                           Section  Version Welcome  B.2 Safety & Warning  B.2 Operating Instructions  B.2 Procedures  B.2 Technical Information  B.2 Guarantee  B.2          Manufactured in the United Kingdom by DIOMED Limited  DIOMED is a Registered Trademark of DIOMED Ltd.   2005 DIOMED Ltd.    DIOMED DELTA 15/30 DIODE LASER OPERATOR MANUAL  Version B.2   July 2005             MAN/26/0068
DELTA 15/30 DELTA 15/30 Operator Manual  Page 1.2  Version B.2  SECTION 1  WELCOME Thank you for purchasing a DIOMED laser. This is a high quality medical instrument that will give many years of service if used and cared for according to the instructions in this operator manual.  Before going any further it is important that the Product Information on page 1.4 is completed at installation. The Guarantee Registration Certificate and Guarantee Form in section 5 must also be filled in and the form returned to DIOMED to complete the registration process.   INTRODUCTION This manual describes the operation of the DIOMED DELTA 15 and the DIOMED DELTA 30, (referred to as DIOMED DELTA 15/30 in this manual). These lasers are to be used only by experienced, trained operators familiar with laser procedures. Before using this instrument for the first time, read the Safety & Warnings section and the Operating Instructions.  The operator must become familiar with all the controls before commencing any therapy.  DIOMED DELTA 15  The DIOMED DELTA15 is a diode laser capable of delivering up to 15W of continuous wave radiation via an optical fiber, or 119J/cm² of pulsed radiation via a spot handpiece, coupled to the laser aperture. The DIOMED DELTA 15 incorporates a Class IV GaAlAs (Gallium Aluminium Arsenide) diode lasers with a wavelength of 810nm (±20nm). The DIOMED DELTA 15 incorporates a visible Class IIIa diode laser aiming beam with a wavelength of 635-660nm and a maximum power output of 5mW.    DIOMED DELTA 30  The DIOMED DELTA 30 is a diode laser capable of delivering up to 30W of continuous wave radiation via an optical fiber, or 400J/cm² of pulsed radiation via a spot handpiece, coupled to the laser aperture. The DIOMED DELTA 30 incorporates Class IV GaAlAs (Gallium Aluminium Arsenide) diode lasers with a wavelength of 810nm (±20nm). The DIOMED DELTA 30 incorporates a visible Class IIIa diode laser aiming beam with a wavelength of 635-660nm and a maximum power output of 5mW.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 1.3  Version B.2  DESCRIPTION OF THE DIOMED DELTA 15/30 The DIOMED DELTA 15/30 has been designed for use with the DIOMED range of procedure kits, such as those for EVLT (EndoVenous Laser Treatment). It can also be used with a wide range of standard fibers and accessories, further details of which are in section 4 of this manual.  The DIOMED DELTA 15/30 consists of three main components:  •  The main enclosure houses the laser module containing the optics, heatsink, microprocessor-based control electronics and power supplies •  The footswitch to activate the laser output when in READY mode •  The fiber or handpiece for delivering the laser radiation to the tissue  Key features of the DIOMED DELTA 15/30 include:  •  15W or 30W power output •  Compact & portable •  Intuitive user interface •  Automatic procedure recognition when used with fibers and accessories conforming to the DIOMED FRS fiber system •  Minimal maintenance & service •  Internal power calibration meter (optional on DIOMED DELTA 15)  ABOUT THIS MANUAL This manual is broken down into five main sections as described below. The contents of each one are detailed at the beginning of the appropriate section.   1  Welcome   2  Safety & Warnings  Explains the general warnings and precautions that must be followed to ensure that the DIOMED DELTA 15/30 is used in a safe manner. 3  Operating Instructions  Detailed instructions on how to install and operate the DIOMED DELTA 15/30 laser. 4  Procedures  Specific information about DIOMED fibers and procedure kits, equipped with FRS. 5  Technical Information  This section explains all the maintenance procedures that can be performed by the user. 6  Guarantee  Contains the DIOMED guarantee policy and your guarantee certificate. Ensure that the accompanying form is returned to DIOMED and that the certificate is kept in a safe place.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 1.4  Version B.2  PRODUCT INFORMATION    Laser Serial Number ...............................................................................................................................    Software Version ....................................................................................................................................  (Note: This information is shown on the screen displayed at start-up.)    Date Installed  .........................................................................................................................................    Installed by ..............................................................................................................................................    Signed  ....................................................................................................................................................    Print Name  .............................................................................................................................................    Organisation ...........................................................................................................................................    For service, parts or repair, contact your local DIOMED representative:
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.1  Version B.2  SECTION 2  SAFETY & WARNINGS  CONTENTS  Section 2 SAFETY & WARNINGS ....................................................................................................... 1 Contents ................................................................................................................................................. 1 Symbols used in this manual.................................................................................................................. 2 Warnings................................................................................................................................................. 3 Safety classifications, hazards and precautions..................................................................................... 5 Eye Injury................................................................................................................................................ 7 Burns ...................................................................................................................................................... 8 Reflection Warning ................................................................................................................................. 8 Explosion Hazard Warning..................................................................................................................... 8 Vapor Plume........................................................................................................................................... 8 Clinical Indications & Contra-Indications................................................................................................ 9 Clinical Warnings.................................................................................................................................. 11 Clinical Precautions.............................................................................................................................. 12 Safety labelling ..................................................................................................................................... 13 Safety features ..................................................................................................................................... 14 EMC Declaration .................................................................................................................................. 15 ESD Precautionary Procedures............................................................................................................ 20 FCC Declaration ................................................................................................................................... 20
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.2  Version B.2  SYMBOLS USED IN THIS MANUAL      This symbol indicates caution should be taken, as there may be a potentially hazardous situation that could result in injury to personnel or damage to the equipment.  This symbol indicates the possibility of a non-radiation hazard that may result in severe injury to personnel within the vicinity of the equipment.  This symbol indicates the possibility of an electrical hazard that could cause injury to personnel within the vicinity of the equipment or damage to the equipment.  This symbol indicates the possibility of exposure to hazardous laser radiation that could cause injury to personnel within the vicinity of the equipment.  This symbol indicates personnel within the vicinity of the equipment should wear appropriate eye protection.   This symbol indicates an important point to be noted.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.3  Version B.2  WARNINGS   US Federal Law restricts the use of this device to sale by or on the order of a physician. Intended for use only by trained physicians/surgeons familiar with laser procedures.    Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.    This product must be stored at temperatures between 0°C and +55°C.  If stored at temperatures below 10°C for a period of time, the l aser requires up to 12 hours to acclimatise, prior to operation. Failure to observe this could result in invalidation of the guarantee.  The laser is not designed to operate at temperatures below 10°C.  The Optical Power Calibration meter, if fitted, is to be used only to calibrate a fiber or spot handpiece in accordance with the instructions detailed in this manual. Under no circumstances may it be used as a beam dump, as this will result in damage to the instrument.  This product contains a lithium battery, which should only be replaced by authorised service personnel. Replace the battery only with the same or equivalent type. Dispose of used batteries according to the manufacturer’s instructions and local disposal requirements.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.4  Version B.2  EMC Warning Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided.  The DIOMED DELTA 15/30 may be interfered with by other equipment, even if that other equipment complies with CISPR emission requirements.  Portable and Mobile RF communications equipment can affect medical electrical equipment.  The DIOMED DELTA 15/30 should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the DIOMED DELTA 15/30 should be observed to verify normal operation in the configuration in which it will be used.  If Electromagnetic interference is experienced, relocate or re-orientate the DIOMED DELTA 15/30 or the other equipment.  Accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by the manufacturer of the DIOMED DELTA 15/30 as replacement parts for internal components, may result in increased emissions or decreased immunity of the DIOMED DELTA 15/30.  ‘Immunity’ is the ability of a device to function normally when operated in the presence of electromagnetic radiation.  The following Cables are compatible with the DIOMED DELTA 15/30:  •  Foot Switch Assembly       Diomed part no.  AS1/A0/0002  •  Remote Interlock Lead (if required). Lemo connector (supplied with laser):  Diomed part no. CON/51/0003  Ferrite sleeve (Farnell part no. 898-454). Screened twisted pair cable 7/0.2 (Farnell part no. 140-457).  Note: The ferrite sleeve should be fitted to the remote interlock cable, at a maximum of 75 mm from the Lemo connector.  •  Remote Interlock Bypass (supplied with laser). Diomed part no. AS1/A3/0024   •  IEC Mains Lead (supplied with Laser) Diomed part numbers:  CBL/02/0002, CBL/02/0040, CBL/02/0042, CBL/02/0046, CBL/02/0051, CBL/02/0063.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.5  Version B.2  SAFETY CLASSIFICATIONS, HAZARDS AND PRECAUTIONS   This product is classified as a Class IV laser product in compliance with FDA 21 CFR 1040.10 and 1040.11, UL 60601-1, EN 60601-1, EN 60601-1-2, EN 60601-2-22 and EN 60825-1. This product conforms to the requirements of Council Directive 93/42/EEC of the Council of European Communities (Medical Devices Directive). Affixing the ‘CE Mark’ to the instrument indicates conformity to this directive.       0086   The local Laser Safety Officer should review all procedures for safety prior to system use.   A Class IV Laser is hazardous to the eye from the direct beam and diffuse reflections. It also presents significant skin and fire hazard.   Avoid eye or skin (except specific treatment) exposure to direct or scattered radiation. Take all necessary protective measures, as explained in the rest of this section, in areas where the laser is being used.   All personnel must wear approved protective glasses appropriate to the wavelength of the DIOMED DELTA 15/30 to reduce the risk of eye damage.   The aiming beam is a Class IIIa Laser and an unprotected eye may view the beam scattered from a non-reflective surface. Do not stare into the aiming beam or view it directly with optical instruments.   Avoid directing the laser beam anywhere other than the treatment area or calibration ports.  Before using a fiber, check it carefully for any signs of damage during storage or transit. Protective caps should be in place over SMA connectors. All screws and ports should be secure. Do not use if there is any sign of damage.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.6  Version B.2   The DIOMED DELTA 15/30 Laser is a portable laser weighing 11kg.  All standard safety procedures for lifting should be applied when moving the instrument.  There are no user serviceable parts in the DIOMED DELTA 15/30 Laser. The exterior cover should only be removed by a trained and authorised laser service technician.  Pins of connectors identified with the Electrostatic Discharge (ESD) warning symbol should not be touched. Connections should not be made to these connectors unless the ESD precautionary procedures detailed in Appendix D are followed. It is recommended that all staff receive an explanation of the ESD warning symbol and made aware of the ESD precautionary procedures described at the end of this section.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.7  Version B.2  EYE INJURY    Extreme  caution  should  be  taken  when  operating  the  DIOMED  DELTA 15/30 near the eyes.  Near infrared light (810nm) from the DIOMED DELTA 15/30 passes through the transparent components of the eye and is focused on the retina at the back of the eye. This light can therefore cause an accidental retinal burn.  All personnel must wear approved protective glasses to reduce the risk of eye damage. The patient should wear protective glasses when not anesthetised. If the patient is anesthetised, the eyelids should be taped shut and covered with moist gauze pads.  The local Laser Safety Officer should review all procedures for safety prior to system use.  All protective glasses should be designed for protection from continuous wave laser radiation in the wavelength range 790 – 830nm.  The  degree  of  optical  filtration  (Optical  Density  or  OD)  depends  on  the application  and  should  be  assessed  and  approved  by  the  appointed  Laser Safety Officer for the establishment. The  recommendations  of  European  Standards  EN  60825-1  or  EN  207  are appropriate  to  assessing  laser  eye  risk.  Note  that  the  standards  assume  a viewing distance from the source of light of more than 100mm. DIOMED supplies laser safety glasses marked in accordance with EN 207 as L5 or greater. Contact your local DIOMED distributor if these are required. The ‘Nominal Ocular Hazard Distance’ is 8 metres. Use  of  optical  accessories  and  viewing  aids,  which  may  increase  the  eye exposure beyond a safe  limit, should be  subject to  the approval  of the  Laser Safety Officer. Never look directly into the laser aperture even if wearing safety glasses. Serious eye injury could result.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.8  Version B.2  BURNS   Irradiation of any substance or material other than the target tissue may result in a laser burn.  REFLECTION WARNING  Avoid placing reflective materials such as glass, metals and polished plastic in the beam. EXPLOSION HAZARD WARNING    Avoid using flammable or explosive anesthetic gases that may be ignited by the laser. Avoid using other flammable or fume-emitting substances (e.g. ether, iodine solution, collodion, and alcohol) in the operative field. VAPOR PLUME  DIOMED recommends that a smoke evacuator or in-line filter be used when lasing. Caution – Laser Plume may contain viable tissue particulates.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.9  Version B.2  CLINICAL INDICATIONS & CONTRA-INDICATIONS  Indications  The DIOMED DELTA 15/30 is intended for the following contact or non-contact laser procedures: •  Incision (only contact) •  Excision (only contact) •  Vaporisation •  Coagulation / haemostasis  - of soft tissue in Endovascular, open and endoscopic procedures in Vascular Surgery,  General  Surgery,  Gynaecology,  Urology,  Otorhinolaryngology (ENT)/Head  &  Neck  Surgery,  Ophthalmology  (oculoplastics), Pulmonology/Thoracic  surgery,  Plastic  Surgery,  Gastroenterology  and Neurosurgery.  Recommended  power  levels  for  the  above  indications  for  contact  fiber accessories are between 5-30W and for non-contact fibers, 10-30W.  Beginning at low power (5-10W) with short pulse duration, the surgeon should note the surgical effect and increase power or pulse duration until the desired surgical effect is obtained.  Generally, the power requirement will vary depending on the contact fiber core and tip size. Less power will be required to obtain tissue reaction with smaller diameter fibers. Recommended power settings are less important than the visual effect. Changes in tissue texture and colour are the best indications of the laser effect. Specific pulse duration is not recommended, but is left to user preference and best medical judgement dependent on the particular application and tissue type.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.10  Version B.2   Contraindications The DIOMED DELTA 15/30 should only be used in conditions where its use is appropriate and of proven efficacy. It should never be operated except under the direct supervision of a trained operator.  The  potential  for  complications encountered  in surgical  laser procedures  will be  the  same  as  those  encountered  in  any  surgical  procedures.    These complications may be serious and could result in death. Complications may include: •  Pain •  Fever and Leucocytosis •  Bleeding •  Sepsis •  Perforation (This is not an exhaustive list.)    Potential complications may be encountered in laser procedures, particularly if inappropriate Fluence settings are used.  Complications in extreme cases may include: •  Pain •  Perforation •  Oedema •  Erythema •  Crusting •  Hyper-pigmentation •  Hypo-pigmentation •  Scarring  (This is not an exhaustive list.)
