Berry Electronic Tech BM1000 Pulse Oximeter User Manual Palm Pulse Oximeter

Shanghai Berry Electronic Tech CO., Ltd Pulse Oximeter Palm Pulse Oximeter

Users Manual

      Operation    Manual Fingertip Pulse Oximeter      Shanghai Berry Electronic Tech Co., Ltd
Instructions This manual provides the instructions necessary to operate Fingertip Pulse Oximeter (hereinafter called as the product) in accordance with its function and intended use. Observance  of  this  manual  is  a  prerequisite  for  proper  performance  and  correct operation, and ensures patient and operator safety. Content of this manual is subject to change without prior notice. Issued date: 2013/11/25 Version:    1.0  Statement The manufacturer is responsible for safety, reliability and performance of this product only in the condition that:  All installation operations, expansions, changes, modifications and repairs of this product are conducted by manufacturer authorized personnel; and    The  electrical  installation  of  the  relevant  room  complies  with  the  applicable national and local requirements; and  This product is operated under strict observance of this manual.  Guarantee Free service scope  The  manufacturer  provides  free  service  to  any  product  which  conforms  to  the warranty regulations. Chargeable service scope  The manufacturer’s obligation or liability under his warranty does not include the service  of  any  factitious  damage,  or  misuse,  or  irresistible  natural  disaster,  or delay resulting from the improper use or application of the product, or repairs by people other than the manufacturer authorized personnel.  Return Policy In the event  that  it  becomes  necessary  to  return a unit  to  the  manufacturer, please obtain a return authorization first. Please contact the manufacturer and provides the model number, serial number, and a brief description of the reason for return. Return shipments will not be accepted if the serial number is not clearly visible. The customer is responsible for freight charges when this product is shipped to the manufacturer for service (including any relevant customs fees or other freight related charges).  I. Safety Information WARING Indicate  a  potential  hazard  situation  or  unsafe  practice  that,  if  not  avoided,  could result in death or serious injury. CAUTION
Indicate a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage. NOTE Provide application tips or other useful information to ensure that you get the most from your product.  WARING  The person who uses the product must receive adequate training before use.  The product is intended  only as adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms. It is not intended as a device used for treatment purposes.  When using the product together with the electrical surgery equipment, the user  should  pay  attention  to  and  guarantee  safety  of  the  patient  being measured.  EXPLOSION  HAZARD:  Do  not  use  the  product  in  the  presence  of flammable anesthetics, explosive substances, vapors or liquids.  Make sure not to use the product during MRI (magnetic resonance imaging) scanning because induced current could potentially cause burns.    The  product is without alarm function.  Continuous  monitoring  for a long time is not suitable.  No modification of this product is allowed.  Use AAA alkaline batteries. Do not use poor quality batteries. Remove the batteries if the product is not to be used for a long time.  For disposal of the Pulse Oximeter and batteries, follow local regulations or your hospital’s policy regarding disposal of such accessories. Do not dispose randomly.  The product  is  commonly  sealed product. Keep its surface dry and clean, and prevent any liquid from infiltrating it.  The device is precision and fragile. Avoid pressure, knock, strong vibration or other mechanical damage. Hold it carefully and lightly. If it is not in use, it should be appropriately placed.  If  patient  is  an  intended  operator,  you  must  read  the  operation  manual carefully  and  understand  deeply  or  consult  with  the  doctor  and manufacturer before using. If you have any discomfort in use, please  stop using immediately and go to the hospital.    The  pulse  oximeter  equipment  is  calibrated  to  display  functional  oxygen saturation. NOTE  Important! Before use, carefully read this manual, all safety information and specifications.
