the ECG will assist with set-up of the system, evaluation  of  operation,  and  troubleshooting.    Marker  Events  are  also  discernible  when  the  audio  output  is enabled.   EVENT MARKER PULSE TONE RV Pace Detect and Transmit Event Full amplitude positive High pitch tone Asynchronous Transmit Event ½ amplitude positive High pitch tone RV Pace Detect without Transmit Event Full amplitude negative Low pitch tone   OR
   26  WiSE CRT System Programmer Model 5100 Instructions for Use     Figure 24: Marker event calibration sequence  3.6 TROUBLESHOOTING ISSUE DESCRIPTION POSSIBLE CAUSE TROUBLESHOOTING AND CORRECTIONS No LV Pacing RV Synchronous mode not programmed • Reprogram Transmitter for synchronous pacing. Co-implant is not pacing • Check  and  reprogram  co-implant  device  mode  and rates. No RV pacing spike is being detected • Check  that  the  co-implant  is  properly  programmed  for WiSE  CRT  RV  spike  detection.  Reinitialize  Pacing  Spike Detection. • Reinitialize  Pacing  Spike  Detection  to  shorter  RV  pulse widths.  • Increase  RV  pace  amplitude  to  test  for  improved detection. Battery depleted, cable fractured, not connected, or connector failed. • Check  system  status  with  the  Programmer.    Replace Battery module when depleted.  Failure of the Electrode  OR  Failure to target the Electrode • Use  Temporary  LVOO  pacing  operation  to  check  for Electrode output (pacing spike) at any energy setting. • Evaluate  locations  of  Electrode  and  Transmitter  on previous  and  new  chest X-rays  to  check  for  movement causing no targeting (no pacing spike).  • Use  Acoustic  Window  set  up  to  check  for  possible movement  of  the  Electrode  or  the  Transmitter  causing poor targeting.  • Use  Acoustic  Window  set  up  to  improve  targeting  by choosing/optimizing [LOCATION SENSING] electrodes. • Use Acoustic Window set up to adjust the [THRESHOLD TUNING]  value  to  set  a  reasonable  signal  to  noise threshold. Transmit Energy set too low • Adjust Transmit Energy (Transmit Level, Pulse Width) to achieve LV pacing capture. RV Pace Detect Only Transmit Event Only RV Pace Detect and Transmit Event
    WiSE CRT System Programmer Model 5100 Instructions for Use  27  ISSUE DESCRIPTION POSSIBLE CAUSE TROUBLESHOOTING AND CORRECTIONS LV Pacing but not Synchronized to RV Pulse widths not programmed correctly in co-implant • Check and reprogram co-implant for correct pulse widths.  Reinitialize Pacing Spike Detection. Noise being sensed as RV pacing spikes. • Investigate noise sources. Reinitialize Pacing Spike Detection. Temporary LVOO mode is active • Check the operational mode with the Programmer. Pacing is synchronized to RA  • Check and reprogram co-implant for correct pulse widths.  Reinitialize Pacing Spike Detection. • The most reliable detection operation is with RA pulse widths >1.0ms. No communication session can be established with the Pulse Generator Programmer Failure • Re-boot  Programmer  to  run  self-test.    Communicate with an alternative Transmitter to check functionality. Interference • Arrange  patient  and  radio  module  to  improve  the signal strength. Weak signal strength • Check  signal  strength  indicator.  Arrange  patient  and radio module to improve the signal strength.   Battery depleted • Check patient records for device longevity history. Battery cable fractured or not connected • X-ray devices and check for cable fracture.   Early Battery Depletion Reduction of the acoustic window Area • Use  Acoustic  Window  set  up  to  adjust  use  of Transmitter aperture. • If necessary, reposition Transmitter in pocket. Acoustic window area larger than needed, resulting in poor efficiency • Use  Acoustic  Window  set  up  to  adjust  use  of Transmitter. Manual pacing parameters set too high • Reprogram.   Poor targeting • Use  Acoustic  Window  set  up  to  adjust  LOCATION SENSING electrodes, DISTANCE TARGET and DISTANCE LIMIT No Marker Events Communication link has stopped  • Check signal strength indicator.   Arrange patient and radio module to improve the signal strength. Cable not connected to Programmer. • Check that analog output cable is connected.  Use the [CALIBRATE  MARKER  OUTPUT]  feature  in  the Programmer Setup. Transmitter programmed off – no pacing or sensing setup • Check and reprogram Transmitter. Communication errors with the Pulse Generator Interference or weak signal strength • Arrange  patient  and  radio  module  to  improve  the signal strength.