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.11  Version B.2  CLINICAL WARNINGS  Diode  laser  radiation,  like  Nd:YAG  laser  radiation,  penetrates  significantly deeper  than  CO2  or  argon  lasers.    Caution  should  be  employed  until  the biological interaction of the laser energy with tissue is fully understood by the operator. Tissue damage could occur if excessive Power/Fluence is used.  Use low power and short pulse duration settings until fully familiar with instrument capabilities and tissue response.        As with any conventional surgical operations, adverse reactions may occur following treatment.      Use  cautiously  with  patients  who  have  had  difficulty  with  previous  laser procedures. The DIOMED DELTA 15/30 should be used only on tissue that is fully observable. Do not use the laser if the desired field is not visible, either directly or via an imaging modality such as ultrasound.   Do not use coaxial gas/air coolant for non-contact fibers when there is a risk of air/gas embolism. Do not use the laser close to large blood vessels or in highly vascularised areas, except when these are the target for the laser treatment. When performing endoscopic surgery it is vital for the surgeon to appreciate that the  view  provided  to  the  surgeon  is  monocular  (not  binocular)  and  depth perception  is  decreased.    Experience  and  training  in  laparoscopic  techniques are strongly recommended prior to clinical use. During ENT procedures, laser safe endotracheal tubes should be used.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 2.12  Version B.2  CLINICAL PRECAUTIONS General Precautions  Only operators who have  been trained in the use of lasers and are thoroughly familiar with this Operator Manual should use the DIOMED DELTA 15/30. The information provided in this section is not intended to be all-inclusive and it is not intended to replace operator training or experience. Please contact DIOMED Ltd. or your DIOMED DELTA 15/30 distributor for training materials available on the use of this equipment. Although it is difficult to specify the effect that the use of the diode laser will have in each therapeutic situation, it is possible to give a general overview as to what the  clinician  might  expect  when  using  the  DIOMED  DELTA  15/30.  The  exact effect depends upon the chosen procedure and, especially when using the laser with manual settings, the Power/Fluence setting, Pulse Duration, Pulse Interval, Spot size (if applicable) and the tissue type being treated.  The diode laser may cause tissue damage if improperly used. Precautions, such as  careful  assessment  of  the  target  tissue  during  treatment  and  the  use  of appropriate  Power/Fluence  and  Pulse  Duration,  should  be  taken.  Use  low Power/Fluence  and  short  Pulse  Duration  settings  until  fully  familiar  with  the instrument’s capabilities. Starting  at  low  powers,  the  operator  should  note  the  effect  on  the  tissue  and increase Power/Fluence, Pulse Duration or treatment time until the desired effect is obtained.  Specific parameters  are not  recommended, but  are  left  to operator  preference and best medical judgement dependent on the particular application. The diode laser may not be effective for coagulation for severe haemorrhages.  The operator must be prepared to control haemorrhages with strident, alternative non-laser techniques.  In contact surgery, the tissue interaction with the DIOMED DELTA 15/30 laser is similar to Nd:YAG laser.  In non-contact surgery the diode laser  wavelength, 810nm,  penetrates less  in most  pigmented tissue  types and blood than the Nd:YAG laser wavelength.  Precautions for use with a handpiece     Starting  at  low  powers,  the  operator  should  note  the  effect  on  the  tissue  and increase Fluence or Pulse Duration until the desired effect is obtained.  The  laser  can  cause  epidermal  injury.  The  risk  increases  with  greater  laser Fluence and skin pigmentation. Specific parameters  are not  recommended, but  are  left  to operator  preference and best medical judgement dependent on the particular application. It is strongly recommended that the physician should carry out a small, discrete test patch at the chosen settings prior to undertaking full treatment. Extreme care should be taken when patients have a recently acquired sun tan, or have a naturally dark skin colour.
 DELTA 15/30 Operator Manual  Page 2.13  Version B.2  SAFETY LABELLING Location of Safety Labelling Safety labels for the DIOMED DELTA 15/30 are positioned as indicated below.              Product Identification Labelling Product identification labelling is located on the rear of the DIOMED DELTA 15/30.  Diomed Delta LaserSerial Number: 000000Date of Manufacture: Sept 2005100-240V500VA  50-60Hz2 x T6.3A  H 250V0086IOIOIMade in the United Kingdom by:Diomed Ltd.Building 2000, Beach Drive,Cambridge Research Park,Waterbeach,Cambridge CB5 9TEDistributed in the USA byDiomed Inc, MA 01810This product complies with21 CFR 1040.10 and 1040.11, UL60601-1,EN60601-1, 60601-2-22, 60825-1IP21This device contains an intentional radiator part number This product complies with part 15 of the FCC Rules.Operation is subject to the following two conditions:(1) This device may not cause harmful interference, and(2) this device must accept any interference received,including interference that may cause undesired operation.This device complies with Canadian RSS-210AS/0670FCC ID: RBQAS0670Canada IC: 4690A-AS0670GROUNDING RELIABILITY CAN ONLY BEACHIEVED WHEN EQUIPMENT ISCONNECTED TO AN EQUIVALENTRECEPTACLE MARKED ‘HOSPITALONLY’ OR ‘HOSPITAL GRADE’PP/1149 STOPEN 60601-2-22 : 1996PP/0299 Iss2DIOMED DELTA 15 Safety Label DIOMED DELTA 30 Safety Label
 DELTA 15/30 Operator Manual  Page 2.14  Version B.2  SAFETY FEATURES  The  DIOMED  DELTA 15/30  includes  a  number  of  safety features,  which  are provided in accordance with the requirements of the appropriate standards. •  protective housing •  remote interlock bypass •  key switch •  laser radiation emission indicator, visible and audible • READY and STANDBY modes •  manual reset mechanism •  shutter (not mechanical) •  emergency switch •  location of controls •  safety labels (Figure 1) •  identification and compliance label (Figure 2) •  internal calibration port (optional on DELTA 15) •  calibration procedure for power measurement •  aiming beam   The  DIOMED  DELTA  15/30  is  equipped  with  the  following  additional  safety features:  •  self test •  laser condition monitoring •  pulse duration monitoring •  power diodes watch-dog •  microprocessor watch-dog •  mains power fail protection •  power supply monitor •  temperature monitors
 DELTA 15/30 Operator Manual  Page 2.15  Version B.2   EMC DECLARATION   Guidance and manufacturer’s declaration – electromagnetic emissions   The DIOMED DELTA 15/30 is intended for use in the electromagnetic environment specified below. The customer or user of the DIOMED DELTA 15/30 should ensure it is used in such an environment.    Emissions test  Compliance  Electromagnetic emissions – guidance   RF emissions CISPR 11   Group 1  The DIOMED DELTA 15/30 uses RF energy only for its internal function.  Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.   RF emissions CISPR 11   Class B   Harmonic emissions IEC 61000-3-2   Class A           Voltage fluctuations / flicker emissions IEC 61000-3-3    Not applicable     The DIOMED DELTA 15/30 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
 DELTA 15/30 Operator Manual  Page 2.16  Version B.2  Guidance and manufacturer’s declaration – electromagnetic immunity  The DIOMED DELTA 15/30 is intended for use in an electromagnetic environment specified below.  The customer or the user of the DIOMED DELTA 15/30 should assure that it is used in such an environment.   Immunity test  IEC 60601 test level  Compliance level  Electromagnetic environment – guidance Electrostatic discharge (ESD) IEC 61000-4-2 +/- 6 kV contact  +/- 8 kV air +/- 6 kV contact  +/- 8 kV air Floors should be wood, concrete or ceramic tile.  If floors are covered with synthetic material, the relative humidity should be at least 30%. Electrical fast transient / burst IEC 61000-4-4 +/- 2 kV for power supply lines  +/- 1 kV for input/output lines  +/- 2 kV for power supply lines  +/- 1 kV for input/output lines Mains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5 +/- 1 kV differential mode  +/- 2 kV common mode  +/- 1 kV differential mode  +/- 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment. Voltage Dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11  <5% UT  (>90% dip in UT) for 0,5 cycle  40 % UT  (90% dip in UT) for 5 cycles  70 % UT  (30 % dip in UT) for 25 cycles  <5% UT  (>95% dip in UT) for 5 sec  <5% UT  (>90% dip in UT) for 0,5 cycle  40 % UT  (90% dip in UT) for 5 cycles  70 % UT  (30 % dip in UT) for 25 cycles  <5% UT  (>95% dip in UT) for 5 sec Mains power quality should be that of a typical commercial or hospital environment.  If the user of the DIOMED DELTA 15/30 requires continued operation during mains interruptions, it is recommended that the DIOMED DELTA 15/30 be powered from an uninterruptible power supply or battery.  Power frequency (50/60 Hz) IEC 61000-4-8  3 A/m  3 A/m  Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.  NOTE    UT   is the a.c. mains voltage prior to application of the test level.
 DELTA 15/30 Operator Manual  Page 2.17  Version B.2   Guidance and Manufacturer's declaration – electromagnetic immunity The DIOMED DELTA 15/30 Laser is intended for use in an electromagnetic environment specified below.  The customer or the user of the DIOMED DELTA 15/30 Laser should assure that it is used in such an environment. Immunity test  IEC 60601 test level  Compliance level  Electromagnetic  environment  - guidance         Conducted RF IEC 61000-4-6  Radiated RF IEC 61000-4-3         3 Vrms 150 kHz to 80 MHz  3 V/m 80 MHz to 2,5 GHz                  3 Vrms   3 V/m Portable and mobile RF communications equipment should be used no closer to any part of the DIOMED DELTA 15/30, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance.  d  = 1.2 √P   d  = 1.2 √P     80MHz to 800 MHz  d  = 2.3  √P      800MHz to 2,5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters as determined by an electronic site survey. a   should be less than the compliance level in each frequency range.  b   Interference may occur in the vicinity of equipment marked with the following symbol:         NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
 DELTA 15/30 Operator Manual  Page 2.18  Version B.2  absorption and reflection from structures, objects and people.    a  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone and land mobile radios, amateur radio, AM and FM radio broadcast cannot be predicted theoretically with accuracy.  To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.  If the measured field strength in the location in which the DIOMED DELTA 15/30 is used exceeds the applicable RF compliance level above, the DIOMED DELTA 15/30 should be observed to verify normal operation.  If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the DIOMED DELTA 15/30. b Over the frequency range 150 kHz to 800 MHz, field strengths should be less than 3 V/m.