II. Product Feature   1. Simple and convenient usage of product, simple one-touch operation.   2. Small volume, light weight, convenient to carry.   3. Lower consumption, original two AAA batteries can continuously work for 24 hours.     4. Low voltage reminder shows in screen when there’s low battery, may influence the normal working.     5. The machine will automatically power off when there’s no signal generated.     6. Daily maintenance and calibration is unnecessary, except changing the batteries.   7.  Communication  can  be  realized  between  Bluetooth  and  computer  or  mobile phone.   8. The product will automatically be powered off when no signal generates in the device for longer than 8 seconds.  III. Applicable people and scope It is suitable for monitoring adults and children. It is used in the hospital’s operation room,  ICU,  clinic  section  office,  out-patient  department,  sickroom,  emergency treatment.  It  can  also  be  used  in  the  recovery  and  health  care  organizations,  the community medical treatments, the oxygen bars, the family nursing, the physical care in sports ( you can use the product before or after the sport, but it is not recommended to use it during the sport).  IV. Intended use This  product  can  be  used  to  monitoring  the  physiological  signal,  included:  arterial oxygen saturation (SpO2), pulse rate (PR) and pulse strength. The product measures these  physiological  signals  of  patients,  and  further  processing,  then  the  numerical results will be displayed in OLED screen or in the computer interface/mobile phone with its wireless Bluetooth. NOTE  If  the  product is  intended  to  allow direct  diagnosis  or  monitoring of  vital physiological processes, then it is likely to result in the immediate danger to the patient.    V. Contraindications The  product  only  applies  to  adults  and  children,  please  don’t  use  the  product  for children under the age of three, infant and neonatal. The damaged skin tissue can’t be measured.  VI Structure and composition This product is composed of one fingertip pulse oximeter.  VII. Measurement Principle SpO2 plethysmogram measurement is employed to determine the oxygen saturation
of hemoglobin in the arterial blood. The SpO2/PLETH parameter can also provide a pulse rate signal, pulse strength and a plethysmogram wave. Arterial  oxygen  saturation  is  measured  by  a  method  called  pulse  oximetry.  It  is  a continuous,  non-invasive  method  based  on  the  different  spectra  absorption  of hemoglobin  and  oxyhemoglobin  (called  spectrophotometer  principle).  It  measures how much light, sent from light sources on the other side. The  sensor  measurement  wavelengths  are  nominally  660nm  for  the  Red  LED  and 940nm for Infrared LED. Maximum optical power output for LED is 4 Mw.  VIII. Cleaning and Disinfection CAUTION  Never immerse or soak the product.  Exercise caution during cleaning/disinfection to avoid wetting the pins.  We  recommend  that  the  product  be  disinfected  only  when  necessary  as determined  by  your  hospital’s  policy,  to  avoid  long  term  damage  to  the product.  Never use cleaning agents/disinfectants other than the recommended.  Never permit high-pressure and high-temperature disinfection of the device. Cleaning 1. Clean the product with cotton or soft cloth moistened with water. 2. After cleaning, wipe off the water with a soft cloth. 3. Allow the product to air dry. Disinfection The recommended disinfectants include: ethanol70%, isopropanol70%, glutaraldehyde (2%)solution disinfectants. 1. Clean the product as instructed above. 2. Disinfect  the  product  with  cotton  or  soft  cloth  moistened  with  one  of  the recommended disinfectants. 3. After disinfection, be sure to wipe off the disinfectant left on the product with a soft cloth moistened with water. 4. Allow the pulse oximeter to air dry.  IX. General Display mode: OLED Size: 58 (H)  × 34 (W)  × 30(D)mm Weight: 50g (include two AAA batteries)  X. Electrical specifications Working voltage: D.C.2.2 V~D.C.3.4V Battery Type: Two common 1.5V AAA alkaline batteries.   Power consumption: smaller than 50mA  XI. Parameter Specification
SpO2   Range: 35~100%   Resolution: 1%   Accuracy: ±2%(80%-100%);    ±3%(70%-79%) PR   Range: 25~250bpm   Resolution: 1bpm   Accuracy: ±2bpm Low perfusion:    Range: 0.5%~20% SpO2 accuracy: ±2% (70%~100%) PR accuracy: 25~250bpm ±2bpm  XII. Environment Temperature Operation: 5-40℃   Transportation and storage: -20-70℃ Humidity   Operation: <80%   Transportation and storage: <93% Altitude   Operation: -390~3012meter (-1254~9882feet)   Transportation and storage: -390~5574meter (-1254~18288feet) Atmospheric pressure   Operation: 86Kpa~106Kpa   Transportation and storage: 50Kpa~106Kpa Enclosure flame class:    V-0  XIII. Packing list The standard configuration Fingertip pulse oximeter 1pc Rope 1pc The Operator manual 1pc  Expected service life: three years  XIV. Operating Instructions CAUTION  Check if the product is in normal condition before monitoring. Do not use the product once it is found damage.  Don’t  put  the  product  on  extremities  with  arterial  catheter  or  venous syringe.  The product can’t be clipped on the edema and tender tissue.