   28  WiSE CRT System Programmer Model 5100 Instructions for Use    ISSUE DESCRIPTION POSSIBLE CAUSE TROUBLESHOOTING AND CORRECTIONS Initializing RV Pacing Spike Detection with the Programmer fails Co-implant not programmed for RV detectable pulse width above 0.35 ms. • Program co-implant. Restart Pacing Spike Detection. Co-implant not pacing. • Check and reprogram co-implant for mode and rates. Acoustic Window assessment identifies small area Transmitter or Electrode has moved. • Use  Acoustic  Window  set  up  to  adjust  use  of Transmitter aperture. • If necessary, reposition Transmitter in pocket. No electrogram or pacing with the Delivery Catheter Connection to the Electrode is disconnected in the Catheter or the Extension Cable is not connected. • Check connections. • Replace cable and retry. • Withdraw Delivery Catheter System and replace. Not connected to recording or pacing instruments • Check connections from Catheter to equipment.
    WiSE CRT System Programmer Model 5100 Instructions for Use  29  4 TECHNICAL SPECIFICATIONS 4.1 PHYSICAL AND ELECTRICAL SPECIFICATIONS PHYSICAL SPECIFICATIONS Tablet Computer  (H x W x T) 20.3 cm x 31.2 cm x 2.4 cm Radio Module Footprint (W x L) / Height 7.4 cm x 9.4 cm / 13.0 cm Mass (Tablet / Radio Module) 1.3 kg / 0.3 kg Radio Module Cable Length  3.0 meters AC/DC Power Adapter/ Cord Length 3.8 meters Marker Events Cable Length 3.0 meters  ELECTRICAL SPECIFICATIONS AC Line Voltage 100-240 VAC Power Input 1.6 A @ 100 VAC AC Frequency 50/60 Hz RF Reception Band 402-405 MHz (Medical Implant Communication Service/MedRadio rules) RF Transmission Bands Frequency: 402.15-404.85 MHz (MICS/MedRadio band) Modulation: 2-FSK @ 200 kbps Effective Radiated Power:  -16 dBm maximum  Frequency: 2.443-2.457 GHz (ISM band) Modulation: Manchester-encoded OOK (100%) Effective Radiated Power:  +20 dBm maximum DC Power Supply Use only the AC power adapter  provided with the Programmer to maintain electrical safety.  ENVIRONMENTAL SPECIFICATIONS Operating temperature range +15˚C to +35˚C Storage temperature range -10˚C to +55˚C Operating Relative humidity (maximum) 10% to 90% (non-condensing) Operating Altitude <2000 meters  4.2 PARAMETER SPECIFICATIONS SETUP PARAMETERS   Values/Range Increments Marker Output Level (mV) 10, 20, 50, 100, 200, 500, 1000  Marker Output Pulse Width (ms) 1.0-20.0 1.0 Display Brightness 1-10 1 Speaker Volume 0-10 1  OPERATIONAL OUTPUT PARAMETERS   Range Tolerance Marker Output Level (mV) 10 - 1000 ±15mV or ±10% Marker Output Pulse Width (ms) 1.0-20.0 0.1
   30  WiSE CRT System Programmer Model 5100 Instructions for Use     4.3 PROGRAMABLE PULSE GENERATOR VALUES  PERMANENT PARAMETERS  Values/Range Mode RV Synchronous, Sense Only, OFF Transmit Level 1, 1.5, 2, 2.5, 3, 4, 5, 6 Transmit Pulse Width (ms) 0.01, 0.02, 0.05, 0.07, 0.1, 0.15, 0.2 to 2.0 in increments of 0.1,  2.5, 3.0, 3.5, 4.4  CO-IMPLANT PARAMETERS   Values/Range Increments RV Pulse Width dual chamber (ms) 0.70 – 0.80 0.05 Atrial Pulse Width dual chamber (ms) 0.35 – 2.00 0.05 RV Pulse Width single chamber (ms) 0.35 – 2.00 0.05  TEMPORARY LVOO PARAMETERS     Values/Range Increments Pacing Rate (bpm) 30 – 150 5 Transmit Level 1, 1.5, 2, 2.5, 3, 4, 5, 6  Pulse Width (ms) 0.01,  0.02,  0.05,  0.07,  0.1,  0.15,  0.2  to 2.0  in  increments  of  0.1,  2.5,  3.0,  3.5, 4.4    4.4 TECHNICAL SERVICE AND SUPPORT WARNING -  No  modification  of  the  Programmer  is  allowed  except  by  authorized  service  personnel  of  EBR Systems.  EBR Systems installs the Model 5100 Programmer in the medical facility and when necessary EBR Systems installs any  new  Software  products  that  become  available  for  the  Programmer.    Maintenance  and  service  for  the Programmer can only be performed by EBR Systems.  Should a Programmer require service it  will be exchanged.  The  Programmer  should  be  tested  biennially  by  an  EBR  Systems  representative.        It  is  the  medical  facility’s responsibility to schedule biennial testing by contacting EBR Systems.   4.5 CLEANING AND DISPOSAL OF DEVICES The  Programmer  may  be  cleaned  by  wiping  the  exposed  surfaces  using  isopropyl  alcohol.    The  Programmer contains  typical  electronic  components  that  require  consideration  for  disposal.    Return  the  Programmer  to  EBR Systems or follow proper e-waste disposal procedures as recommended for personal computers. Contact EBR Systems to be supplied with materials related to returning devices for disposal.