 DELTA 15/30 Operator Manual  Page 2.19  Version B.2   Recommended separation distances between portable and mobile RF communications and the DIOMED DELTA 15/30. The DIOMED DELTA 15/30 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.  The customer or the user of the DIOMED DELTA 15/30 can help prevent electromagnetic interference by maintaining distance between portable and mobile RF communications equipment (transmitters) and the DIOMED DELTA 15/30 as recommended below, according to the maximum output power of the communications equipment.  Separation distance according to frequency of transmitter m Rated maximum output power of transmitter W 150kHz to 80 MHz d = 1.2  √P 80MHz to 800MHz d = 1.2  √P 800 MHz to 2,5 GHz d = 2.3 √P 0.01  0.12  0.12  0.23 0.1  0.38  0.38  0.73 1  1.2  1.2  2.3 10  3.8  3.8  7.3 100  12  12  23  For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.  NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.  NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
 DELTA 15/30 Operator Manual  Page 2.20  Version B.2  ESD PRECAUTIONARY PROCEDURES  ESD (Electrostatic Discharge) occurs in air, causing a spark, when the potential difference between two bodies exceeds the dielectric strength of the air.  The DIOMED DELTA 15/30 has built-in protection from damage due to ESD, but no protection is 100% effective and precautions should be taken to protect the DIOMED DELTA 15/30 and any device connected to it.  When connecting the DIOMED DELTA 15/30 with another device, it is very important for the DIOMED DELTA 15/30, the device, and you to be at or close to the potential of the earth.  (1)  First, momentarily touch a grounded object to remove any existing static charge (2)  Connect one end of the 9-way interface lead to the DIOMED DELTA 15/30, taking care not to touch the pins of the connector (3)  Connect the other end of the 9-way interface lead to the device, taking care not to touch the male pins of the associated connector     FCC DECLARATION This  product  complies  with  part  15  of  the  FCC  Rules.  Operation  is  subject  to  the  following  two conditions:  (1)  This device may not cause harmful interference, and (2)  This  device  must  accept  any  interference  received,  including  interference  that  may  cause undesired operation.  This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.  Operation of this equipment in a residential area may cause harmful interference in which case the user will be required to correct the interference at their expense.  Modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment under FCC rules.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.1  Version B.2  SECTION 3  OPERATING INSTRUCTIONS INTRODUCTION The DIOMED DELTA 15/30 laser is capable of working with a wide range of fibers and accessories. The FRS system fitted as standard also allows it to identify the type of fiber or accessory that is connected. This enables the laser to automatically display suitable preset parameters with the minimum of user intervention. Therefore the menus and prompts displayed on the screen will vary depending on the automatically detected settings and it will not be possible to cover every possible scenario in this operator manual.  Depending on the configuration of the DIOMED DELTA 15/30 some features may be limited or prohibited. In these situations an explanation will normally be displayed on the screen but, if you have any doubt, please contact your DIOMED representative for advice.  The following instructions focus on the most common situations, when the DIOMED DELTA 15/30 is being used with a DIOMED procedure kit, bare fiber or spot handpiece, fitted with the FRS system.  CONTENTS Section 3 OPERATING INSTRUCTIONS ............................................................................................ 1 Introduction............................................................................................................................................. 1 Contents ................................................................................................................................................. 1 Conventions............................................................................................................................................ 2 Summary ................................................................................................................................................ 2 Installation and set-up ............................................................................................................................ 4 Front Panel Controls............................................................................................................................... 6 Rear Panel Controls ............................................................................................................................... 8 Fiber Recognition System ...................................................................................................................... 9 Operating Instructions .......................................................................................................................... 11 Spot Handpiece Mode.......................................................................................................................... 16 Manual Operation ................................................................................................................................. 20 Session Fault Emergency Override...................................................................................................... 23 Set-up Menu ......................................................................................................................................... 24 Session Statistics ................................................................................................................................. 27 Symbols used on Diomed products...................................................................................................... 28
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.2  Version B.2  CONVENTIONS The following naming conventions are used throughout these Operating Instructions  Controls on the front panel are expressed in bold capitals.  SCROLL, STANDBY Functional modes of the DIOMED DELTA 15/30 are expressed in capitals.  STANDBY, READY Menu items are expressed in bold mixed case.  Main Menu, Language  SUMMARY  1.  Place the DIOMED DELTA 15/30 in a convenient position on an instrument table no farther than 1.8 metres from the patient.  Ensure that all controls are within easy reach of the operator.  2.  Ensure that the ventilation holes in the base and rear of the DIOMED DELTA 15/30 are not obstructed.  3.  Connect the electrical power cord to the main power outlet.  4.  Connect the footswitch and place in a convenient position for the operator.  5.  Insert either a remote interlock bypass or, if required, connect the door interlock cable to the remote interlock socket on the rear of the DIOMED DELTA 15/30.    6.  Check that approved safety glasses are available and laser-warning signs are provided at entrances to the treatment room. All personnel present must wear approved safety glasses. DIOMED recommends that the patient’s eyes are taped shut if the patient is unconscious.   7.  Connect the optical fiber or Spot Handpiece to the laser aperture, ensuring that the connector is screwed ‘finger tight’.  8.  Turn on the rear power switch and key switch to activate the DIOMED DELTA 15/30. While the self-test is running, check that the front panel indicators light up and the audible indicator sounds momentarily.  9.  After the self-test, use the SCROLL/CONFIRM control to select the required procedure.  10. If required, follow the on-screen instructions to calibrate the fiber or Spot Handpiece.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.3  Version B.2   TO ENSURE BEST PERFORMANCE AND EFFICACY, ALWAYS USE DIOMED FIBERS, PROCEDURE KITS AND ACCESSORIES.   ONLY 600µµµµm NON-CONTACT FIBERS WITH A COOLING SHEATH CAN BE CALIBRATED.  CONTACT FIBERS DO NOT REQUIRE CALIBRATING.  THE DISPLAYED POWER WITH A CONTACT FIBER OR UNCALIBRATED NON-CONTACT FIBER IS THE POWER LEVEL AT THE LASER APERTURE. IT SHOULD BE ASSUMED THAT THE POWER LEVEL AT TISSUE IS 10-15% LOWER.    11. The system will automatically go to the STANDBY mode, with a set of default operating parameters. If required, adjust these parameters now using the SCROLL/CONFIRM control. The DIOMED DELTA 15/30 is now ready to begin the treatment.  12. To start treatment and delivery of laser energy, press STANDBY/READY, wait for the DIOMED DELTA 15/30 to enter READY mode and depress the footswitch.  During laser radiation the laser emission indicator will light and an audible warning will be heard.  13. To pause treatment, release the footswitch. To continue treatment, press the footswitch. To end treatment, release the footswitch and return the unit to STANDBY.    14. A summary of laser energy delivered may be reviewed if required by selecting Statistics from the Main Menu.  15. To turn the DIOMED DELTA 15/30 OFF turn the key switch and remove the key, then switch off the rear power switch.  If an error message is displayed, refer to section 5 - Technical Information.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.4  Version B.2  INSTALLATION AND SET-UP Installation of the DIOMED DELTA 15/30 can be carried out by the end-user. Inspection Inspect the DIOMED DELTA 15/30 and contents for signs of damage. If the unit is damaged DO NOT USE - contact DIOMED or your local DIOMED representative. If there are no signs of damage and all components are present, assemble the DIOMED DELTA 15/30.  Check that the following components are included in the packaging  DIOMED DELTA 15/30 Laser unit •  Footswitch •  IEC Power cable •  2 x Remote Interlock bypass connectors   •  2 x Keys •  4 x T6.3A fuses •  Operator Manual  The DIOMED DELTA 15/30 will operate at mains voltages between 100 V and 240 V AC without adjustment.  1.  Connect the footswitch to the footswitch socket (line up red dots and insert). 2.  Connect a remote interlock bypass connector to the remote interlock socket (line up red dots and insert).  3.  Connect the optical fiber or Spot Handpiece1 to the laser aperture as described in the section below. 4.  Insert the IEC Power cord into the power inlet socket and connect to the main power supply. 5.  Switch the power switch to ON ( | ).  6.  Insert a key into the key switch on the front of the unit.  7.  The DIOMED DELTA 15/30 is now installed and ready for use.  Connecting to the Laser Aperture Laser energy is delivered to the optical fiber or spot handpiece via the laser aperture located on the front panel of the DIOMED DELTA 15/30. The fiber or handpiece is connected by means of an SMA-905 type optical fiber connector.  To insert the optical fiber connector, first remove the protective cap from the end of the fiber. Then press down on the tab of the spring-loaded dust cover, insert the optical fiber connector into the laser aperture on the DIOMED DELTA 15/30 and turn the gripper clockwise until secured in place (light finger tight only).                                                       1 A Spot Handpiece can only be used with a DIOMED DELTA 15/30 laser fitted with an optical power calibration port. The Ø 2mm Spot Handpiece can be used with either laser but the Ø 4mm Spot Handpiece may only be used with the DIOMED DELTA 30.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.5  Version B.2     It is essential that the exposed end of the optical fiber be kept clean to prevent damage to the DIOMED DELTA 15/30 and optical fiber.    To remove the optical fiber connector, turn the gripper anti-clockwise until fully unscrewed and disconnect from the laser aperture. Dispose of the optical fiber according to institution policy. Alternatively, if the optical fiber is permitted for multiple uses, immediately fit a protective cap over the end of the optical fiber to protect the optical surface from contamination.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.6  Version B.2  FRONT PANEL CONTROLS                   The main operating controls for the DIOMED DELTA 15/30 are located on the lower section of the front panel. The display and other indicators are located in the top section of the panel, as illustrated in the figure above.  Key Switch  The key switch is used to start the DIOMED DELTA 15/30 and is the main control for the device.  The key is removable only in the OFF position and the laser is not operable when the key is removed. DIOMED recommends that the keys are assigned to one or two key-holders, who should keep the keys in a secure place and make them available for scheduled procedures only, thus preventing unauthorised use of the system. DIOMED also recommends that the key is not mixed with other keys on the same ring. Display Screen  This displays all menu options and information. Scroll / Confirm  To enable selection of Menu commands.  Turn the knob left or right to move between commands and press the knob to confirm the selection. Standby / Ready  To select STANDBY or READY mode.  Laser energy delivery is possible only in the READY mode.  When the READY request is made, the READY light flashes for two seconds before the system enters READY mode.  Pressing the button a second time will return the system to STANDBY mode. If the footswitch is pressed when a READY request is made or during the flashing of the READY light, the message ‘Footswitch held down’ is displayed and the footswitch should be released before the operation can continue.  The message will disappear when the footswitch is released. Standby Indicator  This light will be on when the laser is in STANDBY mode. Display Ready Indicator Scroll/Confirm Standby Indicator Standby/Ready Key Switch  Laser Aperture Emergency Stop Calibration Port (if fitted) Laser Emission Indicator Memory Card Slot Key Switch
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.7  Version B.2   Ready Indicator  This light will be on when the laser is in READY mode. Laser Emission Indicator  When laser energy is being delivered to the output port, this light will be on. Emergency Switch  To shut down the laser immediately in case of emergency, press the red button located on the front panel of the main enclosure.  After activation of the emergency switch, the key switch must be used to restart the system. Laser Aperture  Laser energy is delivered to the optical fiber or handpiece via the laser aperture located on the front panel. Slide the cover down to access the aperture. Use only DIOMED approved or DIOMED labelled optical fibers.  Damage caused by use of unapproved handpieces or fibers may invalidate the guarantee. Calibration Port  A calibration port is provided on the front panel of some models. This can be used, with the appropriate adaptor, to calibrate a fixed focus spot handpiece or an optical fiber.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.8  Version B.2  Power Inlet Socket, Switch & Fuseholder Remote Interlock Footswitch Socket Potential Equalisation Point Serial Port REAR PANEL CONTROLS                      Power Inlet Socket  To connect an IEC power cord. Power Switch  To switch the main power to the system on or off. Remote Interlock Socket  To connect the remote interlock cable connector.  This will automatically  switch the system to STANDBY mode in the event of the door being opened during the procedure.  If the remote interlock is connected to a door switch, then the cable used should be shielded and the shield connected to the plug body.  An EMC sleeve (ferrite tube) should also be fitted over the cable adjacent to the connector.  DIOMED can supply these on request.  These precautions will ensure that the possibility of electromagnetic emissions is minimised.    DIOMED supplies two remote interlock bypasses for facilities without or not wishing to use the door switch option.  The DIOMED DELTA 15/30 will not operate without the remote interlock bypass being inserted into the remote interlock socket on the rear of the laser.  Footswitch Socket  To connect the footswitch to the DIOMED DELTA 15/30. Serial Port IOIOI This connection is normally only used for diagnostic purposes by authorised DIOMED personnel.  Potential  To connect a potential equalisation line, for common grounding between Diomed Delta LaserSerial Number: 000000Date of Manufacture: Sept 2005100-240V500VA  50-60Hz2 x T6.3A  H 250V0086IOIOIMade in the United Kingdom by:Diomed Ltd.Building 2000, Beach Drive,Cambridge Research Park,Waterbeach,Cambridge CB5 9TEDistributed in the USA byDiomed Inc, MA 01810This product complies with21 CFR 1040.10 and 1040.11, UL60601-1,EN60601-1, 60601-2-22, 60825-1IP21This device contains an intentional radiator part number This product complies with part 15 of the FCC Rules.Operation is subject to the following two conditions:(1) This device may not cause harmful interference, and(2) this device must accept any interference received,including interference that may cause undesired operation.This device complies with Canadian RSS-210AS/0670FCC ID: RBQAS0670Canada IC: 4690A-AS0670GROUNDING RELIABILITY CAN ONLY BEACHIEVED WHEN EQUIPMENT ISCONNECTED TO AN EQUIVALENTRECEPTACLE MARKED ‘HOSPITALONLY’ OR ‘HOSPITAL GRADE’PP/1149