 If  no  pulse  is  found  or  reading  is  unreasonable,  first  check  the  testee’s condition, and then check the product placement on the finger, finally asking the qualified engineer to check the product for proper functions.  Don’t use the product  to  measure  patients  whose pulse  rate  is  lower than 30bpm, which may cause incorrect results.  Make sure no contamination or scar exists in the size where the product is placed. Otherwise, the measured result may be incorrect because the signal received by the sensor is affected.  When  used  on  different  patients,  the  product  is  prone  to  crossed contamination,  which  should  be  prevented  and  controlled  by  the  user. Disinfection is recommended before using the product on other patients.  The product is not appropriate for the ceaseless monitoring of the patients. Prolonged and continuous monitoring may increase jeopardy of unexpected change of dermal condition such as abnormal sensitivity, rubescence, vesicle, repressive putrescence, and so on. The uncomfortable or painful feeling may appear  if  using  the  device  ceaselessly,  especially  for  the  patient  of  poor perfusion  or  immature  dermographia  by  light  collimation.  It  is recommended that the device should not be applied to the same finger for over 2 hours.  Don’t perform SpO2 monitoring and NIBP measurements on the same arm simultaneously. Obstruction of blood flow during NIBP measurements may adversely affect the reading of the SpO2 value.  Testee’s fingernail can’t be too long. Otherwise the finger can’t be inserted into  the  sensor  to  a  suitable  depth  and  the  SpO2  measurements  may  be inaccurate.  Make sure to place the product on the finger in a correct direction. The LED part  of  the  sensor  should  be  at  the  backside  of  the  patient  hand  and photodetector part at the inside. Make sure to insert the finger to suitable depth  into  the  sensor  so  that  the  fingernail  is  just  opposite  to  the  light emitted from the sensor.  To  acquire  accurate  results,  please  read  data  until  the  product  is  steadily placed. Factors affecting measurement accuracy  The measurements also depend on absorption of special wavelength ray by oxidized hemoglobin and deoxyhemoglobin. Concentration of nonfunctional hemoglobin may affect the accuracy of the measurement,  Shock,  anemia,  hypothermia  and  the  application  of  vasoconstriction  drug may decrease arteria blood flow to an unmeasurable level.  Pigment,  or  deep  color  (for  example:  nail  polish,  artificial  nails,  dye  or pigmented cream) may cause inaccurate measurements.  Interference  of  high  frequency  electricity  and  bright  light  may  affect  the accuracy of the measurement.  Appearance and Display
 Figure 1  Measuring steps 1. Hold the product in one hand with the front panel facing the palm. Put the other hand’s big finger on the battery cabinet lid’s press sign, press downwards and push the lid open at the same time. The battery cabinet is opened as shown in Figure 2. 2. Install batteries into the slots per the “+” and “-” symbols as shown in Figure 3. Cover the lid onto the cabinet and push it upwards to make it close well. 3. Press Clip’s press sign in the figure 1and open the clip. Let the testee’s finger put into the rubber cushions of the clip, make sure the finger is in the right position as shown in Figure 4, and then clip the finger. 4. Press  the  power  and  function  switch  button  on  the  front  panel  to  turn  on  the product. Using  first  finger, middle  finger  or ring finger when doing  test.  Don’t shank the finger and keep the testee at case during the process. The readings will be displayed on the LED screen a moment later as shown in Figure 5. WARRING  When your finger is plugged into the Oximeter, your nail surface must be upward.  The results may be wrong if you did not plug the finger thoroughly in the Oximeter.      Figure 2                                                            Figure 3
 Figure 4                                                        Figure 5  Function instructions a. When the data has been displayed on the screen, change the display direction by pressing the power and function button again. (as shown in Figure 6) b.  When  the  product  is  powered  on,  long  press  the  power  and  function  button, Bluetooth function will be started. The Bluetooth indicator light on the top of display will flicker. (as shown in Figure 7) c. The product will automatically be powered off when no finger is in the device for longer than 8 seconds. And switch to another display mode. (as shown in Figure 8)                  Figure 6                                                          Figure 7                 Figure 8                             Communication with computer   a. Open the Bluetooth of the computer, search for the Bluetooth device, the pulse oximeter’s user name is BerryMed, pairing code is 0000.   b. Check the Bluetooth serial no. when connected, such as com18.   c. Open pc software “SpO2-EN.exe”(as shown in Figure 9)
 Figure 9 d. Choose Bluetooth serial no., such as com18. (as shown in Figure 10)  Figure 10 e. Click to open. (as shown in Figure 11)  Figure 11 f. Enter display interface. (as shown in Figure 12)
