  A310B Fingertip Pulse Oximeter   Operator’s Manual                               File No.: A310B-OPE        V 1.0      20180712
A310B Pulse Oximeter  2 Section 1 Safety                                                   Instructions for the Safe Operation and Use of the A310B Pulse Oximeter  Do not attempt to service the A310B Pulse Oximeter. Only qualified service personnel should attempt any needed internal servicing.  Prolonged  use  or  the  patient’s  condition  may  require changing  the  sensor  site  periodically.  Change  sensor  site and  check  skin  integrity,  circulatory  status  and  correct alignment at least every 2 hours.  SpO2  measurements  may  be  adversely  affected  in  the presence  of  high  ambient  light.  Shield  the  sensor  area    (with a surgical towel, or  direct  sunlight, for example)  if   necessary.  The  following  reason  will  cause  interference  to  the  testing accuracy of the pulse oximeter.  High-frequency electrosurgical equipment.    Placement of a sensor on an extremity with a blood pressure cuff arterial catheter, or intravascular line    The patient has  hypotension severe vasoconstriction severe anemia or hypothermia.    The patient is in cardiac arrest or is in shock.    Fingernail  polish  or  false  fingernails  may  cause  inaccurate SpO2 readings.
A310B Pulse Oximeter  3  1.2    Warnings                                              WARNING: EXPLOSION HAZARD — Do not use the A310B in a flammable atmosphere where concentrations of flammable anesthetics or other materials may occur. WARNING:  Do  not  attempt  to  recharge  normal  dry-cell batteries,  they  may  leak.  And  may  cause  a  fire  or  even explode. And may cause a fire or even explode. WARNING: Do not throw batteries in fire as this may causes them to explode. CAUTION: Keep the operating environment free of dust, vibrations, corrosive, or flammable materials, and extremes of temperature and humidity.   CAUTION: Do not operate the unit if it is damp or wet because of condensation or spills. Avoid using the equipment immediately after moving it from a cold environment to a warm, humid location. WARNING: .Do not use the pulse oximeter in an MRI or CT environment. WARNING:  Do  not  modify  this  equipment  without authorization of the manufacturer. WARNING:  If  this  equipment  is  modified,  appropriate inspection and testing must be conducted to ensure continued safe use of equipment.
A310B Pulse Oximeter  4                           CAUTION: Never use sharp or pointed objects to operate the front-panel switches. CAUTION: The device shall only be used if the battery cover is closed. CAUTION: The batteries must be proper disposed according to local regulation after their use. CAUTION: The batteries must be taken out from the battery compartment if the device will not be used for a long time. CAUTION: The device should keep away from the children, pets and pests to avoid swallowing.
A310B Pulse Oximeter  5 1.3    Definitions and Symbols Symbol Description  Type BF Equipment  Batch code*  Date of manufacture* SN Serial NO*  Information  of  manufacture,  including name and address  Temperature limitation  Bluetooth Indication   When the end-user wishes to discard this product,  it  must  be  sent  to  separate collection  facilities  for  recovery  and recycling
A310B Pulse Oximeter  6  Follow instruction for use IP22   Anti-dust& Anti-water class   Warning: The  information  you  should  know  to protect  patients  and  medical  staff  from possible injury Caution: The  information  you  should  know  to protect  the  equipment  from  possible damage Note: The  important  information  you  should know *Batch code, Date of manufacturer and Serial No are printed on the label on the battery cover.   Section 2 Introduction                   2.1 General This chapter provides a general description of the A310B Pulse Oximeter including:  Brief device description  Product features 2.2 Brief Device Description
A310B Pulse Oximeter  7 The Pulse Oximeter, based on all digital technology, is intended for noninvasive spot-check measurement of functional oxygen saturation of arterial hemoglobin (SpO2). Advanced DSP algorithm can minimize the influence of motion artifact and improve measurement accuracy of low perfusion. The Oximeter can be used to measure human Hemoglobin Saturation and heart rate through finger. The product is suitable for    family, hospital (including clinical use in internist/surgery, Anesthesia, pediatrics, intensive care and etc.) Oxygen Bar, social medical organizations, physical care in sports and etc. 2.3 Product Features  Lightweight for carrying and Easy-To-Use.    Manually adjust the direction of interface .  Color  OLED  display,  simultaneous  display  for  testing    value and plethysmogram.  Low  Perfusion  :  0.2%.  (Advanced  DSP  algorithm  can   improve measurement accuracy, under the condition of low perfusion.)  Visual & Sound alarm function. Real-time spot-checks.  Wireless Bluetooth for data transmission.  Low Battery voltage indicator.  Automatically switch off.  Standard two AAA 1.5V Alkaline Bаttеrу support more than 20 hours continuous work.