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.9  Version B.2  Equalisation Point  equipment(s), if needed.   Fuseholder  2 x T6.3A H 250V   Use of controls or adjustments, and or performance of procedures other than those specified herein may result in hazardous radiation exposure. FIBER RECOGNITION SYSTEM What is the Fiber Recognition System (FRS)? All DIOMED DELTA 15/30 lasers are equipped with the DIOMED Fiber Recognition System (FRS). This system provides a means of identifying the fiber that is connected to a DIOMED DELTA 15/30. DIOMED fibers are normally supplied in a kit of components for a particular procedure. Therefore, by identifying the fiber, the laser can also tell what procedure is going to be performed. So the DIOMED DELTA 15/30 can automatically select suitable preset parameters with the optimum values for that procedure. Compared to setting the parameters manually, this process is much quicker and far less prone to user error.  Some versions of the DIOMED DELTA 15/30 can also operate with non-DIOMED fibers, as long as they are fitted with a DIOMED adapter at the laser end of the fiber. This adapter is available as an accessory from your DIOMED representative and it will allow the DIOMED DELTA 15/30 to be used in Manual mode only. How does FRS work? A miniature Radio Frequency Identification (RFID) device is located inside the gripper at the end of the fiber connected to the laser. This is read by a receiver inside the DIOMED DELTA 15/30 whenever it is switched on and a fiber connected to it. The RFID device contains a memory chip that holds the following information:  The type of fiber  This tells the system what type of fiber or accessory has been connected to it.2 The date when the sterility of the fiber expires If the sterility expiration date is earlier than the current date in the internal clock of the DIOMED DELTA 15/30 then the fiber is invalid and cannot be used. How many times the fiber may be safely used The number of “uses” 3 is decreased each time the system is used with the connected fiber. When the figure reaches zero, the fiber becomes invalid and cannot be used again.                                                       2 The system will read the fiber’s data if the power is on and a FRS fiber is connected or if the DIOMED DELTA 15/30 is powered up with a FRS fiber already connected.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.10  Version B.2   What are the benefits of FRS? •  Single-use fibers cannot be re-used, as sterility and optical performance cannot be assured. •  Fibers  that  have  passed their  sterility  expiration  date  cannot  be  used, minimizing  the  risk  of patient infection. •  The number of available “uses”, subject to satisfactory calibration, for Multiple-use and Spot Handpiece fibers is displayed on the screen, giving a clear indication of when a fiber is due for replacement. •  Operation of the DIOMED DELTA 15/30 is simplified, because the FRS is able to automatically recognize devices independently of the user – for example, the type of Spot Handpiece.  In order to obtain full advantage of DIOMED FRS you should always use fibers supplied by DIOMED or fibersdirect.com. Please contact your DIOMED representative for a list of fibers and procedure kits that are compatible with the FRS system.                                                                                                                                                                         3 The number of available “uses” is only decreased once per procedure and only after the DIOMED DELTA 15/30 has been fired. Subsequent firing of the DIOMED DELTA 15/30 during a continuous procedure will not result in the number of available “uses” being further decreased.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.11  Version B.2  OPERATING INSTRUCTIONS Once the DIOMED DELTA 15/30 has been correctly installed and switched on, it is operated using just two controls:  SCROLL / CONFIRM  Turn the knob in either direction to SCROLL through the available options highlighted on the screen. Press the knob to CONFIRM the selection.  STANDBY / READY  Once the DIOMED DELTA 15/30 is set up and ready for the procedure, press this button to switch between the STANDBY   and READY   modes.  Laser energy delivery is possible only in the READY mode. Switching On Ensure that the DIOMED DELTA 15/30 has been set up correctly, as described above and that the remote interlock, footswitch and fiber connectors are all in place. Switch on the rear panel power switch. The display will show a screen similar to the one on the right.  Turn the key switch clockwise to activate the system. It will now perform a self-test function for a few seconds. While this is running, ensure that the indicators for Laser Emission, Standby and Ready are all illuminated and that the audible indicator sounds briefly.    Setting the language at switch-on The system will now prompt for the language to be used by the user interface. If you wish to change it, turn the SCROLL knob until Language is highlighted and press to CONFIRM. You can now use the SCROLL knob to select the required language from the drop down list. Press again to CONFIRM the selection. When Set at start is set to Yes, this screen will occur every time that the DIOMED DELTA 15/30 is switched on. To prevent this, change the selection to No as follows:  •  Turn the SCROLL knob until Set at start is highlighted. •  Press CONFIRM to toggle between Yes and No. •  Turn the SCROLL knob to select Accept. •  Press CONFIRM.  You will still be able to change the language setting from the Setup menu option on the Main menu.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.12  Version B.2   Interlock Checks The system will now check that all the safety interlocks are properly in place.  If a tick is not shown against one or more of the items in the display, recheck the appropriate connection.    Fiber Identification & Validation After it has finished checking the interlocks, the DIOMED DELTA 15/30 system reads the FRS information stored inside the fiber’s gripper. It will check the type of fiber, how many “uses” are available and whether it is within its sterility expiration date. This information will be displayed on the screen for a few seconds.    If the DIOMED DELTA 15/30 does not detect the presence of a valid DIOMED FRS fiber or adapter then it will not permit a procedure to be carried out. An error message will be displayed on the screen – see Troubleshooting in the Technical Information section.   An Emergency Override option is included to allow the DIOMED DELTA 15/30 to be used once only in the event of a session fault, which has prevented the completion of a treatment. See Session Fault Emergency Override below for instructions on how to use this option.  Once the fiber is validated, the DIOMED DELTA 15/30 is able to determine which procedures may be performed and can set itself up with the appropriate parameter values for power, pulse duration and interval. Depending on the type of fiber connected, a screen similar to the following will be displayed.            Procedure Selection
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.13  Version B.2  In the following example, a fiber from a DIOMED EVLT procedure kit has been connected to the laser. The DIOMED DELTA 15/30 recognises this fiber and loads a list of compatible procedures into the top, as shown on the display4.  The default procedure is shown at the top of the menu. To select a different procedure turn the SCROLL knob until Procedure is highlighted and press CONFIRM. Select the desired procedure with the SCROLL knob and press CONFIRM again.  When the procedure has been selected, highlight Run using the SCROLL knob and press CONFIRM.  Parameter Adjustment At this point, calibration may be required for certain fibers and procedures. In this case, follow the instructions on page 14.  The DIOMED DELTA 15/30 will set the default operating parameters for the selected procedure. Some procedures also allow the default parameters to be varied within lower and upper limits, which are defined in the procedure’s specification.  To adjust a parameter, turn the SCROLL knob to highlight the required parameter (Power, Pulse or Interval) and press CONFIRM. The highlight will now move to the value of the parameter on the right hand side of the display. Turn the SCROLL knob to adjust the value and press CONFIRM again. The highlight will move back to the left hand side of the display. Repeat this process for the other parameters as required.  When you have completed setting up the operating parameters of the DIOMED DELTA 15/30, turn the SCROLL knob to move the highlight to Continue and press CONFIRM.    The method of adjusting settings will vary depending on the procedure selected. Please refer to section 4 entitled DIOMED Procedures for full details of how to use specific procedures.                                                       4 The list of procedures shown on this menu will only include those that are installed in the laser. DIOMED will, over the lifetime of this system, release new procedures that may be compatible with currently available fibers. Please check with your DIOMED representative for the latest information regarding available procedures. Further information on the currently available procedures and the methods of installing them into the DIOMED DELTA 15/30 can be found in section 4.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.14  Version B.2   Run the Procedure The DIOMED DELTA 15/30 is now ready to start the procedure. Press STANDBY/READY and wait for the DIOMED DELTA 15/30 to enter the READY mode. Depress the footswitch. An audible warning will be heard during laser irradiation and the laser emission indicator will be lit. To pause treatment, release the footswitch. Press the footswitch again to continue. To end treatment, release the footswitch and return the unit to STANDBY.  Note: the DIOMED DELTA 15/30 will automatically return to STANDBY if treatment is paused for 3 minutes.  To turn the DIOMED DELTA 15/30 off, turn the key switch anti-clockwise, remove the key and switch off at the rear panel.   Calibration Calibration is not a requirement for all procedures. However, if necessary, calibration is normally performed after the procedure has been selected. If appropriate, calibration can also be performed by selecting the Calibration option at the main menu screen.  This feature is not available on the DIOMED DELTA 15 unless fitted with the optional Optical Power Calibration Port.   This calibration procedure is for 600µm non-contact fibers with a cooling sheath only. Contact fibers cannot be calibrated.  Ensure all personnel are wearing approved safety eyewear.  When calibration is carried out, the displayed power will be the power at the distal end of the fiber.  If calibration is NOT carried out, the displayed power will be the power at the laser aperture.  Instructions on how to perform the calibration routine are displayed on the screen. Selecting Cancel at any time will return the user to the Main Menu leaving the fiber un-calibrated.  1.  Screw in the optical fiber calibration port adapter. (The adapter should be sterilised in order to maintain fiber sterility).
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.15  Version B.2  2.  Guide the fiber into the calibration port using the optical fiber calibration port adapter provided.  3.  Place the DIOMED DELTA 15/30 in READY mode. 4.  Follow the instructions on screen. 5.  Fire the DIOMED DELTA 15/30 by pressing the footswitch until release is indicated on the display. 6.  The screen will display the fiber acceptability and transmission percentage. 7.  To proceed, highlight Continue and press SELECT.  Transmission Acceptability  The DIOMED DELTA 15/30 will calculate the percentage transmission of the fiber system, and the results will be shown on screen as Acceptable or Unacceptable.  •  A percentage rating of 75% - 100% indicates that the fiber is Acceptable, and ‘Cal’ is displayed in the top right corner of the display showing that the fiber has been calibrated.  The power displayed will represent the power output at the distal end of the calibrated fiber and the system will compensate for any fiber transmission losses. •  A percentage rating of 0% - 75% indicates that the fiber is Unacceptable.  In this case the fiber must be replaced with a new fiber and re-calibrated.  Calibration will be cancelled if:  •  The unit is turned off or  •  The fiber port interlock is activated (i.e. the user has changed the fiber).
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.16  Version B.2  SPOT HANDPIECE MODE This section includes special instructions to be followed when using a spot handpiece with the DIOMED DELTA 15/30.  Spot Handpiece mode is only available when the DIOMED DELTA 15/30 is fitted with a calibration port. This port is optional on the DIOMED DELTA 15.  Fixed Focus Spot Handpiece   The handpiece delivery system consists of a sleeved optical fiber, with an optical fiber connector at one end and a handpiece at the other. The handpiece has a user replaceable protective lens.  A Ø 2mm fixed focus Spot Handpiece is available separately for use with the DIOMED DELTA 15/30. The Ø 2mm fixed focus Spot Handpiece, colour-coded blue for identification, delivers a 2mm diameter beam of laser energy.     Ø 2mm fixed focus Spot Handpiece   A Ø 4mm fixed focus Spot Handpiece (not illustrated) is available separately for use with the DIOMED DELTA 30 only. The Ø 4mm fixed focus Spot Handpiece is colour-coded yellow for identification and delivers a 4mm diameter beam of laser energy.  New handpieces are supplied with the fiber fitted and with an integrated FRS gripper. If you have an older Spot Handpiece, without a FRS gripper, please contact your local DIOMED representative who will be able to supply you with a new fiber.                 Great care must be taken in ensuring optical faces are kept clean, particularly at the optical fiber connector end.  A protective cap is provided which should be replaced each time the optical fiber end of the fiber is not connected to the Laser aperture.    Before using the handpiece inspect the fiber and connector for signs of damage and check the handpiece to ensure it is clean, correctly assembled and has no signs of damage.  If there is any evidence of handpiece damage or fiber breakage do not use the handpiece as it may cause injury to the operator or patient. Replace the handpiece or fiber immediately.  (See fixed focus Spot Handpiece Instructions for Use if the fiber requires changing).