 Figure 12  Communication with mobile phone a. Open the mobile phone, and double-click SpO2 software icon “ ”. And enter the following interface. (as shown in Figure 13)  Figure 13 b. Click the “Add new users” and enter the following interface. (as shown in Figure 14)
 Figure 14 c. Input the user name, click the “Details” and jump the following information table. (as shown in Figure 15)  Figure 15 d. Input personal information, then click “Save”  and enter the following interface. (as shown in Figure 16)
 Figure 16 e. Click “berry” and enter the following interface. (as shown in Figure 17)  Figure 17 f. Click“ ”, search equipment and jump the following interface. (as shown in Figure 18)
 Figure 18 g. Wait  to  search  out  the  equipment,  click“BerryMed”  and  enter  the  following interface. (as shown in Figure 19)  Figure 19 h. Click“Pair”, to pair with“BerryMed” and enter the following test interface. (as shown in Figure 20)
 Figure 20 i. If you want to exit the software, click escape key of your mobile phone and jump the  following  window.  And  then  click“OK”,  exit  the  software.  (as  shown  in Figure 21)  Figure 21  XV. Troubleshooting Trouble Possible reason solution The  SpO2  and PR  can’t  be displayed normally. 1. The  finger  is  not  properly positioned. 2. The patient’s SpO2 is too low to be detected. 1. Please  the  finger  properly and try again. 2. Try again; Go to a hospital for  a  diagnosis  if  you  are sure  the  device  works  all right.
The  SpO2  and PR  display instable. 1. The  finger  is  not  placed  inside enough. 2. The  finger  is  shaking  or  the testee is moving. 1. Place  the  finger  properly and try again. 2. Let the testee keep calm. The  device can’t  be  turned on. 1. The  batteries  are  drained  or almost drained. 2. The  batteries  are  not  inserted properly. 3. The device’s malfunction. 1. Change batteries. 2. Reinstall batteries. 3. Please contact the supplier. The  screen  is suddenly off. 1. The  product  is  automatically powered  off  when  no  signal  is detected longer than 10 seconds. 2. Power quantity of the batteries is exhausted. 1. Normal 2. Replace the batteries.   XVI. Symbolic meaning Symbol Meaning  “CAUTIOUS”! Please refer to the operation manual.  Type BF Equipment.  The product does not contain alarm function.  When the end-user wishes to discard this product, it must be sent to separate collection facilities for recovery and recycling.  Information of manufacture, including name and address.  Date of manufacture.  European Union for approval.  Serial Number.  Batch Code.  Type Number.  The European Union authorized. IP21 The product is protected against harmful effects of dripping water per IEC 60529.
IC Warning  This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this  device must  accept  any  interference,  including  interference  that  may  cause  undesired operation of the device.  Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio interference to other users, the antenna type and its gain should be so chosen  that  the  equivalent  isotropically  radiated  power  (e.i.r.p.)  is  not  more  than  that necessary for successful communication. This device complies with Canadian ICES-003 and RSS-210  “Le présent appareil est conforme aux CNR d'Industrie Canada applicable aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, meme si le brouillage est susceptible d'en compromettre le fonctionnement." Cet appareil numérique de la classe B est conforme à la norme NMB-003 du Canada.  FCC Warning  This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:  (1)  This  device  may  not  cause  harmful  interference,  and  (2)  this  device  must accept any interference received, including interference that may cause undesired operation. NOTE 1: This equipment has been tested and found to comply with the limits for a Class B digital device,  pursuant to  part 15  of  the  FCC  Rules. These  limits  are  designed  to provide reasonable protection against harmful interference in a residential installation. This equipment generates,  uses  and  can  radiate  radio  frequency  energy  and,  if  not  installed  and  used  in accordance with the  instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment  does  cause  harmful  interference  to  radio  or  television  reception,  which  can  be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -Reorient or relocate the receiving antenna. -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help.  NOTE  2:  Any  changes  or  modifications  to  this  unit  not  expressly  approved  by  the  party responsible for compliance could void the user’s authority to operate the equipment.
  Shanghai Berry Electronic Tech Co., Ltd Unit C, 1st Floor, 7th Building, No.1188 Lianhang Road, Minhang District, Shanghai, China 201112 TEL: +86-21-5853 1958      FAX: +86-21-5853 0420   www.berry-med.com  Kingsmead Service limited (UK) 145-157 St John Street,London ,EC1V 4PY(UK) TEL:044-20-7193 9159    FAX:044-20-7193 9159 www.kingsmead-service.com  If you need additional information, please contact with the company.