A310B Pulse Oximeter  8                       *DSP algorithm: Digital signal processor algorithm. CAUTION: The device cannot be used to measure the child below 1 years as the test result is not guarantee to accurate. CAUTION: The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms CAUTION:    A function tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.   Clinical testing is used to establish the SpO2 accuracy. The measured arterial hemoglobin saturation value (SpO2) of the sensor is compared to arterial hemoglobin oxygen(SaO2) value, determined from blood samples with a laboratory CO-oximeter.   The accuracy of the sensors in comparison to the CO-oximeter samples measured over the SpO2 range of 70 -100%. Accuracy data is calculated using the root-mean-square(Arms value) for all subjects. Only about two-thirds of PULSE OXIMETER EQUIPMENT measurements can be expected to fall within ±Arms of the value measured by a CO-Oximeter.     Pulse simulator shall be used to assess Pulse rate Accuracy. The measured pulse rate is compared to the preset pulse rate value in simulator. Accuracy data is calculated using the root-mean-square (Arms value) for all subjects.
A310B Pulse Oximeter  9 *Low Perfusion: In physiology, perfusion is the process of a body delivering blood to a capillary bed in its biological tissue. Under the condition of low perfusion, the measurement of non-invasive saturation of pulse-blood oxygen is low-accurate. *Plethysmograph: is an instrument for measuring changes in volume within an organ or whole body (usually resulting from fluctuations in the amount of blood or air it contains).   Section 3 Installation,Setup,and Operation 3.1 Description of the Front Panel (as figure 3.1.1)         Figure 3.1.1 Parts of front & back panel  Table 3.1.1 Part Definition and Description
A310B Pulse Oximeter  10 Item Name Description 1 Power button Turn on the machine 2 OLED Panel Display the SPO2/PR data & Plethysmogram,etc. 3 Battery Compartment    3.2.Display After switch on, the OLED display of A310B is as follows:         Figure 3.2.1 OLED display       Battery Capacity
A310B Pulse Oximeter  11     3.3 Parameter setting When  the  device  is  under  measuring  interface,  press  the direction  button  for  1  second  in  order  to  enter  into  menu page(figure 3.3.1 and figure 3.3.2 ). There are two submenus for choice:   Remind Setup Press  the  direction  button  for  1  second  and  enter  into  the Reminder Setup. User can adjust the setting through moving the ―*‖ symbol to the back of the Sound Reminder, Beep , Bluetooth, Restore or Brightness. Sound Reminder   Press the direction button for 1 second, move the ―*‖ symbol to the back of Sound Reminder, long press the direction button to turn it on/off. Beep Press the direction button for 1 second, move the ―*‖ symbol to  the  back  of  Beep,  long press  the direction  button  to  turn  it on/off. Bluetooth Press the direction button for 1 second, move the ―*‖ symbol to the back of Bluetooth, long press the direction button to turn it on/off.
A310B Pulse Oximeter  12 Restore When the ―*‖ symbol shown behind ―Restore‖, long press the direction  button  can  be  changed  to  ―OK‖,  which  causes  the device to restore factory data setting.  Brightness When the ―*‖ symbol shown on ―Brightness‖, long press the direction button to change the Brightness value from 1 to 5. Limit Value Setting When the * symbol shown on the Reminder Setup, long press the  direction  button  until  enter  into  the  Reminder  Limit  setup menu(figure 3.3.2). User can press the direction button to select the items. And press the direction button for 1 second to change the data you need. On the Reminder Limit setup menu page (figure 3.3.2), when the * symbol shown behind the ―+/-‖. Press direction button for 1 second to change the ―+‖ to ―-‖ or change the ―-‖ to ―+‖.       When ―+‖ shows on the right side, press the direction button for 1 second , move the ―*‖ after the Spo2 Hi or PR Hi setting, can increase the value to a higher value (until it reaches to the highest.)   When ―-‖ shows on the right side, press the direction button for 1  second, move the  ―*‖ after  the Spo2  Lo or PR  Lo value setting, can reduce the value to a lower value (untill it reaches to the lowest).