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.17  Version B.2   The Ø 2mm fixed focus Spot Handpiece delivers power densities four times greater than the Ø 4mm fixed focus Spot Handpiece at the same power/duration settings.  Handpiece Identification After it has finished checking the interlocks, the system will read the FRS information stored inside the handpiece’s gripper. From this the laser is able to determine whether it is a Ø 2mm or Ø 4mm handpiece. Depending on the type of handpiece connected, a screen similar to the following will be displayed.   If the laser does not detect the presence of a valid FRS Spot Handpiece then it will not permit a procedure to be carried out. Please refer to the Troubleshooting section for help in this situation.    Calibration  Calibration of a Spot Handpiece must be performed either each time the unit is switched ON or each time a new handpiece is connected.  This ensures that accurate Fluence is available at the treatment site.  The DIOMED DELTA 15/30 system will always perform a calibration if the handpiece has not been calibrated.  The Spot Handpiece will require re-calibration if any of the following occur during the treatment session: •  Power failure / unit switched off •  Footswitch disconnected •  Emergency switch pressed •  Handpiece is disconnected •  A new handpiece is used   Ensure all personnel are wearing approved safety glasses.  Ensure that the window cell / treatment window of the handpiece to be calibrated is clean.  Instructions on how to perform the calibration routine are displayed on the screen. Selecting Cancel at any time will return the user to the Main Menu leaving the Spot Handpiece un-calibrated.  1.  Screw in the Spot Handpiece calibration port adapter. 2.  Insert the Spot Handpiece into the calibration port adapter. 2mm spot
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.18  Version B.2  3.  Ensure that the probe on the Spot Handpiece lines up with the hole on the adapter and is inserted fully into the calibration port. 4.  Place the DIOMED DELTA 15/30 in READY mode. 5.  Follow the instructions on screen. 6.  Fire the DIOMED DELTA 15/30 by pressing the footswitch until release is indicated on the display. 7.  The screen will display the handpiece acceptability and transmission percentage. 8.  To proceed, highlight Continue and press SELECT.  Transmission Acceptability The DIOMED DELTA 15/30 will calculate the percentage transmission of the handpiece system, and the results will be shown on screen as Acceptable or Unacceptable.  Transmission should be 65% or greater. If less than 65% the result will be shown as Unacceptable and the handpiece will need cleaning or replacing if damaged (see Fixed Focus Spot Handpiece Instructions for Use supplied with the handpiece).  After carrying out the calibration procedure the DIOMED DELTA 15/30 will display that the handpiece has been calibrated.  Fluence After successful calibration of a Spot Handpiece, the Fluence screen will be displayed.  The Fluence function automatically calculates and displays the laser energy per unit area (J/cm2) projected onto the patient. If the Power and Pulse duration are adjusted, the Fluence will be automatically re-calculated and the display updated.  Fluence (J/cm2) is calculated using the formula:      where Spot Radius = 0.1 cm for a Ø 2 mm spot handpiece.  To adjust a parameter, turn the SCROLL knob to highlight the required parameter (Power, Pulse or Interval) and press CONFIRM. The highlight will now move to the value of the parameter on the right hand side of the display. Turn the SCROLL knob to adjust the value and press CONFIRM again. The highlight will move back to the left hand side of the display. Repeat this process for the other parameters as required.  When you have set up the required value of fluence of the DIOMED DELTA 15/30, turn the SCROLL knob to move the highlight to Continue and press CONFIRM.  )()()()(2cmRadiusSpotxsDurationPulsexWPowerFluenceπ=
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.19  Version B.2   The DIOMED DELTA 15/30 is now ready to start the procedure. Press STANDBY/READY and wait for the DIOMED DELTA 15/30 to enter the READY mode. Depress the footswitch. An audible warning will be heard during laser irradiation and the laser emission indicator will be lit. To pause treatment, release the footswitch. Press the footswitch again to continue. To end treatment, release the footswitch and return the unit to STANDBY.  Note: the DIOMED DELTA 15/30 will automatically return to STANDBY if treatment is paused for 3 minutes.  To turn the DIOMED DELTA 15/30 off, turn the key switch anti-clockwise, remove the key and switch off at the rear panel.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.20  Version B.2   MANUAL OPERATION Manual operation is not enabled on all versions of the DIOMED DELTA 15/30. Please consult with your DIOMED representative if you require assistance with setting up your laser.  For best results always use the DIOMED DELTA 15/30 with a DIOMED FRS fiber. This will be detected by the laser, which will automatically select suitable preset parameters. However it is also possible to manually configure the settings DIOMED DELTA 15/30 in the following situations:  •  Some DIOMED FRS fibers allow free setting of parameters. This is enabled by selecting Manual from the Procedure menu. •  DIOMED supply a general purpose bare-ended fiber fitted with FRS, that sets the DIOMED DELTA 15/30 to use Manual settings. •  Third party fibers5 may also be used with the DIOMED DELTA 15/30. But these must first be fitted with a special DIOMED FRS adapter, available from your DIOMED representative.   If a DIOMED FRS fiber is detected but no compatible procedure is installed, please contact your local DIOMED representative who will advise on how you can update the procedures installed in your DIOMED DELTA 15/30.  Parameters Table 1.  The following parameters may be adjusted manually. The available modes are:  Continuous (not with a Spot Handpiece)  Pulse Mode   Repeat Pulse Power (W)  Power may be adjusted in all operating modes:  Fiber: 0.5W to 5W in 0.5W increments 5W to maximum W in 1W increments  Spot Handpiece: 0.5W to maximum W in 0.1W increments  Pulse Duration (s)  The pulse duration may be adjusted in Pulse and Repeat pulse modes:  Fiber:                                                       5 DIOMED do not recommend the use of third party fibers as their quality and efficacy cannot be guaranteed. Any damage caused to your DIOMED DELTA 15/30 by using a fiber not supplied by DIOMED may not by covered under the DIOMED warranty.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.21  Version B.2  0.1 to 9.9 seconds in 0.1 second increments in single pulse mode 0.1 to 1.0 seconds in 0.1 second increments in repeat pulse mode  Spot Handpiece: 50 to 250 ms in 10 ms increments in single or repeat pulse mode  Interval (s)  The interval between the pulses may be adjusted only in the Repeat pulse mode:  0.1 to 1.0 seconds in 0.1 second increments Timer (s)  Provides a countdown timer in Continuous and Repeat Pulse modes:   OFF to 120 seconds  in 10 second increments 120 to 9999 seconds in 30 second increments (final increment of 9 seconds)  Table 2.  The default parameters for each available mode are: Mode  Power  Pulse Duration  Pulse Interval  Timer Continuous  5 W  Not available  Not available  OFF Pulse  5 W  1 sec  Not available  Not available Repeat Pulse  5 W  1 sec  1 sec  OFF  Selecting Manual Operation Manual operation allows the operator complete freedom over the settings of the DIOMED DELTA 15/30 and is enabled by selecting Manual from the Procedure menu.   Setting the Mode The default mode is shown at the top of the menu. To select a different mode turn the SCROLL knob until Mode is highlighted and press CONFIRM. Select the desired mode with the SCROLL knob and press CONFIRM again.  When the mode has been selected, highlight Continue using the SCROLL knob and press CONFIRM.   Setting Power, Pulse Duration and Interval Depending on the Mode selected, the next screen will show a set of default parameters, as described in table 2 above. To adjust a parameter, turn the SCROLL knob to highlight the required parameter and press CONFIRM. The highlight will now move to the value on the right hand side of the display. Turn the SCROLL knob to adjust the value within the range shown in table 1 and press CONFIRM. The highlight will move back to the left hand side of the display. Repeat this process for the other parameters as required.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.22  Version B.2   When you have completed setting up the operating parameters, turn the SCROLL knob to highlight Continue and press CONFIRM.   When the operating Mode is Continuous, the Pulse & Interval parameters are not applicable and cannot be adjusted. Similarly in Pulse mode, the Interval parameter cannot be adjusted and the Timer is not available. Start the Procedure The DIOMED DELTA 15/30 is now ready to start the procedure. Press STANDBY/READY and wait for the DIOMED DELTA 15/30 to enter the READY mode. Depress the footswitch. An audible warning will be heard during laser irradiation and the laser emission indicator will be lit. To pause treatment, release the footswitch. Press the footswitch again to continue. To end treatment, release the footswitch and return the unit to STANDBY.  To turn the DIOMED DELTA 15/30 off, turn the key switch anti-clockwise, remove the key and switch off at the rear panel.  Countdown Timer Countdown mode is enabled by setting the value of the Timer parameter to any value except OFF. This mode allows: •  Longer CW exposure times than the maximum 9.9 seconds allowed when using the Pulse mode •  A train of pulses to be output for a fixed time in Repeat Pulse mode  The Timer is not available if the Pulse mode is selected. The default time is 60 seconds but the time can be set to any value between 10 and 3200 seconds.   When the Timer is active and the footswitch is pressed to start treatment, the display will show a countdown until zero is reached or the footswitch is released. To restart the countdown after the footswitch has been released press the footswitch and the countdown will resume. When the countdown reaches zero the system will automatically enter STANDBY mode.  The countdown time cannot be adjusted after the footswitch has been pressed unless the user returns to the Manual Control menu, or the countdown time has been completed.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.23  Version B.2  SESSION FAULT EMERGENCY OVERRIDE In extreme circumstances, such as a power cut or a technical problem, the DIOMED DELTA 15/30 will need to be restarted and will log a Session Fault. If this occurs during a treatment then the fiber in use may become invalid before the completion of the treatment.  In these circumstances, the Emergency Override may be activated to allow treatment to continue. This option is only available immediately after the Session Fault has been cleared and will not work if an attempt is made to use an expired or otherwise invalid fiber at any other time.  Select Emergency Override to use this option or Main Menu to cancel.   If the Emergency Override has already been used, then this option is not available.   Resetting the Emergency Override The Emergency Override can be reset by means of a special reset device (DIOMED part number AS/598). One of these is supplied with the DIOMED DELTA 15/30.  Each device can only be used once, after which it is advisable to contact your DIOMED representative as soon as possible to obtain a replacement.  To reset the Emergency Override:  •  Switch on the laser •  Attach the device by screwing it on to the laser aperture port, in the same way as you would attach a fiber •  The DIOMED DELTA 15/30 will detect the device and reset the Emergency Override and display a message on the screen
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.24  Version B.2  SET-UP MENU The Set-up Menu is available from the Main Menu. It allows the operator to customise certain properties of the DIOMED DELTA 15/30.  To access any of the functions within the Set-up Menu turn the SCROLL knob until the function is highlighted and then press CONFIRM.  To return to the Set-up Menu from any of these functions, highlight Accept and press CONFIRM.   All of the changes made from this menu will be retained when the DIOMED DELTA 15/30 is switched off.  Language To change the language used for the user interface, turn the SCROLL knob to highlight Language and press CONFIRM. Now use the SCROLL knob to select the required language from the drop down list and press CONFIRM. When Set at start is set to Yes, this language selection screen will occur every time that the DIOMED DELTA 15/30 is switched on. This option can be toggled between Yes and No by using the SCROLL knob to highlight Set at start and pressing CONFIRM.  Audio Volume To adjust the volume of the audible indicator heard when the laser is firing, turn the SCROLL knob to highlight Adjust and press CONFIRM. Now use the SCROLL knob to select the required volume. Press CONFIRM when finished.   If the volume is set to 0% the audible indicator is switched off. In this case the visible laser emission indicator is the sole indication of laser activity  The audible indicator can also operate in two modes, which can be selected by choosing Sound with the SCROLL knob: • Pulsed – the audible indicator pulses in time with the laser output (as long as the pulses are long enough to be discernable) • Continuous – the audible indicator sounds continuously when the laser is firing
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.25  Version B.2   Display Brightness To adjust the brightness of the display, turn the SCROLL knob to highlight Adjust and press CONFIRM. Now use the SCROLL knob to select the required brightness. Press CONFIRM when finished.  Aiming Beam Mode The aiming beam can operate in three modes, which can be selected by choosing Mode with the SCROLL knob: • Off – the aiming beam is switched off • Continuous – the aiming beam is on continuously when the laser is firing • Follow laser - the aiming beam pulses in time with the laser output (as long as the pulses are long enough to be discernable)  The Follow laser mode is a useful aid for the operator as it causes the aiming beam to mimic the characteristics of the procedure. For example, if the laser output is pulsing one second on and one second off then the aiming beam will pulse at the same rate. Similarly, if the intensity of the laser output varies then so will the aiming beam’s intensity. In this mode, the aiming beam’s output will always provide a visual indication that the laser is active, even if the output is very low or very short, infrequent pulses.    If using the laser in non-contact mode, it is advisable to have the aiming beam switched on to identify the target tissue that will be affected by the laser energy.  To observe the different modes of the aiming beam, place the DIOMED DELTA 15/30 into READY mode by pressing the STANDBY/READY button.  The aiming beam will be activated if Continuous or Follow Laser are selected.  For safety, it is not possible to fire the laser in READY state at this menu.  Intensity To adjust the intensity of the aiming beam, turn the SCROLL knob to highlight Intensity and press CONFIRM. Now use the SCROLL knob to select the required intensity. Press CONFIRM when finished.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.26  Version B.2    The intensity adjustment will be ignored if the aiming mode is Off.  To observe the intensity of the aiming beam, place the DIOMED DELTA 15/30 into READY state by pressing the STANDBY/READY key.  The aiming beam will be activated.  For safety, it is not possible to fire the laser in READY state at this menu.  Time & Date To set the internal clock and calendar of the DIOMED DELTA 15/30 turn the SCROLL knob to highlight Adjust and press CONFIRM. Turn the SCROLL knob to change the highlighted value and press CONFIRM to accept and move on.  The order of adjustment is Day, Month, Year, Hours & Minutes.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.27  Version B.2  SESSION STATISTICS Session Statistics can be defined as a summary of the laser energy delivered and are recorded for the time that the DIOMED DELTA 15/30 is switched on.  Session Statistics will be displayed as the amount of joules of energy delivered.   Each time the treatment parameters are changed, the Session Statistics will be shown on a separate line of the screen.  The total amount of laser energy used in this session will be displayed on the screen.   Session Statistics are stored in the DIOMED DELTA 15/30 internal memory. They may also be transferred to a PC using a MMC memory card.   To minimise the risk of error, it is advisable to transfer Session Statistics from the DIOMED DELTA 15/30 to a PC immediately after the procedure has been completed.  The Statistics menu enables the operator to view the session statistics from the current or previous sessions. To access this menu turn the SCROLL knob until Statistics is highlighted in the Main Menu and then press CONFIRM. To access any of the functions within the Session Statistics menu turn the SCROLL knob until the function is highlighted and then press CONFIRM.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 3.28  Version B.2  SYMBOLS USED ON DIOMED PRODUCTS The following symbols are used on the DIOMED DELTA 15/30 laser and on accessories provided by DIOMED.  Refer to instructions  Refer to instructions (alternative)  Power Off  Power Off (only for a part of equipment)  Power On  Type B applied part  Intentional radiator  The component or accessory is non-sterile  Single-use  Do not use if packaging is damaged  Expiry date  Batch number  Product re-order code  Sterile by Ethylene Oxide