    WiSE CRT System Programmer Model 5100 Instructions for Use  31  4.6 EMC INFORMATION This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. WARNING – The Programmer should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the Programmer should be observed to verify normal operation in the configuration in which it will be used.  WARNING – The Programmer may be interfered with by other equipment, even if that other equipment complies with the CISPR emission requirements.   WARNING – Use of accessories, transducers, and cables other than those specified or provided could result in increased electromagnetic emissions or decreased electromagnetic immunity of the Programmer and result in improper operation.   WARNING – Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used not closer than 30 cm (12 inches) to any part of the Programmer, including cables specified for use with the Programmer.
   32  WiSE CRT System Programmer Model 5100 Instructions for Use     5 SYMBOL GLOSSARY  SYMBOL STANDARD REFERENCE STANDARD  TITLE SYMBOL TITLE EXPLANATORY TEXT  ISO 15223-1, Clause 5.1.1 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied. Manufacturer Indicates the medical device manufacture.   ISO 15223-1, Clause 5.1.6 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied. Catalogue or model number Indicates the manufacturer’s catalogue number so that the medical device can be identified.   ISO 15223-1, Clause 5.1.7 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied. Serial number Indicates the manufacture’s serial number so that a specific medical device can be identified.   ISO 15223-1, Clause 5.4.4 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied. Attention: Read all warnings and precautions in instructions for use. Indicates the need for the user to consult the instructions for use for important cautionary information.  ISO 15223-1, Clause 5.3.7 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied. Storage temperature range Indicates the temperature limits to which the medical device can be safely exposed.  IEC 60601-1, Table D.2, Symbol 10 Medical electrical equipment — Part 1: General requirements for basic safety and essential performance. Follow instructions for use Refer to instruction manual/booklet.  IEC 60417-5035  Graphical symbols for use on equipment. Output To identify an output terminal.     Graphical symbols for USB connection  To identify USB port.  IEC 60417-5031 Graphical symbols for use on equipment. Direct current To indicate on the rating plate that the equipment is suitable for direct current only.
  SYMBOL STANDARD REFERENCE STANDARD  TITLE SYMBOL TITLE EXPLANATORY TEXT  AS/NZS 4417.1:2012 Australian Radio Communications Requirements.  Complies with Australian Radio communications requirements.   47 CFR Part 15 Federal Communication Commission Number (FCC ID #)  Complies with United States Radio communication requirements.  765/2008/EC 768/2008/EC MDD 90/385/EEC 199/5/EC The requirements for accreditation and market surveillance relating to the marketing of products; Medical Device Directive.  Signifies conformity to the European Active Implantable Medical Device Directive version M4 90/385/EEC and Radio & Telecommunications Terminal Equipment (199/5/EC).
                             © 2018 EBR Systems, Inc.  All Rights Reserved.  WiSE is a trademark of EBR Systems, Inc.

Navigation menu