A310B Pulse Oximeter  13      Figure3.3.1                  Figure3.3.2 Note: 1.The alarm have 1 second delay after the incorrect result being detected. 2. The customer can preset the alarm value to the 98 or 99 to check whether it is normal for alarm setting. 3.4 Operation 3.4.1 Install battery Installing  two  AAA  batteries  into  battery  cassette  in  correct polarities and cover it.          WARNING: Do not attempt to recharge normal alkaline batteries, they  may leak  and  may  cause a  fire  or  even explode.
A310B Pulse Oximeter  14   3.4.2 Turn the Pulse Oximeter on Put one of fingers into rubber hole of the oximeter (it is best to put the finger thoroughly) with nail surface upward, then releasing the clamp.   Press power button to turn the Pulse Oximeter on. The oximeter will be automatically powered off when no finger in the device for longer than 16 seconds.  3.4.3 Read correspondent data from display screen. 3.4.4 Display Description of OLED The display interface of ―OLED‖ can rotate four directions with six different display modes after pressing the power button for less than 0.5s. It is shown as below:
A310B Pulse Oximeter  15        Note:   1.when  battery  power  is  at  lowest  level,  the  battery  capacity  indicates  symbol  of ― ‖in OLED, remind users of replacement of battery. 2. The pletismogragh can been regarded as correct if the wave is fluctuated regularly.  3.4.5 Date Transmission 1.  The  current  measurement  will  transmit  to  the  App  automatically.  after  data transmission successfully, the measurement flashing for  8s,  then the device will  power  off automatically. If there is  no  Bluetooth connection over 1 min, the device will power off automatically and the data will not be
A310B Pulse Oximeter  16 stored. If there is no reading, the device will power off automatically. Notes: With the Bluetooth 4.0 to transmit the data to App. The transmission distance is 10 m at most.  Section 4 Cleaning and Disinfection 4.1 Cleaning Cleaning the machine once every day. Switch off the power and take  out  the  batteries  before  cleaning,  Cleaning  exterior  surface (OLED  display  screen  included)  of  the  unit  with  a  dry  and  soft cloth. Use 75% density of medical alcohol to clean the surface and use dry fabric with little alcohol to avoid alcohol permeates into the device. 4.2 Disinfection Disinfecting  the  machine after  using  by  the  patient  if  multiple patient use the machine in the hospital.   Use 75% density of  medical  alcohol to clean  the surface  that contacting with the patient.
A310B Pulse Oximeter  17                                           Section 5 Troubleshooting and Maintenance 5.1 Maintenance  Replace the batteries timely when battery indication is low. Clean surface of thе Pulse Oximeter before it is used in diagnosis for patients.  Remove the batteries inside the battery cassette if the oximeter will not be operated for a long time.  It is better to preserve the product in a place where ambient temperature is -10-50℃ and humidity is 15%-80%.  Regular  inspection  to  make  sure  that  no  obvious  damage existed to affect the safety and performance of device.  No  flammable  substance,  overtop  or  lower  temperature  and CAUTION: Don’t use strong solvent. For example, acetone. CAUTION: Never use an abrasive such as steel wool or metal polish. CAUTION: Do not allow any liquid into the product, and do not immerse any parts of the device into any liquids. CAUTION: Avoid pouring liquids on the device while cleaning. CAUTION: Don’t remain any cleaning solution on the surface of the device.