DELTA 15/30 DELTA 15/30 Operator Manual  Page 4.1  Version B.2  SECTION 4  PROCEDURES  The DIOMED DELTA 15/30 lasers are optimised to operate with DIOMED fibers and procedure kits that are equipped with the DIOMED Fiber Recognition System (FRS). For a full description of this system please see section 3 of this manual.  This section describes the FRS fibers, procedure kits and other accessories that are currently available from your DIOMED representative.  DIOMED’s policy of continual product development and improvement means that fibers and procedure kits may be added to the product range at any time. Therefore the information in this section may change without notice. Please contact your DIOMED representative to obtain the latest product information.   CONTENTS  Section 4 PROCEDURES .................................................................................................................... 1 Contents ................................................................................................................................................. 1 Diomed Fibers and Procedures.............................................................................................................. 2 Ordering Information............................................................................................................................... 2 EVLT....................................................................................................................................................... 3 Spot Handpieces .................................................................................................................................... 4 Bare-ended Fiber.................................................................................................................................... 4 Manual Adapter ...................................................................................................................................... 4 Test Fiber................................................................................................................................................ 4 Emergency Override Reset Device ........................................................................................................ 4
DELTA 15/30 DELTA 15/30 Operator Manual  Page 4.2  Version B.2   DIOMED FIBERS AND PROCEDURES When a DIOMED fiber fitted with FRS is connected to a DIOMED DELTA 15/30 laser, the laser will attempt to recognise the type of fiber from the information stored in the FRS gripper. If successful, it will cross-reference the type of fiber with an internal database of compatible procedures. Any procedures that are compatible with the connected fiber are listed in the Procedure menu and may be selected by the operator.  From time to time, Diomed will release new fibers, and procedure kits. In order to use these, your DIOMED DELTA 15/30 will need to be updated with the latest procedure information.   ORDERING INFORMATION The following order codes should be quoted when ordering replacement procedure kits, fibers and accessories for your DIOMED DELTA 15/30.  Description  DIOMED or fibersdirect.com part number EVLT Procedure Kit  EVLT/PK-02 EVLT Procedure Kit – Extra Length  EVLT/PK-02-EL Micro-Introducer Kit  EVLT/PK/MIK 2mm Spot Handpiece replacement fiber only  AS/0618 4mm Spot Handpiece replacement fiber only  AS/0619 Manual Adapter  AS/0686 Test fiber  AS/0604 Emergency Override Reset Device  AS/0598 MMC Multimedia Memory Card  S10/06/0086 USB MMC Card Reader/Writer  S10/06/0087 DIOMED Key Fob / MMC Card Holder  AS/0685  This list was correct at the time of publication but new procedure kits, fibers and accessories may be added at any time. Please check with your local DIOMED representative for the latest information.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 4.3  Version B.2  EVLT There are three procedures that are compatible with the EVLT procedure kit. Two of these are designed to cover the currently-approved methods for performing an EVLT procedure. The third is intended to allow the operator some freedom in setting parameters manually.  Note: If the DIOMED DELTA 15/30 laser is also enabled for Manual operation, then it may also be configured for any parameters within its design specifications.  EVLT Continuous The laser output power is fixed at a continuous 14 W. No adjustments can be made. EVLT Pulsed The laser output power is fixed at 12 W with a 1 second pulse and 1 second interval. No adjustments can be made. EVLT Manual The laser output power may be varied between 5 W and 15 W. The mode may be switched between continuous output and 1 second pulses with 1 second interval.  These three procedures are summarised in the following table:    EVLT Continuous  EVLT Pulsed  EVLT Manual Power (W)  14 W  12 W  5 W to 15 W Pulse Duration (s)  1 second Interval (s)  Continuous 1 second Continuous or 1 second Pulse / 1 second Interval  Selecting the Procedure The default procedure is shown at the top of the menu. To select a different procedure turn the SCROLL knob until Procedure is highlighted and press CONFIRM. Select the desired procedure with the SCROLL knob and press CONFIRM again.  When the required procedure has been selected, highlight Run using the SCROLL knob and press CONFIRM.    Adjusting an EVLT procedure The EVLT Continuous and EVLT Pulsed procedures use preset parameters and cannot be adjusted. The EVLT Manual procedure allows the power to be adjusted between 5 W and 15 W and the mode
DELTA 15/30 DELTA 15/30 Operator Manual  Page 4.4  Version B.2  may be switched between Continuous and Pulsed. The pulse duration and interval are preset to 1 second.  Instructions on how to select mode and adjust power are included in section 3 – Operating Instructions. SPOT HANDPIECES When a DIOMED spot handpiece is connected to the DIOMED DELTA 15/30, FRS will recognise the type of handpiece and automatically displays suitable preset parameters. Follow the operating instructions in section 3. BARE-ENDED FIBER When a DIOMED bare-ended fiber is connected to the DIOMED DELTA 15/30, FRS will recognise the type of fiber and automatically select the manual mode for the laser. Follow the operating instructions in section 3. MANUAL ADAPTER If you wish to use a fiber not supplied by DIOMED then, in certain circumstances, it is possible to use a special adapter to enable the fiber to be accepted by the Fiber Recognition System. The DIOMED DELTA 15/30 will then be enabled for Manual operation.  For details of this adapter please contact your DIOMED representative. TEST FIBER The test fiber is a special type of FRS fiber used during testing and evaluation only. EMERGENCY OVERRIDE RESET DEVICE The Emergency Override Reset Device consists of a special FRS gripper, supplied without a fiber. Instructions on when and how to use this device are in section 3.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.1  Version B.2  SECTION 5  TECHNICAL INFORMATION  The DIOMED DELTA 15/30 has been designed to operate reliably with minimal maintenance and there are no user serviceable parts.  This section includes technical data and describes the routine maintenance procedures that you can perform on the DIOMED DELTA 15/30 and its accessories.     Any attempts to repair, adjust or modify the laser beyond the procedures allowed in this Operator Manual, by any person not authorised by DIOMED, will invalidate the guarantee.  CONTENTS Section 5 TECHNICAL INFORMATION............................................................................................... 1 Contents ................................................................................................................................................. 1 Specifications.......................................................................................................................................... 2 Classification........................................................................................................................................... 3 Cleaning the DIOMED DELTA 15/30 ..................................................................................................... 3 Caring for fibers ...................................................................................................................................... 3 Troubleshooting...................................................................................................................................... 4 Accessories ............................................................................................................................................ 7 Sterilisation of Optical Fibers.................................................................................................................. 8 Surgical Handpieces............................................................................................................................... 8 Fixed Focus Spot Handpiece Maintenance............................................................................................ 9 Fuse Replacement ................................................................................................................................. 9 Disposal.................................................................................................................................................. 9 Software updates.................................................................................................................................... 9 Calibration............................................................................................................................................. 10 Laser Power Output.............................................................................................................................. 13 Technical Description ........................................................................................................................... 15
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.2  Version B.2  SPECIFICATIONS   DIOMED DELTA 15  DIOMED DELTA 30 Laser Type  Diode laser, CW Centre wavelength  810 nm Spectral bandwidth  90% of optical power contained within 810 ± 20 nm Delivery fiber  Minimum 400 µm diameter, 0.37 Numerical Aperture (NA) Fiber connector  Optical SMA-905 to MIL STD Maximum Power  15 W to the laser output port  30 W to the laser output port Minimum Power  0.5 W to the laser output port Set power accuracy  Better than ± 10% of displayed power Output power stability  ± 5% maximum Aiming beam  Red Class 3a diode laser, ≤ 5 mW at port, wavelength 635-655 nm 100 ms – 9900 ms Pulse duration Continuous Wave (CW) Pulse Interval  100 ms – 1000 ms Countdown timer  10 – 3200 seconds Contact fibers, 400µm, 600µm and 1000µm Fibers   Non-contact fibers, 600µm & 1000µm Calibration  Internal power calibration meter for handpiece or fiber calibration (Optional on DIOMED DELTA 15) Spot Handpiece Mode  Spot handpiece mode is available on the DIOMED DELTA 30 and on the DIOMED DELTA 15 when fitted with the optional calibration meter Spot size  Ø 2mm  Ø 2mm , Ø 4mm Single/Repeat Pulse  50ms – 250ms  50ms – 950ms Pulse Interval  100ms – 1000ms  100ms – 1000ms Fluence  Up to 119J/cm2  Up to 400J/cm2 Fiber Recognition System   (FRS) Operating Frequency  125 kHz ± 1 kHz Operating Range  Maximum of 40 mm from the laser output port RFID Modulation Type  A1D Standards  FCC 47 Part 15c, Canadian RSS-210 EN 300 330-1, EN 300 330-2, EN 301 489-1, EN 301 489-3 Cooling  By ambient air, with fan assistance Power Supply  100 – 240 V AC, 50 – 60 Hz, 500 VA max. Dimensions (H x W x D)  150 mm x 370 mm x 325 mm (± 5 mm) Weight  12 kg max. Up to 15W CW     +10ºC to +40ºC Operating Temperature  +10ºC to +40ºC Over 15W CW     +10ºC to +30ºC Operating Humidity  Up to 75% relative humidity, non-condensing Operating Pressure  Normal atmospheric pressure Storage Temperature  0ºC to +55ºC Storage Humidity  Up to 90% relative humidity, including condensing Storage Pressure  500 hPa to 1014 hPa Safety Standards  EN 60601-1, EN 60601-1-2, EN 60601-2-22, EN 60825-1, 21 CFR 1040.10, 1040.11
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.3  Version B.2  CLASSIFICATION Type of protection from electric shock  Class I Degree of protection from electric shock  Type B Degree of protection against ingress of water  IP21  CLEANING THE DIOMED DELTA 15/30 The DIOMED DELTA 15/30 has been designed to operate reliably with minimal maintenance. There are no user-serviceable parts in the DIOMED DELTA 15/30. Any attempts to repair, adjust or modify the system beyond the procedures allowed in this Operator Manual, by any person not authorised by DIOMED, will invalidate the guarantee.  The system enclosure may be wiped down periodically with a cloth dampened with a mild antiseptic solution. Before cleaning always disconnect the DIOMED DELTA 15/30 from the AC supply. Do not use any other solutions, solvents or abrasives. Take care not to get any liquid inside the enclosure.  CARING FOR FIBERS Safety  Carefully read and follow the package insert instructions for use. Handling  Leave  the  fiber  tip  protector  in  place  during  the  uncoiling  and connection process.   To verify the integrity of the fiber, check the fiber for any breaks by overall visual inspection.  For non-contact fibers, ensure the laser is in READY mode, and direct the aiming beam at a flat, white surface positioned 50-70mm away and examine the spot formed. The central spot  should  be  symmetrical  and  the  outer  circle  uniform  in  both intensity and shape. Pre-Carbonisation of sculpted tip (contact) fibers Prior to use on the patient, an enhanced tissue effect can be achieved by establishing a plane of microcarbon on the face (end) of the contact tip.  A  plane  of  carbon  traps  the  laser  energy  within  the  fiber  tip, resulting in increased efficiency at lower powers. 1.  Using  sterile  technique,  darken  a  small  area  (l)  on  a  wooden tongue depressor/spatula with a surgical marker. 2.  Pass the fiber into surgical field. 3.  Set the laser power at 10 Watts continuous. 4.  Observing  the  safety  procedures  described  in  this  Operator Manual, gently touch tip of the fiber to the darkened area on the tongue  depressor  and  operate  laser  only  until  a  small  plume  of smoke appears. 5.  The fiber has  now  been  pre-carbonised and  is  ready  for  clinical use. 6.  Reset  the  laser  to  the  correct  power  setting  and  duration  for desired procedure.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.4  Version B.2  Delivery Fiber Calibration  Non-contact fibers with a cooling sheath can be calibrated using the calibration procedure described in Section 3 – Operating Instructions.  Contact laser fibers do NOT require calibration. Cooling for reusable tip contact and non-contact fibers with gas/fluid cooling Fibers with gas/fluid cooling have a protective catheter with the distal tip secured in a metal ferrule. Fibers with reusable contact tips have threaded ferrules for contact tip connection. On this type of fiber, gas, air or distilled water is introduced near the proximal end through an auxiliary line that has a luer lock connection. The purpose of the cooling is to keep the distal end of fibers clean and cool during use. Gas, air or fluid supply is not provided with the laser system. Typical flow rates for the gas and air are from 0.2 to 1.5 l/min with minimum pressure of 50-60 psi and for the fluid cooling from 2 to 20 ml/min. Application to tissue  When using a conical tip contact fiber, all of the energy is being delivered out of the tip. There is a minimal tissue effect with the side of the tip. The best results will be obtained when holding the tip of the fiber perpendicular to the tissue and applying extremely light pressure with the tip. Haemostasis occurs as the tissue is incised. The orb tip contact fiber is also used for cutting tissue. The laser energy is displaced outwards from the side of the fiber, providing the added advantage of coagulating larger tissue surfaces. Both styles of tips work best when applied lightly to the surface and not buried in the tissue. Cross traction on tissue increases laser effect and extends life of fiber tip.   TROUBLESHOOTING Error Messages The DIOMED DELTA 15/30 is continuously monitoring its operation and performance. Should it detect a problem it will display a short message on the screen. To clear a message, carry out the instructions on the display. Outlined below are some common messages and the appropriate action that the user should undertake.  Remote interlock  The remote interlock has been violated.  Close the entry door or insert the remote  interlock  bypass  into  the  remote  interlock  bypass  socket  on  the rear of the unit. Footswitch not connected  The laser system will not go to READY mode unless the footswitch has been  connected.  Connect  the  footswitch  and  ensure  that  it  is  correctly inserted into the footswitch socket. Fiber not connected The  optical  fiber  or  spot  handpiece  is  disconnected  from  the  laser.  Connect a handpiece or fiber to the laser output port.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.5  Version B.2  Emergency switch pressed.  Use key switch to reset The emergency switch has been pressed. Switch the laser OFF and ON using the key switch.  The laser will carry out a self-test and the message will clear automatically. Fiber calibration failed  An  attempt  has  been  made  to  calibrate  a  non-contact  fiber  and  the calibration has failed (below 75%).  Replace the fiber and re-calibrate. Setting Wavelength  This  message  will  occur  when  the  temperature  of  the  laser  diode  is outside of its specified operating range for the correct diode wavelength. Leave the laser on to allow the temperature to stabilise. This may take a few minutes. Ambient temperature out of range  The laser is being operated outside of its specified ambient temperature range.  Switch  it  off  at  the  mains  inlet  and  allow  it  to  stabilise  at  room temperature. Temperature high  The  laser  is  overheating.  Switch  it  off  at  the  mains  inlet  and  allow  it  to cool. Footswitch held down  There  is  a  two-second  safety  delay  when  the  laser  is  placed  from STANDBY to READY.  Ensure that the footswitch is not held down until after the two-second delay and the audible beep is heard.  The message will clear when the footswitch is released. Diode switched off  The laser has been prevented from firing.  Press SELECT and check the maximum power available. Software loading unsuccessful  A  problem  was  detected  when  loading  new  software.  Switch  the  laser OFF and then ON at the key switch and repeat the software load process. If  the  problem  is  repeated,  call  for  support  from  your  DIOMED representative. Footswitch invalid  The  laser  has  detected  a  problem  with  the  footswitch.  Check  that  the footswitch has been connected correctly. If the message does not clear, switch the laser OFF and then ON at the key switch. If this does not clear the problem, call for support from your DIOMED representative. Panel switch stuck  A switch on the front panel has stuck during switch-on, and the laser will not be able to go into the READY mode. Switch the laser OFF and ON using the key switch.  If this does not clear the message, call for a service engineer. Call service engineer  If this message appears, switch the laser OFF at the key switch and then ON.  If  the  message  does  not  disappear,  call  service  support  from  your DIOMED representative.  FRS Error Messages The following error messages are associated with the Fiber Recognition System.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.6  Version B.2  FRS Error  The  laser  has  detected  a  problem  with  the  FRS  system  for  identifying fibers. Try removing and replacing the fiber connector or connecting a new fiber. If the message does not clear, switch the laser OFF and then ON at the key switch. If this does not clear the problem, call for support from your DIOMED representative. Fiber uses expired  All the  available  uses  of  the connected fiber  have expired. Replace the fiber with a fiber that has uses remaining. Fiber sterility expired  DIOMED fibers are programmed at manufacture with a sterility expiration date. This message is displayed when the expiry date is before the date in the laser’s internal clock. Replace the fiber with one having a valid sterility expiration date. Invalid Fiber  The fiber attached to the DIOMED DELTA 15/30 is either not recognised or not equipped with the FRS system. Replace with an appropriate fiber or, if a Session Fault is indicated, use the Emergency Override option.  FRS Troubleshooting The DIOMED FRS system has been extensively tested. In the unlikely event of a problem being experienced please perform the following checks before contacting your local DIOMED representative for further advice.  “Invalid Fiber” is displayed even with a new unused FRS system fiber (1)  Disconnect and then reconnect the fiber to the DIOMED DELTA 15/30. This  will cause  the  unit  to  try  and  read the fiber’s data again. (2)  Ensure  that  the  DIOMED  DELTA  15/30  is  at  least  2  meters away  from  any  other  electrical  or  electronic  equipment  that might interfere with the FRS system, such as computers or other electronic or medical equipment. I need to use a non-FRS system fiber Some versions of the DIOMED DELTA 15/30 can also operate with non-DIOMED fibers, as long as they are fitted with a special DIOMED adapter at the laser end of the fiber. This adapter is available as an accessory from your DIOMED representative and it will allow the DIOMED DELTA 15/30 to be used in Manual mode only. See your DIOMED representative for further information. The Emergency Override option is not available  After it has been used once, the Emergency Override option must be reset  before  it  can  be  used  again,  as  described  in  ‘Resetting  the Emergency Override’  above.  Contact  your DIOMED representative to obtain a spare reset device. The fiber is still within its sterility date but the laser shows it as expired Check  that  the  date  of  the  internal  clock  in  the  DIOMED  DELTA 15/30  is  set  correctly.  The  procedure  for  setting  the  clock  is described in Section 3 - Operator Instructions.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.7  Version B.2  ACCESSORIES Optical Fibers The DIOMED DELTA 15/30 has an output connector for optical fibers  with  standard  SMA-905  connector.  Only  DIOMED  or fibersdirect.com labelled fibers should be used. A list of fibers available  for  use  with  the  DIOMED  DELTA  15/30  can  be obtained from your DIOMED representative.  Contact Sculpted Tip Fibers Used in contact with the tissue for incision/excision.  Conical Tip Fibers DIOMED conical tip fibers have a tip size of 300µm and should be  selected  where  a  narrow  or  precise  incision/excision  is required. Laser energy is delivered from the end of the sculpted tip  with  minimal  tissue  effect  from  the  side  of  the  tip.  Heat delivered  via  the  tip  performs  the  cutting  and  extremely  light pressure  is  all  that  is  required  for  incising,  excising  and vaporising and coagulating soft tissue. Haemostasis occurs as the tissue is incised.  Orb Tip Fibers DIOMED  orb  tip  fibers  are  available  with  tip  sizes  of  800µm, 1200µm  and  3000µm  and  should  be  selected  where  a  wider incision  or  tissue  vaporisation  is  required.  Laser  energy  is displaced  outwards  from  the  forward  curvature  of  the  fiber which  gives  the  added  advantage  of  vaporising  larger  tissue surfaces.  Heat  delivered  via  the  tip  performs  the  cutting  and extremely  light  pressure  is  all  that  is  required  for  incising, excising  and  vaporising  and  coagulating  soft  tissue. Haemostasis occurs as the tissue is incised.  Bare/Flat end fibers Used in contact with the tissue for incision/ excision and in non-contact with tissue at low power for vaporisation / coagulation.  Non-contact cooled fibers Held at a distance from the tissue for vaporisation / coagulation.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.8  Version B.2   STERILISATION OF OPTICAL FIBERS   Optical  fibers  are  provided  sterile  as  a  disposable,  single-use product.  DO NOT RE-STERILISE THE FIBERS. DO NOT RE-USE THE FIBERS. USE  ONLY  DIOMED  OR  FIBERSDIRECT.COM  LABELLED  OR DIOMED APPROVED FIBERS. Failure to observe this could invalidate the Laser Guarantee. Intra-operative cleaning  If the tip accumulates debris, turn the laser to the STANDBY mode and then carefully wipe the tip clean with a wet sponge/swab. Fiber Disposal  After  use,  the  single-use  optical  fibers  should  be  disposed  of  in accordance with local regulations regarding disposal of contaminated waste.  DIOMED  and  fibersdirect.com  labelled  optical  fibers  have undergone stringent evaluation and testing to ensure that they are  of  the  highest  quality  and  that  they  operate  safely, effectively and efficiently with DIOMED lasers. The exact alignment of the interface between the laser aperture and  the  SMA-905  connector  is  critical.  Misalignment  (as  may occur  with  non-approved  fibers)  can  result  in  damage  to  the laser and poor delivery of laser energy to the patient. Calibration Port Adapter  The optical fiber calibration port adapter provided should be sterilised and used to locate the fiber in the calibration port. Steam Sterilisation  The  calibration  port  adapter  should  be  sterilised  before  use  in accordance with ISO 11134 1993 ‘Sterilisation of Healthcare Products, Requirement for Validation and Routine Control, Industrial Moist Heat Sterilisation. A  validated  cycle  of ≥134°C  (273°F)  for  ≥  3  minutes  sterilising  time should be used to give a sterility Quality Assurance of 106.  SURGICAL HANDPIECES Surgical handpieces with either rigid or malleable cannulae can be obtained from your DIOMED representative. These are available in a range of lengths and internal /external diameters. A full description and list of handpieces is available from your local DIOMED representative.  Guildelines  If the requirement is to use a fiber with a separate handpiece, the following guidelines must be followed: 1.  Remove the fiber from the sterile packaging in accordance with the fiber instructions. 2.  Ensure that the handpiece is sterile, loosen the locking nut at the rear of the handpiece.  To  prevent  any  premature  damage  to  the  fiber  when  using malleable  handpieces,  ensure  that  the  malleable  part  on  the handpiece has been straightened before inserting the fiber. Only
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.9  Version B.2  shape the handpiece after the fiber has been inserted. 3.  Insert the fiber down the handpiece from the rear until the fiber 'tip' protrudes  approximately  10mm  from  the  distal  end  of  the handpiece. 4.  Tighten the lock nut at the rear of the handpiece finger tight. 5.  The fiber is now ready for use. Cleaning and Sterilisation  After  use,  remove  the  fiber  from  the  handpiece,  wipe  down  the outside of the  handpiece with  alcohol. Insert a syringe full of  water into the rear of the handpiece and depress the plunger to flush out any remaining debris. The handpiece can now be sterilised using a validated steam sterilisation cycle. Steam Sterilisation  Reusable handpieces should  be sterilised  before  use in  accordance with  ISO  11134  1993  ‘Sterilisation  of  Healthcare  Products, Requirement for Validation and Routine Control, Industrial Moist Heat Sterilisation.    A  validated  cycle  of  ≥134°C  (273°F)  for  ≥  3  minutes sterilising time should be used to give a sterility Quality Assurance of 106. FIXED FOCUS SPOT HANDPIECE MAINTENANCE For information on the care and maintenance of the Fixed Focus Spot Handpieces, see the Instructions for Use supplied with each handpiece. FUSE REPLACEMENT Spare mains fuses are supplied with the DIOMED DELTA 15/30. Further spares can be obtained from DIOMED. They can be replaced as follows:  1.  Use a small flat-bladed screwdriver to release the fuse compartment from the mains inlet on the rear panel. 2.  Remove the two fuses from the holder and replace with new ones of the same type and rating: T6.3A H 250V. Fuses with a different rating or specification must not be used. DISPOSAL At the end its life the DIOMED DELTA 15/30 should be disposed of according to national environmental requirements or be returned to DIOMED  SOFTWARE UPDATES  From time to time DIOMED may issue new procedures, upgrades and feature enhancements for the DIOMED DELTA 15/30. As a registered owner you will be notified of these when they become available.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.10  Version B.2  CALIBRATION Checking the Calibration of the Internal Power Meter To ensure accurate calibration of the delivery fiber, the internal power meter should be checked at least annually. The LASER SAFETY OFFICER or suitably trained service personnel should carry out this procedure.  The method for carrying out this procedure is described below.  Calibrating the Internal Power Meter THE FOLLOWING INFORMATION IS PROVIDED FOR REGULATORY PURPOSES   INCORRECT CALIBRATION MAY CAUSE INJURY DURING TREATMENT.  THE ENGINEERS INTERFACE MAY BY-PASS MANY OF THE NORMAL SAFETY SYSTEMS.   DO NOT attempt the following procedures unless specifically authorised to do  so  by  DIOMED.  Such  action  may  cause  exposure  to  hazardous  laser radiation and a risk of electrical shock. It may also result in damage to the instrument and invalidate guarantee cover. ONLY  TRAINED  LASER  SERVICE  PERSONNEL  AUTHORISED  BY DIOMED SHOULD PERFORM SERVICE AND MAINTENANCE. DIOMED will not accept liability for the use of this equipment when calibrated by unauthorised personnel. Calibrating the internal power meter To ensure accurate calibration of the delivery fiber, the internal power meter should be checked at least annually and re-calibrated if necessary. This section describes the procedure undertaken to re-calibrate the internal power meter. Equipment required   •  DIOMED Engineer Interface software. •  Traceable power meter capable of recording a 10W, one second pulse of 810nm Infra Red laser light. •  600µm reference fiber.  Procedure  The calibration of the internal power meter is controlled by two electronically controlled potentiometers (EPOTs). These need to be set to scale the output of the power meter correctly. These are normally adjusted for gain and settling time using a production jig. Only the facility to adjust the gain is used here.  Connect the laser to a PC via a null-modem serial cable between the port on the rear panel and a serial port on the PC. Turn on the laser. Load and run the DIOMED Engineer Interface program on the PC. From the Connection menu,  select  Connect  to  Laser,  choose  the  appropriate Comms Port and press Connect.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.11  Version B.2  From the Connection menu select Enter EI and enter the password.  Connect the fiber to the calibration port using a suitable adapter. In the engineer interface, open the ‘Non-volatile details’ tab. Note down pot settings 1 and 2 from the Calibration Port group. Set the laser to continuous mode, 10W and put the laser into the ready state.  In the engineer interface, select Tests/ Read radiometer ADC. A dialog box is shown, headed ‘Reading radiometer’ Fire the laser for not more than 10s. The ADC readings are shown in the dialog. Note down the second ADC reading. This does not have to be very accurate – we are just scaling the radiometer output to fit in the ADC range. Put the value in the table below.  Pot 1 setting  Pot 2 setting  Stage 2 ADC reading                           Change the value of the Pot 1 setting to scale the stage 2 ADC reading to around 700. This is done by closing the ‘Reading radiometer’ window, altering the setting for pot 1, and clicking on the ‘Update’ button. Now repeat the sequence, recording the new settings and ADC reading in the table. Repeat until the ADC reading is between 650 and 750 counts.  