A310B Pulse Oximeter  18 humidity existed in operation conditions.  5.2 Troubleshooting                                                                                           Table 5.2.1 troubleshooting Problems Possible Reason Resolutions Oxyhemoglobin or heart rate can not be shown normally 1. Finger is not plugged correctly. 2. Patient’s perfusion is too low to be measured. 1. Retry by plugging the finger 2. Try some more times, if you can make sure about no problem existing in the product, Please go to a hospital timely for exact diagnosis Oxyhemoglobin or heart rate is shown unstably 1. Finger might not be plugged deep enough 2. Finger is trembling or patient's body is in movement status 1.Retry by plugging the finger   2.Try not to move, Let the patient keep calm. Oxyhemoglobin or heart rate is abnormal and cause alarm   1. Finger is not plugged correctly. 2. Patient’s SPO2&PR is abnormal. 1. 1. Retry by plugging the finger 2. go to the hospital for further examination.
A310B Pulse Oximeter  19 The oximeter can not be powered on 1. Power of batteries might be inadequate or not be there at all 2.Batteries might be installed incorrectly   3.The Oximeter might be damaged 1. Please replace batteries   2.Please reinstall the batteries   3.Please contact with local customer service center he screen are suddenly off 1.The product is automatically powered off when no signal is detected longer than 8 seconds 2.Power quantity of the batteries is exhausted. 1.Normal 2.Replace the batteries   Section 6 Specification Fingertip Pulse Oximeter Specifications: Physical Characteristics Machine: Dimensions -60 mm (L) x 36mm (W) x28mm (D) Weight -approx: 50 g(including 2 x AAA battery)  Color box:
A310B Pulse Oximeter  20 Dimensions-80mm(L)x60mm(w)x55mm(D) Gross Weight: 70g Outer carton: Dimensions-440mm (L) X340mm (W) X290mm (H) Gross Weight: 8.7 kg  Classification : Anti-electric Shock Type:Internally powered equipment Anti-electric Shock Degree:Type BF equipment EMC:Type B Mode of operation: Continuous Operation Enclosure Degree of ingress protection: IP22 IP22 means shell of this product can withstand the water dropping to the surface when the shell deviate 15 degree from horizontal surface.       Power Internal: 2xAAA 1.5v alkaline battery Power Consumption Smaller than 30mA(Normal)
A310B Pulse Oximeter  21  Environmental: Operating Temperature: 5°C to 40°C   Storage Temperature: –10°C to 50°C   Relative Humidity: 15% to 80%   non-condensing Air Pressure 86Kpa-106Kpa  Alarm default value: Parameter Value Hemoglobin saturation: Upper limit: 100/ bottom limit:94 Pulse rate: Upper limit: 130 /bottom limit:50   Electronics Parameters: Parameter Value Hemoglobin saturation display 35-100%   Pulse rate Display 30-250 BPM  Resolution  Hemoglobin Saturation 1% Pulse rate 1 BPM
A310B Pulse Oximeter  22 Measure Accuracy: Hemoglobin Saturation ±3% (70%-100%) Unspecified (<70%) Pulse rate ±1 BPM PI Display 0-20% Resolution 0.1% Measure Accuracy 0-1%: 0.1% 1-20%: 1%  Probe LED Specification;  Wave Length Radiant Power RED 660±2 nm 1.8 mW Infra RED 905±2 nm 2.0 mW
A310B Pulse Oximeter  23  FCC Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -- Reorient or relocate the receiving antenna.
A310B Pulse Oximeter  24 -- Increase the separation between the equipment and receiver.       -- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.     -- Consult the dealer or an experienced radio/TV technician for help. § 15.19 Labeling requirement. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions (1) This device may not cause harmful interference, (2)this device must accept any interference received, including interference that may cause undesired operation. § 15.21 Information to user. Any Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.  Specific Absorption Rate (SAR) information: This product meets the government's requirements for exposure to radio waves. The guidelines are based on standards that were developed by independent scientific organizations through periodic and thorough evaluation of scientific studies.
A310B Pulse Oximeter  25 The standards include a substantial safety margin designed to assure the safety of all persons regardless of age or health.    Body-worn Operation This device was tested for typical body-worn operations. To comply with RF exposure requirements, a minimum separation distance of 5mm for body worn must be maintained between the user’s body, including the antenna. Third-party belt-clips, holsters, and similar accessories used by this device should not contain any metallic components. Body-worn accessories that do not meet these requirements may not comply with RF exposure requirements and should be avoided. Use only the supplied or an approved antenna