Close the ‘Reading radiometer’ window and put the laser into the standby state.  The radiometer amplifier circuit should now be scaled correctly. The next step is to calibrate the radiometer, using calibration points for 1W and 10W, which are stored in the on-board EEPROM. This step will require the calibrated power meter.  With the laser in the standby state, select Set up/ Set up radiometer from the engineer interface menu. The following prompt should be displayed:  ‘Put fiber in external meter and fire laser. Adjust power and press control when set’. Set the power to 10W and fire the laser into the meter. When the power reading has settled make adjustments using the rotary control until the power displayed is exactly 10W.  When the power setting is achieved, release the footswitch and press the rotary control. The following prompt should be displayed: ‘Put fiber in internal meter and fire laser until cal pulse stops.’  Attach the fiber to the calibration port, put the laser into the ready state and fire until the laser stops (allow about 8s). Release the footswitch.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.12  Version B.2   The following prompt should be displayed:  ‘Put fiber in external meter and fire laser. Adjust power and press control when set’.  Adjust the power setting to 1W and put the laser back into the ready state. Fire the laser into the meter and, when the power reading has settled, make adjustments using the rotary control until the power displayed is as close as possible to 1W.  When the power setting is achieved, release the footswitch and press the rotary control.  The following prompt should be displayed: ‘Put fiber in internal meter and fire laser until cal pulse stops.’  Attach the fiber to the calibration port, put the laser into the ready state and fire until the laser stops (allow about 8s). Release the footswitch.  Select ‘Back’  In the engineer interface, update the ‘Non-volatile details’ tab and read the calibration port data. Write the data down in the table.  Pot setting 1   Pot setting 2   ADC at 1W port power   ADC at 10W port power   Checksum    The ‘ADC at 10W port power’ figure is the calibration point for 10W, which we have just obtained with the set-up routine. Compare this with the value noted previously when setting the radiometer gain. That reading was obtained with 10W port power, while the calibration point was taken with 10W distal power, so the calibration point should be 1.1x the previous ADC reading at 10W, allowing for a 90% fiber.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.13  Version B.2  LASER POWER OUTPUT Measuring the Laser Power Output The LASER SAFETY OFFICER or suitably trained service personnel should check the output power of the DIOMED DELTA 15/30 at least annually from the date of installation, by following the procedure described below.  Equipment Required  •  A sampling power meter or an independent energy (integrated power) meter of known calibration •  A bare ended optical fiber  •  Laser unit to be tested Procedure  Calibrate the fiber.  Connect the fiber to the Laser unit output port and present the distal end of the fiber to the external power meter.   1.  Record the Laser unit's actual and displayed outputs at various different power/energy settings e.g. 5W, 10W, 15W etc. Calculate the percentage difference between the displayed and the actual power/energy output as taken from the external power meter. 2.  If calculated disparity exceeds ± 20%, contact DIOMED.  Adjusting Laser Power Output Power output adjustments can only be made by suitably trained DIOMED service personnel. For regulatory purposes, the method for carrying out these adjustments is described below. Please contact your DIOMED representative for further advice.  Adjusting the Laser Power Output THE FOLLOWING INFORMATION IS PROVIDED FOR REGULATORY PURPOSES The procedure below will explain how to adjust the power output of the laser.  Equipment Required  •  A sampling power meter or an independent energy (integrated power) meter of known calibration •  A calibrated bare ended optical fiber  •  Laser unit to be tested Procedure  Connect the laser to a PC via a null-modem serial cable between the port on the rear panel and a serial port on the PC. Connect  the  calibrated  fiber  from  the  laser  unit  output  port  to  the external power meter. Turn on the laser. Load  and  run  the  DIOMED  Engineer  Interface  program  on  the  PC. From  the  Connection  menu,  select  Connect  to  Laser,  choose the appropriate Comms Port and press Connect. From  the  Connection  menu  select  Enter  EI  and  enter  the password. Select the Non-volatile data tab Diode calibration is  stored in both the  Monitor MCU  and  in the laser
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.14  Version B.2  module itself. The values in these locations are shown on the screen – if there are any discrepancies then an error message is displayed. The laser is calibrated at three different power settings: 2W, 5W and 15W, by entering the required diode feedback value (in mV) into the appropriate boxes on the form. Always change the values for both the Monitor params and Diode data together, then press Update to reconfigure the laser.  Start  with  the  2W  setting  by  entering  a  default  value  (no  more  than 150) into the two appropriate boxes and pressing Update. Calculate  the  expected  power  at  the  distal  end  of  the  fiber,  i.e. measured at the power meter,  from the required  port power  and the known efficiency of the fiber.  Fire the laser and record the power measured by the power meter. Adjust the value of the feedback according to this result and enter new values until the correct power is observed when the laser is fired. Repeat this process for the 5W and 15W settings, starting with default values of 400 and 1000 respectively. Finally  check  that  the  laser  outputs  the  correct  power  at  all  three settings.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.15  Version B.2  TECHNICAL DESCRIPTION The DIOMED Delta 15 and Delta 30 lasers contain no user-serviceable components. In the event that repair or service is required please contact your local DIOMED representative. More detailed service instructions, including schematic diagrams, are available only to suitably-qualified technical personnel.  The DIOMED Delta 15/30 lasers are built in a modular fashion, enabling ease of test, assembly and service. The system has been designed so that any one module can be replaced with no performance effect on the other modules or the product as a whole.  The modules present in a complete unit are:  •  Laser Module •  Power Supply / Laser & TEC Driver •  Monitor & Control PCA •  User Interface PCA •  Optical Power Calibration Module (OPCM) Laser Power Control Overview Optical power control is achieved using a monitor photodiode measuring the output of the laser diode. This is used to derive a control signal proportional to the total power output of the laser unit. The feedback signal in this control loop is monitored for errors from the expected value.  In addition to this, an over-current trip circuit will operate to disable the laser driver rapidly should an overcurrent / overpower situation be detected. Diode current is monitored while firing to ensure that it is within acceptable limits. Laser Module The Laser Module consists of a laser diode, TECs, optics and an electronic control module. These components are all mounted on a stable metal platform which doubles as a heatsink. Two fans are mounted at one end of the heatsink to provide air-assisted cooling. The laser and TECs are located inside a hermetically-sealed enclosure.   Semi-transmissive mirrors in the optics path allow for the addition of the visible laser diode and for output power to be measured by means of a photodiode. The laser output port (SMA) incorporates a thermistor, to measure the temperature of the port close to the fiber connection and two micro-switches, which detect when a fiber has been correctly connected to the port. PSU / Laser & TEC Driver The mains power supply, laser driver and TEC driver are contained in their own enclosure. The power supply accepts a mains input from 100V to 240V and provides auxiliary power rails for the system electronics. The drivers are controlled via an interface with the Monitor & Control PCA. Monitor & Control PCA The Monitor & Control PCA consists of the laser power control loop, Control and Monitor microcontrollers, control logic, external interlock interfaces, power supply monitoring, cooling fan control and the RS232 serial port connection. This PCA connects to the Laser Module, the User Interface and the Power Supply.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.16  Version B.2   Two separate microcontrollers are used to independently Control and Monitor the laser diode. Instructions and information are passed from these two devices to the User Microcontroller along a dedicated RS232 serial interface.  Monitoring of the laser diode and peripheral functions such as temperature is performed by the Monitor MCU. The values obtained are also sent to the User Microprocessor. The Monitor MCU additionally monitors the state of many of the Control MCU outputs to check that their status is correct for the operating mode.  The Footswitch and Remote Interlock interface connectors are located on the rear panel. All signals on these connectors are filtered for EMC and protected against ESD. The Remote Interlock connector also includes an electrically-isolated ‘READY OUT’ output.   The two cooling fans attached to the heatsink may be run at four different speeds, depending on the temperature of the heatsink, the power of the laser output and the magnitude of the TEC drive level.   The signals for the RS232 serial port come direct from the User Microcontroller. The pins of the 9-pin D-type connector are electrically isolated from the rest of the circuit using opto-couplers and a transformer.  Ambient temperature is monitored using a digital temperature sensor, which is read by the User Microcontroller over the I2C bus. User Interface PCA The User Interface PCA consists of the User Microcontroller, display and all the controls and indicators required by an operator to use the product.  The User Microcontroller is from the Renesas H8 family. It is supplemented by 256Kb of SRAM memory, 4Mb of Flash program memory and 256Kb of serial FRAM memory for non-volatile data storage. A Real Time Clock (RTC) function provides the time and date features for the user interface. This microcontroller system interfaces to the rest of the product via a Serial Peripheral Interface (SPI) bus, Inter-Integrated Circuit (I2C) bus and RS232 serial interface.  The user interface display is a Vacuum Fluorescent Display (VFD) of 128 x 64 pixels. This is controlled directly by the User Microcontroller. The display is filtered to a mid-blue colour.  The control of the level of TEC drive is a software function within the User Microcontroller. It operates in a feedback loop with the objective of keeping the laser diode at a fixed temperature.  A MultiMedia Card interface allows a MMC card to be inserted into an aperture on the front panel. This card is used to transfer usage data from the laser to the user’s PC. It can also be used to transfer updated software and new procedure information into the laser. Optical Power Calibration Meter The Optical Power Calibration Meter (OPCM) is only fitted as standard to the Delta 30 laser. It is optional on the Delta 15. If the Delta 15 laser is not fitted with an OPCM, then the laser’s software does not permit the use of a spot handpiece.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 5.17  Version B.2  The OPCM consists of a power meter head unit, which has a voltage output proportional to the applied laser power. This is conditioned, amplified and converted to a digital signal. Calibration of the meter is achieved by a digitally-controlled potentiometer and the calibration figure is stored in an EEPROM. Glossary ADC  Analogue to Digital Converter DAC  Digital to Analogue Converter EMC  Electro-Magnetic Compatibility EEPROM  Electrically Erasable Programmable Read Only Memory ESD  Electro-Static Discharge FRS  Fiber Recognition I2C  Inter-Integrated Circuit I/O  Input/Output Kb  kilobits LED  Light Emitting Diode MCU  Microprocessor Control Unit ms  milliseconds OPCM  Optical Power Calibration Meter PCA  Printed Circuit Assembly PCB  Printed Circuit Board PSU  Power Supply Unit PWM  Pulse Width Modulation RAM  Random Access Memory RTC  Real Time Clock SFC  Single Fault Condition SPI  Serial Peripheral Interface TTL  Transistor-Transistor Logic TEC  Thermo Electric Cooling UART  Universal Asynchronous Receiver/Transmitter V  Volts VFD  Vacuum Fluorescent Display
DELTA 15/30 DELTA 15/30 Operator Manual  Page 6.1  Version B.2  SECTION 6  WARRANTY MANUFACTURER’S WARRANTY POLICY  DIOMED guarantees the DIOMED DELTA 15/30 against defects in materials and workmanship for a period of 24 months. The warranty period begins on the date of installation. To enable timely registration of the warranty, the owner/purchaser must complete and return the Warranty registration form within 28 days of installation. The following items are expressly excluded from this Warranty: • Safety Eyewear • All optical fibers, handpieces and accessories • Maintenance Instruments • Footswitch and electrical cables • All other accessories supplied by DIOMED Limited   Any attempt to repair, adjust or modify the system beyond those procedures described in the Operator Manual by any person not authorised by DIOMED, will invalidate the Warranty. WARRANTY CLAIMS To make a warranty claim the purchaser shall, promptly following discovery of the basis of claim, contact your DIOMED distributor in the first instance or DIOMED Ltd. in writing, by telephone, fax or Email at the following address:  DIOMED Inc One Dundee Park Andover, MA 01810 USA  Tel: +1 866 883 8820 Fax: +1 978 475 8488 Email: service@diomedinc.com  DIOMED Limited Building 2000 Beach Drive Cambridge Research Park Waterbeach Cambridge  CB5 9TE United Kingdom  Tel: +44 1223 729372 Fax: +44 1223 729320 E-mail:  service@diomedinc.com http://www.diomedinc.com  DIOMED is a Registered Trademark of DIOMED Limited
DELTA 15/30 DELTA 15/30 Operator Manual  Page 6.2  Version B.2                       THIS PAGE LEFT BLANK
DELTA 15/30 DELTA 15/30 Operator Manual  Page 6.3  Version B.2  WARRANTY REGISTRATION CERTIFICATE  Product Type:  DIOMED DELTA 15/30   Product Serial No:            Registered User Information: Institution:                        Specialist:  Name:          Speciality:          Address:  Street:                         City:           Zip / Post Code:            Country:                    Telephone:             Fax:            Email:                          Installation: Date Installed:                        Installed by:   Signed:         Print Name:          Organisation:                        DIOMED Sales Consultant details:     Distributor Signature:           Print Name:          •  Registered User and Consultant to fully complete this Certificate. •  Registered User must keep this certificate in a safe place for reference. •  The accompanying Registration Form must be completed and sent to DIOMED Inc, One Dundee Park, PO Box 97, Andover, MA 01810, USA, by mail or fax (978) 475 8488 within 28 days of date of Product installation. This Registered User information is essential in the event of a Guarantee Claim.
DELTA 15/30 DELTA 15/30 Operator Manual  Page 6.4  Version B.2                         THIS PAGE LEFT BLANK
DELTA 15/30 DELTA 15/30 Operator Manual  Page 6.5  Version B.2  WARRANTY REGISTRATION FORM  Product Type:  DIOMED DELTA 15/30   Product Serial No:            Registered User Information: Institution:                        Specialist:  Name:          Speciality:          Address:  Street:                         City:           Zip / Post Code:            Country:                    Telephone:             Fax:            Email:                          Installation: Date Installed:                        Installed by:   Signed:         Print Name:          Organisation:                        DIOMED Sales Consultant details:    Distributor Signature:           Print Name:           •  Registered User / Consultant to fully complete, detach and return this Registration form to DIOMED Inc, One Dundee Park, PO Box 97, Andover, MA 01810, USA, by mail or fax (978) 475 8488) within 28 days of date of laser installation. This Registered User information is essential in the event of a Guarantee Claim.